operated under non-flow conditions. Specac advise fitting a pressure
relief valve (P/N GS24160), along with appropriate open/close valve tap
connections (P/N GS24161), or to monitor the pressure to prevent for
any over-pressurisation to the system. (Please check information
found on pages 28 to 32 for Section 7), of the Atmos gas cell
instruction manual 2I-24602-3.
Important Notes For Safety In Operation
Avoid burns! Some parts of the Atmos gas cell and
heating jacket will be hot and should not be handled or
touched with bare hands when the heating jacket is
operating and hot. (e.g. inlet and outlet tubing, the outer
insulating jacket (3) surface and the top section (20) of the optical unit.)
In normal operation at a 200°C setting, the temperature at the top
section (24) of the optical unit box would be about 50°C, the inlet and
outlet tubing would be at 200°C and the insulating jacket (3) would be
circa 85°C; for an ambient temperature maximum of 40°C.
Do not remove the Atmos heating jacket from the gas
cell when hot. If it is necessary to do so, then wear heat
protective gloves and make sure the 4000 series
temperature controller is switched off. Rest the heating
jacket assembly on a bench with none of the metal parts in contact with
the bench. (Use a heat protective surface such as layers of cloth.) Open
both the insulating jacket (3) and inner heating jacket (1) to allow for
faster cooling to return the Atmos heating jacket to room temperature
conditions.
Note: If the Atmos heating jacket is removed from the gas cell during
the experiment for a period of more than 5 minutes, turn the 4000
series temperature controller power switch to off.
The Atmos heating jacket is fitted with a thermal fuse located in the inner
metal heating jacket assembly (1). Should there be a failure of the
thermocouple or if the heating jacket is misused, overheating will occur.