
16 7 VG9F 70 en
❑To conclude press the ?key when the desired
value is shown on the display.
NOTE: If nothing else is defined, please select type 1 in
any class.
❑Select parameters as follows:
4.4.5 Actuator size, ACTS
This parameter defines the actuator size.
❑Look at e.g. the type code on the machine plate
in the Neles B1-series actuators to check the
size. If 3rd party actuator is used, please check
the actuator stroke volume.
❑After selecting ACTS on the display, press the ?
key to enter the edit state and ACTS starts to blink.
❑Select between the following values:
S 1 = Neles B1J8 actuator
(or stroke volume <1 dm3/ <61 in3)
S 3 = B1J10 (1-3 dm3/ 61-183 in3)
S10 = B1J12-16 (3-10 dm3/ 183-610 in3)
S30 = B1J20-25 (10-30 dm3/ 610-1831 in3)
L30 = B1J32- (>30 dm3/ >1831 in3)
Use the +and -keys to change the value.
❑To conclude press the ?key when the desired
value is shown on the display.
4.4.6 Spool type, STYP
This parameter defines the spool type and size in
VG9000F.
❑Look at the machine plate in the device to check
the typecode.
❑After selecting STYP on the display, press the ?
key to enter the edit state and STYP starts to blink.
❑Select between the following values:
15 = VG9_15
35 = VG9235
37 = VG9237
Use the +and -keys to change the value.
❑To conclude press the ?key when the desired
value is shown on the display.
4.4.7 Language selection, LANG
❑Select between three languages EnG, GEr or FrE
(English, German or French) using the +and -keys.
❑To conclude press the ?key when the desired
value is shown on the display.
4.5 Valve travel calibration
Select CAL from the menu by using +or -keys and
press the ?key.
NOTE: When device usage is hybrid the valve needs to
be in the normal position (not fail safe) for calibration to
be started.
4.5.1 AUTO CAL calibration function
During calibration process a blinking text "CAL run" will
be show on the display. If calibration ends successfully,
a text "CALIBRATION SUCCESSFUL" will be shown. Cali-
bration can be cancelled with the =key, which will
show a text "CALIBRATION CANCELLED". If calibration
fails, the reason will be shown, eg. "CALIBRATION
START FAILED", "POSITION SENSOR RANGE TOO
SMALL", "CALIBRATION TIMEOUT" or "CALIBRATION
FAILED". After calibration the device will return to the
main menu (measurement monitoring).
4.6 Testing, TESt
❑Select between two tests Partial Stroke Test
(PARt tst) or Pneumatics test (PNEU tst).
❑To conclude press the ?key when the desired
value is shown on the display.
4.6.1 Partial Stroke Test, PARt tst
Table 4 Selecting extra pneumatics instrumentation
parameters
Instrumentation type Parameter (EXTI)
Volume booster (VB) Booster type 1 (bo1)
Quick exhaust (QEV) Quick exhaust 1 (qE1)
Combination of VB and QEV Combination type 1 (co1)
NOTE:
When QEV or volume booster is used, VG with small
spool valve (VG_15_) is required.
NOTE:
When QEV or volume booster is used, VG with small
spool valve (VG_15_) is required.
CAUTION:
It is important to select the correct actuator size
because this parameter is used in device control.
Erroneous value may cause instability.
CAUTION:
It is important to select the correct spool type
because this parameter is used in device control.
Erroneous value may cause instability.
WARNING:
Automatic calibration drives the valve against the
mechanical open and closed travel limits of the
valve-actuator assembly and a tuning procedure is
performed. Make sure that these procedures can
be safely executed.
NOTE:
Make sure to run the pneumatics test before the auto-
matic travel calibration.
Fig. 14 Calibration selection
NOTE:
Valve position needs to be in the normal operating
position, supply pressure needs to be in valid range,
no supply pressure drop is allowed and any test can-
not be active when calibration is started.
Pneumatics test needs to be performed succesfully
before the calibration.
WARNING:
Partial Stroke Test moves the valve according to
the stroke size. Make sure that this procedure can
be safely executed.