WHALETEQ DFS360 User manual

WHALETEQ
DFS360 / DFS360D
User Manual
PC Software Version 1.0.0.28
Revision 2021-10-22

DFS360|User Manual
1
Copyright © 2013-2021, All Rights Reserved.
WhaleTeq Co. LTD
No part of this publication may be reproduced, transmitted, transcribed,
stored in a retrieval system, or translated into any language or computer
language, in any form, or by any means, electronic, mechanical, magnetic,
optical, chemical, manual or otherwise, without the prior written
permission of WhaleTeq Co. LTD.
Disclaimer
WhaleTeq Co. LTD. provides this document and the programs "as is"
without warranty of any kind, either expressed or implied, including, but
not limited to, the implied warranties of merchantability or fitness for a
particular purpose.
This document could contain technical inaccuracies or typographical errors.
Changes are periodically made to the information herein; these changes will
be incorporated in future revisions of this document. WhaleTeq Co. LTD. is
under no obligation to notify any person of the changes.
The following trademarks are used in this document:
is a registered trademark of WhaleTeq Co. LTD
All other trademarks or trade names are property of their respective
holders.

DFS360|User Manual
2
Contents
1Introduction ..................................................................................... 6
1.1 Concept ............................................................................................ 6
1.2 Applications...................................................................................... 6
2Specification..................................................................................... 7
2.1 General Specifications...................................................................... 7
2.2 Compliance Specifications................................................................ 7
2.3 Energy Measurement Specifications................................................ 8
2.4 ECG Signal Specifications.................................................................. 8
3Instrument Familiarization................................................................ 9
3.1 Top.................................................................................................... 9
3.2 Front............................................................................................... 10
3.3 Rear ................................................................................................ 10
3.4 Left and Right Side.......................................................................... 11
3.5 Wiring Diagram .............................................................................. 12
3.5.1 Connection Diagr am and Precautions....................................... 12
4Getting Started................................................................................14
4.1 PC System Requirements ............................................................... 14
4.2 Software Installation ...................................................................... 14
4.3 USB Driver Installation ................................................................... 15
4.4 Microsoft .Net Framework Installation.......................................... 16
4.5 PC Software and Firmware Update................................................ 16
5PC Software Operation ....................................................................17
5.1 Software Function Introduction..................................................... 17
5.1.1 Customize ECG Waveform.......................................................... 19
5.2 PC Software Operation Instructions .............................................. 23
5.3 Over Heat Warning and Function Protection ................................ 25
5.3.1 AP Display Status ........................................................................ 25
5.3.2 DFS Status LED ............................................................................ 26
5.3.3 Fan Behavior............................................................................... 26

DFS360|User Manual
3
5.3.4 Over Heat Protection.................................................................. 26
5.4 Optional Software Development Kit (SDK) .................................... 26
5.4.1 Command Function .................................................................... 26
6Standalone Operation......................................................................28
7Calibration and Validation ...............................................................28
8Troubleshooting ..............................................................................29
9Cautions ..........................................................................................29
10 Ordering Information.......................................................................30
11 Package Contents ............................................................................32
12 Revision History...............................................................................32
13 Contact WhaleTeq ...........................................................................33

DFS360|User Manual
4
List of Tables
Table 1: General Specifications ......................................................... 7
Table 2: Compliance Specifications ................................................... 7
Table 3: Energy Measurement Specifications ................................... 8
Table 4: ECG Signal Specifications..................................................... 8
Table 5: AP Interface Function Introduction ................................... 17
Table 6: Right-Click Function List Introduction................................ 19
Table 7: Customize ECG Waveform Function Introduction............. 20
Table 8: Result Page Function Introduction .................................... 24
Table 9: Command Function............................................................ 26
Table 10: Ordering info. for DFS360................................................ 30
Table 11: Ordering info. for DFS360D ............................................. 30
Table 12: Optional DFS360/DFS360D Paddle Box Set..................... 31
Table 13: Optional Software Add-On Pack...................................... 31
Table 14: Optional Calibration Service and Warranty Extension.... 31
Table 15: Package Contents for DFS360.......................................... 32
Table 16: Package Contents for DFS360D ....................................... 32
Table 17: Revision History ............................................................... 32

DFS360|User Manual
5
List of Figures
Figure 1: Top Side of DFS360 / DFS360D........................................... 9
Figure 2: Front Side of DFS360 / DFS360D ...................................... 10
Figure 3: Rear Side of DFS360 / DFS360D........................................ 10
Figure 4: Left and Right Side of DFS360 / DFS360D ........................ 11
Figure 5: Wiring Diagram................................................................. 12
Figure 6: Correct Wiring Diagram of the AED Cable........................ 12
Figure 7: Correct Wiring Diagram of the USB Cable........................ 13
Figure 8: AP Interface ...................................................................... 17
Figure 9: Right-Click Function List.................................................... 18
Figure 10: Customize ECG Waveform Interface.............................. 19
Figure 11: Complete Settings for Customized Waveform............... 23
Figure 12: ECG Patterns Conversion................................................ 23
Figure 13: Result Page ..................................................................... 24
Figure 14: High-Temperature Alert ................................................. 25
Figure 15: DFS360 / DFS360D Return to Normal Temperature ...... 25

DFS360|User Manual
6
1 Introduction
1.1 Concept
WhaleTeq DFS360 / DFS360D is a multifunction testing device,
functioning as an ECG signal simulator and energy measurement
equipment for the AED and defibrillator manufacturing environment.
For ECG signal simulation, DFS360 / DFS360D simulates NSR as well as
various arrhythmias that may or may not need AED defibrillation.
Arrhythmia convert combination (ECG waveform before vs. after
defibrillation) is also available.
For energy measurement calculation, DFS360 / DFS360D distinguishes
multiple kinds of biphasic waveforms and monophasic waveforms,
which are what the majority of current AEDs and defibrillators deliver.
Capturing the characteristic points of each kind of waveform, DFS360
/ DFS360D is also capable of displaying the real defibrillation
waveforms.
These features can be consolidated into the computer software of
AED or defibrillator manufacturers. What’s more, DFS360D is designed
to withstand more intense defibrillation shocks over longer stretches; it
takes shocks at 15-second intervals, meaning a boost of testing capacity. All
in all, DFS360 / DFS360D is a durable tester to speed up AED or
defibrillator automatic production.
1.2 Applications
WhaleTeq DFS360 / DFS360D is suitable for AED or defibrillator
manufacturing automation. With SDK provided, User can integrate
the functions of DFS360 / DFS360D into the manufacturing execution
system of the AED or defibrillator manufacturer’s choice. Test results
from DFS360 / DFS360D are thus combined into the production
history of each AED.

DFS360|User Manual
7
2 Specification
2.1 General Specifications
Table 1: General Specifications
Function
Specifications
Temperature
Operating: 0 °C to + 50 °C (+ 32 °F to + 122 °F)
Storage: -20 °C to + 60 °C (-4 °F to + 140 °F)
Humidity
10 % to 90 % non-condensing
Power
USB 3.0 Bus Power
Communications
USB Type C connector
Operation
Modes
Remote (PC Software via API and SDK)
Mechanical
Housing: ABS Plastic
Size (L x W x H): 19.0 x 12.5 x 6.0 (cm)
Weight: DFS360: 665g / DFS360D: 815g
2.2 Compliance Specifications
Table 2: Compliance Specifications
Function
Specifications
Safety
Standards
CE: IEC/EN61010-1:2010 + A1:2016; Pollution
degree 2
61010-2-030:2017
Electromagnetic
Compatibility
Standards (EMC)
CE EN61326-1
FCC: EMC P15B

DFS360|User Manual
8
2.3 Energy Measurement Specifications
Table 3: Energy Measurement Specifications
Function
Specifications
Maximum Energy
Up to 600 J
Accuracy
±1 % of reading ±1 J for up to 360J;
±1 % of reading ±1 J, typical for 360J ~ 600J
Load Resistance
Resistance: 50Ω
Accuracy: ±1 %, non-inductive (<2 μH)
Pulse Width
Range: 1.0 to 48.0 ms
Accuracy: ±0.05 ms
Voltage
Range: Up to 3000 V
Accuracy: 1 % ±1V for 450V ~ 3000V & -450V ~ -3000V;
1 % ±2V for -450V ~ 450V
Sample rate
250 kHz (4 μs sample)
Maximum Average
Power
Model 360: 12 W, for 1 defib. pulse of 360 J every 30 sec.
Model 360D: 24 W, for 1 defib. pulses of 360 J every 15 sec.
2.4 ECG Signal Specifications
Table 4: ECG Signal Specifications
Function
Specifications
ECG waves
Ventricular Fibrillation - Coarse
Vp-p = 3 mV
Ventricular Fibrillation - Fine
Vp-p = 0.5 mV
Ventricular Tachycardia
Vp-p = 3 mV, ≒180BPM
Ventricular Tachycardia Plus
Vp-p = 3 Mv,≒234 BPM
Atrial Fibrillation
Vp-p = 1.2 mV
Asystole
Vp-p = 0.03 mV
(Optional) User defined pattern: max. 100 seconds / up to 10 sets
(Max. 5K Sample Rate)
Note:
Amplitude low range: VF at maximum peak to peak of 200μV
Normal Sinus Rhythm (NSR): 30 to 240 (increment by 1) BPM
Amplitude range Vp-p 0.5 ~ 5.0mV, step 0.5 mV

DFS360|User Manual
9
3 Instrument Familiarization
3.1 Top
Figure 1: Top Side of DFS360 / DFS360D

DFS360|User Manual
10
3.2 Front
Figure 2: Front Side of DFS360 / DFS360D
A. AED Input: AED interface cable extends from here, connecting
device under test (AED or defibrillator) with DFS360 / DFS360D.
3.3 Rear
Figure 3: Rear Side of DFS360 / DFS360D
B. USB Port: Use to supply power. Also, it can be used to connect
to the computer for command sending and data transferring.
C. DFS Status LED: Indicate the working state of DFS360 / DFS360D .
Mode
Normal
Heat Alert
Standalone Operation
Green light blinks
every 2 seconds.
Red light blinks
every 2 seconds.
AP Connected (ECG
Playing)
AP Connected (ECG
NOT Playing)
Green light always
on.
Red light always
on.
A
B
C

DFS360|User Manual
11
3.4 Left and Right Side
Figure 4: Left and Right Side of DFS360 / DFS360D
D. Air Inlet: Allow ventilation during long hours of operation.
E. Air Outlet: Allow ventilation during long hours of operation.
D
E

DFS360|User Manual
12
3.5 Wiring Diagram
Figure 5: Wiring Diagram
3.5.1 Connection Diagram and Precautions
1. Connect the AED cable to DFS360 / DFS360D.
。Do not put the AED cable on DFS360 / DFS360D main console.
。The AED cable should not twist together.
Figure 6: Correct Wiring Diagram of the AED Cable

DFS360|User Manual
13
2. After connecting the AED cable, connect DFS360 / DFS360D and
PC/ Laptop with the USB cable.
。The USB cable and PC/ Laptop should not be placed near the
AED cable.
。The USB cable should not twist with the AED cable.
。The USB cable should not be placed on DFS360 / DFS360D main
console.
Figure 7: Correct Wiring Diagram of the USB Cable
3. Make sure the output voltage of the USB cable falls between
4.75V ~ 5.25V (The minimum input voltage that DFS360 /
DFS360D can work is 4.5V).
4. Open AP “WhaleTeq DFS360 Test System” on PC/ Laptop .

DFS360|User Manual
14
4 Getting Started
4.1 PC System Requirements
DFS360 software uses a Windows PC to connect and control DFS360 /
DFS360D via USB module.
PC requirements:
- Windows PC (Windows 7 or later, suggest to use the genuine
version)
- Microsoft .NET 4.0 or higher
- Administrator access (essential for installing software, driver, and
Microsoft .Net Framework)
- 1.5 GHz CPU or higher
- 1GB RAM or higher
1
- USB port
4.2 Software Installation
Please follow the steps below to download and execute DFS software.
1. Download DFS360 software from WhaleTeq’s website.
2. Open the file explorer and select the download location.
3. Double click on WhaleTeqDFS360.exe to execute DFS360 PC
software.
Note:
If DFS software can’t be executed properly, or this is the first time
using WhaleTeq’s product, please refer to the two sections below (4.3
and 4.4) to confirm that the USB driver and Microsoft .Net Framework
4.0 are both installed.
1
Relative to normal PC processing, there is no special use of PC speed. However, there has been
noted a slow increase in system RAM usage over long periods of time up to 30-40MB (related to MS
Windows “garbage collection”). PCs with only 512MB or less installed and are running several other
programs (in particular, Internet Explorer), may exceed the available RAM, requiring access to the
hard drive and dramatically impacting speed. In this case, streaming interruptions and other
problems may occur.

DFS360|User Manual
15
4.3 USB Driver Installation
If Windows Device Manager can’t recognize WhaleTeq’s product,
please follow the instructions below to install Microchip® USB Driver.
For Microsoft Windows 10
⚫As Windows 10 has built-in Microchip® USB Driver, there is no
need to install any drivers. It just takes a while for Windows
Device Manager to recognize and install the driver.
For Microsoft Windows 8 and Windows 8.1
⚫If Windows 8 and Windows 8.1 can’t recognize DFS360 /
DFS360D unit, please download “mchpcdc.inf” from WhaleTeq’s
website. This driver is provided by Microchip®, and is intended to
be used with PIC microprocessors which have built-in USB
functions.
⚫As mchpcdc.inf provided by Microchip® does not contain a digital
signature, please disable the driver signature enforcement in
Windows 8 and Windows 8.1. Please click here to watch the
tutorial video.
⚫When the USB module is connected for the first time, select
Manual Installation, and point to the folder containing the above
file. Then, continue to follow the instructions to finish the
installation. There may be a warning that the driver is not
recognized by Windows®, and this can be ignored. Please click
here to watch the tutorial video.
For Microsoft Windows 7
⚫If Windows 7 can’t recognize DFS360 / DFS360D unit, please
download “mchpcdc.inf” from WhaleTeq’s website. This driver is

DFS360|User Manual
16
provided by Microchip®, and is intended to be used with PIC
microprocessors which have built-in USB functions.
⚫When the USB module is connected for the first time, select
Manual Installation, and point to the folder containing the above
file. Then, continue to follow the instructions to finish the
installation. There may be a warning that the driver is not
recognized by Windows®, and this can be ignored. Please click
here to watch the tutorial video.
4.4 Microsoft .Net Framework Installation
WhaleTeq’s software is developed based on Microsoft .Net
Framework 4.0. If DFS software fails to execute properly, please check
whether Microsoft .Net Framework 4.0 or a higher version was
installed in the operating system.
If the PC hasn’t installed Microsoft .Net Framework 4.0 or a higher
version, please download it from Microsoft’s website. Please click
here to watch the tutorial video (from 2:17).
4.5 PC Software and Firmware Update
User can download the latest software from WhaleTeq’s website. If
User cannot update the DFS software, please contact WhaleTeq’s

DFS360|User Manual
17
5 PC Software Operation
5.1 Software Function Introduction
When opening DFS360 AP, User will see the figure below on the
screen.
Figure 8: AP Interface
Table 5: AP Interface Function Introduction
Item
Name
Description
A
ECG Pattern (Before
defib.)
To set the ECG pattern before the
defibrillation.
B
ECG Pattern (After
defib.)
To set the ECG pattern after the
defibrillation.
C
Mode
To switch the mode to adult or child.
Note:
To ensure the accuracy of the result, it
is recommended to use the child mode
if the defibrillation is under 100 joules
or 1.5kV.
D
Start button
To start the test.
A
B
C
D
E
F

DFS360|User Manual
18
Item
Name
Description
E
Energy Joules
To display the joules of the
defibrillation.
F
Icons
The Top Icon: The shortcut to the file
storage in the local computer.
The Middle Icon: To display the
position of the characteristic points on
the defibrillation waveform.
The Bottom Icon: Allow User to report
issues in case of software error.
Note:
。The BPM and amplitude can be changed for NSR.
。The provided ECG patterns are listed below:
Ventricular Fibrillation –Coarse Atrial Fibrillation
Ventricular Fibrillation –Fine Asystole
Ventricular Tachycardia NSR
Ventricular Tachycardia Plus (the faster VT)
Right-click the AP title bar and the functions below will show on the screen.
Figure 9: Right-Click Function List
G
H
J
K
I

DFS360|User Manual
19
Table 6: Right-Click Function List Introduction
Item
Name
Description
G
Check for Updates
To check whether the DFS360 AP is the
latest version.
H
Customize ECG
Waveform
Settings for customized ECG
waveforms. (For more information,
please look at sec. 5.1.1)
I
Help
Format information for customized
waveform is listed here.
J
Language
To switch the language of the DFS360
AP.
Note:
The languages include traditional
Chinese, simplified Chinese, and
English.
K
Shock Display
To change the display unit of the
parameters of the result.
5.1.1 Customize ECG Waveform
When purchasing this function, User will receive a set of activation
password. After inputting the password, User will see Figure 10
shown on the screen.
Figure 10: Customize ECG Waveform Interface
L
M
N
O
Other manuals for DFS360
2
This manual suits for next models
1
Table of contents
Other WHALETEQ Test Equipment manuals

WHALETEQ
WHALETEQ CMRR 3.0E User manual

WHALETEQ
WHALETEQ CMRR 2.0 User manual

WHALETEQ
WHALETEQ DFS360 User manual

WHALETEQ
WHALETEQ SECG 4.0 User manual

WHALETEQ
WHALETEQ DFS360 User manual

WHALETEQ
WHALETEQ SEEG 100E User manual

WHALETEQ
WHALETEQ CMRR 3.0 User manual

WHALETEQ
WHALETEQ WECG400 User manual

WHALETEQ
WHALETEQ WAIOT-N User manual

WHALETEQ
WHALETEQ DFS100 User manual