manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. MKS
  6. •
  7. Measuring Instrument
  8. •
  9. MKS Ophir Nova User manual

MKS Ophir Nova User manual

To use the Nova to measure laser energy and frequency:
1. In measurement mode, verify that the units are mJ, µJ etc. If not, press
the left most soft key.
2. Press the menu button once then press “range” then the “up” “down”
soft keys until the proper range is highlighted. The correct range is the
lowest one that is larger than the pulse energy measured. See section
6.4 of the main manual for details.
3. Press “exit” then press the menu button twice to return to the bargraph
measurement screen. If you wish to have the frequency displayed,
press the menu button only once.
To use the Nova to measure laser power:
1. Verify that the display shows power units in W, mW etc.
2. If the display shows energy units of J, mJ etc. then press “power” to
switch to the power measurement mode.
To expand the bargraph scale fivefold about the present reading:
1. From the bargraph energy or power measurement screen press the
menu button then press “zoom”, then press the menu button twice, or
just press the center button from the bargraph screen.
2. Press “zoom” again to return to full scale.
To measure total energy exposure:
1. Press the menu button until “more” appears. Press “more”.
2. Press “select” until “exposure” is highlighted. Press “go”.
3. Select parameters then press “go”. Press go again. The Nova will
start summing laser energy exposure and the legend will change to
“stop”. When you wish to stop measuring, press “stop”. If you wish to
reset reading to zero before another reading, press “reset”.
4. To return to the bargraph screen, press “exit”.
1Getting Started
The Nova is equipped with “soft keys.” That is, the functions of
the keys change as indicated by the legend above each key.
When the Nova is first switched on with a sensor attached, the
first screen usually has a digital display with a bargraph at the
bottom. In order to access the soft keys, press the menu button,
located in the second row on the right of the panel. Pressing
the menu button again will access more functions. Pressing it
yet again will bring it back in a cyclical manner to the original
bargraph screen.
To connect sensor to the Nova meter:
1. Insert the D type connector of the measuring sensor cable
into the socket marked “Head Input” on the rear panel of the
Nova meter.
To switch the Nova on:
1. Push up the slide switch on the left side of the Nova meter.
The unit will switch on, and the display will appear. Note that
the sensor must be plugged in before the unit is switched on.
To switch the Nova off:
1. Return the slide switch to its original position.
2. If you wish to save the current Nova configuration, use the
“configure” function before switching off. See below,
section 2 or refer to the main manual for details.
To set line frequency:
1. Disconnect the sensor and switch off then on again. “Head
Disconnected” will appear.
2. Press “select” until “line” is highlighted.
3. Press “go” then change to correct frequency.
4. Press “exit” and the change will be saved.
To zero instrument:
1. Disconnect sensor, turn off then on again. “Head
Disconnected” will be displayed.
2. Press “select” until “zero” is highlighted.
3. Press “go”.
4. Make sure instrument is not in an electrically noisy
environment and is undisturbed. Press “go” and wait for
message, “zeroing completed”. Press “exit”.
5. For thermal heads, zeroing with the head may also be
necessary. See section 3.5.2 in the full manual.
Nova Quick Reference
NOVA
Quick Reference
Warning:
Incorrect readings will result if pulse length is not set up correctly.
Note: Some sensors have only one time setting for all pulse lengths. In that
case “N.A.” appears.
4.2 Setting Startup Configuration
1. From the bargraph energy measurement mode, press the menu button
until “more” appears. Press “more”.
2. Press “select” until “config” is highlighted. Press “go”.
3. Press “value” to choose “power” or “energy” for the desired startup
screen.
4. Press “select” until the laser wavelength is highlighted. Now press
“value” to select the laser type you wish to be the default.
5. Press “select” and “value” to choose the pulse length you wish to be
the default.
6. Now press “select” and “value” again to choose the default energy
range.
7. Press “exit” then “all” to save. For further details on configuration see
sections 6.2 and 6.3 of the main manual.
4.3 Setting up PE-DIF Diffuser Sensors to Diffuser IN or OUT Setting
1. Press the menu button twice and press “setup”.
2. Press “select” until the diffuser setting is highlighted and select “IN”
or “OUT” as desired. Make sure the diffuser is physically installed or
not installed on the sensor. Press exit.
3. Press the menu button twice and press “laser” until the correct
wavelength is selected. Note that only wavelengths compatible with
the diffuser setting are visible. (All wavelengths for diffuser IN have a D
suffix, e.g. 106D = 1.06µm, diffuser in).
4. If you wish to save these settings as the defaults, from the main
display, press the menu button twice then press “more”. Select
“config” and press “go”. Press “exit” and “all” to save all present
settings including diffuser IN or OUT.
4.4 Energy or Average Power Measurement
Warning:
Do not exceed maximum sensor limits for power, energy, power density
and energy density as listed in tables 6 and 7 section 9.2 of the main
manual. Otherwise, there is a risk of damaging the absorber. Use the
supplied test slide to test damage if you are in doubt. If the slide is
damaged, then either enlarge your beam or lower the laser energy until
damage is no longer seen.
4.5 To Choose Energy or Average Power Measurement
1. To go from the bargraph energy measurement screen to power
measurement, press the menu button then press “power”. Alternately,
you can press the leftmost soft key directly from the bargraph screen
to go to power measurement.
2. To go from the bargraph power measurement screen to energy
measurement, press the menu button then press “energy”.
Alternately, you can press the leftmost soft key directly from the
bargraph screen to go to energy measurement. Ophir Optronics Solutions LTD.
Science Based Industry Park P.O.B. 45021, Jerusalem 9145001, Israel
Phone +972-2-5484444
https://www.ophiropt.com/photonics
rev 3.27-3/17.Feb.19/be Ophir P/N 1J06014
2Thermal Sensors
2.1 Use of Nova with Thermal Type Sensors
To set type of laser being used:
1. While the Nova is off, plug in the sensor then switch it on again.
2. From bargraph measurement screen, press the menu button twice and
press “laser” until the correct laser type is displayed.
3. Return to bargraph screen by pressing the menu button again.
To choose manual or automatic ranging in power measurement:
1. From the bargraph measurement screen, press the menu button once,
then press “range” or simply press the right most soft key directly from
the bargraph screen.
2. Select the appropriate manual range or autorange.
3. Press “exit” and then press the menu button twice to return to the
bargraph measurement screen.
To choose power or energy measurement:
1. To go from the bargraph power measurement screen to the energy
measurement screen, press the menu button then press “energy”.
2. Alternatively, you can press the leftmost soft key directly from the
bargraph screen to go to energy measurement.
3. To go from the energy measurement screen to power measurement
press the “power” soft key.
2.2 Setting and Saving Startup Configuration
1. From the power measurement mode, press the menu button located on
the right side until “more” appears. Press “more”.
2. Press “select” until “configure” is highlighted. Press “go”.
3. Press “value” to choose “power” or “energy” for the desired startup
screen.
4. Press “select” until the laser type is high-lighted. Now press “value” to
select the laser type you wish to be the default.
5. Press “select” and “value” to choose the manual power range you wish
to be the default or choose autorange.
6. Now press “select” and “value” again to choose the default energy
range.
7. Press “exit” then “all” to save all present settings.
Further details on configuration in section 4.3. of main manual.
2.3 Power or Single Shot Energy Measurement
Warning:
Do not exceed maximum sensor limits for power, energy, power density and
energy density as listed in tables 6 and 7 in section 9.2. in the main manual.
Otherwise, there is a risk of damaging the absorber.
To use the Nova to measure laser power:
1. Verify that the display shows power units in W, mW etc.
2. If the display shows energy units of J, mJ etc. then press “power” to
switch to the power measurement mode.
To expand the bargraph scale ±10% about the present reading:
1. From the bargraph power measurement screen press the center
button.
2. Press the center “zoom” button again to return to full scale.
To subtract background and set current reading to zero:
1. From the bargraph power measurement screen press the menu button
twice then press “offset”. Press the menu button once to return to the
bargraph screen.
2. Press “offset” again to cancel. See section 4.4.2.2. of the main manual
for full details.
To use the Nova to fine tune laser power:
1. From the bargraph power measurement screen press the menu button
twice then press “more”.
2. Press “select” until “tune” is highlighted. Press “go”.
3. Set the percentage range of the power scale to be displayed by
repeatedly pressing the left key.
4. Set the horizontal sweep time using the middle soft key. See section
4.4.2.3. of the main manual for full details.
To use the Nova to measure laser energy:
1. In measurement mode, verify that the units are J, mJ etc. If not, press
the menu key until “energy” appears. Press that key to switch to energy
measurement mode. Alternately, press the left soft key directly.
2. The energy mode is manual ranging. Press “range” then the “up”
“down” soft keys until the proper range is highlighted then press “exit”.
The correct range is the lowest one that is larger than the pulse energy
measured.
3. When the Nova screen flashes “ready,” on and off, fire the laser.
See section 4.5 of the main manual for full details.
To save configuration:
To save configuration, follow directions in section 2 above.
3Photodiode Sensors
3.1 Use of Nova With Photodiode Type Sensors
To set type of laser being used:
1. From bargraph measurement screen, press the menu button twice and
press “laser” until the correct laser wavelength is displayed.
2. Return to bargraph screen by pressing the menu button again.
To choose manual or automatic ranging or dBm in power measurement:
1. From the bargraph measurement screen, press the menu button once.
2. Press “range” and then select the appropriate manual range, autorange
or dBm (logarithmic scale).
3. Press “exit” and then press the menu button twice to return to the
bargraph measurement screen. Alternately, range is accessed directly
by pressing the right most soft key from the bargraph screen.
3.2 Selecting Chosen Wavelengths
1. From the power measurement mode, press the menu button until
“more” appears. Press “more”.
2. Press “select” until “wavelength” is highlighted. Press “go”.
3. Press “change” then “up” and “down” to select the first wavelength.
Repeat steps 2 and 3 for other wavelengths desired. Up to 6
wavelengths may be selected.
3.3 Setting Startup Configurations
1. From the power measurement mode, press the menu button until
“more” appears. Press “more”.
2. Press “select” until “configure” is highlighted. Press “go”.
3. Now press “value” to select filter in or out to be the default.
4. Press “select” and “value” to choose the manual power range you
wish to be the default or choose autorange.
5. Now press “select” and “value” again to choose the startup laser
wavelength. For further details on configuration see section 5.3. of the
main manual.
3.4 Power Measurement
Warning:
Do not exceed maximum sensor limits for power, energy, power density
and energy density as listed in tables 6 and 7 section 9.2 of the main
manual. Otherwise, there is a risk of damaging the absorber.
To expand the bargraph scale 10% about the present reading:
1. From the bargraph power measurement screen press the menu
button. Then press “zoom”. Now press the menu button twice
again. Alternatively press the center button from the bargraph screen
without the menu button.
2. Press “zoom” again to return to full scale.
To offset current reading and set to zero:
1. From the bargraph power measurement screen press the menu button,
press “offset” then press menu twice to return to the bargraph screen.
Alternatively press the left upper button from the bargraph screen.
2. Press “offset” again to cancel. See section 5.4.5. of the main manual
for more details.
To use the Nova to fine tune laser power:
1. From the bargraph power measurement screen press the menu button
twice, then press “more”.
2. Press “select” until “tune” is highlighted. Press “go”.
3. With the left soft key set the percentage range of the power scale to be
displayed.
4. Set the horizontal sweep time using the middle soft key. See section
4.4.2.3. of the main manual for full details.
4
Pyroelectric or Photodiode Energy Sensors
4.1 Selecting Chosen Wavelengths
1. While the instrument is switched off, plug in the sensor, then switch on.
2. From the bargraph screen which appears, press the menu button until
“more” appears. Press “more”.
3. Press “select” until “wavelength” is highlighted. Press “go”.
4. Press “select” then “change” then “up” and “down” to change the
first wavelength. Now press “done”. Repeat steps 2 and 3 for other
wavelengths desired up to 6.
5. Save as described in section 2.4.4. of the main manual.
To set type of laser being used:
1. From bargraph measurement screen, press the menu button twice and
press “laser” until the correct laser type or wavelength is displayed.
2. Return to bargraph screen by pressing the menu button again.
To set laser pulse length and average:
1. From the bargraph measurement screen press the menu button twice
then press “setup”.
2. Press “value” to choose 50µs or 1ms depending whether the pulses of
your laser are less than or greater than 50 micro-seconds.
3. Press “select” then “value” to choose the time period over which
you wish to average energy readings. Press “exit”.
4. Save as described in section 2.4.4. of the main manual.

Other manuals for Ophir Nova

1

Other MKS Measuring Instrument manuals

MKS RGA Series User manual

MKS

MKS RGA Series User manual

MKS Spectra-Physics User manual

MKS

MKS Spectra-Physics User manual

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS Ion Systems 775 User manual

MKS

MKS Ion Systems 775 User manual

MKS HPQ2 User manual

MKS

MKS HPQ2 User manual

MKS Ophir LaserStar User manual

MKS

MKS Ophir LaserStar User manual

MKS Vision 2000-C User manual

MKS

MKS Vision 2000-C User manual

MKS Ophir IPM User manual

MKS

MKS Ophir IPM User manual

MKS Baratron 622D User manual

MKS

MKS Baratron 622D User manual

MKS Spectra-Physics HighQ-2 User manual

MKS

MKS Spectra-Physics HighQ-2 User manual

MKS Novx 5000 Series User manual

MKS

MKS Novx 5000 Series User manual

MKS 146C User manual

MKS

MKS 146C User manual

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS Ophir NOVA II User manual

MKS

MKS Ophir NOVA II User manual

MKS Ophir Spiricon BeamWatch Standard Quick start guide

MKS

MKS Ophir Spiricon BeamWatch Standard Quick start guide

MKS e-Vision+ User manual

MKS

MKS e-Vision+ User manual

MKS Microvision2 User manual

MKS

MKS Microvision2 User manual

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS Ophir Centauri User manual

MKS

MKS Ophir Centauri User manual

MKS DX 345 User manual

MKS

MKS DX 345 User manual

MKS Ophir Nova User manual

MKS

MKS Ophir Nova User manual

MKS Baratron DMA Series User manual

MKS

MKS Baratron DMA Series User manual

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS

MKS Granville-Phillips Mini-Convectron 275... User manual

MKS Ophir BeamWatch AM Quick start guide

MKS

MKS Ophir BeamWatch AM Quick start guide

Popular Measuring Instrument manuals by other brands

TPI 706 manual

TPI

TPI 706 manual

Badger Meter Blancett B2800 Standard Programming manual

Badger Meter

Badger Meter Blancett B2800 Standard Programming manual

OWL +USB user manual

OWL

OWL +USB user manual

Digatron DT- 51K instructions

Digatron

Digatron DT- 51K instructions

Ruskin RRS ROOFTOP installation instructions

Ruskin

Ruskin RRS ROOFTOP installation instructions

Dräger X-node Technical manual

Dräger

Dräger X-node Technical manual

Esam E1200 user guide

Esam

Esam E1200 user guide

PEWA WIKA CPH6300 operating instructions

PEWA

PEWA WIKA CPH6300 operating instructions

Metrohm 7010 TOC user manual

Metrohm

Metrohm 7010 TOC user manual

Woodley Skyla VB1 Quick reference guide

Woodley

Woodley Skyla VB1 Quick reference guide

Eaton Power Xpert PXM 4000 User and installation manual

Eaton

Eaton Power Xpert PXM 4000 User and installation manual

FLIR Extech 380941 user guide

FLIR

FLIR Extech 380941 user guide

Hanna Instruments HI98161 instruction manual

Hanna Instruments

Hanna Instruments HI98161 instruction manual

Pfeiffer Vacuum GSD 350 OMNISTAR operating instructions

Pfeiffer Vacuum

Pfeiffer Vacuum GSD 350 OMNISTAR operating instructions

YOKOGAWA PZ4000 user manual

YOKOGAWA

YOKOGAWA PZ4000 user manual

horiba F-70 Series instruction manual

horiba

horiba F-70 Series instruction manual

Sealey TA102 instructions

Sealey

Sealey TA102 instructions

RaycoWylie i4000 Calibration instructions

RaycoWylie

RaycoWylie i4000 Calibration instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.