Mitutoyo HTD-R User manual

Mitutoyo Corporation 20-1, Sakado 1-Chome, Takatsu-ku, Kawasaki-shi, Kanagawa 213-8533, Japan Date of publication: December 1, 2021
Printed in Japan
1.  Names of Components
④
⑤⑥⑨⑩
⑪⑫
⑬
⑦⑧
①②
③
①Measuring head
②LCD body
③Display unit
④Contact point
⑤Connection area
⑥Display unit (LCD)
⑦Output connector
⑧Cover
⑨Sleeve
⑩Ratchet
⑪Thimble
⑫Setting screw
⑬Battery compartment cover
■Display unit (LCD)
⑱
⑲
⑳
⑮
⑯
⑰
⑭
⑭[HOLD] key
⑮[ZERO/ABS] key
⑯[PRESET] key
⑰[in/mm] key (in/mm products only)
⑱ Sign display
⑲ Function Lock display
⑳Hold display
㉑Low voltage display
㉒INC display
㉓Preset display
㉔Unit display
2.  Installing the Battery
When attaching the battery compartment cover, be sure to attach the gasket properly. The product
may display an error or malfunction if the battery compartment cover or gasket is not mounted
correctly.
Tips
• Be sure to use SR44 (silver oxide button battery part No. 938882) for the battery.
• Do not rotate the thimble until the count is displayed. Initial setting of the electrical components
may fail, or the product may not count normally. If you mistakenly move the thimble, reinstall the
battery.
• The battery supplied is for confirming the functions and performance of the product. Note that
this battery may not fulfill the predetermined life.
• Malfunction or damage due to depleted batteries, etc. is not covered by the warranty.
• Follow local rules and regulations regarding battery disposal.
The battery is not installed into the product at purchase. Install the battery as follows.
1Use the supplied Phillips screwdriver (No.05CAA952) to loosen and remove the battery
compartment cover fixing screws (M1.7 x 0.35 x 4/No.04AAB541).
2If replacing an existing battery, remove the old battery.
3Insert the battery (SR44) with the positive side facing upward.
Confirm that the gasket (No.05SAA372) is installed correctly in the proper position.
4
Place the battery compartment cover over the battery compartment and hold down the edge by
hand while making sure there is no gap between the cover and body, and then tighten it using
the screws.
》 The "- - - - - -" display blinks.
5Press the [PRESET] key.
》Count display appears and counting starts.
PRESET
3
4
1
5
①
+ -
+ terminal
- terminal
SR44
②
SR44
Tips
• Re-installing the battery will erase the PRESET value (reference point) position. Perform
reference point setting again (refer to "4. Reference Point Setting").
• If an abnormal display is shown, such as an error display or not counting, etc., try removing the
battery and reinstalling.
Safety Precautions
To ensure operator safety, use this product in conformance with the directions, functions
and specifications given in this User's Manual.
Use under other conditions may compromise safety.
• Always keep batteries out of reach of children, and if swallowed, consult a physician
immediately.
• Batteries should never be short-circuited, disassembled, deformed or come in contact with
extreme heat or flames.
• If battery alkaline liquid comes in contact with the eyes, flush eyes immediately with clean
water and consult a physician. If battery alkaline liquid comes in contact with the skin,
flush the exposed area thoroughly with clean water.
• Never attempt to charge the primary battery or reverse the positive-negative terminals
when mounting. Improper battery handling or mounting may cause the battery to explode,
cause battery leakage and/or serious bodily injury or malfunctioning.
• Always handle the sharp measuring faces of this product with care to avoid injury.
• Do not disassemble or modify.
• Do not use or store the product in a place with sudden temperature changes. Adapt the
product to ambient temperature before use.
• Do not store the product in a place with high humidity or a lot of dust.
• Firmly close the battery compartment cover if the product is used in a place where it is
directly exposed to splashes of coolant, etc. When mounting the output cable and cover,
firmly tighten the fixing screws so that there is no gap. As well, clean and apply anti-rust
treatment after use. Rust may cause malfunction.
• Do not use even waterproof types submerged, as coolant ingress cannot be completely
prevented. Complete prevention of coolant ingress, etc., may also not be possible if the
product is used in locations exposed to direct jets of liquid.
• Do not apply excessive force or subject to sudden impacts such as dropping.
• If oil or cutting chips become attached to the sliding portion of the contact point, a
malfunction may result. Wipe off any oil or cutting chips after use.
• When cleaning, wipe this product with a soft cloth moistened with diluted neutral
detergent. Do not use an organic solvent such as thinner, which may cause the product to
deform or malfunction.
• Do not write numbers, etc. with an electric pen.
• If the product is to be out of use for three months or more, remove the battery before storage.
Liquid leakage from the battery may damage the product.
• Do not pry out or suspend the product while it is inserted inside the workpiece.
• Use the product only with the supplied contact point.
• Do not remove the contact points; otherwise damage may result.
Digimatic Holtest
HTD-R
User's Manual No. 99MAB026A1
=
> 2 s
=
< 1 s
Key operation icon
Contents
1. Names of Components ...........................................................................................Page 1
2. Installing the Battery ...............................................................................................Page 1
3. Precautions for Use ................................................................................................Page 2
4. Reference Point Setting..........................................................................................Page 2
5. Measurement Method.............................................................................................Page 2
6. Key Functions .........................................................................................................Page 2
7. Function Lock Function (Preventing Accidental Operation)....................................Page 2
8. Errors and Troubleshooting.....................................................................................Page 3
9. Installation/Removal Method for Measuring Head and Extension Rod...................Page 3
10. Specifications..........................................................................................................Page 3
11. Output Function ......................................................................................................Page 3
12. Options....................................................................................................................Page 3
13. Off-Site Repairs (Subject to Charge) ......................................................................Page 3
en
1
Mitutoyo Corporation 20-1, Sakado 1-Chome, Takatsu-ku, Kawasaki-shi, Kanagawa 213-8533, Japan Date of publication: December 1, 2021
Printed in Japan
3.  Precautions for Use
■Measuring Force
• Use the ratchet to ensure consistent measuring force.
• The appropriate measuring force is achieved with the following procedure: make light contact
between the measurement surfaces and the workpiece, stop momentarily, and then manually turn
the ratchet about five to six times. Note that excessive measuring force may cause errors.
■Precautions after Use
• After use, clean the entire product and check that none of the parts are damaged.
•
If using in places exposed to water-based cutting fluid, always apply anti-rust treatment after cleaning.
• If the product is to be out of use for three months or more, remove the battery before storage.
4.  Reference Point Setting
Set an arbitrary preset value (reference point registration) prior to setting the reference point (reference point setting).
•
Be sure to follow the procedure below to confirm and set the reference point prior to measuring.
•
When setting the reference point for this product, make sure to use a calibrated gage (setting ring, etc.).
• Remove any dirt or oil from the measuring surfaces of the gage and product prior to
setting the reference point.
• Due to the product mechanism, the measured value will differ depending on whether
the entire surface of the contact point or only its edge is used for measurement. When
measuring, use the same conditions when setting the reference point.
• Use the same orientation and conditions when measuring and setting the reference
point. (Refer to Figure 1 if measuring with the contact point edge in a blind hole.)
Do not ground the bottom of the measuring head when setting the reference point or
measuring (Figure 2).
Figure 1
Reference point setting
Workpiece
Figure 2
1) Reference point registration
Register (preset) the gage dimensions to the product. Two preset values (P1 and P2) can be
registered to the product.
Tips
Press and hold the [HOLD] key to switch between P1 and P2.
<Example> Registering 125.000 mm to P1
1 Briefly press the [PRESET] key.
》
The previously registered number is displayed and "P1" blinks.
Tips
•
Zero will be displayed immediately after replacing the battery.
• If "P2" is blinking, press and hold the [HOLD] key to
cause "P1" to blink.
2Press and hold the [PRESET] key.
》 The sign blinks.
Tips
Briefly press the [PRESET] key to switch between "+" and "-".
3Press and hold the [PRESET] key.
》 The number on the left blinks.
Tips
The numbers will switch in order from "0" to "1" to "2" up to "9"
and then "0" each time the [PRESET] key is briefly pressed.
4 Briefly press the [PRESET] key until "1" is displayed.
5Press and hold the [PRESET] key.
》 The number in the next digit blinks.
3
5
4
6
7
8
2
1
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
6Repeat steps 4and 5, so that "2," "5," and "0" are
displayed in each digit.
7Press and hold the [PRESET] key until "P1" blinks.
8 Briefly press the [PRESET] key.
》 "P1" goes out and registration is complete.
2) Reference point setting
1Remove any dirt or dust from the measuring surfaces of
the gage and the product.
2Set the dimension of the product slightly smaller than
that of the gage, and then slowly insert it into the gage.
3Rotate the thimble and bring the contact points gently
into contact with the inside of the gage.
4 Rotate the ratchet five to six times to apply the proper
measuring force.
5 Briefly press the [PRESET] key.
》 "P1" or "P2" blinks, and the registered preset value
(zero if not registered) is displayed.
Tips
• Press and hold the [HOLD] key to switch between P1
and P2.
• To change the preset value, refer to steps 2through
8in "1) Reference point registration".
6 Briefly press the [PRESET] key.
》 "P1" or "P2" goes out.
To use thimble graduations as well, set using the procedure
below.
7Perform steps 1through 4, and then use the supplied
hex wrench to loosen the setting screw.
8Slightly rotate the sleeve and align it with the proper
indicated value.
9Use the supplied hex wrench to tighten the setting
screw and fix the sleeve.
10
Perform steps
1
through
4
, and confirm that the proper
number is displayed.
If the proper number is displayed, setting is complete. If it is
not displayed, repeat steps 7through 9.
Tips
• The display of this product automatically turns off if not
used for 20 minutes or more. To display again, either
rotate the thimble or press the [ZERO/ABS] key.
• If the [PRESET]key is accidentally pressed during
measurement, press the [ZERO/ABS]key to return to
the former state. If this does not enable the product to
recover, perform the reference point setting procedure
once more.
• Once the product has been inserted, do not move it
until the reference point has been set.
2 3
7
8
9
5.  Measurement Method
Be sure to perform reference point setting before measurement.
Using the same orientation and procedure used during reference point setting, insert the product
into the workpiece, and then read the display value.
6.  Key Functions
■[ZERO/ABS] Key
• Briefly press the [ZERO/ABS] key.
》 "INC" is displayed, and the display
is set to zero.
• Press and hold the [ZERO/ABS] key
(for at least 2 seconds).
》
"INC" goes out, and the length from
the reference point (anvil measuring
surface) is displayed.
■[HOLD] Key
• Press the [HOLD] key.
》 "H" is displayed, and the display value is held.
Press the key again to stop holding the value.
■[in/mm] Key (in/mm products only)
• Press the [in/mm] key.
》 "in" and "mm" switch back and forth each time
the key is pressed.
7.  Function Lock Function (Preventing Accidental Operation)
This product has the Function Lock function, which disables the PRESET function and ZERO/ABS
function in order to avoid accidental changes to the reference point position.
Setting the Function Lock causes []on the LCD to light up and disables the [PRESET] key,
[ZERO/ABS] key, and [in/mm] key (in/mm products only), with only the hold operation function
enabled.
1First press and hold the [HOLD] key, and then additionally press and hold the [ZERO/ABS] key
(for at least 2 seconds).
》 [H] display and [ ] display light up in sequence ([H] turns off first).
2Perform the same operation to release the Function Lock.
21
+
ZERO/ABSHOLD
+
ZERO/ABSHOLD
=
> 2 s
=
< 1 s
Key operation icon
ZERO/ABS ZERO/ABS
HOLD
in / mm
3
5
4
6
7
8
2
1
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
PRESET
2
Mitutoyo Corporation 20-1, Sakado 1-Chome, Takatsu-ku, Kawasaki-shi, Kanagawa 213-8533, Japan Date of publication: December 1, 2021
Printed in Japan
8.  Errors and Troubleshooting
• " " display
The battery voltage is low. Replace the battery promptly.
• "Err-oS" display
A counting error has occurred due to excessive speed or noise. Try removing the battery and
reinstalling.
• "Err-S" display
Initial setting of the electrical components failed, or a counting error has occurred due to a sensor
signal error. Try removing the battery and reinstalling.
9.  Installation/Removal Method for Measuring Head and Extension Rod
For interchangeable-head sets, it is necessary to replace the measuring head according to the size of
the workpiece.
An extension rod (optional for models other than interchangeable-head sets) can be used to measure
deep holes.
Do not hold the display unit while using the wrench to loosen or tighten the connection area.
Otherwise, damage may result.
• Be sure to set the reference point after installing or removing the measuring head or
extension rod.
• Do not misassemble the measuring head and display unit (for models other than the
interchangeable-head sets). These parts are incompatible and accuracy cannot be
guaranteed if they are used.
Use the following procedure to install and remove the measuring head and extension rod.
1
Set the supplied wrench to the groove in the
connection area. Fix the display unit in place
while rotating the measuring head by hand to
detach it.
2To replace the measuring head, use the
wrench to hold the display unit in place,
replace the measuring head, and then rotate
the new head by hand to tighten it.
3To install the extension rod, use the wrench
to hold the display unit in place, and then turn
the extension rod and then the measuring
head (in that order) by hand to tighten
them.
Use the same procedure to remove the
extension rod.
Display unit
Measuring head
Extension rod
Wrench
Loosen
Tighten
10.  Specications
■ Common Specications
Display : LCD (6-digit and minus sign)
Power supply : Button type silver-oxide battery (SR44 No.938882), x1
Battery life : 1.2 years
Operating temperature : 5 °C to 40 °C
Storage temperature : -10 °C to 60 °C
Standard accessories :
Single item Non-interchangeable-head sets Interchangeable-head sets
Wrench P P P
Hex wrench P P P
Phillips screwdriver
(No.05CAA952) P P P
Setting ring - P P
Extension rod - - P
IP protection level :
IP65 (refer to IEC60529 for details).
Dust-proof (level 6) : No ingress of dust allowed.
Protection against water jets (level 5) : Protects the equipment against water jets from any direction.
■ Individual Specications
Maximum measuring length Maximum permissible error
J
MPE*1 Resolution
8 - 12 mm ±2 µm (maximum difference 2 µm)
0.001 mm
16 - 63 mm ±3 µm (maximum difference 3 µm)
75 - 100 mm ±4 µm (maximum difference 4 µm)
125 - 200 mm ±5 µm (maximum difference 5 µm)
225 - 300 mm ±6 µm (maximum difference 6 µm)
0.35 - 0.5 in ±0.0001 in (maximum difference 0.0001 in)
0.00005 in0.65 - 2.5 in ±0.00015 in (maximum difference 0.00015 in)
3 - 4 in ±0.0002 in (maximum difference 0.0002 in)
5 - 8 in ±0.00025 in (maximum difference 0.00025 in) 0.0001 in
9 - 12 in ±0.0003 in (maximum difference 0.0003 in)
*1: Maximum permissible error for indicated value via contact with full measuring face
J
MPE (20 °C).
11.  Output Function
■Display Value External Output
The display value can be output to a device by connecting the product and the external device with
a connection cable (option).
■Connection Cable Installation Method
• Always use the 0-size Phillips screwdriver (No.05CZA619) supplied with the connection cable
(option) when installing/removing screws, and tighten to a torque of 5 to 8 cN・m or so.
Otherwise, it may cause damage.
• When connecting the connection cable, ensure that the connector gasket does not protrude.
If the connector gasket is not installed properly, waterproof functionality may decrease and lead
to malfunctions.
•
When connecting the connection cable, pay attention to the connector direction.
Otherwise, damage may result.
1
Use the Phillips screwdriver supplied with the connection cable to remove the cover fixing screws
(M1.7 x 0.35 x 2.5, No.04AAB543).
2Remove the cover.
3 Check that the connector gasket (No.09GAA374) is correctly installed at the proper position
(do not remove the connector gasket).
4Mount the connection cable plug.
5 While holding in the connection cable plug by hand, tighten the fixing screws.
Tips
•
Ensure that there is no gap between the connection cable plug and the body connector.
• Held display values (refer to "6. Key Functions ■[HOLD] Key") will be released if output to an
external device.
4
1
5
2
■Output Data Format
(1)
Output order
(2) All "F"
(3) Sign
(4)
Measured value
(5)
Decimal point
(6) Unit
■Timing Chart
200 ms
0.14 ms (Typ:0.122 ms)
0.14 ms (Typ:0.122 ms)
0.27 ms (Typ:0.244 ms)
80 ms
0 ms
0.11 ms
0.11 ms
0.22 ms
T1
T2
T3
T4
T5
*1: DATAsw is LOW while the data output key is being pressed.
*2: The time T5 until DATAsw goes to the LOW level and REQ is input is determined by the data processing
device performance.
12.  Options
• Connection cable (1 m) : No.05CZA662
• Connection cable (2 m) : No.05CZA663
• Extension rod
Length 100 mm, measuring range 6 mm to 12 mm (0.275 in to 0.5 in) : No.952322
Length 150 mm, measuring range 12 mm to 20 mm (0.5 in to 0.8 in) : No.952621
Length 150 mm, measuring range 20 mm to 50 mm (0.8 in to 2 in) : No.952622
Length 150 mm, measuring range 50 mm to 300 mm (2 in to 12 in) : No.952623
For options other than the above, refer to the General Catalog.
13.  O󰀨-Site Repairs (Subject to Charge)
Off-site repair (subject to charge) is required in the case of the following malfunctions. Contact your
nearest dealer or our sales office.
• The operation of the contact point is poor and the thimble rotation is sluggish.
Operation will worsen if there is oil or rust on the sliding portion of the contact point.
(1)
MSD LSD
(5)
(2)
(1111)
(4)
(3)
+:0(0000)
–:8(0001)
3(1100)
4(0010)
5(1010)
(6)
mm:0(0000)
in:1(1000)
3

Other Mitutoyo Measuring Instrument manuals

Mitutoyo CONTRACER CV-2100 User manual

Mitutoyo

Mitutoyo CONTRACER CV-2100 User manual

Mitutoyo 530-101 User manual

Mitutoyo

Mitutoyo 530-101 User manual

Mitutoyo HDM-30AX User manual

Mitutoyo

Mitutoyo HDM-30AX User manual

Mitutoyo BA9700 User manual

Mitutoyo

Mitutoyo BA9700 User manual

Mitutoyo 99MAG014M User manual

Mitutoyo

Mitutoyo 99MAG014M User manual

Mitutoyo QM-Data 300 User manual

Mitutoyo

Mitutoyo QM-Data 300 User manual

Mitutoyo 572-200-20 User manual

Mitutoyo

Mitutoyo 572-200-20 User manual

Mitutoyo ID-C X series Service manual

Mitutoyo

Mitutoyo ID-C X series Service manual

Mitutoyo MDH User manual

Mitutoyo

Mitutoyo MDH User manual

Mitutoyo LGB User manual

Mitutoyo

Mitutoyo LGB User manual

Mitutoyo ID-SX2 User manual

Mitutoyo

Mitutoyo ID-SX2 User manual

Mitutoyo 500-150-30 User manual

Mitutoyo

Mitutoyo 500-150-30 User manual

Mitutoyo DMS User manual

Mitutoyo

Mitutoyo DMS User manual

Mitutoyo 700 Series User manual

Mitutoyo

Mitutoyo 700 Series User manual

Mitutoyo 129 Series User manual

Mitutoyo

Mitutoyo 129 Series User manual

Mitutoyo ID-C125XB User manual

Mitutoyo

Mitutoyo ID-C125XB User manual

Mitutoyo ID-CNX Series User manual

Mitutoyo

Mitutoyo ID-CNX Series User manual

Mitutoyo 209 Series User manual

Mitutoyo

Mitutoyo 209 Series User manual

Mitutoyo IT-012U User manual

Mitutoyo

Mitutoyo IT-012U User manual

Mitutoyo ID-C112GXB User manual

Mitutoyo

Mitutoyo ID-C112GXB User manual

Mitutoyo LG200 User manual

Mitutoyo

Mitutoyo LG200 User manual

Mitutoyo QM-Height QMH-600AX User guide

Mitutoyo

Mitutoyo QM-Height QMH-600AX User guide

Mitutoyo D/E User manual

Mitutoyo

Mitutoyo D/E User manual

Mitutoyo ID-U User manual

Mitutoyo

Mitutoyo ID-U User manual

Popular Measuring Instrument manuals by other brands

YOKOGAWA TA520 user manual

YOKOGAWA

YOKOGAWA TA520 user manual

Dytran 3196C operating guide

Dytran

Dytran 3196C operating guide

MULTISPAN MFM 142DC-M1 quick start guide

MULTISPAN

MULTISPAN MFM 142DC-M1 quick start guide

Bowers Metrology Ultima operating instructions

Bowers Metrology

Bowers Metrology Ultima operating instructions

Extech Instruments AN100 user manual

Extech Instruments

Extech Instruments AN100 user manual

Lovibond PFX995 Series Operator's instruction manual

Lovibond

Lovibond PFX995 Series Operator's instruction manual

YaliTech MINI RAYTEMP operating instructions

YaliTech

YaliTech MINI RAYTEMP operating instructions

Hioki MEMORY HiCORDER MR8847-01 instruction manual

Hioki

Hioki MEMORY HiCORDER MR8847-01 instruction manual

FIS E Series user manual

FIS

FIS E Series user manual

Bender RCMB-35-30 Series quick start guide

Bender

Bender RCMB-35-30 Series quick start guide

Scientific Atlanta Quan-Tech 3158 instruction manual

Scientific Atlanta

Scientific Atlanta Quan-Tech 3158 instruction manual

FLUS ET-951W user manual

FLUS

FLUS ET-951W user manual

Kimo VT110 user manual

Kimo

Kimo VT110 user manual

Fluke Biomedical RaySafe Xi 8201011 user manual

Fluke Biomedical

Fluke Biomedical RaySafe Xi 8201011 user manual

HW Group Poseidon2 3266 user manual

HW Group

HW Group Poseidon2 3266 user manual

Stanley IntelliMeasure 0-77-018 user manual

Stanley

Stanley IntelliMeasure 0-77-018 user manual

Amprobe AT-2005-A quick guide

Amprobe

Amprobe AT-2005-A quick guide

Xylem OMNI sensus C2 user guide

Xylem

Xylem OMNI sensus C2 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.