WARNING
Electric shock hazard
Voltage and current can cause electric shock.
Only skilled, authorized people may carry out maintenance work.
►Insulate and lock the power supply circuit by setting the switch to O.
►Disconnect the power supply cable from all power sources before working on the product and/or
removing the covers.
WARNING
Risk of electric shock in case of contact with products that are not electrically isolated
When powering off _mains switch to O_, certain components located between the mains connection
and the switch will still contain an electric charge (live). There is a risk of electric shock in case of
contact.
►Make sure that the mains connection is always visible and accessible so that it can be unplugged
at any time.
►Disconnect the mains cable from the electrical network before working on the product.
►Wait 5 minutes after power-off before working on the product and/or removing the cover(s).
WARNING
Health risk related to residual traces on the parts tested
A leak detection operation must be carried out under environmental conditions that do not present
any risks to the operator and the equipment. The user and/or integrator of the product are fully re-
sponsible for the operational safety conditions of the equipment.
►Do not test parts or equipment with traces of harsh, chemical, corrosive, inflammable, reactive,
toxic, or explosive substances, nor condensable vapors, even in small amounts.
►Apply the relevant safety instructions in accordance with local regulations.
WARNING
Risk of serious injury due to falling objects
When transporting parts/components and during product maintenance, there is a danger of injury
caused by loads slipping or falling.
►Carry small and medium-size components with both hands.
►Transport components that weigh more than 20 kg using suitable lifting gear.
►Wear safety shoes with steel toes in accordance with directive EN 347.
WARNING
Risk of burns in case of contact with hot surfaces
For the operator’s safety, the products are designed to avoid thermal risk. However, specific operat-
ing conditions may exist that require extra caution on the part of the operator due to the high temper-
atures (surfaces > 70 °C for parts inside the cover(s)).
►Wait for the product to fully cool down before working on it.
►Protective gloves must be worn in accordance with standard EN ISO 21420.
WARNING
Risk of injury from contact with pressurized neutral gas
The product uses a pressurized inert gas (e.g. nitrogen) as the purge gas. Installations that do not
comply with or are not configured according to professional standards can endanger the user's life.
►Install a manual valve on the circuit at a distance of 3 m from the product, so that the neutral gas
supply can be locked out.
►Observe the recommended supply pressure.
►Always lock out and disconnect the neutral gas circuit before working on the product.
►Regularly check the condition of the pipework and supply circuit connections.
Safety