Hameg HM5033 User manual

®
Instruments
ENGLISH
MANUAL • HANDBUCH • MANUEL
3.3 GHz Portable
Spectrum Analyzer
HM5033

2Subject to change without notice

3
Subject to change without notice

4Subject to change without notice
Before Starting to Use the Unit
When you use the unit, please observe the following notes
listed on the rear of the body.
For you to use it safely
1) When abnormal sounds, abnormal smell and smoke were
confirmed, remove the battery and AC adapter and stop the use.
2) Never use with hands that got wet, because doing so may
cause damage, fire and electric shock to the unit.
3) Never use it under the thunder.There is a possibility of receiving
a thunderbolt.
4) Never use an AC adapter other than the one specified, because
doing so may cause damage to the unit. For static electricity
protection, ground the unit by connecting the three cores if
possible. Not grounding the unit can damage it and the object
measured.
5) Never use a battery other than the one specified, because
doing so may cause damage to the unit. When removing or
installing the battery, be sure to do it after you turn off the unit
and disconnect the AC adapter.
6) When replacing the fuse, disconnect the AC adapter, open the
battery cover on the back and remove battery, and then take
sufficient care to perform the replacement. Use 5 A / 250 V
fuse (IEC127-2 sheet 3, slow-blow type).
Never use a fuse not specified because doing so may cause
damage to the unit.
Guarantee of quality
Guarantee period
Guarantees that it will repair any failure free of charge if it
occurs because of our responsibility within one year after
delivery. However, the above guarantee does not apply to such
a failure that:
1) is caused by a fire, natural disasters, etc.
2) is caused by inappropriate handling of the unit, such as dropping
it while moving it after purchasing.
3) is caused by handling counter to the instructions or precautions
listed in the operating manual.
4) is caused by modifying the unit or by being considered to be
your responsibility because of inappropriate use.
We will not be responsible for direct or indirect damage caused
by use of this product or by a failure of this product.
Warm-up time
In order to stabilize the electric performance at the time of turning
on the unit, please perform warming-up for at least 10 minutes.
Precautions for storage
1) Strictly observe the storage conditions specified for this unit,
such as avoiding direct sunlight and dust.
2) Store this unit in a place where –20 °C to 60 °C, less than
60 °C / 70 %RH, variations in temperature and humidity are
small.
After service
If you have any question about the contents of this product or
how to operate it, please contact us at:
HAMEG GmbH
Industriestr. 6
D-63533 Mainhausen
Tel.: ++49 (0) 6182 89090
URL: http://www.hameg.de
E-mail: [email protected]

5
Subject to change without notice
General information regarding the CE marking
HAMEG instruments comply with the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual
generic- and product standards. In cases where different limit values are applicable, HAMEG applies the severer standard. For
emission the limits for residential, commercial and light industry are applied. Regarding the immunity (susceptibility) the limits for
industrial environment have been used.
The measuring- and data lines of the instrument have much influence on emmission and immunity and therefore on meeting the
acceptance limits. For different applications the lines and/or cables used may be different. For measurement operation the following
hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) sufficiently screened
cables must be used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline
must be less than 3 meters and not be used outside buildings. If an interface has several connectors only one connector must have
a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L
from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction
in the manual for a shorter length, signal lines must be less than 3 meters and not be used outside buildings.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators
double screened cables (RG223/U, RG214/U) must be used.
3. Influence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fields, even with careful setup of the measuring equipment an
influence of such signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the
instruments specifications may result from such conditions in individual cases.
4. RF immunity of Spectrum Analyser
4.1 Electromagnetic RF field
Although the interior of the spectrum analyser is screened by the cabinet, the influence of electromagnetic RF fields may become
visible, if the field intensity under the present environmental condition is high. The device under test as well as the measuring cable
may also receive such signals.
HAMEG GmbH
KONFORMITÄTSERKLÄRUNG
DECLARATION OF CONFORMITY
DECLARATION DE CONFORMITE
Instruments
®
Herstellers HAMEG GmbH
Manufacturer Kelsterbacherstraße 15-19
Fabricant D - 60528 Frankfurt
Die HAMEG GmbH bescheinigt die Konformität für das Produkt
The HAMEG GmbH herewith declares conformity of the product
HAMEG GmbH déclare la conformite du produit
Bezeichnung / Product name / Designation:
Spektrum-Analysator/Spectrum Analyzer/Analyseur de spectre
Typ / Type / Type: HM5033
mit / with / avec: -
Optionen / Options / Options: -
mit den folgenden Bestimmungen / with applicable regulations / avec les
directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
Angewendete harmonisierte Normen / Harmonized standards applied / Normes
harmonisées utilisées
Sicherheit / Safety / Sécurité
EN 61010-1: 2001 / IEC (CEI) 1010-1: 2001
Messkategorie / Measuring category / Catégorie de mesure: I
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility /
Compatibilité électromagnétique
EN 61326-1/A1 :1997 + A1:1998 + A2 :2001/IEC 61326 :1997 + A1 :1998 + A2 :2001
Störaussendung / Radiation / Emission:Tabelle / table / tableau 4; Klasse / Class /Classe B.
Störfestigkeit / Immunity / Imunitee:Tabelle / table / tableau A1.
EN 61000-3-2/A14
Oberschwingungsströme / Harmonic current emissions / Émissions de courant
harmonique: Klasse / Class / Classe D.
EN 61000-3-3
Spannungsschwankungen u. Flicker / Voltage fluctuations and flicker / Fluctuations
de tension et du flicker.
Datum / Date / Date Unterschrift / Signature / Signatur
25.6.2003
G. Hübenett
Product Manager

6Subject to change without notice

7
Subject to change without notice
1. Outlines ............................................................................... 8
1.1 Product outlines .......................................................... 8
1.2 Standard accessories .................................................. 8
1.3 Optional accessories .................................................. 8
2. Specifications ..................................................................... 8
2.1 Performances .............................................................. 8
2.2 Outline ....................................................................... 10
3. Description of Panel ........................................................ 10
4. Description of Screen ...................................................... 11
5. Function Key Menu ......................................................... 12
5.1 List of the function key menus ................................ 12
5.2 Menu tree ................................................................. 12
6. Preparing for Operation .................................................. 14
6.1 Stand ......................................................................... 14
6.2 Connection to power supply .................................... 14
6.3 Replacing the fuse .................................................... 14
6.4 Installing the battery ................................................. 14
6.5 Soft carrying case ..................................................... 14
7. Center Frequency <FREQ> ............................................. 15
7.1 Setting with the step keys ........................................ 15
7.2 Setting with the encoder .......................................... 15
7.3 Setting with the numeric keys.................................. 15
7.4 According to the marker position ............................. 15
8. Frequency Span <SPAN> ................................................ 15
9. Reference Level <REFER> ............................................... 16
9.1 Setting the reference level ....................................... 16
9.2 Switching units of amplitude axis ............................ 16
9.3 Reference level setting range for each unit ............. 16
9.4 Relation between the reference level ...................... 16
and ATT·AMP
9.5 Setting the offset level.............................................. 16
9.6 Setting the input impedance .................................... 17
10. Display Scale <SCALE> ................................................... 17
10.1 Setting with the keys ................................................ 17
10.2 Setting with the encoder .......................................... 17
11. Resolution Bandwidth <RBW> ...................................... 17
11.1 MANUAL mode......................................................... 17
11.2 AUTO mode ............................................................... 17
11.3 ALL AUTO mode ....................................................... 17
12. Video Bandwidth <VBW> ............................................... 17
12.1 MANUAL mode ......................................................... 17
12.2 AUTO mode ............................................................... 17
12.3 ALL AUTO mode ....................................................... 18
13. Sweep Axis · Detection mode <SWEEP> ..................... 18
13.1 MANUAL mode ......................................................... 18
13.2 AUTO mode ............................................................... 18
13.3 ALL AUTO mode ....................................................... 18
13.4 Setting the detection mode ...................................... 18
14. AUTO Tuning <AUTO TUNE> .......................................... 18
15. Hold/Run <HOLD/RUN> ................................................. 18
16. Calculation Function <CALC> ........................................ 18
16.1 NORM mode ............................................................. 18
16.2 MAX HOLD mode ..................................................... 18
16.3 MIN HOLD mode ...................................................... 19
16.4 AVERAGE mode ........................................................ 19
16.5 OVER WRITE mode ................................................... 19
17. Marker · Peak Search <MKR> ......................................... 19
17.1 Moving the marker.................................................... 19
17.2 Setting the peak search ............................................ 19
<PEAK SEARCH>
17.3 Changing the unit of marker point ........................... 19
18. Save/Load <SAVE/LOAD> .............................................. 19
18.1 Setting the location to store the trace ..................... 19
18.2 Setting the location to store the parameter ............ 20
18.3 Saving the data ......................................................... 20
18.4 Loading the data ....................................................... 20
18.5 Clearing the loaded trace .......................................... 20
18.6 Presetting (Initialization) ............................................ 20
19. Measuring Function <MEAS> ........................................ 20
19.1 Channel power measurement .................................. 20
<Ch Power>
19.2 Adjacent channel leakage power measurement ..... 21
<Adj Ch Pw>
19.3 Occupied frequency bandwidth measurement ....... 22
<Occ Bw>
19.4 Electric field strength measurement ....................... 22
<E/F ANT>
19.5 Magnetic field strength measurement .................... 25
<M/F PROBE> (optional)
20. Screen Control <DSPL> .................................................. 26
20.1 Adjusting the contrast .............................................. 26
20.2 Switching ON and OFF the LCD backlight ............... 26
20.3 Adjusting the brightness of the LCD backlight ........ 26
20.4 Inverting the display .................................................. 26
20.5 Enable or disabling the beep .................................... 26
21. Printing <PRINT> (optional) ........................................... 26
21.1 Hard copy of the screen ........................................... 26
22. Data Output <RS232C> ................................................... 26
22.1 Selecting the trace to transfer.................................. 26
22.2 Selecting the communication speed (baud rate) ..... 26
22.3 Transfer the data ....................................................... 26
23. RS-232C Interface ............................................................ 28
23.1 RS-232C specifications ............................................. 28
23.2 How to connect ........................................................ 28
23.3 Command description .............................................. 28
23.4 Input the frequency .................................................. 28
23.5 Writing of original correction data ............................ 28
23.6 Sample Programs...................................................... 32
24. PC Software (optional) ................................................... 32
25. Basis Performance Test ................................................... 33
25.1 Frequency characteristics ......................................... 33
25.2 Accuracy of reference level ...................................... 35
25.3 The display accuracy of the center frequency ......... 35
25.4 The display accuracy of the frequency span ........... 35
25.5 Linearity of the amplitude axis ................................. 36
Table of contents

8Subject to change without notice
1. Outlines
1.1 Product outlines
HM5033 is an authentic spectrum analyzer providing
performance and functions that are comparable to those of large-
size bench type equipment, in a compact, lightweight and
inexpensive model.
1) Compact and lightweight, 1.7 kg
The external dimensions are as small as 162 (W) × 70 (H) ×
260 (D) (mm), and the weight is only 1.7 kg including the
battery. It is very convenient for outdoor use and while on
business trips.
2) Measuring frequency bandwidth 50 kHz to 3.3 GHz
This bandwidth covers those of W-CDMA, CDMA, PDC, PHS,
GSM, 2.4 GHz band wireless LAN, Bluetooth, etc.
3) Operation with battery for 120 minutes
When battery MB300 is fully charged, HM5033 works for about
120 minutes (with the backlight turned off). It is extremely
convenient for outdoor use and for use in the survey of wireless
LAN installation environment.
4) Performance that is comparable to that of large-size bench
type equipment
HM5033 guarantees a highly stable frequency axis by PLL
synthesizer system. The center frequency setup resolution is
100 kHz. Furthermore, the mean noise level is -110 dBm or less.
Thus, a broad dynamic range is secured and the reference level
can be set in 1 dB steps.
5) Abundant functions
Measuring functions: Channel power measurement,
Adjacent channel leakage power measurement, Occupied
frequency bandwidth measurement, Electric field strength
measurement.
Electric field strength measurement: Optimum for
measurement of cellular phone and wireless LAN working
environment.
Magnetic field strength measurement: Optimum for EMI
design of printed circuit boards and for evaluation of signal quality.
Calculation functions: MAX HOLD, MIN HOLD, AVERAGE,
OVER WRITE
Marker & peak search
Save/Load
6) Auto tuning
The center frequency is set at the spectrum of the
maximum level in the 3.3 GHz band, and in addition,
optimum reference level, resolution bandwidth, video
bandwidth and sweep time are set when the AUTO TUNE
key is pressed. This function is very convenient for
measurement of an unknown signal.
7) Auto range motion
The resolution bandwidth, video bandwidth and sweep time are
set automatically based on the set frequency span. It is also
possible to set auto range motion only one out of resolution
bandwidth, video bandwidth and sweep time.
8) Hard copy of the image
Connect a printer (optional) and press the [PRINT] key on
HM5033. The image on the screen is printed as it is.
9) High resolution display on the PC screen
The trace is displayed at high resolution, 1001 points in the
horizontal axis, on the PC screen when „PC Software MAS300"
(optional) is used.
1.2 Standard accessories
1. AC adaptor MA300
2. Soft carrying case
3. Accessory pouch
4. Operating manual
5. Ni-MH battery MB300 (Refer to„6.4 Installing the battery“ for details.)
1.3 Optional accessories
1. PC software MAS300 (Refer to „24. PC Software“ for details).
2. Electric field antenna M301, M302, M303, M304, M305 (Please
note „19.4 Electric field strength measurement“).
3. Magnetic field probe CP-2S with a dedicated double shielded
coaxial cable (Please note „19.5 Magnetic field strength
measurement“).
4. Ni-MH battery MB300 (Refer to„6.4 Installing teh battery“ for details).
5. Battery charger MBC300
6. Printer with AC adaptor, 4 pcs. of AA batteries and one paper
roll (Please note „21. Printing“)
7. Paper pack (10 rolls)
Outlines
2. Specifications
2.1 Performance
Frequency section
Frequency range: 50 kHz to 3.3 GHz
Center frequency
Setting resolution: 100 kHz, allows Rotary encoder,
numeric key and function key
Accuracy (kHz): within ±(30 + 100 x t) ±1 dot
t: Sweep time (s)
(frequency span: 200 kHz to 10 MHz,
RBW: 30 kHz, 23 °C ±5 °C)
within ±(100 +700 x t) ±1 dot
t: Sweep time (s)
(frequency span: 20 MHz to 3.3 GHz,
RBW: 100 kHz, 23 °C ±5 °C)
RBW frequency error: within ±6 % of RBW
(RBW: 3 kHz, 30 kHz)
within ±30 % of RBW
(RBW: 100 kHz to 3 MHz)

9
Subject to change without notice
RBW switching error: within ±0.6 dB
Display resolution: 0.4 dB (10 dB/div)
0.08 dB (2 dB/div)
Display dot number: 200 dots
Display Scale:
Scale: 10 dB / div, 2 dB/div
Accuracy: within ±0.2 dB / 2 dB ±1 dot
within ±0.8 dB / 10 dB ±1 dot
within ±1.6 dB / 70 dB ±1 dot
Input damage level: +20 dBm (CW average power), 25 VDC
Sweep section
Sweep time 10 ms to 30 s
(1-3step, frequency span: 0 to 2 GHz) and AUTO
Setting range: 30 ms to 30 s
(1-3step, frequency span: full span) and AUTO
Accuracy: within ±0.1 % ±1 dot
(frequency span: 0 to 2 GHz)
within ±1.5 % ±1 dot
(frequency span: full span)
Trigger mode: AUTO (frequency span: zero span)
Detection mode: Positive peak, Negative peak, Sample
(When sweep time is 10 ms or 30 ms,
only Sample can be set)
Functions
Marker:
NORM: displays frequency (7 digits max)
and level (4 digits max) at marker point
.
DELTA: displays differential frequency
and level between 2 markers.
Peak search:
NORM: searches a peak point within 10 div.
Available NEXT peak (10 max).
ZONE: searches a peak point within a zone
designated by center and width.
Marker moves to a peak point each sweep.
Calculation: NORM, MAX HOLD, MIN HOLD,
AVERAGE, OVER WRITE
MAX/MIN HOLD: 2 to 1024 times
AVERAGE: 2 to 256
Measuring:
Channel power, Adjacent channel leakage power, Occupied
frequency bandwidth, Electric field strength (optional
antenna),Magnetic field strength (needs optional magnetic field
probe) measurement, Frequency counter.
Frequency span
Setting range: 0 Hz (zero span),
200 kHz to 2 GHz (1-2-5step),
3.3 GHz (full span)
Accuracy (kHz): within [±3 % (20 x t)] ±1 dot
(frequency span: 200 kHz to 10 MHz, 23 °C ±5 °C)
within [±3 % ±(200 x t)] ±1 dot
(frequency span: 20 MHz to 3.3 GHz, 23 °C ±5 °C)
t: Sweep time (s)
Display resolution:
LCD: Frequency span/250
PC Monitor (max.): Frequency span/1000 (via RS-232C)
Display dot number:
LCD: 251 dots,
PC Monitor (max.): 1001 dots (via RS-232C)
(The unit displays data as 251 horizontal dots,
but it internally captures the signal as in 1001 dots)
Resolution bandwidth: 3 dB bandwidth
Setting range: 3 kHz to 3 MHz (1-3step) and AUTO
Accuracy: within ±20 %
Selectivity: 1 : 12 (typical, 3 dB : 60 dB)
Video bandwidth: 100 Hz to 300 kHz (1-3step), OFF and AUTO
SSB phase noise: -90 dBc/Hz (typical, 100 kHz offset,
RBW: 3 kHz, VBW: 100 Hz, Sweep time: 0.3 s)
Spurious response: less than -60 dBc
Harmonics: less than -40 dBc (50 kHz to 100 MHz)
less than -45 dBc (100 Mhz to 3.3 GHz)
Amplitude section
Reference level
Setting range: +10 to -60 dBm (1dB step)
Accuracy: within ±0.8 dB ±1 dot
(center frequency: 100 MHz, RBW: 3 MHz,
VBW: 1 MHz, ATT: 0 dB, 23 °C ±5 °C)
Unit: dBm, dBV, dBmV, dBµV, dBµV/m, dBµA/m
(dBµV/m and dBµA/m is used the measuring function)
Average noise level: -110 dBm
(typical, center frequency: 100 MHz, RBW: 3 kHz, VBW: 100 Hz)
Frequency Characteristic:
within ±2.0 dB ±1 dot (50 kHz to 100 MHz)
within ±1.0 dB ±1 dot (100 Mhz to 3.3 GHz)
Input impedance: 50 Ω
Input VSWR: less than 2.0
Input attenuator:
Operating range: 0 to 25 dB (1 dB step),
coupled with reference level
Switching error: within ±0.6 dB
Specifications

10 Subject to change without notice
AUTO tuning:
When pushing AUTO TUNE key, the maximum level spectrum
within 3.3 GHz bandwidth is adjusted to center, and reference
level, RBW, VBW and sweep time are adjusted to optimum values.
Save/Load
Save: Saves 100 traces and 100 setups
Load: Loads 1 trace and 1 setup
General
Input connector: SMA(J)
Communication:
Interface: RS-232C
Baud rate: 2400 to 38400 bps
Hard copy: Allows direct hard copy with an optional printer.
Display: LCD
Backlight: CFL backlight
Resolution: 320 (H) × 240 (V) dots
Power source:
Battery: Ni-MH battery
External DC source: Pin jack, DC 5 V / 4 A
Other
Operating temperature: 0 °C to 40 °C
(Guaranteed at 23 °C ±10 °C, without soft carrying case)
Operating humidity: less than 40 °C / 80 % RH
(Guaranteed at less than 33 °C / 70 % RH,
without soft carrying case)
Storage temperature: -20 °C to 60 °C,
less than 60 °C / 70 % RH
Dimensions (W x H x D):162 × 70 × 260 mm
Weight: approx. 1.5 kg (without battery)
2.2 Outline
HAMEG & MICRONIX reserves the right to make changes in
design, specification and other information without prior notice.
Description of Panel
3. Description of Panel
1) LCD screen
This is a large liquid crystal display with 240 (V) × 320 (H)
dots. It simultaneously displays traces (8 ×10 div), various
setting values, measured values, etc.
2) Input connector
SMA (J) connector.
3) Input connector for DC power source
Connects AC adaptor MA300.
4) RS-232C connector
Connects PC and printer, by using RS-232C cable MI180.
5) Function keys (F1 to F6)
Functions change according to operation. Have functions cor-
responding to the on-screen displays.
6) Center frequency key
Use this key to set the center frequency. It can set between
0 to 3.3 GHz (100 kHz step).
7) Frequency span key
Use this key to set the frequency span. It can set between
200 kHz to 2 GHz, ZERO SPAN and FULL SPAN (3.3 GHz).
8) Reference level key
Set the reference level, etc. Reference level can set between
+10dBm and -60dBm (1dB step).
9) Resolution bandwidth key
Use this key to set the resolution bandwidth. It can set
between 3 kHz and 3 MHz.
10) Video bandwidth key
Use this key to set the video bandwidth. It can set between
100 Hz and 1 MHz .
11) AUTO tuning key
Tune up to the maximum level in 3.3 GHz zones, and display
by the optimal setup. This does not operate normally when
the signal level is lower than -40 dBm , or when the input
frequency is below 50 MHz, or when the frequency span is
ZERO SPAN or FULL SPAN.

11
Subject to change without notice
Description of Screen
12) Measuring function key
Available for Channel power, Adjacent channel leakage power,
Occupied frequency bandwidth, Electric field strength and
Magnetic field strength measurement (optional), Frequency
counter (factory option).
13) Calculation function key
Available for Max hold, Min hold, Average and Over write.
14) Display scale key
Use this key to select the display scale of amplitude axis
from 2 dB/div or 10 dB/div.
15) Sweep key
Use this key to set the sweep time between 10 ms to 30 s
or set the detection mode.
16) Hold/Run key
Stops or restarts the measurement.
17) Marker & Peak search key
Use this key to set and move a marker.
18) Save/Load key
Saves 100 traces and 100 setups, and loads 1trace and 1setup.
19) Print key
When pressing this key, the image is printed with a printer
(optional) as it is.
20) RS-232C key
Sets baud rate and transfers a current or saved trace.
21) Display control key
Sets contrast, backlight ON/OFF, brightness of backlight,
invert display and buzzer ON/OFF.
22) Rotary encoder
Use this to make various settings.
23) Power switch
Use this to turn the power ON or OFF.
4. Description of Screen
1) Trace display
8 div × 10 div.
2) Frequency axis setting values display
Center frequency, Frequency span,Resolution bandwidth,
Video bandwidth.
3) Amplitude axis setting values display
Reference level, Input attenuator, Display scale.
4) Sweep axis setting values display
Sweep time, Detection mode.
5) Calculation function display
6) Measuring function display
7) Operating information display
8) Loaded trace information display
This is also used as a setting values display when the
measuring function is used.
9) Display for function keys
10) Measured values display

12 Subject to change without notice
Function Key Menu
5. Function Key Menu
5.1 List of the Function key menus
The types of function keys are shown in the table below. For
description of each function, see the detailed pages. For the flow
of change in the function key display, refer to „5.2 Menu tree“.
P) PARAM SAVE/LOAD→F2 20
PEAK SEARCH CNTR MKR→(F3)→F4 19
PEAK SEARCH NEXT MKR→(F3)→F5 19
PEAK SEARCH NORM MKR→(F3)→F3 19
PEAK SEARCH PEAK MKR→(F3)→F4 19
PEAK SEARCH WIDTH MKR→(F3)→F5 19
PEAK SEARCH ZONE MKR→(F3)→F3 19
PRE SET SAVE/LOAD→F6 20
PROBE MEAS→(F6)→(F5)→F1 25
R) RATIO MEAS→(F6)→(F3)→F2 22
RBW ALL RBW→F3 17
RBW AUTO RBW→F2 17
RBW MANU RBW→F1 17
REFERENCE CNTR MEAS→(F6)→(F2)→F4 21
REFERENCE WIDTH MEAS→(F6)→(F2)→F5 21
S) SCALE 10dB SCALE→F1 17
SCALE 2dB SCALE→F2 17
SET MKR FREQ→F6 15
SWEEP ALL SWEEP→F3 18
SWEEP AUTO SWEEP→F2 18
SWEEP MANU SWEEP→F1 18
T) TRACE SAVE/LOAD→F1 19
RS232C→F1 26
U) UNIT REFER→F1 to 4 16
V) VBW ALL VBW→F3 18
VBW AUTO VBW→F2 17
VBW MANU VBW→F1 17
Function key menu Key flow Page
A) Adj Ch OFS MEAS→(F6)→(F2)→F2 21
Adj Ch Pw MEAS→(F6)→F2 21
Adj Ch WIDTH MEAS→(F6)→(F2)→F3 21
ANT MEAS→(F6)→(F5)→F1 22
AVER CALC→F4 19
B) B.L. DSPL→F2 26
BACK SPACE FREQ→F5→F6 15
BAND CNTR MEAS→(F6)→(F1)→(F1)→F2 21
BAND WIDTH MEAS→(F6)→(F1)→(F1)→F3 21
BAUD RS232C→F2 26
BLCTR DSPL→F3 26
BUZZR DSPL→F5 26
C) CENTER FREQ FREQ→F1 15
CENTER FREQ FREQ→F2 15
Ch Power MEAS→(F6)→F1 20
CLEAR FREQ→F5→F5 15
CONV MKR→F6 19
CTRS DSPL→F1 26
D) DET SWEEP→F4 18
E) E/F ANT MEAS→(F6)→(F3)→F4 22
EncST FREQ→F4 15
EXEC RS232C→F3 26
EXECUTE DEL SAVE/LOAD→F5 20
EXECUTE LOAD SAVE/LOAD→F4 20
EXECUTE SAVE SAVE/LOAD→F3 20
F) Freq COUNT MEAS→F6 20
I) IMP REFER→F6 17
INVT DSPL→F4 26
K) KEYST FREQ→F3 15
M) M/F PROBE MEAS→(F6)→F5 25
MAXHD CALC→F2 18
MEAS OFF MEAS→(F1 to 5)→F6 25
MINHD CALC→F3 19
MKR DELTA MKR→F2 19
MKR NORM MKR→F1 19
MODE MEAS→(F6)→(F1 to 3)→F1 21/22
N) NORM CALC→F1 18
NUM FREQ→F5 15
O) Occ BW MEAS→(F6)→F3 22
OFSdB REFER→F5 16
OVRWR CALC→F5 19
5.2 Menu tree
The displayed items on the bottom of the screen correspond to
the function keys under them, as shown in the figure below:
„Displayed items on the bottom of the screen“
KeyST
100M
EncST
0.1M NUM
SET
MKR
CLEAR
BACK
SPACE
FREQ Refer to "7. Center Frequency" for details
Set the center frequency
Ch
Power
F1
Adj
Ch Pw
F2
Occ
BW
F3
E/F
ANT
F4
M/F
PROBE
F5
Freq
COUNT
F6

13
Subject to change without notice
Function Key Menu
Ch
Power
Adj
Ch Pw
Occ
BW
E/F
ANT
M/F
PROBE
Freq
COUNT
BAND
CNTR WIDTH
MODE
BAND
MEAS
OFF
MEAS 1
st
Refer to "19. Measuring Function" for details
Adj Ch Pw
OFS WIDTH
MODE
TOTAL
MEAS
OFF
REFERENCE
CNTR WIDTH
RATIO
99.5%
MODE
N%
MEAS
OFF
ANT
M301
MEAS
OFF
Set the parameter
Set the parameter
Set the parameter
Input the probe ID (first time only) –
Refer to "19.5 Magnetic field strength measurement" for details
Ch
Power
Adj
Ch Pw
Occ
BW
E/F
ANT
M/F
PROBE
Freq
COUNT
PROBE
CP-2S
MEAS
OFF
MEAS 2
nd
NORM MAXHD
**
MINHD
**
AVER
256
OVRWR
CALC Refer to "16. Calculation functions" for details
Set the number
RBW
RBW Refer to "11. Resolution Bandwidth" for details
Set the RBW
MANU AUTO ALL
SCALE
SCALE Refer to "10. Display Scale" for details
Set the display scale
10 dB 2 dB
VBW
VBW Refer to "12. Video Bandwidth" for details
Set the VBW
MANU AUTO ALL
SWEEP
SWEEP Refer to "13. Sweep Axis" for details
Set the sweep time
MANU AUTO ALL
MKR Refer to "17. Marker / Peak Search" for details
Move the marker position
(NORM mode)
F3: Changing the marker mode
PEAK SEARCH
NORM PEAK NEXT
MKR
NORM DELTA
Set the zone center frequency (ZONE mode)
PEAK SEARCH
ZONE CNTR WIDTH
MKR
NORM DELTA
PRE
SET
Refer to "18. SAVE/LOAD" for details
Set the adress to store the trace
EXECUTE
SAVE LOAD DEL
TRACE
00
SAVE/LOAD
00
PARA M
EXEC
Refer to "22. Data Output" for details
Set the trace to transfer
TRACE
CURR
RS 232 C
38400
BAUD
BLCTR
200
INVT
OFF
BUZZR
ON
Refer to "20. Screen Control" for details
Set the contrast
CTRS
140
DSPL
ON
B.L.
UNIT OFSdB
0.0
REFER Refer to "9. Reference Level" for details
Set the reference level
dBm dBµV dBmV dBV

14 Subject to change without notice
Preparing for Operation
6. Preparing for Operation
6.1 Stand
Utilize the stand on the back to use the screen in an easier-to-
see angle on the desk.
6.2 Connection to power supply
The MA300 AC adapter is both for the use with AC power supply
and for charging the MB300 built-in battery (charge is started
automatically if AC adapter is connected).
Connect the adapter as in the figure below and connect the AC
plug to the power line (100 to 240 VAC, 50/60 Hz). For static
electricity protection, ground the unit by connecting the three
cores if possible. Not grounding the unit can damage it and the
object measured. Do not use an AC adapter other than the MA300
supplied with the unit. Using an AC adapter other than the MA300
may cause damage to the unit.
Battery full charge time: approx. 8 hours
Battery operate time: the longest 120 min (back light OFF)
Battery full charge time with MBC300:approx.120 to 150 min
When it is not operated at normal temperature and setting
parameters is the initialization.
If the voltage of a battery becomes low at the time of battery
operation, it will be displayed on a screen as „Low Batt“,
and a buzzer will sound (it sounds, even if it is set up so
that a buzzer may not sound), and a power supply will be
shut off within several minutes. At that time, since the
switch is the position of „ON“, please push once and return
to the position of „OFF“. If it is with the position of „ON“
after a power supply is shut off, it discharges inside, will be
in electric overdischarge state, and becomes the cause of
contracting the life of a battery. Please take care.
Moreover, under low temperature (near 0 ºC), since a battery
performance falls and voltage becomes low, even when capacity
remains enough, it may display on a screen as „Low Batt“.
When a battery repeats charge and electric discharge, the
fall (the fall of capacity and increase in internal resistance) of
a battery performance begins from about 200 times, and
capacity falls to the original half by about 500 times also
under good conditions. On bad conditions (high temperature,
etc.), the life of battery will be shorter than this.
6.3 Replacing the fuse
5 A / 250 V fuse (IEC127-2 sheet3, slow-blow type) is used for
the battery power supply. When replacing it, turn the power off
first, disconnect the AC adapter, remove the battery cover and
on the back as shown in the figure below, remove the battery,
and then take sufficient care to perform the replacement.
Be sure to use the fuse supplied with the unit, or specified one.
6.4 Installing the battery
When installing the battery, turn the power off first, disconnect
the AC adapter, open the battery cover on the back of the unit
after removing the two screws as shown in the figure below,
and then take sufficient care to perform the installation. Be sure
to use the specified battery, MB300.
6.5 Soft carrying case
When carrying the unit or using it outdoors, the soft carrying
case is convenient. You can also carry the AC adapter and printer
with it, putting them in the accessory pouch.
Avoid using the unit in the soft carrying case in places where
temperature is high because, with the soft carrying case,
the temperature inside becomes higher than the ambient
temperature.

15
Subject to change without notice
Center Frequency - Frequency Span
7. Center Frequency <FREQ>
Center frequency can be set between 0 to 3.3 GHz.Center
frequency may shift for the time being (1 to 10 sec.), after
changing a setting.
7.1 Setting with the step keys ([F1], [F2])
1. Each time
F1
is pressed, the center frequency decreases
in the set step size.
2. Each time
F2
is pressed, the center frequency increases
in the set step size.
3. Setting the step size:
Each time
F3
is pressed, it is set in the following order:
7.2 Setting with the encoder
1. When is turned, the center frequency changes in the set
step size.
2. Setting the step size:
Each time
F4
is pressed, it is set in the following order:
7.3 Setting with the numeric keys
1. Press
F5
to enter into the numeric key input mode.
[F5] functions as the <CLEAR> key. [F6] functions as the
<BACK SPACE> key.
In this mode, setting with [F1], [F2] or the encoder is not
accepted.
2. The center frequency can be directly input according to the
„Numeric Key Mapping Diagram“.
3. The value is entered by pressing a unit key, [MHz (RS232C)] or
[GHz (DSPL)].
Any figures below the resolution (100 kHz) will be
discarded.
4. Changing the setting:
F5
: Deletes the entire value and allows you to input
one from the beginning.
F6
: Deletes the last input figure.
KeyST
100M
EncST
0.1M NUM
SET
MKR
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press
FREQ
0,1MHz 1MHz 10MHz 100MHz
0,1MHz 1MHz 10MHz 100MHz
5. Canceling the numeric key mode:
FREO
: Enables setting with step keys ([F1], [F2]) or the
encoder again.
„Numeric Key Mapping Diagram“
7.4 According to the Marker position
1. When
F6
is pressed, the center frequency is set
according to the frequency of current marker position.
Any figures below the resolution (100 kHz) will be discarded.
This does not operate when the marker is not displayed
(and the function display disappears).
8. Frequency Span <SPAN>
Press
SPAN
and use to set the frequency span.
The frequency span can be set only with the encoder.
Function keys are not available.
1. When is turned, the frequency span changes in the
specified step.
RBW
SCALE
PRINT
VBW
SWEEP
MEAS
MKR
CALC
SAVE
LOAD
RS232C
AUTO
TUNE
HOLD
RUN
DSPL
1
6
2
7
3
8
MHz GHz
4
9
5
0
ZERO 200k 500k 1M 2M 5M 10M
20M 50M 100M 500M 1G 2G FULL (3.3G)
[Hz]

16 Subject to change without notice
Reference Level
UNIT OFSdB
0.0
dBm dBµV dBmV dBV
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press
REFER
9. Reference Level <REFER>
9.1 Setting the Reference level
1. When is turned, the reference level changes.
(Refer to „9.3 Reference level setting range for each unit“ for
details.)
9.2 Switching units of amplitude axis
(dBµV/m and dBµA/m are optional. Refer to „19.4 Electric field
strength measurement“ and „19.5 Magnetic field strength
measurement“ for details.)
1. Press
F1
to switching units to dBm.
Press
F2
to switching units to dBµV
Press
F3
to switching units to dBmV
Press
F4
to switching units to dBV
9.3 Reference level setting range for each unit
UNIT dBm dBµµ
µµ
µV dBmV dBV
MAXIMUM 10 117 57 -3
MINIMUM -40 67 7 -53
MINIMUM* -60 47 -13 -33
(
*
shifted trace data)
„Unit that is able to use it with the measuring function“
UNIT dBµµ
µµ
µV/m** dBµµ
µµ
µA/m***
Setting M301 M302 M303 M304 M305 CP2S
MAXIMUM 143 146 148 150 137 160 to 203
MINIMUM 93 96 98 100 87 110 to 153
MINIMUM* 73 76 78 80 67 90 to 133
(
*
shifted trace data
(
*
*Electric filed strength measurement)
(
***
Magnetic field strength measurement)
When the reference level is set between the „MINIMUN“
and „MINIMUN (shifted trace data)“, the trace in
„MINIMUM“ is shifted and displayed on a screen. When
the reference level is set below to the „MINIMUM“, the
ATT display area is displayed as „S/W AMP“.
Calculation expression (conversion formula to and from dBm):
A [dBµµ
µµ
µV] = 107+X [dBm]
B [dBmV] = 47+X [dBm]
C [dBV] = -13+X [dBm]
D [dBµµ
µµ
µV/m] = 68.8/λ×√(X/Gar) [dBm]
λ: Wavelength[m]
Gar: Antenna absolute gain [times]
E [dBµµ
µµ
µA/m] = 180+X+F [dBm]
F: probe calibration coefficient
changes by frequency
9.4 Relation between the reference level and
ATT · AMP (in dBm indication)
The programmable attenuator (ATT) and the input amplifier (AMP)
inside HM5033 are automatically set according to the setting
value of the reference level (REFER). ATT cannot be set
independently.
When the input signal level is higher than the suitable level
for 1st mixer’s terminal, it generates harmonics distortion
and spurious. HM5033 is designed so the input signal level
of 1st mixer is determined between -20 and
-25 dBm by the reference level. An input level to 1st mixer
is calculated for by the following formula.
Input level of 1st mixer [dBm] = Input level [dBm] - (Fixed ATT
+ Insertion loss [dB]) – ATT [dB] + Input AMP [dB]
„(Fixed ATT + Insertion loss [dB])“ is always 5 dB.
9.5 Setting the offset level
1. Press
F5
and use to set the offset of reference level.
When amplifier and attenuator are used externally, display level
can be matched by offset.
The setting range is from -50.0 to 50.0 dB (0.1 dB step).
Offset is calculated to the reference level, and it is displayed.
When offset is set, it is displayed on LEVEL display area
as „OFS“.
Furthermore, the value of a marker point is displayed reflecting
the calculated offset.
Offset of dB
µ
V, dBmV, dBV, W, etc. are changed
automatically.
REFER ATT AMP
(dBm) (dB) (dB)
10 25 0
9240
8230
7220
6210
5200
4190
3180
2170
1160
0150
-1 14 0
-2 13 0
-3 12 0
-4 11 0
-5 10 0
-6 9 0
-7 8 0
-8 7 0
-9 6 0
-10 5 0
-11 4 0
-12 3 0
-13 2 0
-14 1 0
-15 0 0
REFER ATT AMP
(dBm) (dB) (dB)
-16 20 21
-17 19 21
-18 18 21
-19 17 21
-20 16 21
-21 15 21
-22 14 21
-23 13 21
-24 12 21
-25 11 21
-26 10 21
-27 9 21
-28 8 21
-29 7 21
-30 6 21
-31 5 21
-32 4 21
-33 3 21
-34 2 21
-35 1 21
-36 5 26
-37 4 26
-38 3 26
-39 2 26
-40 1 26

17
Subject to change without notice
Display Scale - Resolution Bandwidth - Video Bandwidth
9.6 Setting the input impedance
1. Press
F6
and use to select the input impedance
compensation.
50ΩΩ
ΩΩ
Ω75ΩΩ
ΩΩ
Ω
When coaxial adaptor MA301 (50 Ω/75 Ωimpedance converter)
is attached, and choose „75 Ω“, then offset is calculated to the
reference level, and it changes for the measured value as 75 Ω
system, and display it.
When „75
Ω
“ is selected, „75
Ω
“ is displayed in the
LEVEL area on the screen. When „75
Ω
“ is selected, the
offset is set to 5.7 dB (insertion loss of MA301).
Moreover, can set offset. Moreover, while setting the unit
of the marker point to [W, V, V/m] etc, it changes from
dBm correctly.
When you set it as „75
Ω
“, please be sure to attach coaxial
adapter MA301 (50
Ω
/75
Ω
impedance converter).
10. Display Scale <SCALE>
10.1 Setting with the keys ([F1], [F2])
1. Press
F1
to set the 10 dB/div display scale.
2. Press
F2
to set the 2 dB/div display scale.
10.2 Setting with the encoder
1. Turn to switch between the 10 dB/div and 2 dB/div dis-
play scale.
10dB 2dB
In 2 dB/div, display level may not become smaller than fixed
level, by frequency compensation.
11. Resolution Bandwidth <RBW>
Any selected parts of MANU, AUTO and ALL become
inverted display.
11.1 MANUAL mode
1. Press
F1
or turn the to enter MANUAL mode.
Use to set the RBW.
3kHz 10kHz 30kHz 100kHz 300kHz
1MHz 3MHz
11.2 AUTO mode
1. When
F2
is pressed, optimum RBW is set according to
the settings of SPAN and SWEEP.
Since „
*
“ is displayed on the right end of RBW setting value
display portion of a screen when set as AUTO mode, it can
check being set as AUTO mode.
11.3 ALL AUTO mode
1. When
F3
is pressed, optimum RBW, VBW and SWEEP
are set according to the setting of SPAN.
Since „*“ will be displayed on the right end of each setting
value display portion of RBW, VBW, and SWEEP if ALL AUTO
mode is set up, it can check being set as ALL AUTO mode.
When RBW is set as 3 kHz or 10 kHz, selectivity (60 dBc)
becomes larger than an actual value, by influence of SSB
phase noise.
12. Video Bandwidth <VBW>
Any selected parts of MANU, AUTO and ALL become
inverted display.
12.1 MANUAL mode
1. Press or turn the to enter MANUAL mode.
Use to set the VBW.
12.2 AUTO mode
1. When
F2
is pressed,VBW is set according to the settings
of SPAN and SWEEP.
Since „*“ is displayed on the right end of VBW setting value
display portion of a screen when set as AUTO mode, it can
check being set as AUTO mode.
SCALE
SCALE
to switch over to the function screen shown below:
10 dB 2 dB
F1 F2 F3 F4 F5 F6
Press
RBW
MANU AUTO ALL
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press RBW
VBW
MANU AUTO ALL
VBW
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press
100Hz 300Hz 1kHz 3kHz 10kHz 30kHz
100kHz 300kHz 1MHz

18 Subject to change without notice
Sweep Axis - Auto Tuning - Hold/Run - Calculation Function
12.3 ALL AUTO mode
1. When
F3
is pressed, RBW, VBW and SWEEP are set
according to the setting of SPAN.
Since „*“ will be displayed on the right end of each setting
value display portion of RBW, VBW, and SWEEP if ALL AUTO
mode is set up, it can check being set as ALL AUTO mode.
13. Sweep Axis · Detection Mode <SWEEP>
Any selected parts of MANU, AUTO and ALL become
inverted display. When [F4] is pressed, the part of DET
become inverted display.
13.1 MANUAL mode
1. Press
F1
or turn the to enter MANUAL mode.
Use to set the SWEEP.
10ms 30ms 0.1s 0.3s 1s 3s 10s 30s
Can’t set 10ms at the FULLSPAN.
13.2 AUTO mode
1. When
F2
is pressed, SWEEP is set according to the
settings of SPAN and RBW.
Since „*“ is displayed on the right end of SWEEP setting value
display portion of a screen when set as AUTO mode, it can
check being set as AUTO mode.
13.3 ALL AUTO mode
1. When
F3
is pressed, RBW, VBW and SWEEP are set
according to the setting of SPAN.
Since „*“ will be displayed on the right end of each setting
value display portion of RBW, VBW, and SWEEP if ALL AUTO
mode is set up, it can check being set as ALL AUTO mode.
13.4 Setting the Detection mode
1. Pressing
F4
allows you to change the method to capture
the trace.
PosPK (Positive Peak): Traces the maximum value of the sample points.
SMPL (Sample): Traces the momentary value of the sample points.
NegPK (Negative Peak): Traces the minimum value of the sample points.
When sweep time is 10ms or 30ms, detection mode is set
to SMPL.
SWEEP
MANU AUTO ALL
SWEEP
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press
PosPK SMPL NegPK
14. AUTO Tuning <AUTO TUNE>
1. When
AUTOTUNE
is pressed, center frequency is set at the
spectrum of the maximum level in the 3.3 GHz band, and in
addition, optimum reference level, RBW, VBW and SWEEP are
set according to the setting of SPAN.
The AUTO tuning does not operate normally, at the time of
the following 4 conditions.
1) ZERO SPAN
2) FULL SPAN
3) Signal level is –40 dBm or lower
4) Signal frequency is 50 MHz or lower
15. Hold/Run <HOLD/RUN>
1. Press
HOLD/RUN
to switch to between sweep halt and
continuous sweep.
This operates only with the key press, with no function
indication.
16. Calculation Function <CALC>
After sweeps stops, press
HOLD/RUN
to restart sweep.
Press [F1] to [F5] to set the CALC mode.
Use
to set the number of sweeps.
16.1 NORM mode
1. Press
F1
Calculation is not performed in this mode.
The number of sweeps is always unlimited.
Usually, please choose this mode.
„NORMAL“ is displayed in the CALC area on the LCD
screen. (Refer to „4. Description of Display“ for details)
16.2 MAX HOLD mode
1. Press
F2
ans use to set the number of sweeps.
2. Sweeps are performed the set number of times, the maximum
value of each point of trace data is displayed as a trace, and
then sweep is halted.
„MAX --- (number of sweeps)“ is displayed in the CALC
area on the LCD screen. (Refer to „4. Description of Display“
for details)
NORM MAXHD
**
MINHD
**
AVER
256
OVRWR
CALC
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press
2 4 8 16 32 64 128
256 512 1024
**
(unlimited)

19
Subject to change without notice
16.3 MIN HOLD mode
1. Press
F3
and use to set the number of sweeps.
2. Sweeps are performed the set number of times, the minimum
value of each point of trace data is displayed as a trace, and
then sweep is halted.
„MIN — (number of sweeps)“ is displayed in the CALC area on
the LCD screen. (Refer to „4. Description of Display“ for details)
16.4 AVERAGE mode
1. Press
F4
Use to set the number of sweeps.
2. Sweeps are performed the set number of times, average value
of each point of trace data is displayed as a trace, and then
sweep is halted.
2 4 8 16 32 64 128 256
„AVG — (number of sweeps)“ is displayed in the CALC area on
the LCD screen. (Refer to „4. Description of Display“ for details)
16.5 OVER WRITE mode
1. Press
F5
to enter into the OVER WRITE mode, where traces
are written one over another. The number of sweeps is unlimited.
„OVER WR“ is displayed in the CALC area on the LCD
screen. (Refer to „4. Description of Display“ for details)
Only the last one trace is saved.
17. Marker · Peak Search <MKR>
17.1 Moving the marker
Use
F1
and to move the marker.
Use
F2
to put DELTA REF at the current marker position.
2 4 8 16 32 64 128
256 512 1024
**
(unlimited)
PEAK SEARCH
ZONE CNTR WIDTH
MARKER
NORM DELTA
PEAK SEARCH
NORM PEAK NEXT
MARKER
NORM DELTA
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press
F1 F2 F3 F4 F5 F6
MKR
The display when an NORM marker is selected.
The marker is manually moved at NORM mode. Peak search
function, NEXT peak function is available.
The display when an ZONE marker is selected.
The marker moves to the highest paek position automatically at
ZONE mode, inside specified zone.
17.2 Setting the peak search <PEAK SEARCH>
NORM mode (Use
F3
to select NORM.)
Use
F4
to move the marker to the maximum peak
position.
Use
F5
to move the marker successively from higher to
lower peak positions other than the maximum peak. (The
marker moves to 10 peaks.)
When you move the marker to the 10th peak or moving the
marker, the NEXT peak search function stops and the
function display disappears.
ZONE mode (Use
F3
to select ZONE.)
Use
F4
and to move the center position.
Use
F5
and to change the width.
17.3 Changing the unit of marker point
Press
F6
to change the unit of marker point.
When unit of reference level is dBm, the unit is changed from
[dBm] to [W].
When unit of reference level is dBµV, the unit is changed from
[dBµV, dBmV, dBV] to [V].
When unit of reference level is dBµV/m, the unit is changed
from [dBµV/m] to [V/m].
When unit of reference level is dBìA/m, the unit is changed
from [dBµA/m] to [A/m].
Moreover, according to each unit, it is displayed as follows.
[W] [W, mW, µW, nW, pW, fW]
[V] [V, mV, µV, nV]
[V/m] [V/m, mV/m, µV/m, nV/m]
[A/m] [A/m, mA/m, µA/m, nA/m]
18. Save/Load <SAVE/LOAD>
18.1 Setting the location to store the trace
1. Pressing
F1
allows you to set the number of location to
store the trace.
2. Use to set the number of location.
00 01 02 03 04 ... 98 99
The part ofTRACE become inverted display after it is selected.
PRE
SET
EXECUTE
SAVE LOAD DEL
TRACE
00 00
PARA M
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press SAVE/LOAD
Marker Peak Search - SAVE/LOAD

20 Subject to change without notice
18.2 Setting the location to store the parameter
1. Pressing
F2
allows you to set the number of location to
store the parameter.
2. Use to set the number of location.
00 01 02 03 04 ... 98 99
The part of PARAM become inverted display after it is
selected.
18.3 Saving the data
1. Press
F3
to save the data at the set number.
This saves the trace when TRACE is selected, or the setting
parameters when PARAM is selected.
„
*
“ is displayed on the right end of the number of location
place at which data is saved.
It can be overwritten, too.
18.4 Loading the data
1. Press
F4
to read out the data at the set number.
This reads out the trace whenTRACE is selected.The setting
parameter of the loaded trace is displayed in the loaded
trace information display area. (Refer to „4. Description of
Display“ for details)
This reads out the setting parameters when PARAM is
selected.
When you load a trace, the current trace disappears, the
HOLD state is set, and the loaded trace is displayed. For
the loaded trace, you can use the marker, but cannot use a
measuring function. When you press the HOLD/RUN key,
the loaded and the current traces are displayed overlapping
each other.
„*“ is displayed on the right end of the number of location
place at which data is saved.
When you search the trace or setting parameters to be read
out, repeat
F4
F4
..., and load the
trace or setting parameters in turn.
18.5 Clearing the loaded trace
1. Press
F5
to clear the loaded trace that has been
displayed.
18.6 Presetting (Initialization)
1. Press
F6
to preset the setting parameters as the
Initialization shown below:
„Initialization“
Items Parameter Items Parameter
Center frequency 1 GHz Sweep time 0.3 s
Frequency span 20 MHz Detection mode Positive peak mode
Reference level 10 dBm RBW 100 kHz
Offset 0.0 dB VBW 10 kHz
Impedance 50 ΩDisplay scale 10 dB/div
19. Measuring Function<MEAS>
Select the measuring function:
F1
Ch Power: Channel power measurement
F2
Adj Ch Pw: Adjacent channel leakage power
measurement
F3
Occ BW: Occupied frequency bandwidth measurement
F4
E/F ANT: Electric field strength measurement
F5
M/F PROBE: Magnetic field strength measurement
(optional)
F6
Invalid
Once you select the measuring function, pressing
MEAS
next time will directly bring up the function
screen for the function you selected the last time. If you
want to stop the measuring function, or if you want to
select another measuring function, press [F6] (MEAS
OFF). This stops the measuring function and switches to
the above screen, which allows you to select the
measuring function.
The measuring function is stops, when push
MKR
while these 4 functions (Channel power ,Adjacent channel
leakage power, Occupied frequency bandwidth,
Frequency counter) are selected. Because each 4
functions and marker operation cannot be used
simultaneously. Similarly, the function of the marker
stops, when the functions of these 4 measurements are
selected while using the marker.
The unit displays data in 251 horizontal dots, but it
internally captures the trace and calculates the measured
value (Channel power measurement, Adjacent channel
leakage power measurement and Occupied frequency
bandwidth measurement) in 1001 dots.
19.1 Channel power measurement
<Ch Power>
F1
Measures the sum of the power in the zone specified.Two modes,
TOTAL and BAND, are available.
TOTAL mode [Use
F1
(MODE) to select TOTAL.]
Measure the sum of the power in the zone specified by the
center frequency and frequency span (whole range of the
screen).
It is shown on MEAS area of LCD as „CP TOTAL“
The measured value is displayed at the right lower corner
on the screen.
Ch
Power
Adj
Ch Pw
Occ
BW
E/F
ANT
M/F
PROBE
Freq
COUNT
to switch over to the function screen shown below:
F1 F2 F3 F4 F5 F6
Press
MEAS
Measuring Function
Table of contents
Other Hameg Measuring Instrument manuals