WIKA TR31-3 User manual

DE
EN
Operating instructions
Betriebsanleitung
Model TR31-3 Model TR31-K
OEM miniature resistance thermometer
Models TR31-3, TR31-K
OEM-Miniatur-Widerstandsthermometer
Typen TR31-3, TR31-K
70018194

2
14096788.07 03/2018 EN/DE
WIKA operating instructions models TR31-3, TR31-K
DE
EN
Operating instructions models TR31-3, TR31-K
Page 3 - 34
Betriebsanleitung Typen TR31-3, TR31-K
Seite 35 - 66
Further languages can be found at www.wika.com.
© 07/2014 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
3
Contents
Contents
Declarations of conformity can be found online at www.wika.com.
1. General information 4
2. Safety 6
3. Specications 10
4. Design and function 14
5. Transport, packaging and storage 17
6. Commissioning, operation 18
7. Conguration 24
8. CongurationsoftwareWIKAsoft-TT 25
9. Connecting PU-548 programming unit 28
10. Maintenance and cleaning 30
11. Faults 30
12. Dismounting, return and disposal 32
Appendix:CSAcontroldrawing 34

EN
14096788.07 03/2018 EN/DE
4
WIKA operating instructions models TR31-3, TR31-K
1. General information
■
The resistance thermometer described in the operating instructions
has been designed and manufactured using state-of-the-art
technology. All components are subject to stringent quality and
environmental criteria during production. Our management systems
are certied to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on
handling the instrument. Working safely requires that all safety
instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and
general safety regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept
in the immediate vicinity of the instrument and readily accessible to
skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the
operating instructions prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage caused
by using the product contrary to its intended use, non-compliance
with these operating instructions, assignment of insuciently qualied
skilled personnel or unauthorised modications to the instrument.
■
The general terms and conditions contained in the sales
documentation shall apply.
■
Subject to technical modications.
■
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: TE 60.31
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
1. General information

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
5
Explanationofsymbols
WARNING!
... indicates a potentially dangerous situation that can result
in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result
in light injuries or damage to equipment or the environment,
if not avoided.
Information
... points out useful tips, recommendations and information
for ecient and trouble-free operation.
DANGER!
... identies hazards caused by electrical power. Should the
safety instructions not be observed, there is a risk of serious
or fatal injury.
WARNING!
... indicates a potentially dangerous situation that can result
in burns, caused by hot surfaces or liquids, if not avoided.
Abbreviations
2-wire The lead resistance is recorded as an error in the measurement.
3-wire With a cable length of 30 m or longer, measuring errors can
occur.
4-wire The lead resistance can be neglected.
1. General information

EN
14096788.07 03/2018 EN/DE
6
WIKA operating instructions models TR31-3, TR31-K
2. Safety
WARNING!
Before installation, commissioning and operation, ensure
that the appropriate resistance thermometer has been
selected in terms of measuring range, design, specic
measuring conditions and appropriate wetted parts' materials
(corrosion).
Non-observance can result in serious injury and/or damage
to the equipment.
Depending on the type of application, the electrical
connection must be protected against mechanical damage.
Further important safety instructions can be found in the
individual chapters of these operating instructions.
2.1 Intendeduse
The model TR31 resistance thermometer is used as a general-
purpose thermometer for the measurement of temperatures from
-50 … +150 °C or -58 ... +302 °F (without neck tube) and -50 … +250 °C
or -58 ... +482 °F (with neck tube) in liquid and gaseous media.The
version with mineral-insulated sheathed cable and neck tube allows
to measure temperatures up to 300 °C or 572 °F. It can be used for
pressures up to 140 bar with 3 mm sensor diameters and up to 270 bar
with 6 mm sensor diameters, dependent on the instrument version.
The instrument has been designed and built solely for the intended use
described here, and may only be used accordingly.
The technical specications contained in these operating instructions
must be observed. Improper handling or operation of the instrument
outside of its technical specications requires the instrument to be taken
out of service immediately and inspected by an authorised WIKA service
engineer.
2. Safety

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
7
If the instrument is transported from a cold into a warm environment, the
formation of condensation may result in instrument malfunction. Before
putting it back into operation, wait for the instrument temperature and the
room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on
operation contrary to the intended use.
2.2 Personnelqualication
WARNING!
Riskofinjuryshouldqualicationbeinsucient!
Improper handling can result in considerable injury and
damage to equipment.
■
The activities described in these operating instructions
may only be carried out by skilled personnel who have the
qualications described below.
■
Keep unqualied personnel away from hazardous areas.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based
on their technical training, know-how and experience as well as their
knowledge of country-specic regulations, current standards and directives,
are capable of carrying out work on electrical systems and independently
recognising and avoiding potential hazards.The skilled electrical personnel
have been specically trained for the work environment they are working
in and know the relevant standards and regulations.The skilled electrical
personnel must comply with current legal accident prevention regulations.
Special operating conditions require further appropriate knowledge, e.g.
of aggressive media.
2. Safety

EN
14096788.07 03/2018 EN/DE
8
WIKA operating instructions models TR31-3, TR31-K
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene,
ammable or toxic gases or liquids, and refrigeration plants,
compressors, etc., in addition to all standard regulations,
the appropriate existing codes or regulations must also be
followed.
WARNING!
Protection from electrostatic discharge (ESD) required!
The proper use of grounded work surfaces and personal
wrist straps is required when working with exposed circuitry
(printed circuit boards), in order to prevent static discharge
from damaging sensitive electronic components.
To ensure safe working on the instrument, the operating
company must ensure
■
that suitable rst-aid equipment is available and aid is
provided whenever required.
■
that the operating personnel are regularly instructed in all
topics regarding work safety, rst aid and environmental
protection and know the operating instructions and in
particular, the safety instructions contained therein.
DANGER!
Danger of death caused by electric current
Upon contact with live parts, there is a direct danger of death.
■
Electrical instruments may only be installed and
connected by skilled electrical personnel.
■
Operation using a defective power supply unit (e.g. short
circuit from the mains voltage to the output voltage) can
result in life-threatening voltages at the instrument!
2. Safety

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
9
WARNING!
Residual media in dismounted instruments can result in a
risk to persons, the environment and equipment.
Take sucient precautionary measures.
Do not use this instrument in safety or emergency stop
devices. Incorrect use of the instrument can result in injury.
Should a failure occur, aggressive media with extremely high
temperature and under high pressure or vacuum may be
present at the instrument.
2.4 Labelling, safety marks
Productlabel(examples)
2. Safety
Model
Date of manufacture (Year-Month)
Approval logos
Information on version (measuring element, output signal, measuring range...)
■
Thermometer with transmitter and 4 ... 20 mA output signal
■
Thermometer with direct sensor output with Pt100 and Pt1000
Serial number, TAG no.
Before mounting and commissioning the instrument,
ensure you read the operating instructions!

EN
14096788.07 03/2018 EN/DE
10
WIKA operating instructions models TR31-3, TR31-K
3. Specications
Thermometerwithtransmitterandoutputsignal4...20mA
(modelTR31-x-x-TT)
Temperature range Without neck tube -30 ... +150 °C (-22 ... +302 °F)
With neck tube -30 ... +250 °C (-22 ... +482 °F) 1) 2)
Measuring element Pt1000
Connection method 2-wire
Tolerancevalueofthe
measuring element
Class A
(per IEC 60751)
Measuringdeviationofthe
transmitter
±0.25 K
(per IEC 60770)
Totalmeasuringdeviationin
accordancewithIEC60770
Measuring deviation of the measuring element +
the transmitter
Measuring span Minimum 20 K, maximum 300 K
Basicconguration Measuring range 0 ... 150 °C (-32 ... +302 °F),
other measuring ranges are adjustable
Analogueoutput 4 ... 20 mA, 2-wire
Linearisation Linear to temperature per IEC 60751
Linearisation error ±0.1 % 3)
Switch-on delay, electrical Max. 4 s (time before the rst measured value)
Warming-up period After approx. 4 minutes, the instrument will
function to the specications (accuracy) given in
the data sheet.
Current signal for fault
signal
Congurable in accordance with NAMUR NE43
downscale ≤ 3.6 mA upscale ≥ 21.0 mA
Sensor short-circuit Not congurable, in accordance with
NAMUR NE43 downscale ≤ 3.6 mA
Sensor current < 0.3 mA (self-heating can be neglected)
Load RARA ≤ (UB- 10 V) / 23 mA with RA in Ω and UB in V
Eectofload ±0.05 % / 100 Ω
Power supply UBDC 10 ... 30 V
Max.permissibleresidual
ripple
10 % generated by UB< 3 % ripple of the output
current
Power supply input Protected against reverse polarity
Powersupplyeect ±0.025 % / V (depends on the power supply)
3.Specications

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
11
Thermometerwithtransmitterandoutputsignal4...20mA
(modelTR31-x-x-TT)
Inuenceoftheambient
temperature
0,1 % of the span / 10 K Ta
Electromagnetic
compatibility(EMC)5)
EN 61326 emission (group 1, class B) and
interference immunity (industrial application) 4),
conguration at 20 % of the full measuring range
Temperature units Congurable °C, °F, K
Infodata TAG No., descriptor and user message can be
stored in transmitter
Congurationand
calibration data
Permanently stored
Electrical connection
■
M12 x 1 circular connector (4-pin)
■
Directly connected cable
Readings in % refer to the measuring span
For a correct determination of the overall measuring deviation, both sensor and transmitter
measuring deviations have to be considered.
1)
The temperature transmitter should therefore be protected from temperatures over 85 °C (185 °F).
2) Version with mineral-insulated sheathed cable can be used up to 300 °C (572 °F).
3) ±0.2 % for measuring ranges with a lower limit less than 0 °C (32 °F)
4) Use resistance thermometers with shielded cable and ground the shield at at least one end
of the lead if the lines are longer than 30 m or leave the building. Operate the instrument in a
grounded state.
5) During transient interferences (e.g. burst, surge, ESD) take into account an increased measuring
deviation of up to 2 %.
3.Specications

EN
14096788.07 03/2018 EN/DE
12
WIKA operating instructions models TR31-3, TR31-K
Thermometer with direct sensor output with Pt100
(modelTR31-x-x-Px)andPt1000(modelTR31-x-x-Sx)
Temperature range
■
Class A
■
Class B
Without neck tube -30 ... +150 °C (-22 ... +302 °F)
With neck tube -30 ... +250 °C (-22 ... +482 °F) 6)
Without neck tube -50 ... +150 °C (-58 ... +302 °F)
With neck tube -50 ... +250 °C (-58 ... +482 °F) 6)
Temperature at the
connector or at the directly
connected cable
Max. 85 °C (185 °F)
Measuring element
■
Pt100 (measuring current: 0.1 ... 1.0 mA)
■
Pt1000 (measuring current: 0.1 ... 0.3 mA)
Connection method
■
2-wire
■
3-wire
■
4-wire
Tolerancevalueofthemeasu-
ringelementperIEC60751
■
Class A
■
Class B at 2-wire
Electrical connection
■
M12 x 1 circular connector (4-pin)
■
Directly connected cable
6) Version with mineral-insulated sheathed cable can be used up to 300 °C (572 °F).
For detailed specications for Pt sensors, see Technical information IN 00.17 at
www.wika.com.
Case
Material Stainless steel
Ingressprotection
■
Case with connected connector 7) 8)
or directly connected cable
■
Coupler connector, not connected
IP67 and IP69 per EN/IEC 60529,
IP69K per ISO 20653
IP67 per EN/IEC 60529
Weight in kg Ca. 0.2 ... 0.7
(depending on version)
Dimensions See “Dimensions in mm”
7) The stated ingress protection only applies when plugged in using mating connectors that have
the appropriate ingress protection.
8) Not tested at UL
3.Specications

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
13
Ambientconditions
Ambienttemperaturerange
■
M12 x 1 circular connector
Model TR31-3-x-TT
Models TR31-3-x-Px, TR31-3-x-Sx
■
Directly connected cable
(model TR31-K-x-xx)
-40 ... +85 °C (-40 ... +185 °F)
-50 ... +85 °C (-58 ... +185 °F)
-20 ... +80 °C (-4 ... +176 °F)
Storage temperature range
■
M12 x 1, 4-pin circular connector
(model TR31-3-x-xx)
■
Directly connected cable
(model TR31-K-x-xx)
-40 ... +85 °C (-40 ... +185 °F)
-20 ... +80 °C (-4 ... +176 °F)
ClimateclassperIEC60654-1
■
M12 x 1 circular connector
Model TR31-3-x-TT
Models TR31-3-x-Px, TR31-3-x-Sx
■
Directly connected cable
(model TR31-K-x-xx)
Cx (-40 ... +85 °C or -40 ... +185 °F,
5 ... 95 % r. h.)
Cx (-50 ... +85 °C or -58 ... +185 °F,
5 ... 95 % r. h.)
Cx (-20 ... +80 °C or -4 ... +176 °F,
5 ... 95 % r. h.)
Maximumpermissiblehumidityper
IEC60068-2-30var.2
Relative humidity 100 %, condensation
allowed
Maximumoperatingpressure9)10) 140 bar with 3 mm sensor diameter
270 bar with 6 mm sensor diameter
VibrationresistanceperIEC60751 10 ... 2,000 Hz, 20 g 9)
Shock resistance
perIEC60068-2-27
50 g, 6 ms, 3 axis, 3 faces, 3 times for
each face
Salt fog IEC 60068-2-11
9) Dependent on the instrument version
10) Reduced operating pressure when using a compression tting:
■
Stainless steel
■
PTFE
max. 100 bar
max. 8 bar
3.Specications

EN
14096788.07 03/2018 EN/DE
14
WIKA operating instructions models TR31-3, TR31-K
Patents/property rights
M12x1adaptertoangular
connectorDINEN175301-803
No. 001370985
Conditionsforoutdooruse(forULapprovalonly)
■
The instrument is suitable for applications with pollution degree 3.
■
The power supply must be suitable for operation above 2,000 m
should the temperature transmitter be used at this altitude.
■
The instrument shall be installed in locations sheltered from the
weather.
■
The instrument shall be installed “sun/UV radiation protected”.
For further specications see WIKA data sheet TE 60.31 and the order
documentation.
4. Design and function
4.1 Description
The model TR31 resistance thermometer consists of a thermowell with
a xed process connection and is screwed directly into the process.
It is designed to be impact and vibration resistant and all electrical
components are protected against humidity (IP67 or IP69K). The
vibration resistance conforms to IEC 60751 (20 g, dependent on the
instrument version).The impact resistance of all versions meets the
requirements of IEC 60751.The electrical connection is made via an
M12 x 1 circular connector or via the directly connected cable.
An adapter for electrical connection with angular connector per
DIN EN 175301-803 is optionally available for the M12 x 1 circular
connector version.
3.Specications/4.Designandfunction

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
15
X
X
X
X
Thread Height process connection X
G 1/2 11
G 3/8 11
G 1/4 10
M12 11
M20 11
1) For process temperature > 150 °C (302 °F) a neck length N (MH) of 70mm is required,
otherwise N (MH) selectable (55, 65 or 70 mm).
140069565.02
without
neck tube
without process
connection
with compression
tting
with
neck tube
Legend:
A (U1) Insertion length
N (MH) Neck length
X Height process connection
with direct cable outlet
connecting cable
14097792.01
4.2 Dimensions in mm
■
Process connection with parallel thread (or without process connection)
4. Design and function
Ød Sensor diameter
W Length of directly connected cable
L Length of bare wires

EN
14096788.07 03/2018 EN/DE
16
WIKA operating instructions models TR31-3, TR31-K
X
X
X
■
Process connection with tapered thread
4. Design and function
with direct cable outlet
connecting cable
14097792.01
140069565.02
without
neck tube
with compression
tting
with
neck tube
4.3 Scopeofdelivery
Cross-check scope of delivery with delivery note.
Thread Height process connection X
1/4NPT 15
1/2NPT 19
1) For process temperature > 150 °C (302 °F) a neck length N (MH) of 70mm is required,
otherwise N (MH) selectable (55, 65 or 70 mm).
Legend:
A (U2) Insertion length
N (MH) Neck length
X Height process connection
Ød Sensor diameter
W Length of directly connected cable
L Length of bare wires

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
17
5. Transport, packaging and storage
5.1 Transport
Check the instrument for any damage that may have been caused by
transport. Obvious damage must be reported immediately and damaged
instruments must not be used.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport
(e.g. change in installation site, sending for repair).
5.3 Storage
Permissibleconditionsattheplaceofstorage:
■
Storage temperature:
- M12 x 1, 4-pin circular connector: -40 ... +85 °C (-40 ... +185 °F)
- Directly connected cable: -20 ... +80 °C (-4 ... +176 °F)
■
Humidity: 5 ... 95 % r. h.
Avoidexposuretothefollowingfactors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Potentially explosive environments, ammable atmospheres
Store the instrument in its original packaging in a location that fulls the
conditions listed above. If the original packaging is not available, pack
and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument along with shock-absorbent material in the
packaging.
3. If stored for a prolonged period of time (more than 30 days), place a
bag containing a desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove
any residual media.This is of particular importance if
the medium is hazardous to health, e.g. caustic, toxic,
carcinogenic, radioactive, etc.
5. Transport, packaging and storage

EN
14096788.07 03/2018 EN/DE
18
WIKA operating instructions models TR31-3, TR31-K
6. Commissioning, operation
WARNING!
Avoid putting any mechanical loading on the electrical
connections and on the enclosures. Connections must only
be opened once the instrument has been depressurised and
has cooled down.
Maximumpermissibletemperatures:
■
At case with transmitter: 85 °C (185 °F)
■
With directly connected cable: -20 ... +80 °C (-4 ... +176 °F)
■
Class A:
Without neck tube -30 ... +150 °C (-22 ... +302 °F)
With neck tube -30 ... +250 °C (-22 ... +482 °F) 1)
■
Class B:
Without neck tube -50 ... +150 °C (-58 ... +302 °F)
With neck tube -50 ... +250 °C (-58 ... +482 °F) 1)
1) Version with mineral-insulated sheathed cable can be used up to 300 °C (572 °F).
6.1 Mounting
These resistance thermometers are designed for screw-tting directly
into the process. The insertion length, along with the ow velocity and
viscosity of the process media, may reduce the max. loading on the
thermowell.
The housing must be grounded against electromagnetic elds and
electrostatic discharge. It is not necessary to connect the housing
separately to the equipotential bonding system, provided that it has a
xed and secure contact to the metallic vessel, its components or pipes,
and that these are connected to the equipotential bonding system.
When there is a non-metallic contact with the vessel, or with its
structural components or piping, the instrument must be provided with
equipotential bonding.
6. Commissioning, operation

EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
19
Installation on pipes
a on elbows
b in smaller pipe, inclined
c perpendicular to ow direction
Insertion length A
WARNING!
Neither repairs nor structural modications are permitted,
and any would void the guarantee.
Installationexamples
For information on tapped holes, refer to DIN 3852 or for NPT threads to
ANSI B 1.20.
6.1.1Tighteningtorquesforcompressionttings
Sealing Turns Max.pressureinbar
Stainless steel ferrule 1 ¼ ... 1 ½ 100
Stainless steel compression ring 1 ¼ ... 1 ½ 100
PTFE ferrule 1 ¼ ... 1 ½ 8
6.1.2 Tightening torque for M12 mating connector or M12 adapter
Choose a tightening torque of 0.6 Nm.
6. Commissioning, operation

EN
14096788.07 03/2018 EN/DE
20
WIKA operating instructions models TR31-3, TR31-K
6.2 Electrical connection
The electrical connection is made via a circular connector.
■
Output signal Pt100 and Pt1000 (standard)
M12 x 1 circular connector (4-pin) Directly connected cable
6. Commissioning, operation
2-wire
red
white
white
white
white
red
red
red
red
3-wire
4-wire
mA
4 ... 20 mA
M12x1connector Angularconnector
Accessories:M12x1PtadaptertoDINEN175301-803angularconnector
Alternative pin assignments possible.
For further information see order documentation.
Other manuals for TR31-3
1
This manual suits for next models
1
Table of contents
Languages:
Other WIKA Accessories manuals