WIKA N-10 User manual

SL-1
2450335.11USA 07/2008
N-10, N-11
USA
Pressure Transmitter
Operating instructions
N-10 N-11
WIKA Instrument Corporation
A1000 Wiegand Boulevard
Lawrenceville, GA 30043
Phone (770) 513 8200
Fax (770) 338 5118
E-Mail [email protected]
www.wika.com

2450335.11 USA 07/2008
Operating instructions
N-10,
N-11
Pressure transmitter
N-11
N-10
WIKA Instrument Corporation
A1000 Wiegand Boulevard
Lawrenceville, GA 30043
Phone (770) 513 8200
Fax (770) 338 5118
E-Mail [email protected]
www.wika.com

2450335.11 USA 07/2008
2WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
3
WIKA Operating instructions N-10, N-11
Contents Page
1. Important details for your information 3-4
2. A quick overview for you 4
3. Signs, symbols and abbreviations 4-5
4. Function 5
5. For your safety 5-6
6. Packaging 6
7. Starting, operation 7-14
8. Adjustment of zero point / span 15-16
9. Maintenance, accessories 16
10. Trouble shooting 16-18
11. Storage, disposal 18
12. Control drawing FM, CSA 19
WIKA Global 20
USA
USA
1. Important details for your information
Use of the product in accordance with the intended use N-10, N-11:
Use the non-incendive pressure transmitter to transform the pressure into an electrical signal
in hazardous areas.
1. Important details for your information
Read these operating instructions before installing and starting the pressure transmitter. Keep
the operating instructions in a place that is accessible to all users at any time.
The following installation and operating instructions have been compiled by us with great care
but it is not feasible to take all possible applications into consideration. These installation and
operation instructions should meet the needs of most pressure measurement applications. If
questions remain regarding a specific application, you can obtain further information:
Via our Internet address www.wika.de / www.wika.com
The product data sheet is designated as APE N-10
Contact WIKA for additional technical support (770) 513 8200
If the serial number and/or the 2D code on the hexagon gets illegible (e.g. by mechanical
damage or repainting), the retraceability of the instrument is not possible any more.
WIKA pressure transmitters are carefully designed and manufactured using state-of-the-art
technology. Every component undergoes strict quality and environmental inspection before
assembly and each instrument is fully tested prior to shipment. Our environmental manage-
ment system is certified to DIN EN ISO 14001.
Certificate FM/CSA:
Pressure transmitter for operation in hazardous areas in compliance with the respective certifi-
cate (see Control drawing No. 2245906, section 12).
FM / CSA Approval ratings: Non-incendive
Non-incendive for Class I Division 2 Groups A, B, C and D
Dust-ignitionproof for Class II, Division 1, Groups E, F, and G.
Current terms and conditions apply.
Details are available on
www.wika.com
Contents

2450335.11 USA 07/2008
4WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
5
WIKA Operating instructions N-10, N-11
USA
USA
3. Signs, symbols and abbreviations
Potential danger of life or of
severe injuries.
Notice, important informa-
tion, malfunction.
FM
Factory Mutual
The product was tested and
certified by FM Approvals. It
complies with the applicable
US-American standards on
safety (including explosion
protection).
Instructions for hazardous
areas: Potential danger of
life or of severe injuries.
1. Important details for your information / 2. A quick overview for you / 3. Signs and symbols
Knowledge required
Install and start the pressure transmitter only if you are familiar with the relevant regulations
and directives of your country and if you have the qualification required. You have to be
acquainted with NEC. Depending on the operating conditions of your application you have to
have the corresponding knowledge, e.g. of aggressive media.
2. A quick overview for you
If you want to get a quick overview, read Chapters 3, 5, 7 and 11. There you will get some
short safety instructions and important information on your product and its starting. Read
these chapters in any case.
Potential danger of life or
of severe injuries due to
catapulting parts.
Potential danger of burns
due to hot surfaces.
CSA
Canadian Standard Association
The product was tested and
certified by CSA International.
It complies with the applicable
Canadian and US-American
standards on safety (including
explosion protection).
3. Signs, symbols and Abbreviations / 4. Function and accessories / 5. For your safety
4. Function and accessories
N-10: Pressure connection (non-incendive) with internal diaphragm (standard version).
N-11: Pressure connection with flush diaphragm (non-incendive) for highly viscous or
solids entrained media which might clog the pressure port.
2-wire Two connection lines are intended for the voltage supply.
The supply current is the measurement signal.
3-wire Two connection lines are intended for the voltage supply.
One connection line is intended for the measurement signal.
Function: The pressure prevailing within the application is transformed into a standardised
electrical signal through the deflection of the diaphragm, which acts on the sensor element
with the power supply fed to the transmitter. This electric signal changes in proportion to the
pressure and can be evaluated correspondingly.
5. For your safety
Select the appropriate pressure transmitter with regard to scale range,
performance and specific measurement conditions prior to installing and
starting the instrument.
Observe the relevant national regulations (e.g.: EN 50178) and observe
the applicable standards and directives for special applications (e.g. with
dangerous media such as acetylene, flammable gases or liquids and toxic
gases or liquids and with refrigeration plants or compressors). If you do not
observe the appropriate regulations, serious injuries and/or damage
can occur!
Open pressure connections only after the system is without pressure!
Please make sure that the pressure transmitter is only used within the overload threshold
limit all the time!
Observe the ambient and working conditions outlined in section 7 „Technical data”.
Ensure that the pressure transmitter is only operated in accordance with the provisions i.e.
as described in the following instructions.
!
Warning
!
Warning
Warning
Warning
Caution

2450335.11 USA 07/2008
6WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
7
WIKA Operating instructions N-10, N-11
Remove the protection cap only just before installation and absolutely avoid any
damage to the diaphragm during installation as well (N-11).
USA
USA
5. For your safety / 6. Packaging
5. For your safety
Do not interfere with or change the pressure transmitter in any other way
than described in these operating instructions.
Remove the pressure transmitter from service and mark it to prevent it from
being used again accidentally, if it becomes damaged or unsafe for opera-
tion.
Take precautions with regard to remaining media in removed pressure transmitter.
Remaining media in the pressure port may be hazardous or toxic!
Have repairs performed by the manufacturer only.
6. Packaging
Has everything been supplied?
Check the scope of supply:
Completely assembled pressure transmitters; with flush version N-11 including pre-
assembled sealings and protection cap.
Inspect the pressure transmitter for possible damage during transportation. Should
there be any obvious damage, inform the transport company and WIKA without delay.
Keep the packaging, as it offers optimal protection during transportation (e.g. chan-
ging installation location, shipment for repair).
Ensure that the pressure connection thread and the connection contacts will not be
damaged.
In order to protect the diaphragm, the pressure connection of the instrument N-11 is provided
with a special protection cap.
Remove this protection cap only just before installing the pressure transmitter in order
to prevent any damage to the diaphragm or the thread.
Keep the protection cap of the pressure connection thread and the diaphragm for
later storage or transport.
Mount the protection cap when removing and transporting the instrument.
7. Starting, operation
7. Starting, operation
Required tools: wrench (flats 27), screw driver
Diaphragm test for your safety
It is necessary that before starting the pressure transmitter you test the diaphragm, as this is
a safety-relevant component.
Pay attention to any liquid leaking out, for this points to a diaphragm
damage.
Check the diaphragm visually for any damage (N-11).
Use the pressure transmitter only if the diaphragm is undamaged.
Use the pressure transmitter only if it is in a faultless condition as far as the
safety-relevant features are concerned.
P #
If there is no serial number on the product label, the number on the hexagon will apply.
Mechanical connection Product label (example)
Coded manufacture date
Pressure range
Signal
Power Supply
Product No.
Serial No.
PIN assignment
Special instructions for non-incendive
S #
Use to install
the transmitter
process
connection
max. 50 Nm
!
Warning
!
Warning

2450335.11 USA 07/2008
8WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
9
WIKA Operating instructions N-10, N-11
For Model N-10 you have to provide for a sealing element; exceptions are instruments
with self-sealing threads (e.g. NPT thread).
For Model N-11 the sealing ring is included in delivery.
Please refer to our data sheet “Pressure gauge sealing washers AC 09.08” in WIKA’s
product catalog Pressure and Temperature Measurement or our website www.wika.de
for details about sealing washers.
When mounting the instrument, ensure that the sealing faces of the instrument and
the measuring point are clean and undamaged.
Screw in or unscrew the instrument only via the flats using a suitable tool and the
prescribed torque. The appropriate torque depends on the dimension of the pressure
connection and on the sealing element used (form/material). Do not use the case as
working surface for screwing in or unscrewing the instrument.
When screwing the transmitter in, ensure that the threads are not jammed.
For tapped holes and welding sockets please see Technical Information IN 00.14 for
download at www.wika.de -Service
USA
USA
7. Starting, operation
Protect the diaphragm against any contact with abrasive substances
and pressure peaks and do not touch it with tools. If you damage the
diaphragm, non-incendive can not be guaranteed (FM, CSA)!
Electrical connection
Do not exceed the permissible surface temperatures applicable for this
range according to the defined temperature classes.
!
Warning
7. Starting, operation
Do not disconnect the electrical connection while under voltage.
Always connect the case to ground to protect the pressure transmitter
against electromagnetic fields and electrostatic charges.
!
Warning
Electrical installation shall be prepared in accordance with the control
drawing no. 2245906 (section 12).
Use a power supply not exceeding the non-incendive field wiring
parameters (30 V DC max, 120 mA max.)
Attempting to remove the cable connection will damage the transmitter and
void the factory warranty and FM + CSA approval.
The electrical connection provided on the transmitter should be used as originally supplied
and not bypassed or modified (other than cable length). Improper installation or modifica-
tion on the electrical connection will void the hazardous area approval rating.
Switch on the operating voltage only after establishing the electrical connection in order to
avoid any spark formation.
Connect the cable screen to ground. Provide a potential equalization for long cables or
where required by the installation conditions.
Cover flying leads with fine wires by an end splice (cable preparation).
Operate the pressure transmitter with a shielded cable and ground the shield at least
on one side of the cable, if the cable is longer than 30m (2-wire) or 3m (3- or 4-wire),
or if it is run outside of the building.
Ingress protection per IEC 60529 (The ingress protection classes specified only apply
while the pressure transmitter is connected with female connectors that provide the
corresponding ingress protection).
Ensure that the cable diameter you select fits to the cable gland of the connector
Ensure that the cable gland of the mounted connector is positioned correctly and
that the sealings are available and undamaged. Tighten the threaded connection and
check the correct position of the sealings in order to ensure the ingress protection.
Please make sure that the ends of cables with flying leads do not allow any ingress of
moisture.
Warning
Warning

2450335.11 USA 07/2008
10 WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
11
WIKA Operating instructions N-10, N-11
USA
USA
7. Starting, operation
Load (e.g. display)
Power supply
UB+/Sig+ Positive supply / measurement connection
OV/Sig- Negative supply / measurement connection
Cable with 1,6 m (6 foot) length,
IP 67 (NEMA 4x)
Order code: 2X
L-Connector,
DIN EN 175301-803,
IP 65 (NEMA 5)
Order code: AX
3-wire2-wire
(D) Mounting hole
1. Loosen the screw (1).
2. Loosen the cable gland (2).
3. Pull the angle housing (5), with the
terminal block (6) inside, away from the
instrument.
4. Using the head of a small screwdriver in
the mounting hole (D), lever the terminal
block (6) out of the angle housing (5).
In order not to damage the sealing of the angle housing, do not try to push the terminal
block (6) out using the screw hole (1) or the cable gland (2).
5. Ensure that the conductor outer diameter you select is matched to the angle housing’s
cable gland. Slide the cable through the cable gland nut (2), washer (3), gland seal (4) and
angle housing (5).
6. Connect the flying leads to the screw terminals on the terminal block (6) in accordance
with the pin-assignment drawing.
7. Press the terminal block (6) back into the angle housing (5).
8. Tighten the cable gland (2) around the cable. Make sure that the sealing isn’t damaged
and that the cable gland and seals are assembled correctly in order to ensure ingress
protection.
9. Place the flat, square gasket over the connection pins on the top of the instrument
housing.
10. Slide the terminal block (6) onto the connection pins.
11. Secure the angle housing (5) and terminal block (6) to the instrument with the screw (1).
Assembly of L-connector DIN EN 175301-803
(6)
(5)
(1)
(2) (3) (4)
Clamping nut,
Male connector,
Case with
pressure connection
Sealing
Female
connector
7. Starting, operation
brown
brown
green
white
green
screen / case screen / case

2450335.11 USA 07/2008
12 WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
13
WIKA Operating instructions N-10, N-11
Specifications Model N-10 / N-11
Pressure ranges psi 5 10 15 25 30 60 100 200
Over pressure safety psi 29 58 72 145 145 240 500 1160
Burst pressure psi 35 69 87 170 170 290 600 1390
Pressure ranges psi 300 500 1000 1500 2000 3000 5000
Over pressure safety psi 1160 1160 1740 2900 4600 7200 11,600
Burst pressure psi 1390 5800 7970 11,600 14,500 17,400 24,650 2)
Pressure ranges psi 8000 1) 10000 1) 15000 1)
Over pressure safety psi 17,400 17,400 21,750
Burst pressure psi 34,800 2) 34,800 43,500
{Vacuum, gauge pressure, compound range, absolute pressure are available}
1) Only Model N-10.
2) For model N-11: the value specified in the table applies only when sealing is rea-
lised with the sealing ring underneath the hex. Otherwise max. 21,000 psi applies.
Materials
Wetted parts(other materials see WIKA diaphragm seal program)
Model N-10»Stainless steel (> 300 psi stainless steel and Elgiloy®)
Model N-11»Stainless steel
O-ring: NBR {FPM/FKM}
CaseStainless steel
Internal transmission fluid Synthetic oil (not for N-10 with pressure ranges > 300 psi)
Power supply UBUBin VDC 10 < UB≤30 with signal output 4 ... 20 mA, 2-wire
6 < UB≤30 with signal output 1 ... 5 V, 3-wire low power
Signal output and 4 … 20 mA, 2-wire, RA≤(UB – 10 V) / 0.02 A
maximum ohmic load RARAin Ohm 1 … 5 V, 3-wire, RA> 10000
± 10 via potentiometers inside the instrument
Response time (10 ... 90 %) ms ≤1 (≤ 10 ms at medium temperatures below < -22 °F / -30 °C
for pressure ranges up to 300 psi or with flush diaphragm)
Specifications Model N-10 / N-11
Insulation voltage VDC 500
Accuracy % of span ≤0.25 (BFSL)
≤0.5 3)
3) Including non-linearity, hysteresis, zero point and full scale error (corresponds to
error of measurement per IEC 61298-2).
Adjusted in vertical mounting position with lower pressure connection.
Non-linearity % of span ≤0.2 (BFSL) according to IEC 61298-2
Non-repeatability % of span ≤0.1
1-year stability % of span ≤0.2 (at reference conditions)
Permissible temperature of
Medium 4) -22 ... +212 °F {-40 … +221 °F} -30 ... +100 °C {-40 … +105°C}
Ambience 4) -22 ... +212 °F {-40 … +221 °F} -30 ... +100 °C {-40 … +105°C}
Storage 4) -40 ... +221 °F {-58 … +221 °F} -40 ... +105 °C {-50 … +105°C}
4) Also complies with EN 50178, Tab. 7, Operation (C) 4K4H, Storage (D) 1K4,
Transport (E) 2K3
Compensated temp range 0 ... +80 °C 32 ... +176 °F
Temperature coefficients within
compensated temp range
Mean TC of zero% of span ≤0.2 / 10 K (< 0,4 for pressure range ≤250 mbar)
Mean TC of range% of span ≤0.2 / 10 K
Approval FM, CSA Non-incendive for Class I, Division 2, Groups A, B, C and D
Dust ignitionproof for: Class II, Division 1, Groups E, F and G
Maximum electrical ratings 30 V, 20 mA
RF-immunity V/m 10
Burst kV 4
Shock resistance g 1000 according to IEC 60068-2-27 (mechanical shock)
Vibration resistance g 20 according to IEC 60068-2-6 (vibration resonance)
USA
USA
7. Starting, operation 7. Starting, operation

2450335.11 USA 07/2008
14 WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
15
WIKA Operating instructions N-10, N-11
Specifications Model N-10 / N-11
Wiring protection
Short-circuit proofnessSig+ towards UB-
Reverse polarity protectionUB+ towards UB-
Weight lb Approx. 0.4
USA
USA
When designing your plant, take into account that the stated values (e.g.burst pressure,
over pressure safety) apply depending on the material, thread and sealing element used.
7. Starting, operation
{ } Items in curved brackets are optional extras for additional price.
Functional test
Open pressure connections only after the system is without pressure!
Observe the ambient and working conditions outlined in section 7 „Technical
data.
Please make sure that the pressure transmitter is only used within the over-
load threshold limit at all times!
When touching the pressure transmitter, keep in mind that the surfaces of
the instrument components might get hot during operation.
The output signal must be proportional to the pressure. If not, this might point to a
damage of the diaphragm. In that case refer to chapter 10 „Troubleshooting“.
Z = Zero
S = Span
Female
connector
Sealing
Clamping
nut
Male
connector
Sealing
Case with
pressure
connection
8. Adjustment of zero point / span (only for pressure transmitter with
clamping nut)
A
B
We do not recommend to adjust the span potentiometer. It is used for adjustment ex
factory and should not be adjusted by you unless you have adequate calibration equip-
ment at your disposal (at least three times more accurate than the instrument being
tested).
Make sure wires are not cut or pinche
during disassembly and reassembly of the
connector.
Remove the female connector. Open
the pressure transmitter by detaching
the clamping nut (see Fig. A ). Carefully
remove the male connector from the case.
Adjust the zero point (Z) (see Fig. B ) by
generating the lower limit of the pressure
range.
Adjust the span (S) by generating the
higher limit of the pressure range.
Check the zero point.
If the zero point is incorrect, repeat proce-
dure as required.
Reassemble the instrument carefully.
Make sure all sealings and o-rings are not
damaged and correctly installed to assure
the rated moisture ingress protection.
8. Adjustment of zero point / span
Do not open the instrument (e.g. for adjustment) in potentially explosive
atmospheres!
Warning
Caution
Warning

2450335.11 USA 07/2008
16 WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
17
WIKA Operating instructions N-10, N-11
Failure Possible cause Procedure
No output signal No/incorrect voltage supply or current
spike
Adjust the voltage supply to correspond
with the Operating Instructions *)
Cable break Check connections and cable
No/False output signal Incorrectly wired (e.g. Connected as
2-wire instead of 3-wire system)
Follow pin assignment (see Instrument
Label / Operating Instructions)
Output signal unchanged after change
in pressure
Mechanical overload through over-
pressure
Replace instrument; if failure reoccurs,
consult the manufacturer *)
Output signal unchanged after change
in pressure Wrong supply voltage or current spike Replace instrument
Signal span dropping off/too small
Diaphragm is damaged, e.g. through
impact, abrasive/agressive media;
corrosion of diaphragm/pressure
connector; transmission fluid missing.
Contact the manufacturer and replace
the instrument
Signal span too small Power supply too high/too low Correct the power supply in line with
the Operating Instructions
Mechanical overload through over-
pressure Re-calibrate the instrument *)
Signal span drops off Moisture present (e.g. at the cable tail) Install the cable correctly
Signal span erratic Electromagnetic interference source in
the vicinity, e.g. inverter drive
Shield the device; shield the cables;
remove the interference source.
Instrument not grounded Ground instrument
Strong fluctuations in the power supply Stabilise the power supply; smooth it
(i.e.; remove interferences)
Violent fluctuations in the process
media pressure Damping; consult with manufacturer
USA
USA
Recommended recalibration cycle: 1 year For further information (770) 513 8200
8. Adjustment of zero point / span / 9. Maintenance, accessories / 10. Trouble shooting
9. Maintenance, accessories
Accessories
For details about the accessories (e. g. connectors), please refer to WIKA‘s price list, WIKA‘s
product catalog on CD or or contact our sales department.
10. Trouble shooting
Open pressure connections only after the system is without pressure!
WIKA pressure transmitters require no maintenance.
Have repairs performed by the manufacturer only.
Do not insert any pointed or hard objects into the pressure port for cleaning to
prevent damage to the diaphragm of the pressure connection.
Take precautions with regard to remaining media in removed pressure trans-
mitters. Remaining media in the pressure port may be hazardous or toxic!
Remove the pressure transmitter from service and mark it to prevent it from
being used again accidentally, if it becomes damaged or unsafe for opera-
tion.
Have repairs performed by the manufacturer only.
1) Switch off the operating voltage.
2) Pull the connector or disconnect the cable.
Shutdown
10. Trouble shooting
In case of unjustified reclamation we charge the reclamation handling expenses.
*) Make sure that after the setting the unit is working properly. In case the error continues to exist send in the instrument for reparation (or
replace the unit).
Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the
right voltage supply and the right type of wiring (2-wire/ 3-wire) has been chosen?
Warning
!
Warning

2450335.11 USA 07/2008
18 WIKA Operating instructions N-10, N-11
2450335.11 USA 07/2008
19
WIKA Operating instructions N-10, N-11
USA
USA
10. Trouble shooting / 11. Storage, disposal
If the problem persists, contact our sales department.
USA, Canada
If the problem continues, contact WIKA or an authorized agent for assistance.
If the pressure transmitter must be returned obtain an RMA (return material authorization)
number and shipping instructions from the place of purchase. Be sure to include detailed
information about the problem. Pressure transmitters received by WIKA without a valid RMA
number will not be accepted.
Process material certificate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our employees
and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with which
the instrument has come into contact, either through installation, test purposes, or cleaning.
You can find the Product Return Form on our internet site (www.wika.de / www.wika.com).
11. Storage, disposal
When storing or disposing of the pressure transmitter, take precautions with
regard to remaining media in removed pressure transmitters.
We recommend cleaning the transmitter properly and carefully.
Remaining media in the pressure port may be hazardous or toxic!
Mount the protection cap when storing the pressure transmitter in order to prevent any
damage to the diaphragm (N-11).
Dispose of instrument components and packaging materials in accordance with the
respective waste treatment and disposal regulations of the region or country to which
the instrument is supplied.
Storage
Disposal
WIKA reserves the right to alter these technical specifications.
12. Control Drawing (FM, CSA)
Control drawing S-No. 2245906.03
This Equipment is suitable for Locations:
Class I, Division 2, Groups A, B, C, D
Class II, Division 1, Groups E, F, G
BARE
WHITE
GREEN
BROWN
4
3
2
1
1-5V, 0.5-4.5V RM
0V / Sig-0V / Sig-
4-20 mA
CONTROL EQUIPMENT
NON-HAZARDOUS LOCATIONHAZARDOUS LOCATION
SERIES N-10, N-11
APPROVED/CERTIFIED
NONINCENDIVE TRANSMITTER
UB+UB+ Sig+
Sig+
Notes U.S. Installations
1 Install per NATIONAL ELECTRICAL
CODE and local codes,
as applicable.
2 To maintain type 4X sealing of series
N-10 / N-11-transmitter, use thread
sealant when installing conduit.
12. Control Drawing (FM, CSA)
!
Warning

WIKA Global
Argentina WIKA Argentina S.A.
Phone: (+54) 11/47 30 18 00
Australia WIKA Australia Pty. Ltd.
Phone: (+61) 2-88 45 52 22
Austria WIKA Messgerätevertrieb
Ursula Wiegand GmbH & Co. KG
Phone: (+43) 1/8 69 16 31
Benelux WIKA Benelux
Netherlands Phone: (+31) 4 75/53 55 00
Brazil WIKA do Brasil Ind. e Com. Ltda.
Phone: (+55) 15-32 66 16 55
Canada WIKA Instruments Ltd.
Phone: (+1) 780/463-7035
China WIKA International Trading
(Shanghai) Co., Ltd.
Phone: (+86) 21 53 85 25 73
Finland WIKA Finland Oy
Phone: (+358) 9/6 82 49 20
France WIKA Instruments s.a.r.l.
Phone: (+33) 1/34 30 84 84
Germany WIKA Alexander Wiegand GmbH Co. KG
Phone: (+49) 93 72/13 20
India WIKA Instruments India Pvt. Ltd.
Phone: (+91) 20/ 6629 3200
Italy WIKA Italiana SRL
Phone: (+39) 02/93 86 11
Japan WIKA JAPAN K. K.
Phone: (+81) 3/54 39-66 73
Kazakhstan TOO WIKA Kazakhstan
Phone: (+7) 32 72/33 08 48
Korea WIKA Korea Ltd.
Phone: (+82) 2/8 69 05 05
Malaysia WIKA Instrumentation (M) Sdn. Bhd.
Phone: (+60) 3 56 36/88 58
Poland WIKA Polska S.A.
Phone: (+48) 54 230 11 00
Russia ZAO WIKA MERA
Phone: (+7) 495 - 648 01 80
Serbia WIKA Merna Tehnika d.o.o.
Phone: (+381) 11 2763 722
Singapore WIKA Instrumentation Pte. Ltd.
Phone: (+65) 68 44 55 06
South Africa WIKA Instruments (Pty.) Ltd.
Phone: (+27) 11/6 21 00 00
Spain Instrumentos WIKA S.A.
Phone: (+34) 902 902 577
Switzerland Manometer AG
Phone: (+41) 41/9 19 72 72
E-Mail: info@manometer.ch
Taiwan WIKA Instrumentation Taiwan Ltd.
Phone: (+886) 34 20 60 52
Ukraine WIKA Pribor GmbH
Phone: (+38) 062 345 34 16
United Arab WIKA Middle East FZE
Emirates Phone: (+971) 4/883 90 90
United Kingdom WIKA Instruments Ltd
Phone: (+44) 1737 644 008
U.S.A. WIKA Instrument Corporation
Phone: (+1) 770 / 5 13 82 00
www.wika.de
Other manuals for N-10
1
This manual suits for next models
1
Table of contents
Other WIKA Transmitter manuals