Uniks ISO-5 User manual

Operating Manual
ISO-5
5 kV Digital Insulation
Tester
ISO-10
10 kV Digital Insulation
Tester
d
i
g
i
t
a
l

2

3
CHAPTER 1 GENERAL ASPECTS........................................................................ 5
1.1
Features of the meter
................................................................................... 5
CHAPTER 2...................................................................................................... 7
SYMBOLS ON THE INSTRUMENT:....................................................................... 8
CHAPTER 3...................................................................................................... 9
3.2
BASIC INDICATIONS AND THEIR MEANING .......................................... 10
CHAPTER 4.................................................................................................... 11
CHAPTER 5 MEASUREMENT MODES OF THE METER........................................ 13
5.1
IR Mode: ................................................................................................. 13
5.2
DAR Mode: .............................................................................................. 15
5.3
PI Mode:.................................................................................................. 18
5.4
Step Mode: .............................................................................................. 21
5.5
Ramp
Mode:............................................................................................. 25
5.6
Dielectric
Discharge
Mode:
......................................................................... 28
5.7
Voltage Mode: .......................................................................................... 32
CHAPTER 6 GRAPH WINDOW ........................................................................ 33
CHAPTER 7 FILE EXPLORER........................................................................... 36
CHAPTER 8.................................................................................................... 38
CHAPTER 9 SETTINGS OF THE METER............................................................ 39
9.1
SLEEP Mode Settings: ............................................................................... 39
9.2
Time/Date Settings: .................................................................................. 40
9.3
Customer Information Settings: ................................................................. 41

4
9.4
Brightness Settings:.................................................................................. 42
9.5
Short Circuit Settings:............................................................................... 43
9.6
Go No Go Settings:................................................................................... 44
9.7
Other Settings: ........................................................................................ 45
CHAPTER 10 METER SPECIFICATIONS........................................................... 47
CHAPTER 11 .................................................................................................. 51
11.1
Battery charging process: ................................................................... 51
11.2
Battery alarms: ................................................................................. 51
CHAPTER 12 .................................................................................................. 52
12.1
TRANSFORMERS: ................................................................................... 52
12.2
CIRCUIT BREAKERS:............................................................................... 53
12.3
CABLES: ............................................................................................... 53
12.4
CURRENT TRANSFORMERS: ..................................................................... 54
12.1
MOTORS: .............................................................................................. 54
CHAPTER 13 .................................................................................................. 52
CHAPTER 14 .................................................................................................. 54
CHAPTER 15 .................................................................................................. 58

5
CHAPTER 1
GENERAL ASPECTS
1.1
Features of the meter
⚫
Selectable Test Voltage up to 5000 V (ISO 5 model) o 10.000V (ISO 10
model): Test voltages can be set to any desired value from 100V to 5000V ISO 5
Model o 100V to 10.000V ISO 10 model with steps of 10V up to 1000V and steps of 25V
above 1000V.
⚫
High Insulation resistance measurement: Insulation resistance measurement from
50kΩ to 10TΩ (ISO 5) o 20TΩ (ISO 10) .
⚫
Noise rejection - 8mA:
High Noise immunity allows accurate measurement under noisy
environment.
⚫
Polarization Index: Polarization index (PI) testing is an extension of the insulation
resistance test and is designed to check specific issues such as moisture and insulation
deterioration.
⚫
Dielectric Absorption Ratio: DAR measurement is a diagnostic test similar to the
Polarization Index (PI), but DAR takes the ratio of the Insulation Resistance usually
measured at 30 sec and 1 min.
⚫
Dielectric Discharge Test: Dielectric Discharge(DD) test is a diagnostic insulation test
that allows aging and deterioration to be assessed.
⚫
Step Voltage Test: Step Voltage Test is designed as a controlled over-voltage or proof
test to provide an additional evaluation of the insulation system integrity.
⚫
Ramp Diagnostic Test:
The Ramp Test is performed with a slowly rising voltage.
⚫
Software Selectable Filter: User selectable software filter can be used depending on
noise levels. This help in accurate reading in noisy environments.
⚫
Capacitance
Measurement:
Capacitance
can
be
measured
from
1nf
to
50uF.

6
⚫
Go- NoGo Function: User can set an insulation resistance limit below which user will be
alarmed about the quality of insulation.
⚫
Insulation Graph Display: Insulation Resistance trend can be displayed over the time in
graphical form.
⚫
Communication Interface: Bluetooth 2.0 & 4.0 Class II and electrically isolated USB 2.0
communication.
⚫
Software and Mobile Application: It has data logging and monitoring software for
window system and an interactive mobile application for android.
⚫
File Explorer: It has file explorer to view the datalog file on TFT display.
⚫
Audio
Read-Out:
Audible
test
result
on
completion
of
test
for
awkward
locations.
⚫
Rechargeable
Battery:
Li-Ion
3
Cell
Battery
with
charging
current
of
1A.
⚫
Internal Data logging: It has internal data logging feature. It can log up to 2000 test
files with customer information and time stamp.
⚫
Selectable test time: User can set the test time to any desired test time from 45 sec to
99 min 59 sec.
The home screen of the meter is as shown below:

7
CHAPTER 2
USER AND BATTERY SAFETY
USER SAFETY:
Following
safety
measures
should
be
taken
while
using
the
instrument:
The circuit under test should be turned off /denergized(dead), all its terminals should be
isolated, the circuit should also be discharged and checked before the insulation is tested.
During the test please make sure that you do not touch any of the test leads. The caution
symbol on the measurement screen indicates that hazardous voltage is present at the
test leads.
After test completion you should make sure that the load is completely discharged
because this charge can be dangerous.
The voltage indicator and internal discharge should be treated like additional safety
devices and not a substitute for the general practices.
Insulation testing in wet conditions can be hazardous. It is recommended that the
instrument should not be used under these circumstances.
During control from bluetooth or USB the test can be started at any time. Please make
sure the connections are proper before turning on the test.
Please use only the leads supplied with the meter for testing the device.
NOTE: INSTRUMENT SHOULD BE OPERATED BY SUITABLY TRAINED AND
COMPETENT PERSONS ONLY
BATTERY SAFETY:
Following
safety
measures
should
be
followed
while
handling
the
battery:
Do not disassemble or change the battery. This battery consists of safety and protection
devices which if damaged can cause harm to the battery.
Do not heat the battery in any way.
Do not pierce the battery.
Do not expose the battery to water.
Do not subject the battery to shocks.

8
Never
short
circuit
the
battery
pack.
In case the battery starts leaking, do not allow the liquid to come in contact with skin or
eyes. If by mistake you make contact, wash the affected area with water and contact
doctor.
Keep the battery away from children.
Do not expose the battery to extreme temperatures. Extreme temperature will reduce
the lifetime of the lithium ion battery.
SYMBOLS ON THE INSTRUMENT:
Caution: Please refer user manual
Caution: Risk of electrical shock
Equipment complies with current EU directives
Equipment protected throughout double insulation
Bluetooth
Do not dispose in normal waste
Audio Readout
Mains connected
Universal Serial Bus (USB)
600V
CAT IV
Overvoltage category IV (equipment installed at or near the
origin of the electrical supply to a building) & 600V refers to
the rms phase to earth voltage that instruments can
withstand to the overvoltage category IV rating

9
5
CHAPTER 3
BASIC COMPONENTS AND INDICATIONS IN THE METER
7
3.1
LEGENDS AND THEIR MEANING
1. Negative(-) terminal
2. Guard(G) terminal
3. Positive(+) terminal
4. LED indicating presence of mains
5. USB slot for device connection
4
6. Speaker
7. Protective cover
2 8. TFT touch screen display
3 6 9. Power supply socket
10. Fuse
11. On/Off Switch
8
910 11
1

10
3.2
BASIC INDICATIONS AND THEIR MEANING
Mains
connected
Breakdown mode
Guard fuse blown
Battery
charging
Burn mode
USB mode
Bluetooth mode
Pre-alarm for
battery charging
Danger High Voltage at
leads
Speaker mute
Speaker unmute
Current limit warning
Battery fully
charged indication

11
CHAPTER 4
CONNECTION OF TEST LEADS TO METER
The meter has 3 probes guard, positive and negative. Each probe consists of 2 ends.
One of these ends is the probe handle which has to be connected to the meter and the other
end will depend upon the type of accessories ordered. Figure below shows the probe handle
for the single probes(Guard and positive) and double probe which is connected to the
negative terminal. Both the handles have a rotor for locking the probe to the meter
Probe rotor
Single probe handle Double probe handle
NOTE: Insulation of the DUT should be fully discharged before connecting the leads
to the DUT.
Due to the locking mechanism of the probes the probes remain firmly connected to the
meter during the test. Following are the steps to connect the probe to the meter:
⚫
Insert the probes into the meter as shown in the below diagram:
Negative
Guard
Positive

12
⚫Then push the probes into the meter and rotate the probe rotor as shown below so that
the probe gets locked to the meter.
NOTE: Do not disconnect the test leads of the meter till the test has been completed
and the DUT has been discharged as the leads can cause electrical hazard to the
user when the test is being conducted or if the DUT is still charged.

13
CHAPTER 5
MEASUREMENT MODES OF THE METER
5.1
IR Mode:
This mode is used to measure the insulation resistance of a device for a fixed amount of
time. The default time of this test is 45 sec. The steps to do a test in the IR mode are as follows
⚫
To enter the IR mode press icon on the home screen. The main measurement
screen will appear.
⚫
To setup the basic settings of the mode press the icon. The control window will
appear as shown below.
⚫
Here you can set the voltage using the voltage up and down in steps of 10V below 1000V
and in steps of 25V above 1000V. There is also a standard voltage button to set standard
testing voltages (100V, 250V, 500V, 1000V, 2500V, 5000V).
min sec
button button
Software filter
button

14
⚫
You can also change the filter settings in this window (hardware and software filter).
These filters are very useful in noisy environments. The software filter can be selected
from the drop down list which will appear after pressing the software filter button.
⚫
You can also set the test time in this mode by pressing the second or minute button. The
keyboard will appear and then you can enter the desired time between 45 sec and 99 min
59 sec and then press the button on the keyboard to set the time.
⚫
After setting up all the basic settings press the button to go back to the main
measurement screen. All these control settings are stored by the meter so that the next
time you start the meter you do not need to do these settings again.
⚫
Then connect the test leads to the equipment whose Insulation Resistance has to be
measured.
⚫
On the measurement screen press and hold the button for about 3 seconds and
then release it to start the test.
⚫
The speaker will give 3 beeps in order to indicate that the test has started. After the test
has been completed all the parameters will be displayed(according to the meter setting)
like insulation resistance, capacitance, DAR, PI, voltage, current as shown below.

15
⚫
Because there are a lot of parameters to be displayed they have been divided into three
sub screens which can be scrolled using the scrolling arrows. Figure shows the sequence
of the screens along with the parameters displayed on the corresponding screen.
SCREEN 1
VOLTAGE
AT LEADS
/DUT
SCREEN 2
VOLTAGE
DURING
TEST
SCREEN 3
SET
VOLTAGE
DAR
INFORMATION
PI
INFORMATION
⚫
You can save the data of this test by pressing the
button. After pressing this button
a message will appear to tell you about the file number. You can press the OK button on
the message box in order to go back to the measurement screen.
5.2
DAR Mode:
DAR(Dielectric Absorption Ratio) is defined as the ratio of insulation resistance at 1
minute divided by insulation resistance at 30 seconds. It is very useful to know the condition
of the insulation. It can be expressed in formula as:

16
Software filter
button
DAR = IR60s / Ir30s
DAR of less than 1 indicates that the insulation is in poor condition, DAR of 1-1.4
indicates that the insulation is in good condition and DAR of greater than 1.4 indicates that
the insulation is in an excellent condition. The steps to take a test in DAR mode are as follows:
⚫
To enter the DAR mode press the icon on the home screen and following screen will
be displayed.
⚫
To setup the basic settings of the mode press the icon. The control window will
appear as shown below.

17
⚫
Here you can set the voltage using the voltage up and down in steps of 10V below 1000V
and in steps of 25V above 1000V. There is also a standard voltage button to set standard
testing voltages (100V, 250V, 500V, 1000V, 2500V, 5000V ISO 5 o 10.000V ISO 10).
⚫
You can also change the filter settings in this window (hardware and software filter).
These filters are very useful in noisy environments. The software filter can be selected
from the drop down list which will appear after pressing the software filter button.
⚫
As we know DAR tests are measurements of resistance over time expressed as a ratio
of resistance at time t2 divided by resistance at time t1. The t1 and t2 both can be set in
the control window with the help of their corresponding arrows. The up arrow will
increase the time and the down arrow will decrease the time.
⚫
After setting up all the basic settings press the button to go back to the main
measurement screen. All these control settings are stored by the meter so that the next
time you start the meter you do not need to do these settings again.
⚫
Then connect the test leads to the equipment whose Insulation Resistance has to be
measured.
⚫
On the measurement screen press and hold the button for about 3 seconds and
then release it to start the test.
⚫
The speaker will give 3 beeps in order to indicate that the test has started. After the test
has been completed all the parameters will be displayed(according to the meter setting)
like insulation resistance, capacitance, DAR, PI, voltage, current.

18
⚫
Because there are a lot of parameters to be displayed they have been divided into three
sub screens which can be scrolled using the scrolling arrows. Figure shows the sequence
of the screens along with the parameters displayed on the corresponding screen.
SCREEN 1
VOLTAGE
AT LEADS
/DUT
SCREEN 2
VOLTAGE
DURING
TEST
SCREEN 3
SET
VOLTAGE
DAR
INFORMATION
PI
INFORMATION
⚫
You can save the data of this test by pressing the
button. After pressing this button
a message will appear to tell you about the file number. You can press the OK button on
the message box in order to go back to the measurement screen.
5.3
PI Mode:
PI(Polarization Index) is defined as the ratio of insulation resistance at 10 minute
divided by insulation resistance at 1 minute. It is very useful to know the condition of the
insulation. It can be expressed in formula as:
PI = IR10min / IR1min

19
Software filter
button
PI of less than 1-2 indicates that the insulation is in poor condition, PI of 2-4 indicates
that the insulation is in good condition and PI of greater than 4 indicates that the insulation is
in an excellent condition. The steps to take a test in PI mode are as follows:
⚫
To enter the PI mode press the
icon on the home screen and the following screen will
be displayed:
⚫
To setup the basic settings of the mode press the icon. The control window will
appear as shown below.

20
⚫
Here you can set the voltage using the voltage up and down in steps of 10V below 1000V
and in steps of 25V above 1000V. There is also a standard voltage button to set standard
testing voltages (100V, 250V, 500V, 1000V, 2500V, 5000V ISO 5 o 10.000V ISO 10).
⚫
You can also change the filter settings in this window (hardware and software filter).
These filters are very useful in noisy environments. The software filter can be selected
from the drop down list which will appear after pressing the software filter button.
⚫
As we know PI tests are measurements of resistance over time expressed as a ratio of
resistance at time t2 divided by resistance at time t1. The t1 and t2 both can be set in the
control window with the help of their corresponding arrows. The up arrow will increase
the time and the down arrow will decrease the time.
⚫
After setting up all the basic settings press the button to go back to the main
measurement screen. All these control settings are stored by the meter so that the next
time you start the meter you do not need to do these settings again.
⚫
Then connect the test leads to the equipment whose Insulation Resistance has to be
measured.
⚫
On the measurement screen press and hold the button for about 3 seconds
and then release it to start the test.
⚫
The speaker will give 3 beeps in order to indicate that the test has started. After the test
has been completed all the parameters will be displayed(according to the meter setting)
like insulation resistance, capacitance, DAR, PI, voltage, current.
This manual suits for next models
1
Table of contents
Other Uniks Test Equipment manuals
Popular Test Equipment manuals by other brands

TX Testing Instruments
TX Testing Instruments PHB-3000 user manual

FLIR
FLIR Extech Instruments 382252 manual

Tektronix
Tektronix 2445B Operator's manual

Gefen
Gefen GTB-HD-SIGGEN user manual

BGS technic
BGS technic 40105 instruction manual

Meilhaus Electronic
Meilhaus Electronic MEphisto Scope UM202 quick start guide