Hioki P9000-01 User manual

DIFFERENTIAL PROBE
Instruction Manual
P9000-01
P9000-02
EN
Dec. 2019 Revised edition 1
P9000A981-01 19-12H
Warranty
Warranty malfunctions occurring under conditions of normal use in
conformity with the Instruction Manual and Product Precautionary
Markings will be repaired free of charge. This warranty is valid for a
period of three (3) years from the date of purchase. Please contact
the distributor from which you purchased the product for further infor-
mation on warranty provisions.
Introduction
Thank you for choosing the Hioki P9000-01,P9000-02 Differential
Probe. Preserve this manual carefully and keep it handy to make full
use of this product for along time.
Overview
The product is a differential probe that is connected to a Memory
HiCorder input unit. The probe divides two high voltages input to its
high and low terminals by 100 or 1000 and then outputs the differ-
ence as a signal relative to ground. The P9000-02 includes AC RMS
value output functionality.
Power can be supplied via the optional AC adapter or a commer-
cially available USB cable.
Verifying Package Contents
When you open the package,carefully inspect the product to ensure
that everything is in good condition,and that no damage occurred
during shipping. Carefully check the accessories and connectors.
If the product seems to have been damaged or does not work as
specied,contact your authorized Hioki distributor or reseller.
Cleaning
If the product becomes dirty,slightly moisten a soft cloth moistened
with water or a neutral detergent and wipe the product clean.
IMPORTANT
Never use solvents such as benzene,alcohol,acetone,ether,
ketones,thinners or gasoline,as they can deform and discolor the
case.
Repair
If the product seems to be malfunctioning, conrm that the AC adapt-
er and the USB cable are connected properly before contacting your
authorized Hioki distributor or reseller.
WARNING
Touching any of the high-voltage points inside the
product is very dangerous. Do not attempt to mod-
ify,disassemble,or try to repair the product. Doing
so may cause a re,electric shock,or injury.
Calibration
IMPORTANT
Periodic calibration is necessary in order to ensure that the in-
strument provides correct measurement results of the specied
accuracy.
The calibration period varies with the conditions and environment of
use. It is recommended to determine a calibration period based on
those factors and to have the product regularly calibrated by Hioki.
Please contact your Hioki distributor to have your product periodi-
cally calibrated.
Precautions during shipment
Be sure to follow these precautions when transporting the product:
•Remove accessories and optional equipment from the product
in order to avoid damage. Additionally,be sure to double box the
product. Accidental damage suffered in transit is not covered by the
warranty.
•Attach a description of the issue when sending out your product for
repair.
Disposal
Dispose of the product in accordance with local regulations.
Safety Notes
This product is designed to conform to IEC 61010 Safety Standards,
and has been thoroughly tested for safety prior to shipment. How-
ever,using the product in a way not described in this manual may
negate the provided safety features.
Carefully read the following safety notes before using the product.
WARNING
Mishandling product could result in bodily injury or
even death,as well as damage to the product. Fa-
miliarize yourself with the instructions and precau-
tions in this manual before using the product.
WARNING
•Protective Gear
Performing measurement using this product in-
volves live-line work. To prevent an electric shock,
use appropriate protective insulation and adhere
to applicable laws and regulations.
•Electricity can potentially cause serious events
such as an electric shock,heat generation, re,
and an arc ash due to a short-circuit. If you have
not used electrical measuring instrument before,
you should be supervised by a technician who
has experience in electrical measurement.
Notations
In this document, the severity levels of risk and hazard are classied
as follows.
DANGER
Indicates an imminently hazardous situation
that will result in death of or serious injury to the
operator.
WARNING
Indicates a potentially hazardous situation that
may result in death of or serious injury to the
operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunc-
tion.
IMPORTANT
Indicates information or content that is particu-
larly important from the standpoint of operating
or maintaining the product.
Indicates cautions and hazards. When the sym-
bol is printed on
the product,refer to a corresponding topic in
the Instruction Manual.
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates an action that must not be performed.
Indicates action that must be performed.
Indicates that the product conforms to regula-
tions required by the EU Directive.
Indicates the Waste Electrical and Electronic
Equipment Directive (WEEE Directive) in EU
member states.
Measurement categories
To ensure safe operation of measuring instruments,IEC 61010
establishes safety standards for various electrical environments,cat-
egorized as CAT II to CAT IV,and called measurement categories.
WARNING
•Using a measuring product in an environment
designated with a higher-numbered category than
that for which the product is rated could result in
a severe accident,and must be carefully avoided.
•Never use a measuring product that lacks cat-
egory labeling in a CAT II to CAT IV measurement
environment. Doing so could result in a serious
accident.
This product conforms to the safety requirements for CAT III 1000 V,
CAT IV 600 V measuring instruments.
CAT II: When directly measuring the electrical outlet recep-
tacles of the primary electrical circuits in equipment
connected to an AC electrical outlet by a power cord
(portable tools,household appliances,etc.)
CAT III: When measuring the primary electrical circuits of heavy
equipment (xed installations) connected directly to
the distribution panel,and feeders from the distribution
panel to outlets
CAT IV: When measuring the circuit from the service drop to the
service entrance,and to the power meter and primary
overcurrent protection device (distribution panel)
T Outlet
CAT II
Internal wiring
Distribution panel
Service entrance
Service drop
CAT IV
Power meter
CAT III
Fixed installation
Usage Notes
Observe the following precautionary information to ensure that the
product can be used safely and in a manner that allows it to perform
as described in its specications.
WARNING
To prevent an electric shock, conrm that the white
or red portion (insulation layer) inside the cable is
not exposed. If a color inside the cable is exposed,
do not use the cable.
WARNING
Do not allow the product to get wet,and do not take
measurements with wet hands. This may cause an
electric shock.
CAUTION
•To avoid damage to the product,avoid subjecting it to
vibration or mechanical shock during transportation
and handling. Exercise particular care to avoid sub-
jecting the product to mechanical shock,for example
by dropping.
•Do not store or use the product where it could be ex-
posed to direct sunlight,high temperatures or humid-
ity,or condensation. Under such conditions,the prod-
uct may be damaged and insulation may deteriorate
so that it no longer meets specications.
•This product is not designed to be entirely waterproof
or dustproof. Do not use it in an especially dusty en-
vironment,nor where it might be splashed with liquid.
This may cause damage.
This product complies with EN 61326 Class A.
This product may cause interference if used in residential areas.
Such use must be avoided unless the user takes special measures
to reduce electromagnetic emissions to prevent interference to the
reception of radio and television broadcasts.

Parts Names
Step
6
BNC connector
Power jack
Connect a model Z1008
AC adapter.
Black lead
(LOW)
Alligator clip
Maximum rated voltage
to earth:
1000 V (CAT lll)
600 V (CAT IV)
POWER switch
Selects the voltage division
ratio (power on).
Turns off the product.
POWER LED
Lights up when
the product is
receiving power.
OVER LED
Lights up when
excessively high
input is received.
USB-microB jack
Connect a USB cable
(commercially available).
Connected instrument (pictured: Model MR8875)
To measurement target
Red lead
(HIGH)
Output selector switch
(Model P9000-02 only)
Step
4
Step
7
Step
5
Step
2
Step
1
Step
3
For more information about
devices that can be connect-
ed,contact your authorized
Hioki distributor or reseller.
Range Scaling setting
(pictured: P9000-02)
Inspection Before Use
To ensure that the product is properly operating,conduct an inspec-
tion and check product operation before using product to ensure that
no damage has occurred during storage or transport. If there is any
damage to the product,contact your authorized Hioki distributor or
reseller for repair.
WARNING
Conrm that the insulation on the using the probes
are undamaged and that no bare conductors are im-
properly exposed before using the product. If there
is any damage to the product that leads to an elec-
tric shock,contact your authorized Hioki distributor
or reseller for repair.
Measurement Method
WARNING
•For your safe operation,do not connect the prod-
uct to the primary of the distribution panel.
•Do not connect the product to a voltage or current
that exceeds the rating indicated on the product
labeling or the measurement range indicated in its
specications. Doing so may damage the product
or cause it to become hot,resulting in electric
shock or bodily injury.
•To avoid short circuits and electric shock,do not
short-circuit two wires to be measured by bring-
ing the metal part of the test lead into contact with
them. Never touch the metal end.
WARNING
When supplying power through the USB-micro
B jack,make sure that the USB ground pin of the
power source is connected to the ground. Using an
ungrounded USB terminal could cause an electric
shock.
CAUTION
To prevent damage to the connector,be sure to release
the locking mechanism,grip the head of the connector
(not the cord),and pull it out.
1
Select which of the following two sources you will
use to power the product:
Method 1 Connect a model Z1008 AC adapter
To commercial
power supply
Model Z1008
Method 2 Connect a USB cable (commercially available,
USB (type A male)↔USB (type B female))
To USB terminal on
connected instrument
2
Connect the product’s BNC connector to the input
terminal on the connected instrument.
(2) Lock
(1) Line up
Line up
3
Connect the alligator clips to the input leads.
Purchase the tip option that is best suited to the
measurement target. These options can be pur-
chased via authorized Hioki distributor or reseller.
CAUTION
Precautions when using the L4932 Test Pin Set
To prevent a short circuit accident,be sure to use the
test leads with the sleeves attached when performing
measurements in the CAT III measurement category.
(For the measurement categories,see “Measurement
categories”.)
If the sleeves are inadvertently removed during mea-
surement,stop the measurement.
1000V
600V
1000V
600V
IMPORTANT
Use only the specied tip options. Using a non-specied tip option
may result in incorrect measurements due to poor connection or
other reasons.
4
Select the voltage division ratio with the POWER
switch.
Voltage
division ratio Description
1/1000 Divides the high signal and low signal by 1,000
and then outputs the difference between the
two signals as a signal relative to ground.
1/100 Divides the high signal and low signal by 100
and then outputs the difference between the
two signals as a signal relative to ground.
5
Select the output type (P9000-02 only).
Output Description
Wave Outputs the signal using the selected voltage
division ratio. Choose this setting when you
wish to view the power supply waveform.
AC-RMS Outputs the RMS value using the selected
voltage division ratio. Input is AC coupled.
6
Set the input range scaling for the connected in-
strument.
7
Connect the alligator clips to the measurement target.
8
After measurement is complete,turn off the POW-
ER switch.
Specifications
Product warranty
period
3 years (except for breaks in cables caused by use)
Accuracy guaran-
tee period
1 year
DC output accu-
racy
±0.5% f.s.
(f.s.=1.0 V (ratio 1000:1),3.5 V (ratio 100:1))
Temperature
characteristic
±0.05% f.s./C
Frequency
characteristic
DC to 100 kHz,-3 dB
Voltage division
ratio
1000:1,100:1 (user-switchable)
Common-mode
rejection ratio
80 dB or more (50/60 Hz shorted input)
Input resistance 10.5 MΩ± 1% (between HIGH and LOW terminals)
Input capacitance 5 pF or less (at 100 kHz)
Input jacks Banana plug
Output connector Insulated BNC
Measurable range 1000 V AC/DC (ratio 1000:1)
250 V AC,350 V DC (ratio 100:1)
Maximum input
voltage
1000 V AC/DC
Maximum rated
voltage to earth
1000 V AC/DC,measurement category III
anticipated transient overvoltage: 8000 V
Operating temper-
ature and humidity
Temperature: -40C to 80C/ -40F to 176F
Humidity: -40C to 40C [non-inclusive]
/ -40F to 104F,80% RH or less
40C to 45C [non-inclusive]
/ 104F to 113F,60% RH or less
45C to 80C/ 113F to 176F,
50% RH or less (no condensation)
Storage tempera-
ture and humidity
Temperature: -40C to 80C/ -40F to 176F
Humidity: 80% RH or less (no condensation)
Operating
environment
Indoors,Pollution Degree II,altitude up to 2000 m (6562
ft.)
Dimensions Approx. 128.0W×36.0H×22.0D mm
/ Approx. 5.04"W×1.42"H×0.87"D
Mass Approx. 170 g/ 6.0 oz.
Power supply (1) Model Z1008 AC adapter
(12 V DC,DC jack φ5.5×2.1 mm)
Rated supply voltage: 100 V AC to 240 V AC
(Voltage uctuations of ±10% from the rated supply
voltage are taken into account.)
Rated supply frequency: 50 Hz/60 Hz
Anticipated transient overvoltage: 2500 V
(2) USB bus power (5 V DC USB-microB connector)
(3) External power supply,2.7 V DC to 15 V DC
Maximum rated
power
(1) When using a model Z1008 AC Adapter:
6 VA (model P9000-01/-02 with a AC adapter)
0.9 VA (model P9000-01/-02 only)
(2) When using a USB bus power: 0.8 VA
(3)
When using a 15 V DC external power supply: 1 VA
Effect of radiated radio-frequency electromagnetic eld
±5% f.s. max. at 3 V/m
Effect of conducted radio-frequency electromagnetic eld
±5% f.s. max. at 3 V
Applicable
Standards
EMC EN61326 Class A
Safety EN61010
Accessories Instruction manual ×1
Alligator clips×1 set (red and black/ one each)
Carrying case×1
Options
The following options are available for the product. To purchase an option,
please contact your authorized Hioki distributor or reseller.
Options are subject to change. Please check Hioki’s website for the latest
information.
Z1008 AC Adapter
9243 Grabber Clip (CAT III 1000 V)
L4936 Bus Bar Clip Set (CAT III 600 V)
L4937 Magnetic Adapter Set (CAT III 1000 V)
L4931 Extension Cable Set (For extending the length of the test leads,1.5 m)
L4932 Test Pin Set
(CAT IV 600 V,CAT III 1000 V)
L4934 Small Alligator Clip Set (CAT III 300 V,CAT II 600 V)
(L4932 is required when using L4934.)
Model P9000-02 only
Measurement modes Waveform,AC-RMS (selected with a slide switch)
Convertible voltage 1000 V AC (ratio 1000:1),250 V AC (ratio 100:1)
RMS value measure-
ment accuracy
±1% f.s. (30 Hz to 1 kHz [non-inclusive],sine wave)
±3% f.s. (1 kHz to 10 kHz,sine wave)
(f.s.=1.0 V (ratio 1000:1),2.5 V (ratio 100:1))
RMS value measure-
ment response time
300 ms
Rising 0% f.s.→100% f.s.
Response time for input from 0% f.s. to 90% f.s.
600 ms
Falling 100% f.s.→0% f.s.
Response time for input from 100% f.s. to 10% f.s.
f.s. : maximum display value or scale length (This is usually the max-
imum value of the currently selected range.)

DIFFERENTIAL PROBE
Instruction Manual
P9000-01
P9000-02
EN
Dec. 2019 Revised edition 1
P9000A981-01 19-12H
Warranty
Warranty malfunctions occurring under conditions of normal use in
conformity with the Instruction Manual and Product Precautionary
Markings will be repaired free of charge. This warranty is valid for a
period of three (3) years from the date of purchase. Please contact
the distributor from which you purchased the product for further infor-
mation on warranty provisions.
Introduction
Thank you for choosing the Hioki P9000-01,P9000-02 Differential
Probe. Preserve this manual carefully and keep it handy to make full
use of this product for along time.
Overview
The product is a differential probe that is connected to a Memory
HiCorder input unit. The probe divides two high voltages input to its
high and low terminals by 100 or 1000 and then outputs the differ-
ence as a signal relative to ground. The P9000-02 includes AC RMS
value output functionality.
Power can be supplied via the optional AC adapter or a commer-
cially available USB cable.
Verifying Package Contents
When you open the package,carefully inspect the product to ensure
that everything is in good condition,and that no damage occurred
during shipping. Carefully check the accessories and connectors.
If the product seems to have been damaged or does not work as
specied,contact your authorized Hioki distributor or reseller.
Cleaning
If the product becomes dirty,slightly moisten a soft cloth moistened
with water or a neutral detergent and wipe the product clean.
IMPORTANT
Never use solvents such as benzene,alcohol,acetone,ether,
ketones,thinners or gasoline,as they can deform and discolor the
case.
Repair
If the product seems to be malfunctioning, conrm that the AC adapt-
er and the USB cable are connected properly before contacting your
authorized Hioki distributor or reseller.
WARNING
Touching any of the high-voltage points inside the
product is very dangerous. Do not attempt to mod-
ify,disassemble,or try to repair the product. Doing
so may cause a re,electric shock,or injury.
Calibration
IMPORTANT
Periodic calibration is necessary in order to ensure that the in-
strument provides correct measurement results of the specied
accuracy.
The calibration period varies with the conditions and environment of
use. It is recommended to determine a calibration period based on
those factors and to have the product regularly calibrated by Hioki.
Please contact your Hioki distributor to have your product periodi-
cally calibrated.
Precautions during shipment
Be sure to follow these precautions when transporting the product:
•Remove accessories and optional equipment from the product
in order to avoid damage. Additionally,be sure to double box the
product. Accidental damage suffered in transit is not covered by the
warranty.
•Attach a description of the issue when sending out your product for
repair.
Disposal
Dispose of the product in accordance with local regulations.
Safety Notes
This product is designed to conform to IEC 61010 Safety Standards,
and has been thoroughly tested for safety prior to shipment. How-
ever,using the product in a way not described in this manual may
negate the provided safety features.
Carefully read the following safety notes before using the product.
WARNING
Mishandling product could result in bodily injury or
even death,as well as damage to the product. Fa-
miliarize yourself with the instructions and precau-
tions in this manual before using the product.
WARNING
•Protective Gear
Performing measurement using this product in-
volves live-line work. To prevent an electric shock,
use appropriate protective insulation and adhere
to applicable laws and regulations.
•Electricity can potentially cause serious events
such as an electric shock,heat generation, re,
and an arc ash due to a short-circuit. If you have
not used electrical measuring instrument before,
you should be supervised by a technician who
has experience in electrical measurement.
Notations
In this document, the severity levels of risk and hazard are classied
as follows.
DANGER
Indicates an imminently hazardous situation
that will result in death of or serious injury to the
operator.
WARNING
Indicates a potentially hazardous situation that
may result in death of or serious injury to the
operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunc-
tion.
IMPORTANT
Indicates information or content that is particu-
larly important from the standpoint of operating
or maintaining the product.
Indicates cautions and hazards. When the sym-
bol is printed on
the product,refer to a corresponding topic in
the Instruction Manual.
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates an action that must not be performed.
Indicates action that must be performed.
Indicates that the product conforms to regula-
tions required by the EU Directive.
Indicates the Waste Electrical and Electronic
Equipment Directive (WEEE Directive) in EU
member states.
Measurement categories
To ensure safe operation of measuring instruments,IEC 61010
establishes safety standards for various electrical environments,cat-
egorized as CAT II to CAT IV,and called measurement categories.
WARNING
•Using a measuring product in an environment
designated with a higher-numbered category than
that for which the product is rated could result in
a severe accident,and must be carefully avoided.
•Never use a measuring product that lacks cat-
egory labeling in a CAT II to CAT IV measurement
environment. Doing so could result in a serious
accident.
This product conforms to the safety requirements for CAT III 1000 V,
CAT IV 600 V measuring instruments.
CAT II: When directly measuring the electrical outlet recep-
tacles of the primary electrical circuits in equipment
connected to an AC electrical outlet by a power cord
(portable tools,household appliances,etc.)
CAT III: When measuring the primary electrical circuits of heavy
equipment (xed installations) connected directly to
the distribution panel,and feeders from the distribution
panel to outlets
CAT IV: When measuring the circuit from the service drop to the
service entrance,and to the power meter and primary
overcurrent protection device (distribution panel)
T Outlet
CAT II
Internal wiring
Distribution panel
Service entrance
Service drop
CAT IV
Power meter
CAT III
Fixed installation
Usage Notes
Observe the following precautionary information to ensure that the
product can be used safely and in a manner that allows it to perform
as described in its specications.
WARNING
To prevent an electric shock, conrm that the white
or red portion (insulation layer) inside the cable is
not exposed. If a color inside the cable is exposed,
do not use the cable.
WARNING
Do not allow the product to get wet,and do not take
measurements with wet hands. This may cause an
electric shock.
CAUTION
•To avoid damage to the product,avoid subjecting it to
vibration or mechanical shock during transportation
and handling. Exercise particular care to avoid sub-
jecting the product to mechanical shock,for example
by dropping.
•Do not store or use the product where it could be ex-
posed to direct sunlight,high temperatures or humid-
ity,or condensation. Under such conditions,the prod-
uct may be damaged and insulation may deteriorate
so that it no longer meets specications.
•This product is not designed to be entirely waterproof
or dustproof. Do not use it in an especially dusty en-
vironment,nor where it might be splashed with liquid.
This may cause damage.
This product complies with EN 61326 Class A.
This product may cause interference if used in residential areas.
Such use must be avoided unless the user takes special measures
to reduce electromagnetic emissions to prevent interference to the
reception of radio and television broadcasts.

Parts Names
Step
6
BNC connector
Power jack
Connect a model Z1008
AC adapter.
Black lead
(LOW)
Alligator clip
Maximum rated voltage
to earth:
1000 V (CAT lll)
600 V (CAT IV)
POWER switch
Selects the voltage division
ratio (power on).
Turns off the product.
POWER LED
Lights up when
the product is
receiving power.
OVER LED
Lights up when
excessively high
input is received.
USB-microB jack
Connect a USB cable
(commercially available).
Connected instrument (pictured: Model MR8875)
To measurement target
Red lead
(HIGH)
Output selector switch
(Model P9000-02 only)
Step
4
Step
7
Step
5
Step
2
Step
1
Step
3
For more information about
devices that can be connect-
ed,contact your authorized
Hioki distributor or reseller.
Range Scaling setting
(pictured: P9000-02)
Inspection Before Use
To ensure that the product is properly operating,conduct an inspec-
tion and check product operation before using product to ensure that
no damage has occurred during storage or transport. If there is any
damage to the product,contact your authorized Hioki distributor or
reseller for repair.
WARNING
Conrm that the insulation on the using the probes
are undamaged and that no bare conductors are im-
properly exposed before using the product. If there
is any damage to the product that leads to an elec-
tric shock,contact your authorized Hioki distributor
or reseller for repair.
Measurement Method
WARNING
•For your safe operation,do not connect the prod-
uct to the primary of the distribution panel.
•Do not connect the product to a voltage or current
that exceeds the rating indicated on the product
labeling or the measurement range indicated in its
specications. Doing so may damage the product
or cause it to become hot,resulting in electric
shock or bodily injury.
•To avoid short circuits and electric shock,do not
short-circuit two wires to be measured by bring-
ing the metal part of the test lead into contact with
them. Never touch the metal end.
WARNING
When supplying power through the USB-micro
B jack,make sure that the USB ground pin of the
power source is connected to the ground. Using an
ungrounded USB terminal could cause an electric
shock.
CAUTION
To prevent damage to the connector,be sure to release
the locking mechanism,grip the head of the connector
(not the cord),and pull it out.
1
Select which of the following two sources you will
use to power the product:
Method 1 Connect a model Z1008 AC adapter
To commercial
power supply
Model Z1008
Method 2 Connect a USB cable (commercially available,
USB (type A male)↔USB (type B female))
To USB terminal on
connected instrument
2
Connect the product’s BNC connector to the input
terminal on the connected instrument.
(2) Lock
(1) Line up
Line up
3
Connect the alligator clips to the input leads.
Purchase the tip option that is best suited to the
measurement target. These options can be pur-
chased via authorized Hioki distributor or reseller.
CAUTION
Precautions when using the L4932 Test Pin Set
To prevent a short circuit accident,be sure to use the
test leads with the sleeves attached when performing
measurements in the CAT III measurement category.
(For the measurement categories,see “Measurement
categories”.)
If the sleeves are inadvertently removed during mea-
surement,stop the measurement.
1000V
600V
1000V
600V
IMPORTANT
Use only the specied tip options. Using a non-specied tip option
may result in incorrect measurements due to poor connection or
other reasons.
4
Select the voltage division ratio with the POWER
switch.
Voltage
division ratio Description
1/1000 Divides the high signal and low signal by 1,000
and then outputs the difference between the
two signals as a signal relative to ground.
1/100 Divides the high signal and low signal by 100
and then outputs the difference between the
two signals as a signal relative to ground.
5
Select the output type (P9000-02 only).
Output Description
Wave Outputs the signal using the selected voltage
division ratio. Choose this setting when you
wish to view the power supply waveform.
AC-RMS Outputs the RMS value using the selected
voltage division ratio. Input is AC coupled.
6
Set the input range scaling for the connected in-
strument.
7
Connect the alligator clips to the measurement target.
8
After measurement is complete,turn off the POW-
ER switch.
Specifications
Product warranty
period
3 years (except for breaks in cables caused by use)
Accuracy guaran-
tee period
1 year
DC output accu-
racy
±0.5% f.s.
(f.s.=1.0 V (ratio 1000:1),3.5 V (ratio 100:1))
Temperature
characteristic
±0.05% f.s./C
Frequency
characteristic
DC to 100 kHz,-3 dB
Voltage division
ratio
1000:1,100:1 (user-switchable)
Common-mode
rejection ratio
80 dB or more (50/60 Hz shorted input)
Input resistance 10.5 MΩ± 1% (between HIGH and LOW terminals)
Input capacitance 5 pF or less (at 100 kHz)
Input jacks Banana plug
Output connector Insulated BNC
Measurable range 1000 V AC/DC (ratio 1000:1)
250 V AC,350 V DC (ratio 100:1)
Maximum input
voltage
1000 V AC/DC
Maximum rated
voltage to earth
1000 V AC/DC,measurement category III
anticipated transient overvoltage: 8000 V
Operating temper-
ature and humidity
Temperature: -40C to 80C/ -40F to 176F
Humidity: -40C to 40C [non-inclusive]
/ -40F to 104F,80% RH or less
40C to 45C [non-inclusive]
/ 104F to 113F,60% RH or less
45C to 80C/ 113F to 176F,
50% RH or less (no condensation)
Storage tempera-
ture and humidity
Temperature: -40C to 80C/ -40F to 176F
Humidity: 80% RH or less (no condensation)
Operating
environment
Indoors,Pollution Degree II,altitude up to 2000 m (6562
ft.)
Dimensions Approx. 128.0W×36.0H×22.0D mm
/ Approx. 5.04"W×1.42"H×0.87"D
Mass Approx. 170 g/ 6.0 oz.
Power supply (1) Model Z1008 AC adapter
(12 V DC,DC jack φ5.5×2.1 mm)
Rated supply voltage: 100 V AC to 240 V AC
(Voltage uctuations of ±10% from the rated supply
voltage are taken into account.)
Rated supply frequency: 50 Hz/60 Hz
Anticipated transient overvoltage: 2500 V
(2) USB bus power (5 V DC USB-microB connector)
(3) External power supply,2.7 V DC to 15 V DC
Maximum rated
power
(1) When using a model Z1008 AC Adapter:
6 VA (model P9000-01/-02 with a AC adapter)
0.9 VA (model P9000-01/-02 only)
(2) When using a USB bus power: 0.8 VA
(3)
When using a 15 V DC external power supply: 1 VA
Effect of radiated radio-frequency electromagnetic eld
±5% f.s. max. at 3 V/m
Effect of conducted radio-frequency electromagnetic eld
±5% f.s. max. at 3 V
Applicable
Standards
EMC EN61326 Class A
Safety EN61010
Accessories Instruction manual ×1
Alligator clips×1 set (red and black/ one each)
Carrying case×1
Options
The following options are available for the product. To purchase an option,
please contact your authorized Hioki distributor or reseller.
Options are subject to change. Please check Hioki’s website for the latest
information.
Z1008 AC Adapter
9243 Grabber Clip (CAT III 1000 V)
L4936 Bus Bar Clip Set (CAT III 600 V)
L4937 Magnetic Adapter Set (CAT III 1000 V)
L4931 Extension Cable Set (For extending the length of the test leads,1.5 m)
L4932 Test Pin Set
(CAT IV 600 V,CAT III 1000 V)
L4934 Small Alligator Clip Set (CAT III 300 V,CAT II 600 V)
(L4932 is required when using L4934.)
Model P9000-02 only
Measurement modes Waveform,AC-RMS (selected with a slide switch)
Convertible voltage 1000 V AC (ratio 1000:1),250 V AC (ratio 100:1)
RMS value measure-
ment accuracy
±1% f.s. (30 Hz to 1 kHz [non-inclusive],sine wave)
±3% f.s. (1 kHz to 10 kHz,sine wave)
(f.s.=1.0 V (ratio 1000:1),2.5 V (ratio 100:1))
RMS value measure-
ment response time
300 ms
Rising 0% f.s.→100% f.s.
Response time for input from 0% f.s. to 90% f.s.
600 ms
Falling 100% f.s.→0% f.s.
Response time for input from 100% f.s. to 10% f.s.
f.s. : maximum display value or scale length (This is usually the max-
imum value of the currently selected range.)
This manual suits for next models
1
Other Hioki Measuring Instrument manuals

Hioki
Hioki 3506-10 Parts list manual

Hioki
Hioki VT1005 User manual

Hioki
Hioki 3196 User manual

Hioki
Hioki CM4003 User manual

Hioki
Hioki PW3336-01 User manual

Hioki
Hioki RM3544 User manual

Hioki
Hioki SM9001 User manual

Hioki
Hioki LUX HiTESTER 3423 User manual

Hioki
Hioki 9651 User manual

Hioki
Hioki FT3424 User manual

Hioki
Hioki 9666 User manual

Hioki
Hioki SM-8216 User manual

Hioki
Hioki SM7420 User manual

Hioki
Hioki SP7001 User manual

Hioki
Hioki RM3542-50 User manual

Hioki
Hioki PW3335 User manual

Hioki
Hioki MR8740T Instruction manual

Hioki
Hioki PW3390 User manual

Hioki
Hioki PW3335 User manual

Hioki
Hioki IM3533-01 Use and care manual