Hioki IM7580A-1 Use and care manual

EN
Communication Instruction Manual
IMPEDANCE ANALYZER
Sept. 2016 Revised edition 2
IM7585A987-02 16-09H
IM7581
IM7580A
IM7585
IM7583 IM7583-01
IM7583-02
IM7585-01
IM7585-02
IM7580A-1
IM7580A-2
IM7581-01
IM7581-02
IM7587 IM7587-01
IM7587-02

i
Contents
Introduction.............................................................................................................................1
Safety Information..................................................................................................................1
1 Specications 3
1.1 RS-232CSpecications ................................................................................................3
Handshake (About Buffer Flow Control)............................................................................................... 3
1.2 GP-IBSpecications.....................................................................................................4
1.3 USBSpecications .......................................................................................................4
1.4 LANSpecications .......................................................................................................4
2 Connection and Setting 5
2.1 Overview of Communication........................................................................................5
Screen Displayed while Setting Interfaces ........................................................................................... 6
2.2 USBSettingsandConnection .....................................................................................7
Setting USB.......................................................................................................................................... 7
Connecting the USB Cable................................................................................................................... 8
2.3 LANSettingsandConnection .....................................................................................9
LAN Settings......................................................................................................................................... 9
Connecting a LAN Cable.................................................................................................................... 13
2.4 GP-IB Connection and Settings
(whentheZ3000isconnected) ..................................................................................14
Connecting the GP-IB Cable .............................................................................................................. 14
Setting GP-IB...................................................................................................................................... 14
2.5 RS-232CConnectionandSettings
(whentheZ3001isconnected) ..................................................................................16
Connecting the RS-232C Cable ......................................................................................................... 16
Setting RS-232C................................................................................................................................. 17
2.6 Remote Mode...............................................................................................................18
Canceling Remote Mode.................................................................................................................... 18
Appendix A1
Appx.1 CheckingtheUSBVirtualCOMPort A1
IM7585A987-02
10
9
8
7
6
5
4
3
2
1
付録索引

Contents
ii

1
Introduction
Introduction
This instruction manual provides details on the communication interfaces of the IM7580A series,IM7581
series,IM7583 series,IM7585 series and IM7587 series Impedance Analyzer.
In this document,the “instrument” means the IM7580A series,IM7581 series,IM7583 series ,IM7585
series and IM7587 series.
Safety Information
This manual contains information and warnings essential for safe operation of the instrument and for
maintaining it in safe operating condition. Before using it,be sure to carefully read the following safety
precautions.
Notation
In this manual, the risk seriousness and the hazard levels are classied as follows.
WARNING Indicates a potentially hazardous situation that may result in death or serious
injury to the operator.
CAUTION Indicates a potentially hazardous situation that may result in minor or moderate
injury to the operator or damage to the instrument or malfunction.
IMPORTANT Indicates information related to the operation of the instrument or maintenance
tasks with which the operators must be fully familiar.
Indicates prohibited actions.
Indicates the action which must be performed.
*Indicates the action which must be performed.
[ ] Menus,commands,dialogs,buttons in a dialog,and other names on the screen
and the keys are indicated in brackets.
CURSOR
( Bold character) Bold characters within the text indicate operating key labels.
Windows Unless otherwise specied,“Windows” represents Windows Vista,Windows 7,
Windows 8 or Windows 10.
Dialogue Dialogue box represents a Windows dialog box.
Mouse Operation
Click: Press and quickly release the left button of the mouse.
Right-click: Press and quickly release the right button of the mouse.
Double click: Quickly click the left button of the mouse twice.
Drag: While holding down the left button of the mouse,move the mouse and then release the
left button to deposit the chosen item in the desired position.
10
9
8
7
6
5
4
3
2
1
付録索引

2
Safety Information

3
1Specications
1.1 RS-232C Specications
Transmission Method Communication method: Full duplex
Synchronous method: Start-stop synchronization
Transmission Speed 9600 bps,19200 bps,38400 bps,57600 bps
Data Bits 8 bits
Parity None
Stop bit 1 bits
Message terminator (delimiter) CR+LF,CR
Flow control Software (XON/XOFF control)
“Handshake (About Buffer Flow Control)” (p. 3)
Electrical
Specications
Input voltage level 5 to 15 V ON
-15 to -5 V OFF
Output voltage level 5 to 9 V ON
-9 to -5 V OFF
Handshake (About Buffer Flow Control)
Control during Receiving
Input Buffer
Usage
Amount
85%
75%
25%
Buffer Space
Send XON
Send XOFF
XON/XOFF control:
•When the data in the receive buffer exceeds 75% of the
buffer,XOFF(13H) is sent and the controller is notied
that there is not much space remaining in the buffer.
•Processing of the data in the buffer continues,and then
XON(11H) is sent and the controller is notied that there
is sufcient remaining space in the buffer when the
amount of data becomes less than 25%.
Control during Sending
XON/XOFF control:
•When XOFF is received,the sending of data is halted. When XON is received,the sending of data is
resumed.
1 Specications
1

4
GP-IB Specications
1.2 GP-IB Specications
SH1 Supports all source handshake functions.
AH1 Supports all acceptor handshake functions.
T6 Supports standard talker functions.
Supports serial poll functions.
Talk only mode is not supported.
Supports the talker cancel function by MLA (My ListenAddress).
L4 Supports standard listener functions.
Listener only mode is not supported.
Supports the listener cancel function by MTA(My Talk Address).
SR1 Supports all service request functions.
RL1 Supports all remote/local functions.
PP0 Parallel poll functions are not supported.
DC1 Supports all device clear functions.
DT1 Supports all device trigger functions.
C0 Controller functions are not supported.
Code used: ASCII code
1.3 USB Specications
Connector Series B receptacle
Compliance tandard USB2.0 (Full Speed/High Speed)
No. of ports 1
Class Communication class
Supported OS Windows Vista,Windows 7,Windows 8,Windows 10
1.4 LAN Specications
Connector RJ-45 connector × 1
Compliance
standard
IEEE 802.3-compliant Ethernet
Transfer system 10BASE-T/ 100BASE-TX/ 1000BASE-TAuto detected
Protocol TCP/IP
Function Command control

5
2Connection and Setting
2.1 Overview of Communication
You can control the instrument with communication commands from a computer via the USB,GP-IB,RS-
232C and LAN interfaces.
There are the following four communication methods. To enable communication,the communication
conditions need to be set on the instrument.
USB communication (p. 7)
The instrument is communication class compatible.
LAN communication (p. 9)
Command control using the TCP/IP protocol is possible.
GP-IB communication (when the Z3000 is connected) (p. 14)
•Commands common to IEEE-488-2 1987 (requirement) can be used.
•The instrument has been designed with reference to the following standard.
(Reference standard: IEEE-488.2 1987)
RS-232C communication (when the Z3001 is connected) (p. 16)
WARNING
•Always turn both devices OFF when connecting and disconnecting an interface
connector. Otherwise,an electric shock accident may occur.
•To avoid damage to the instrument,do not short-circuit the terminal and do not
input voltage to the terminal.
•Failure to fasten the connectors properly may result is sub-specication
performance or damage to the equipment.
2 Connection and Setting
2

6
Overview of Communication
Screen Displayed while Setting Interfaces
When you set an interface,the icon for the set interface is displayed on the right side of the screen.
When USB is set
When GP-IB is set
When RS-232C is set
When LAN is set

7
USB Settings and Connection
2.2 USB Settings and Connection
To connect the instrument to a computer the rst time,a dedicated USB driver must be installed.
Before connecting the instrument to the computer,install the USB driver.
The USB driver can be downloaded from the bundled CD,or our web site. (http://www.hioki.com)
The USB driver is compatible with the Windows Vista (32-bit,64-bit version),Windows 7 (32-bit,64-bit
version),Windows 8 (32-bit,64-bit version) and Windows 10 (32-bit,64-bit version) operating systems.
Additionally,do not put the computer into the sleep state while the instrument is connected to the
computer.
Setting USB
Procedure
You can congure the setting from any of mode and mode.
1
LCR Measurement Screen Interface Settings
2
USB Setting
Press .
2

8
USB Settings and Connection
3
USB Setting
Select the terminator setting.
CR+LF
CR
4
Press to conrm the setting.
Connecting the USB Cable
Connect a USB cable (commercially available USB cable) to the USB port of the instrument.
USB cable (commercially
available product)
USB interface port of
computer
Type B
CAUTION
•To avoid faults,do not disconnect or reconnect the USB cable during instrument
operation.
•Connect the instrument and the computer to a common earth ground. Using different
grounds could result in potential difference between the instrument and the computer.
Potential difference on the USB cable can result in malfunctions and faults.
•Do not disconnect or connect other USB devices from or to,respectively,PC during
USB communications between instrument and computer. The USB communications
between instrument and PC may stop.

9
LAN Settings and Connection
2.3 LAN Settings and Connection
LAN Settings
You can perform command control using the TCP/IP protocol.
Set the instrument to match your network environment in advance.
•Make these settings before connecting to a network. Changing settings while connected can duplicate
IP addresses of other network devices,and incorrect address information may otherwise be presented
to the network.
•The instrument does not support DHCP (automatic IP address assignment) on a network.
Setting Items
IP address
Identies each device connected on a network.
Each network device must be set to a unique address.
The instrument supports IP version 4,with IP addresses indicated as four decimal octets,
e.g.,“192.168.0.1”.
Subnet mask This setting is for separating the IP address into the network address that indicates the
network and the host address that indicates the instrument. On this instrument,the subnet
mask is represented as four decimal numbers separated by “. “ such as “255.255.255.0.”
Default Gateway
When the computer and instrument are on different but overlapping networks (subnets),this
IP address species the device to serve as the gateway between the networks.
If the computer and instrument are connected one-to-one,no gateway is used,and the
instrument’s default setting “0.0.0.0” can be kept as is.
Network Environment Conguration
Example 1. Connecting the instrument to an existing network
Example 1. Connecting the instrument to an existing network
When connecting the instrument to an existing network, the network settings need to be conrmed
in advance.
An IP address which is not the same as that of another network device needs to be assigned.
Conrm the following items with the network administrator,and write them down.
IPAddress _________._________._________._________
Subnet Mask _________._________._________._________
Default Gateway _________._________._________._________
Example 2. Connecting multiple instruments to a single computer using a hub
When building a local network with no outside connection,the following private IP addresses are
recommended.
Example of private IP address:
IPAddress.....................Computer: 192.168.0.100
Instrument: 192.168.0.1,192.168.0.2,192.168.0.3...
(Set an IP address that differs from that of other network devices.)
Subnet Mask..................255.255.255.0
Default Gateway.............OFF(0.0.0.0)...
Example 3. Connecting one instrument to a single computer using the 9642 LAN Cable
The 9642 LAN Cable can be used with its supplied connection adapter to connect one instrument
to one computer,in which case the IP address is freely settable. Use the recommended private IP
addresses.
IPAddress.....................Computer: 192.168.0.100
Instrument: 192.168.0.1 (Set to a different IP address than the computer.)
Subnet Mask..................255.255.255.0
Default Gateway....OFF(0.0.0.0)
2

10
LAN Settings and Connection
Procedure
You can congure the setting from any of mode and mode.
1
LCR Measurement Screen Interface Settings
2
LAN Settings
Press .
3
LAN Settings
Select the IP address.
4
IP address Settings
or to set the IP address.
Press to conrm the setting.

11
LAN Settings and Connection
5
LAN Settings
Select the subnet mask.
6
Subnet mask Settings
Use or to set the subnet mask,
and press to conrm the setting.
Any of the following 30 subnet masks can be set for the instrument.
128.000.000.000 255.128.000.000 255.255.128.000 255.255.255.128
192.000.000.000 255.192.000.000 255.255.192.000 255.255.255.192
224.000.000.000 255.224.000.000 255.255.224.000 255.255.255.224
240.000.000.000 255.240.000.000 255.255.240.000 255.255.255.240
248.000.000.000 255.248.000.000 255.255.248.000 255.255.255.248
252.000.000.000 255.252.000.000 255.255.252.000 255.255.255.252
254.000.000.000 255.254.000.000 255.255.254.000
255.000.000.000 255.255.000.000 255.255.255.000
(Initial setting)
7
LAN Settings
Select the default gateway.
If the default gateway does not need to
be set,for example,when connecting the
instrument and computer on a one-to-one
basis using a cross cable,leave this set
to OFF.
2

12
LAN Settings and Connection
8
Default gateway Settings
or to set the default gateway.
Press to conrm the setting.
9
LAN Settings
Select the port number.
10
Port number Settings
or to set the port number to
use for communication commands.
Settable range : 1024 to 65535
Press to conrm the setting.
11
LAN Settings
Select the terminator setting.
CR+LF
CR
12
Press to conrm the setting.

13
LAN Settings and Connection
Connecting a LAN Cable
Use a LAN cable to connect the instrument and computer.
Required items:
When connecting the instrument to an existing network (prepare any of the following):
•Straight-through Cat 5,1000BASE-T-compliant Ethernet cable (up to 100 m,commercially available).
For 100BASE/10BASE communication,a 100BASE-TX/10BASE-T-compliant cable may also be used.
•Hioki 9642 LAN Cable (option)
(A cross adapter cannot be used.)
When connecting one instrument to a single computer (prepare one of the following):
•1000BASE-T-compliant cross-over cable (up to 100 m)
•1000BASE-T-compliant straight-through cable with cross-over adapter (up to 100 m)
•Hioki 9642 LAN Cable (option)
SPEED LED
On (Green) :1000BASE
On (Orange) :100BASE
Off:10BASE
LINK UP LED
On: Link up
Off: Link down
Blinking: Sending or
receiving data
The MAC address of the LAN is displayed below the serial number.
You can also check it on the instrument screen.
See: “Checking the Version of the Instrument” in the instruction manual.
When connecting the instrument to a single computer (connect the instrument to the computer)
Connecting with the 9642 LAN Cable and crossover adapter (supplied with the 9642)
Connect the LAN Cable to the
crossover adapter.
LAN interface
Crossover adapter
Connect the LAN Cable to the
100BASE-TX connector on the PC.
Connect the crossover adapter
to the LAN interface on the
instrument.
1
2
3
2

14
GP-IB Connection and Settings (when the Z3000 is connected)
2.4 GP-IB Connection and Settings
(when the Z3000 is connected)
Connecting the GP-IB Cable
Connect the GP-IB cable to the GP-IB connector.
Recommended cable:
9150-02 GP-IB connection cable (2 m)
Setting GP-IB
Procedure
You can congure the setting from any of mode and mode.
1
LCR Measurement Screen Interface Settings
2
GPIB Setting
Press .

15
GP-IB Connection and Settings (when the Z3000 is connected)
3
GPIB Setting
Use or to set the GP-IB address.
Select the terminator setting.
LF with EOI
LF with CR+EOI
4
Press to conrm the setting.
2

16
RS-232C Connection and Settings (when the Z3001 is connected)
2.5 RS-232C Connection and Settings
(when the Z3001 is connected)
Connecting the RS-232C Cable
Connect the RS-232C cable to the RS-232C connector.
(Recommended cable: 9637 RS-232C cable)
Male 9-pin D-sub
#4-40 attaching
screws
To connect the instrument to a controller (DTE),use a crossover cable compatible
with the connectors on both the instrument and the controller.
The I/O connector is a DTE (Data Terminal Equipment) conguration.
1 2 3 4 5
6 7 8 9
Connector (D-sub)
Pin No. Interchange Circuit
Nam
e
CCITT
Circuit No. EIA
Abbreviation JIS
Abbreviation Common
Abbreviation
1Unused
2Received Data 104 BB RD RxD
3Transmitted Data 103 BA SD TxD
4
Data Terminal Ready
108/2 CD ER DTR
5Signal Ground 102 AB SG GND
6Unused
7Unused
8Unused
9Unused
Example: Connecting to a DOS/V PC
Specication: D-sub 9-pin female and female connector,reverse connection

17
RS-232C Connection and Settings (when the Z3001 is connected)
Setting RS-232C
Procedure
You can congure the setting from any of mode and mode.
1
LCR Measurement Screen Interface Settings
2
RS-232C Settings
Press .
3
RS-232C Settings
Select the terminator setting.
CR+LF
CR
Select the baud rate setting.
Select the handshake setting.
No ow control
Software
(XON/XOFF control)
4
Press to conrm the setting.
2
This manual suits for next models
9
Table of contents
Other Hioki Measuring Instrument manuals

Hioki
Hioki MR8875 User manual

Hioki
Hioki 9695-02 User manual

Hioki
Hioki MR8847A User manual

Hioki
Hioki IM3533 User manual

Hioki
Hioki MR8740T Instruction manual

Hioki
Hioki 3664 User manual

Hioki
Hioki CM4375 User manual

Hioki
Hioki LR8450 User manual

Hioki
Hioki PW3336-01 User manual

Hioki
Hioki 3532-50 User manual

Hioki
Hioki SM-8213 User manual

Hioki
Hioki PW3360-21 User manual

Hioki
Hioki 9196 User manual

Hioki
Hioki PW9100-03 User manual

Hioki
Hioki 8730-10 User manual

Hioki
Hioki IM7585 Series Use and care manual

Hioki
Hioki PW3198 User manual

Hioki
Hioki CT6500 User manual

Hioki
Hioki LR5041 User manual

Hioki
Hioki PW6001 User manual