User Manual
6
the aperture (2), and by rotation of the graduated wheel (3), the correct
angle of polarized light from the polarized grid orientation laid onto the
substrate material of the polarizer itself, can be read from the angle
viewing window (4).
The front of the polarizer rotator 3” x 2” slide mount (1) carries the two
groove slots (5) that allow for the mounting of an FTIR accessory that
carries a 3” x 2” slide mounting plate itself for installation into a
spectrometer. The 3” x 2” slide mount plate of the Accessory simply
slides into the two grooves (5) and in this way a polarizer can be
introduced into any beam path of light to or from the Accessory (and a
sample) to discriminate for polarized light.
When a 3” x 2” slide mounted Accessory is installed for use into the
polarizer rotator 3” x 2” slide mount (1) using the groove slots (5), the
angle viewing window (4) will be obscured from reading a specific
angle that may be set on the GS12000 Series polarizer itself. However,
there is an additional marked notch (6) on the side of the rotator mount
(1) that can be used to align to any of the angle marks on the
graduated wheel (3) for reading if required.
Note: he angle reading from alignment of a graduated mark on the
wheel (3) with the notch (6) will always be a difference of 90° to
the angle reading as displayed at the viewing window (4).
In operation, as shown from the rear side view (Fig 2 ), the polarizer
rotator 3” x 2” slide mount (1) (ideally with a GS12000 polarizer already
fitted) is installed into the 3” x 2” slide mount plate of a spectrometers
mount system using the slide mount runners (7) at the side edges of
the rear plate assembly. Also, at the rear, there is a tightening thumb
screw (8). This can be used to secure a 3” x 2” slide mounted
Accessory when attached to the rotator mount (1) via the groove slots
(5) as described. Turning the screw (8) clockwise will apply pressure to
the 3” x 2” slide plate of the Accessory forcing the plate overall against
the groove slots (5) for a tighter fit with no lateral movement of the
Accessory. The screw (8) needs to be removed completely if the 3” x
2” slide mount plate system of the spectrometer does not allow for a
correct beam height fit of the polarizer rotator 3” x 2” slide mount (1)
when the screw (8) is in place.