manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. GE
  6. •
  7. Medical Equipment
  8. •
  9. GE AKTAprocess User manual

GE AKTAprocess User manual

Other manuals for AKTAprocess

2

Other GE Medical Equipment manuals

GE Dash 2500 User manual

GE

GE Dash 2500 User manual

GE Voluson 730Pro User manual

GE

GE Voluson 730Pro User manual

GE LOGIQ V3 User manual

GE

GE LOGIQ V3 User manual

GE Healthcare Portascan User manual

GE

GE Healthcare Portascan User manual

GE PDM User manual

GE

GE PDM User manual

GE Invenia ABUS 2.0 User manual

GE

GE Invenia ABUS 2.0 User manual

GE SENOGRAPHE DMR+ User manual

GE

GE SENOGRAPHE DMR+ User manual

GE GEH-ECG 1200 User manual

GE

GE GEH-ECG 1200 User manual

GE Voluson i BT06 User manual

GE

GE Voluson i BT06 User manual

GE Corometrics 170 Series User manual

GE

GE Corometrics 170 Series User manual

GE DASH 3000 V5 User manual

GE

GE DASH 3000 V5 User manual

GE MAC 1600 User manual

GE

GE MAC 1600 User manual

GE MAC 800 User manual

GE

GE MAC 800 User manual

GE AMX 4+ User manual

GE

GE AMX 4+ User manual

GE Modular MAC Trolley User manual

GE

GE Modular MAC Trolley User manual

GE LOGIQ S8 User manual

GE

GE LOGIQ S8 User manual

GE Invenia ABUS User manual

GE

GE Invenia ABUS User manual

GE Vivid iq User manual

GE

GE Vivid iq User manual

GE Optima XR200amx User manual

GE

GE Optima XR200amx User manual

GE Vivid 3N Pro Series User manual

GE

GE Vivid 3N Pro Series User manual

GE Innova 2100-IQ User manual

GE

GE Innova 2100-IQ User manual

GE Vivid P3 Installation and user guide

GE

GE Vivid P3 Installation and user guide

GE MAC 800 User manual

GE

GE MAC 800 User manual

GE Voluson P8 BT16 User manual

GE

GE Voluson P8 BT16 User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

28-9266-70 AA Sightglass cover assembly instruction
GE Healthcare
ÄKTAprocess™ Air Trap sightglass cover
Air traps with part numbers listed in the table at the end of this document must be fitted with a metal cover for added protection in case the
air trap is subjected to a sudden overpressure.
Follow the instructions in this document to mount the cover assembly correctly on the air trap.
Note: For added protection, replace allthe sightglass mounting bolts with the enclosed 20 mm bolts as instructed below, before mounting the
sightglass cover assembly.
Note: After the sightglass cover assembly is mounted, the air trap should be tested for leakage and/or damage by applying an overpressure of
1.43 times the rated maximum pressure. This test shall be performed with the air trap filled with liquid. Inspect the air trap for any sign of
deformation before the system is taken into operation.
Sightglass cover assembly kit - parts list
Note: There are different variants of the mounting details and supports connecting the air trap to the ÄKTAprocess system frame. You may find
that the illustrations below do not exactly correspond to the mounting details on your specific system. Please contact GE Healthcare
service if you need advice or assistance when mounting the sightglass cover.
Assembly mount procedure - adjust the sightglass assemblies
Part No. of pcs Part number
Sightglass cover – for 6 mm and 0.5 inch air traps 2 28-9260-08
Sightglass cover – for 1 inch air traps 2 28-9260-36
Bolt M6S M6X14 2 28-9260-71
Nut M6M M6 2 59-1582-00
Washer DIN 125 6.4x12x1.6 4 59-0762-00
Bolt MC6S M5x20 12 56-2210-00
Align and tighten the sightglass assemblies
• Removeoneoftheoriginalbolts(E) fromthetopsightglass
assembly.
• Move the washer (F) from the original bolt to one of the
replacement bolts enclosed with the cover assembly kit.
Note: It is important that the new, longer bolts are used.
• Insert the replacement bolt through the sightglass bracket
(A)andtightenthefirstturnscarefullyby hand. Ensurethat
the bolt is positioned straight.
• Repeat the steps above for the other bolts, cross-wise.
• Ensurethatthesightglassassembly fits snuglyagainstthe
O-ring (D) in the air trap housing and that the
polypropylene spacer (B) and sightglass C are properly
aligned.
• Using a torque setting of 0.5 Nm, tighten the bolts cross-
wise, slowly and carefully.
• Verify that the sightglass bracket (A) is properly aligned
with the air trap housing.
• Increase the torque setting to 1 Nm and tighten the bolts
cross-wise.
Note: To avoid damage to the threads, it is important that the
procedureaboveisfollowed,with propertorquesettings.
2
28-9266-70 AA Sightglass cover assembly instruction
Assembly mount procedure – 6 mm and 1/2 inch air traps
1 Disconnect the air trap and remove the bottom nuts
• Remove the TC coupling (A) and evacuation hose from the
top valve.
• Unscrew and disconnect the cable contacts (B) from the
level sensors.
• Disconnect the airtrap inlet and outlet from the system
valves.
• Loosen andremovefourbottomnuts(E) and washersfrom
the connecting rods that are fixed to the support (D).
Note: The nut on the fifth connecting rod does not need to be
removed.
Note: The washers from the two front connecting rods will not
be used when the cover is mounted.
• Lift the air trap from the support (D) and move it to a
suitable place to mount the cover assembly.
2 Pull out the connecting rods
• Remove the bottom nuts (C) from the two front connecting
rods.
• Pull the connecting rods (F) out of the air trap. Remove the
washers from the rods.
Note: Donotremovethedomenuts.Ensurethatthedomenuts
still are securely fastened. If loose, the nuts should be
fastened and locked into place using Loctite™ 243.
3 Mount the top sightglass cover part
• Place the top cover (G) on the air trap, with the opening
aligned with the sight glass.
• Re-insert the front two connecting rods (F) through the top
holes in the cover.
• Insert a plain washer and a lock washer on each of the
two front connecting rods underneath the air trapand
mount the nuts (C).
Note: The lock washer should be closest to the nut.
• Tighten the nuts using atorque of 3 Nm (±0.3Nm). Verify
also that the other three nuts are tightened, using the
same torque value.
• Return the air trap to the support (D), aligning the bottom
part of each connecting rod through the holes in the
support.
• Connect the air trap inlet and outlet to the system valves.
A
B
E
D
C
F
F
G
28-9266-70 AA Sightglass cover assembly instruction
3
Assembly mount procedure – 1 inch air traps
4 Mount the bottom sightglass cover part
• Slide the bottom cover (H) over the lower ends of the two
front connecting rods (F) under the support (D) and align
the corresponding holes on both cover parts.
• Fasten the bottom cover with two nuts (E) and tighten
securely. Use a suitable wrench to hold the bottom nut (C)
steady while tightening, to avoid increasing the strain on
the connecting rod.
Note: Do not use washers on the front connecting rods.
• Fasten the other connecting rods to the support (D).
5 Connect the cover parts
• Fasten the two cover parts together using bolts (I),
washers (J) and nuts (K).
•Tightensecurely.
6 Final connections
• Reconnect the TC coupling (A) and evacuation hose to the
top valve.
• Attach the cable contacts (B) to the level sensors.
1 Disconnect the evacuation hose and remove the
bottom nuts
• RemovetheTCcouplingandevacuationhosefromthetop
valve (A).
• Loosen and remove the bottom nuts(B) and washers from
the two front connecting rods.
Note: The washers from the two front connecting rods will not
be used when the cover is mounted.
2 Pull out the connecting rods
• Pull the connecting rods (C) out of the air trap. Remove the
washers from the rods.
Note: Donotremovethedomenuts.Ensurethatthedomenuts
still are securely fastened. If loose, the nuts should be
fastened and locked into place using Loctite™ 243.
H
E
I
J
A
B
C
28-9067-71 AA 07/2007
For contact information for your local office, please visit
www.gelifesciences.com/contact
GE Healthcare Bio-Sciences AB
Björkgatan 30
751 84 Uppsala
Sweden
www.gelifesciences.com/protein-purification-bioprocess
GE, imagination at work, and GE monogram are trademarks of General Electric Company.
Drop Design, ÄKTA and ÄKTAprocess are trademarks of GE Healthcare companies.
All third party trademarks are the property of their respective owners.
© 2007 General Electric Company—All rights reserved
First published Jul. 2007
All goods and services are sold subject to the terms and conditions of sale of thecompany within GE
Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact
your local GE Healthcare representative for themost current information.
GE Healthcare Limited
Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK
GE Healthcare Bio-Sciences Corp.
800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327, USA
GE Healthcare Europe GmbH, Munzinger Strasse 5
D-79111 Freiburg, Germany
GE Healthcare Bio-Sciences KK
Sanken Bldg., 3-25-1, Hyakunincho, Shinjuku-ku, Tokyo, 169-0073 Japan
28-9266-70 AA Sightglass cover assembly instruction
Air trap parts number list
Air traps with the part numbers listed below must be fitted with the sightglass cover assembly:
3 Mount the sightglass covers
• Place the top cover (D) on the air trap, with the opening
aligned with the sight glass.
• Re-insertthefronttwoconnectingrods(C)through thetop
holes in the cover.
• Slide the bottom cover (E) over the lower ends of the two
front connecting rods (C) and align the corresponding
holes on both cover parts.
• Replace the bottom nuts (B) and washers on the front
connecting rods underneath the air trap.
• Tighten the nuts using atorque of 3 Nm (±0.3Nm). Verify
also that the other three nuts are tightened, using the
same torque value.
• Fasten the two cover parts together using bolts, washers
and nuts (F) as described in the mount procedure for 6
mm/0.5 inch systems.
4 Final connections
• ReconnecttheTCcouplingand evacuationhoseto the top
valve (A).
C
D
E
B
F
Dimension Material Part number
6 mm PEEK/EPDM 03-0111-41
6 mm PEEK/FEP 03-0111-42
1/2 inch PEEK/EPDM 03-0111-39
1/2 inch PEEK/FEP 03-0111-40
6 mm PP/EPDM 03-0111-45
6 mm PP/FEP 03-0111-46
Dimension Material Part number
1/2 inch PP/EPDM 03-0111-43
1/2 inch PP/FEP 03-0111-44
1 inch PEEK/EPDM 28-4051-68
1 inch PEEK/FEP 28-4051-70
1 inch PP/EPDM 28-4051-69
1 inch PP/FEP 28-4051-71