SERIAL NUMBERS
This manual applies to all instruments.
ABCDE
HP Part No. 81536-90011
Printed in Germany
Second Edition
E0796
Control Serial Number: Edition 2 applies directly to all instruments.
Edition 1 : 1st May 1990 : 81536-90011 : E0590
1st December 1990 : 81536-90011 : E1290
Edition 2 : 1st April 1992 : 81536-90011 : E0492
1st March 1993 : 81536-90011 : E0393
1st December 1994 : 81536-90011 : E1294
1st July 1996 : 81536-90011 : E0796
Safety Considerations
Before operation, you should review the instrument and manual, including the
red safety page, for safety markings and instructions.You must follow these to
ensure safe operation and to maintain the instrument in safe condition.
Initial Inspection
Warning
To avoid hazardous electrical shock, do not perform
electrical tests when there are signs of shipping damage to
any portion of the outer enclosure (covers, panels, etc.).
Power Requirements
The HP 81536A will operate when installed into the HP 8153A Optical
Multimeter mainframe.
Input/Output Signals
Caution
A maximum of 15V can be applied as an external voltage to any
BNC connectors.
iii
Contents
C. HP 81536A Specications
Supplementary Performance Characteristics . . . . . . . . . . . C-3
Analog output: .... .. .. .. .. .. .. .. .. .. . C-3
D. HP 81536A Performance Tests
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Equipment Required . . . . . . . . . . . . . . . . . . . . . D-1
Test Record .......................... D-2
Test Failure .......................... D-3
Instrument Specication .... .. .. .. .. .. .. .. .. D-3
Performance Test .... .. .. .. .. .. .. .. .. .. . D-3
Accuracy and Linearity Test .... .. .. .. .. .. .. . D-4
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . D-4
I. Accuracy Test ...................... D-5
II. Linearity Test...................... D-5
III. Noise Test .. .. .. .. .. .. .. .. .. . .. .. . D-7
E. Cleaning Procedures
The Cleaning Kit . . . . . . . . . . . . . . . . . . . . . . . . E-1
Other Cleaning Tools . . . . . . . . . . . . . . . . . . . . . E-3
Preserving Connectors . . . . . . . . . . . . . . . . . . . . . E-4
Cleaning Instrument Housings ................. E-4
Correct Cleaning Procedure .... .. .. .. .. .. .. .. E-5
Cleaning Cable Connectors . . . . . . . . . . . . . . . . . . E-5
Cleaning Connector Adapters . . . . . . . . . . . . . . . . . E-6
Cleaning Connector Interfaces ................ E-6
How to clean instruments with a recessed lens interface . . . . E-7
Light dirt .. .. .. .. .. .. .. .. .. .. . .. .. E-7
Heavy dirt ........................ E-7
Index
Contents-1
Figures
D-1. Accuracy and Linearity Test Setup .............. D-4
D-2. Addition of results (example) . . . . . . . . . . . . . . . . . D-7
Contents-2
C
C
HP 81536A Specications
Specications describe the instrument's warranted performance. Supplementary
performance characteristics describe the instrument's non-warranted typical
performance.
Because of the modular nature of the instrument, these performance
specications apply only to this module.You should insert these pages into the
appropriate section of the manual.
HP 81536A Specications C-1
C
HP 81536A Specications
Sensor Element
InGaAs
Wavelength range
800-1700nm
Power range
+3 to -70dBm
Display resolution
dBm, dB 0.001
0.0001 on printout
Watt 0.01 to 10
(depending on power range)
Applicable ber type
9/125

m - 100/140

m, NA

0.3
Uncertainty (Accuracy)
6
2.5% (1000-1650nm)
[1]
Total Uncertainty
6
5%
6
50pW (1000-1650nm)
[2]
Linearity
18

Cto28

C const. temp
6
0.015dB
6
30pW
0

Cto55

C const. temp
6
0.05dB
6
50pW
Noise
<
50pW (1200-1600nm)
peak-peak, avg. time 1sec
Dimensions
75mm H, 32mm W, 335mm D (2.8
"
2
1.3
"
2
13.2
"
)
Weight
net 0.6kg (1.3lbs), shipping 1kg (2.2lbs)
Recalibration period
2 years
Warmup time
20 min.
The display may vary by
6
1 count.
Information on the traceability of power meters is available on request
[1] at the following reference conditions:
Power level 10

W (-20dBm), Continuous Wave (CW).
Fiber 50

m graded index, NA=0.2, fully excited.
C-2 HP 81536A Specications
C
Ambient temperature 23

C
6
5K.
Connector Diamond HMS-10/HP.
At day of calibration (add 0.3% for aging over one year, add 0.6% over
two years).
Spectral width of source
<
10nm.
[2] at the following operating conditions:
Fiber

50

m, NA

0.2.
For NA
>
0.2 add 1%.
Ambient temperature 0 to 55

C, non-condensing.
Within 1 year after calibration, add 0.3% for second year.
For Biconic Connector add
6
1%.
Supplementary Performance Characteristics
Add 1% to total uncertainty for the full wavelength range.
Outside the specied wavelength range, the noise increases by up to 5 times
the value shown above.
Analog output:
Bandwidth:

DC,

300 to 400Hz, depending on range and sensor module.
Output voltage: 0 to 2V into open.
Output impedance: 600 typical.
Max. input voltage:
6
10V.
HP 81536A Specications C-3
D
D
HP 81536A Performance Tests
Introduction
The procedures in this section test the performance of the instrument.
The complete specications to which the HP 81536A is tested are given in
Appendix C. All tests can be performed without access to the interior of the
instrument. The test equipment given corresponds to tests carried out with
Diamond HMS-10/HP conectors.
Equipment Required
Equipment required for the performance test is listed in the table below. Any
equipment which satises the critical specications of the equipment given in
the table, may be substituted for the recommended models.
HP 81536A Performance Tests D-1
D
Instrument/Accessory Recommended Model Required
Characteristics
Power Meter Standard HP 8153A Mainframe with Uncertainty
#C01 HP 81533A Optical Head
6
1.5%
Interface Module with
HP 81524A Optical Head
Multimeter Mainframe HP 8153A
CW Laser Sources HP 81552SM and HP 81553SM or Wavelength
HP 81554SM Uncertainty
6
1.5nm
Optical Attenuator HP 8156A Option 101
Connector Adapter (head) HP 81000AA
Connector Interface HP 81000AI 4ea (08154-61701)
Singlemode Fiber HP 81101AC 2ea
Plastic Cap 5040-9351 2ea
Test Record
Results of the performance test may be tabulated on the Test Record provided
at the end of the test procedures. It is recommended that you ll out the Test
Record and refer to it while doing the test. Since the test limits and setup
information are printed on the Test Record for easy reference, the record can
also be used as an abbreviated test procedure (if you are already familiar with
the test procedures). The Test Record can also be used as a permanent record
and may be reproduced without written permission from Hewlett-Packard.
D-2 HP 81536A Performance Tests
D
Test Failure
If the HP81536A fails any performance test, return the instrument to the
nearest Hewlett-Packard Sales/Service Oce for repair
.
Instrument Specication
Specications are the performance characteristics of the instrument which
are certied. These specications, listed in Appendix C, are the performance
standards or limits against which the HP 81536A can be tested. Appendix C
also lists some supplemental characteristics of the HP 81536A. Supplemental
characteristics should be considered as additional information.
Any changes in the specications due to manufacturing changes
, design, or
traceability to the National Institute of Standards and T
echnology (NIST), will be
covered in a manual change supplement, or revised manual. Such specications
supercede any that were previously published.
Performance Test
The performance test given in this section includes the Accuracy Test, the
Linearity Test, and the Noise Test. Perform each step in the order given, using
the corresponding test equipment.
Note
Make sure that all optical connections of the test setups given
in the procedure are dry and clean. DO NOT USE INDEX
MATCHING OIL.
The optical cables from the laser source to and from the HP
8156A Attenuator to the power meter must be xed on the
table to ensure minimum cable movement during the tests.
HP 81536A Performance Tests D-3
D
Accuracy and Linearity Test
Specications HP 81536A
Uncertainty :
6
2.8%
y
(1000-1650nm)
Linearity :
6
0.015dB
6
30pW
y
includes aging.
Test Setup
1. Make sure that cable connector, detectors and adapters are clean.
2. Connect the equipment as shown in Figure D-1. Ensure that the cables to
and from the attenuator are xed on the table and that both the optical head
and the DUT are close together so that minimum cable movement is required
when connecting the cable to the head or to the DUT.
Figure D-1. Accuracy and Linearity Test Setup
3. Turn the instruments on, enable the laser source and allow the instruments
to warm up for at least 20 minutes.
Note
The linearity test must only be performed at either 1300nm or
1550nm. The accuracy test must be performed in the -20dBm
range at 10

W at both 1300nm and 1550nm.
D-4 HP 81536A Performance Tests
D
I. Accuracy Test
1. On both power meters:
a. Set the calibration factor to zero.
b. Set the wavelength to the wavelength of the laser source.
c. Set the instruments into MEAS mode. Make sure AUTOranging is o and
select the -20dBm range.
d. Set the averaging time to 500ms.
2. Make sure that the HP 8156A output is disabled. ZERO the DUT.
3. Connect the HP 8156A output cable to the standard head and ZERO the
standard meter.
4. Enable the HP 8156A output. Change the HP 8156A attenuation until the
power meter standard displays 10.00

W.
5. Connect the attenuator output cable to the DUT. Check the DUT display and
note the result in the test record.
6. Repeat steps 1 to 5 at the second wavelength with the corresponding source.
II. Linearity Test
1. Make sure that the HP 8156A output is disabled. ZERO the DUT
.
2. Enable the HP 8156A output.
3. On the DUT, switch o autoranging and select the -20dBm range. Set the
display to show results in dBm.
4. Alter the attenuation until the DUT displays -17.4dBm.
5. Press
4
Disp
!
Ref
5
, then
4
dB
5
.
6. Press
4
Up
5
to select the -10dBm range.
7. Note the deviation, displayed in dB, as R1 on the test record.
a. In the -10dBm range. Set the display to show results in dBm.
b. Alter the attenuation until the DUT displays -7.4dBm.
c. Press
4
Disp
!
Ref
5
, then
4
dB
5
.
d. Select the 0dBm range.
e. Note the deviation, displayed in dB, as R2 on the test record.
HP 81536A Performance Tests D-5
D
8. Repeat the steps a to e at the all the range and level settings shown in the
test record.
9. Calculate the non-linearity using the formulae in the test record.
Example Results for the Linearity test
Test Min Max imum Measurement
No.Test Description Spec. Result Spec. Uncertainty
II.
Linearity Test
dB
Range Power (dBm)
-20dBm -17.4 Disp
!
Ref
-10dBm
-0.004
=R1
-10dBm -7.4 Disp
!
Ref
0dBm
-0.008
=R2
-20dBm -27.4 Disp
!
Ref
-30dBm
-0.008
=R3
Example Calculated Results for the Linearity Test
Test Minimum Maximum Measurement
No.Test Description Spec. Result Spec. Uncertainty
Non-Linearity
Range Formula
0dBm R1+R2 -0.015dB
-0.012
dB +0.015dB
6
0.006dB
-10dBm R1 -0.015dB
-0.004
dB +0.015dB
6
0.006dB
-20dBm 0.000dB
6
0.00%
-30dBm R3 -0.015dB
-0.008
dB +0.015dB
6
0.006dB
Shown graphically, these results look as shown in the following gure:
D-6 HP 81536A Performance Tests
D
Figure D-2. Addition of results (example)
III. Noise Test
Note
The noise measurement must be performed with either a
module or a blank panel in the second channel position.
Connect the equipment as shown in Figure D-1.
Disable the HP 8156A output.
On the DUT, select AUTOranging, and display in Watts. Set the averaging time
to 1s.
Press
4
Zero
5
.
Select the
MENU
mode.
Press
4
Record
5
, to select
Stability
.
Press
4
Edit
5
and select the
T_TOTAL
parameter. Set
T_TOTAL
to 1 minute.
Press
4
Edit
5
, and then
4
Exec
5
to run the noise measurement.
When the test has completed, press
4
More
5
to select the
SHOW
application. Press
4
Edit
5
and then
4
Next
5
/
4
Prev
5
to display the
DIFF
result.
Note the DIFF result as the noise value in the test record.
HP 81536A Performance Tests D-7
D
Performance Test for the HP 81536A
Page 1 of 4
Test Facility:
Report No.
Date
Customer
Tested By
Model HP81536A Sensor Module
Serial No. Ambient temperature

C
Options Relative humidity %
Firmware Rev. Line frequency Hz
Special Notes:
D-8 HP 81536A Performance Tests
D
Performance Test for the HP 81536A
Page 2 of 4
Test Equipment Used:
Description Model No. Trace No. Cal. Due Date
1. Lightwave Multimeter (Std.) HP 8153A
2. Opt. Head Interface Module HP 81533A
3. Optical Head HP 81524A
4a1. CW Laser Source HP 81552SM
4a2. CW Laser Source HP 81553SM
4b. CW Laser Source HP 81554SM
5. Optical Attenuator HP 8156A Option 101
6. Lightwave Multimeter (DUT) HP 8153A
7. Connector Interface HP 81000AI N/A N/A
8. Connector Adapter HP 81000AA N/A N/A
9. Singlemode Fiber (2ea) HP 81101AC N/A N/A
12.
13.
14.
15.
16.
17.
18.
HP 81536A Performance Tests D-9
D
Performance Test for the HP 81536A
Page 3 of 4
Model HP 81536A Sensor Module No. Date
Test Minimum Maximum Measurement
No.Test Description Spec. Result Spec. Uncertainty
I.
Accuracy Test

W
measured at nm
Output Power 9.72

W 10.28

W
6
2%
measured at nm
Output Power 9.72

W 10.28

W
6
2%
II.
Linearity Test
dB
Range Power
(dBm)
-20dBm -17.4 Disp
!
Ref
-10dBm =R1
-10dBm -7.4 Disp
!
Ref
0dBm =R2
-20dBm -27.4 Disp
!
Ref
-30dBm =R3
D-10 HP 81536A Performance Tests

Other HP Control Unit manuals

HP 83400 Series User manual

HP

HP 83400 Series User manual

HP 83487A User manual

HP

HP 83487A User manual

HP VXI E1476A Operating instructions

HP

HP VXI E1476A Operating instructions

HP A12500 Assembly instructions

HP

HP A12500 Assembly instructions

HP 6600-A FIP-600 FlexIntfPltfrm Rtr Mod Assembly instructions

HP

HP 6600-A FIP-600 FlexIntfPltfrm Rtr Mod Assembly instructions

HP 98624A Quick start guide

HP

HP 98624A Quick start guide

HP HP 81551MM User manual

HP

HP HP 81551MM User manual

HP HPE FlexNetwork 7500 User manual

HP

HP HPE FlexNetwork 7500 User manual

HP HP 81552SM User manual

HP

HP HP 81552SM User manual

HP 3484A Service manual

HP

HP 3484A Service manual

HP E3456A User manual

HP

HP E3456A User manual

HP ProCurve Wireless 170wl User manual

HP

HP ProCurve Wireless 170wl User manual

HP 81541MM Owner's manual

HP

HP 81541MM Owner's manual

HP SIC-4G-LTE-V User manual

HP

HP SIC-4G-LTE-V User manual

HP E2261A User manual

HP

HP E2261A User manual

HP 572018-B21 User manual

HP

HP 572018-B21 User manual

HP 7500 48p 1000BASE-T w/PoE+ SE Mod Installation instructions

HP

HP 7500 48p 1000BASE-T w/PoE+ SE Mod Installation instructions

HP JG589A Assembly instructions

HP

HP JG589A Assembly instructions

HP 70427A User manual

HP

HP 70427A User manual

HP 83481A User manual

HP

HP 83481A User manual

HP 75000 B Series User manual

HP

HP 75000 B Series User manual

HP R6000 - UPS - Lead Acid Expandable User manual

HP

HP R6000 - UPS - Lead Acid Expandable User manual

HP ProCurve PCM+ Agent on ONE zl Module Assembly instructions

HP

HP ProCurve PCM+ Agent on ONE zl Module Assembly instructions

HP Synergy 480 Gen9 User manual

HP

HP Synergy 480 Gen9 User manual

Popular Control Unit manuals by other brands

Tektronix VX4286 user manual

Tektronix

Tektronix VX4286 user manual

ArduCAM 8MP IMX179 quick start guide

ArduCAM

ArduCAM 8MP IMX179 quick start guide

REXROTH IndraControl L25 operating instructions

REXROTH

REXROTH IndraControl L25 operating instructions

Protectoseal 7800 Series Installation & maintenance

Protectoseal

Protectoseal 7800 Series Installation & maintenance

ASCO Numatics 651 Series Installation and maintenance instructions

ASCO Numatics

ASCO Numatics 651 Series Installation and maintenance instructions

Agilent Technologies N5532A User's and service guide

Agilent Technologies

Agilent Technologies N5532A User's and service guide

Vestamatic Vesta Building Controller Installation and operating instructions

Vestamatic

Vestamatic Vesta Building Controller Installation and operating instructions

Atmel ATBTLC1000 BluSDK user guide

Atmel

Atmel ATBTLC1000 BluSDK user guide

Spectron MV3 Series operating manual

Spectron

Spectron MV3 Series operating manual

DeLOCK 95254 user manual

DeLOCK

DeLOCK 95254 user manual

Mo-vis P015-61 installation manual

Mo-vis

Mo-vis P015-61 installation manual

Siemens MICROMASTER Vector 6SE32 Getting started

Siemens

Siemens MICROMASTER Vector 6SE32 Getting started

rotork IQT3 Original instructions

rotork

rotork IQT3 Original instructions

Visonic POWERMAX PRO installation instructions

Visonic

Visonic POWERMAX PRO installation instructions

EWM Expert XQ 2.0 operating instructions

EWM

EWM Expert XQ 2.0 operating instructions

Orbit B-hyve quick start guide

Orbit

Orbit B-hyve quick start guide

DFI WL968 user manual

DFI

DFI WL968 user manual

King Pigeon S273 user manual

King Pigeon

King Pigeon S273 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.