WIKA S-10 User manual

Operating instructions
Betriebsanleitung
Mode d'emploi
Manual de instrucciones
EN
DE
FR
ES
Druckmessumformer, Typ S-10
Pressure transmitter, model S-10
Pressure transmitter, model S-10
Transmetteur de pression, type S-10
Transmisor de presión, modelo S-10

2WIKA operating instructions pressure transmitter, model S-10
EN
DE
FR
1604457.16 09/2017 EN/DE/FR/ES
ES
© 2002 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d‘emploi avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!
Betriebsanleitung Typ S-10 Seite 27 - 50
Operating instructions model S-10 Page 3 - 26
Mode d‘emploi type S-10 Page 51 - 74
Manual de instrucciones modelo S-10 Página 75 - 99

EN
3WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
Contents
Declarations of conformity can be found online at www.wika.com
Contents
1. General information 4
2. Design and function 5
3. Safety 6
4. Transport, packaging and storage 8
5. Commissioning, operation 9
6. Adjustment of zero point and span 12
7. Faults 13
8. Maintenance and cleaning 15
9. Dismounting, return and disposal 16
10. Specications 18

EN
4WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
1. General information
1. General information
■
The instrument described in the operating instructions has been designed and manufactured using state-of-the-
art technology. All components are subject to stringent quality and environmental criteria during production. Our
management systems are certied to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument.Working safely requires that
all safety instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range
of use.
■
The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and
readily accessible to skilled personnel at any time. Pass the operating instructions onto the next operator or owner of
the instrument.
■
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modications.
■
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: PE 81.01
- Application consultant: Tel.: +49 9372 132-8976
Fax: +49 9372 132-8008976

EN
5WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
2. Design and function
2. Design and function
2.1 Overview
Electrical connection (depending on version)
Case; product label
Process connection, spanner flats
Process connection, thread
2.2 Scope of delivery
■
Pressure transmitter
■
Operating instructions
Cross-check scope of delivery with delivery note.

EN
6WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
3. Safety
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to equipment or
the environment, if not avoided.
Information
... points out useful tips, recommendations and information for ecient and trouble-free operation.
3.2 Intended use
The pressure transmitter is used for measuring pressure.The measured pressure is output as an electrical signal.
Only use the pressure transmitter in applications that lie within its technical performance limits (e.g. max. ambient
temperature, material compatibility, ...).The pressure transmitter is suitable for applications with pollution degree 3 and
an ambient temperature of -20 ... +80 °C.
→ For performance limits, see chapter 9 “Specications”.
The instrument has been designed and built solely for the intended use described here, and may only be used
accordingly.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.

EN
7WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
3. Safety
3.3 Personnelqualication
Skilled personnel
Skilled personnel, authorised by the operator, are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowledge of country-specic
regulations, current standards and directives, are capable of carrying out the work described and independently
recognising potential hazards.
3.4 Labelling, safety marks
Product label
S# Serial no., P# Model code Accuracy
Power supply Pin assignment
Output signal
Measuring range
Symbols
Before mounting and commissioning the instrument, ensure you read the operating instructions!

EN
8WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the pressure transmitter for any damage that may have been caused during transportation.
Obvious damage must be reported immediately.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site, sending for
repair).
Permissible conditions at the place of storage:
■
Storage temperature: -40 ... +100 °C
■
Humidity: 45 ... 75 % relative humidity (no condensation)

EN
9WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
5. Commissioning, operation
5. Commissioning, operation
5.1 Mounting the instrument
Prior to commissioning, the pressure transmitter must be subjected to a visual inspection.
■
Leaking uid is indicative of damage.
■
Only use the pressure transmitter if it is in perfect condition with respect to safety.
Requirements for mounting point
The mounting point must meet the following conditions:
■
Ambience corresponds to a max. pollution degree of 3.
■
Humidity ≤ 80 % r. h. (non-condensing).
■
Protected from weather inuences.
■
Sealing faces are clean and undamaged.
■
Sucient space for a safe electrical installation.
■
For information on tapped holes and welding sockets, see Technical information IN 00.14 at www.wika.com.
■
Permissible ambient and medium temperatures remain within the performance limits. Consider possible restrictions
on the ambient temperature range caused by mating connector used.
→ For performance limits, see chapter 9 “Specications”
Sealing variants
Parallel threads
Seal the sealing face with a flat gasket, lens-type
sealing ring or WIKA profile sealing.
per EN 837 per DIN 3852-E
Tapered threads
Wrap threads with sealing material (e.g. PTFE
tape).
NPT, R and PT

EN
10 WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
Mounting the instrument
The max. torque depends on the mounting point (e.g. material and shape). If you have any questions,
please contact our application consultant.
→ For contact details, see chapter 1 “General information” or the back page of the operating instructions.
1. Seal the sealing face (→ see “Sealing variants”).
2
.
At the mounting point, screw the pressure transmitter in hand-tight.
3
.
Tighten with a torque spanner using the spanner ats.
5.2 Connecting the instrument to the electric system
Requirements for voltage supply
→ For power supply, see product label
The power supply for the pressure transmitter must be made via an energy-limited electrical circuit in accordance
with section 9.4 of UL/EN/IEC 61010-1, or an LPS per UL/EN/IEC 60950-1 / CSA C22.2 no. 60950-1, or class 2 in
accordance with UL1310/UL1585 (NEC or CEC). The voltage supply must be suitable for operation above 2,000 m
should the pressure transmitter be used at this altitude.
Requirements for electrical connection
■
Cable diameter matches the cable bushing of the mating connector.
■
Cable gland and seals of the mating connector are correctly seated.
■
With cable outlets, no humidity can ingress at the cable end.
Requirement for shielding and grounding
The pressure transmitter must be shielded and grounded in accordance with the grounding concept of the plant.
Connecting the instrument
1. Assemble the mating connector or cable outlet.
→ For pin assignments, see product label
2
.
Establish the plug connection.
5. Commissioning, operation

EN
11WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
5. Commissioning, operation
5.3 Fitting a DIN 175301-803 angular connector
1. Loosen the screw (1).
2. Loosen the cable gland (2).
3. Pull the angular connector (5) + (6) away from the instrument.
4. CAUTION!
Improper mounting
The seal of the angle housing will be damaged.
▶
Do not try to push the terminal block (6) out using the
screw hole (1) or the cable gland (2).
Via the mounting hole (D), lever the terminal block (6) out of the angle
housing (5).
5. Slide the cable through the cable gland (2), the ring (3), the sealing (4) and
the angle housing (5).
6. Connect the cable ends to the terminal blocks (6) in accordance with the
connection diagram.
7. Press the angle housing (5) onto the terminal block (6).
8. Make sure that the seals are not damaged and that the cable gland and
seals are correctly seated in order to ensure ingress protection.
9. Tighten the cable gland (2) around the cable.
10. Place the at gasket (7) over the instrument's connection pins.
11. Push the angular connector (5) + (6) onto the instrument.
12. Tighten the screw (1).
(D)
(7)
(5)
(6)
(8)
(1)
(2)
(3)
(4)

EN
12 WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
6. Adjustment of zero point and span
Only adjust the span-setting potentiometer if calibration equipment is available which has at least three
times the accuracy of the pressure transmitter.
7.1Preparation(gureA)
To gain access to the potentiometers, open the instrument as follows:
■
Disconnect the electrical connection from the instrument.
■
Remove the clamping nut .
■
Carefully pull the instrument connector from the instrument.
■
Connect the instrument connector to the power supply and a display unit
(e.g. ammeter, voltmeter) according to the connection diagram.
7.2Adjustmentofzeropoint(gureB)
■
Go to the start of the measuring range.
■
Using potentiometer "Z", adjust the minimum output signal (e.g. 4 mA)
7.3Settingthespan(gureB)
■
Go to the end of the measuring range.
■
Using potentiometer "S", adjust the maximum output signal (e.g. 20 mA)
■
Check the zero point and if there is any deviation, re-adjust it.
■
Repeat the procedure until the zero point and the span are set correctly.
A
B
6. Adjustment of zero point and span

EN
13WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
7. Faults
7. Faults
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the pressure transmitter must be taken
out of operation immediately.
▶
Ensure that pressure or signal is no longer present and protect against accidental commissioning.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter 8.2 “Return”.
WARNING!
Physical injuries and damage to property and the environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, ammable or toxic substances), harmful
media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compres-
sors, there is a danger of physical injuries and damage to property and the environment.
▶
Should a failure occur, aggressive media with extremely high temperature and under high pressure
or vacuum may be present at the instrument.
▶
For these media, in addition to all standard regulations, the appropriate existing codes or regulations
must also be followed.
▶
Wear the requisite protective equipment (see chapter 3.4 “Personal protective equipment”).
For contact details, see chapter 1 “General information” or the back page of the operating instructions.
In the event of any faults, rst check whether the pressure transmitter is mounted correctly, mechanically and electrically.
If complaint is unjustied, the handling costs will be charged.

EN
14 WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
Faults Causes Measures
No output signal Cable break Check the continuity
Deviating zero point signal Overload safety exceeded Observe the permissible overload safety
Deviating zero point signal Too high/low working temperature Observe the permissible temperatures
Constant output signal upon change in
pressure
Mechanical overload caused by overpres-
sure
Replace instrument; if it fails repeatedly,
contact the manufacturer
Signal span varies EMC interference sources in the environ-
ment; for example, frequency converter
Shield instrument; cable shield; remove
source of interference
Signal span varies/inaccurate Too high/low working temperature Observe the permissible temperatures
Signal span drops/too small Mechanical overload caused by overpres-
sure
Replace instrument; if it fails repeatedly,
contact the manufacturer
If complaint is unjustied, we will charge you the complaint processing fees.
7. Faults

EN
15WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
8. Maintenance and cleaning
8. Maintenance and cleaning
8.1 Maintenance
This pressure transmitter is maintenance-free.
Repairs must only be carried out by the manufacturer.
8.2 Cleaning
CAUTION!
Unsuitable cleaning agents
Cleaning with unsuitable cleaning agents may damage the instrument and the product label.
▶
Do not use any aggressive cleaning agents.
▶
Do not use any hard or pointed objects.
▶
Do not use any abrasive cloths or sponges.
Suitable cleaning agents
■
Water
■
Conventional dishwashing detergent
Cleaning the instrument
1. Depressurise and de-energise the pressure transmitter.
2. Wipe the instrument surface using a soft, damp cloth.

EN
16 WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
9. Dismounting, return and disposal
9. Dismounting, return and disposal
9.1 Dismounting
WARNING!
Physical injuries and damage to property and the environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, ammable or toxic substances), harmful
media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compres-
sors, there is a danger of physical injuries and damage to property and the environment.
▶
Should a failure occur, aggressive media with extremely high temperature and under high pressure
or vacuum may be present at the instrument.
▶
Wear the requisite protective equipment.
Dismounting the instrument
1. Depressurise and de-energise the pressure transmitter.
2
.
Disconnect the electrical connection.
3
.
Unscrew the pressure transmitter with a spanner using the spanner ats.
9.2 Return
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances (acids, bases, solutions, etc.)
and must therefore be cleaned before being returned.
WARNING!
Physical injuries and damage to property and the environment through residual media
Residual media in the dismounted instrument can result in a risk to persons, the environment and
equipment.
▶
With hazardous substances, include the material safety data sheet for the corresponding medium.
▶
Clean the instrument, see chapter 7.2 “Cleaning”.

EN
17WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
9. Dismounting, return and disposal
When returning the instrument, use the original packaging or a suitable transport packaging.
Information on returns can be found under the heading “Service” on our local website.
9.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally compatible way and in accordance
with the country-specic waste disposal regulations.

EN
18 WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
10. Specications
Measuringrangesandoverloadsafety(gaugepressure)
bar Measuring range 0 ... 0.01 0 ... 0.16 0 ... 0.25 0 ... 0.4 0 ... 0.6 0 ... 1 0 ... 1.6
Overload safety 1 1.5 224510
Measuring range 0 ... 2.5 0 ... 4 0 ... 6 0 ... 10 0 ... 16 0 ... 25 0 ... 40
Overload safety 10 17 35 35 80 50 80
Measuring range 0 ... 60 0 ... 100 0 ... 160 0 ... 250 0 ... 400 0 ... 600 0 ... 1,000
Overload safety 120 200 320 500 800 1,200 1,500
psi Measuring range 0 ... 5 0 ... 10 0 ... 15 0 ... 20 0 ... 25 0 ... 30 0 ... 50
Overload safety 29 29 72.5 145 145 145 240
Measuring range 0 ... 60 0 ... 100 0 ... 150 0 ... 160 0 ... 170 0 ... 200 0 ... 250
Overload safety 240 500 500 1,160 1,160 1,160 1,160
Measuring range 0 ... 300 0 ... 400 0 ... 500 0 ... 600 0 ... 750 0 ... 800 0 ... 1,000
Overload safety 1,160 1,160 1,160 1,160 1,740 1,740 1,740
Measuring range 0 ... 1,500 0 ... 1,600 0 ... 2,000 0 ... 3,000 0 ... 4,000 0 ... 5,000 0 ... 6,000
Overload safety 2,900 4,600 4,600 7,200 7,200 7,200 11,600
Measuring range 0 ... 7,500 0 ... 8,000 0 ... 10,000 0 ... 15,000
Overload safety 17,400 17,400 17,400 21,700
10.Specications

EN
19WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
Measuringrangesandoverloadsafety(absolutepressure)
bar Measuring range 0 ... 0.25 0 ... 0.4 0 ... 0.6 0 ... 1 0 ... 1.6 0 ... 2.5 0 ... 4
Overload safety 2 2 4 5 10 10 17
Measuring range 0 ... 6 0 ... 10 0 ... 16 0 ... 25 0.8 ... 1.2
Overload safety 35 35 80 80 5
psi Measuring range 0 ... 15 0 ... 25 0 ... 50 0 ... 100 0 ... 250
Overload safety 72.5 145 240 500 1,160
Measuringrangesandoverloadsafety(vacuumand+/-measuringrange)
bar Measuring range -0.6 ... 0 -0.4 ... 0 -0.25 ... 0 -0.16 ... 0 -0.1 ... 0 -1 ... 0 -1 ... +0.6
Overload safety 4 2 2 1.5 1 510
Measuring range -1 ... +1.5 -1 ... +3 -1 ... +5 -1 ... +9 -1 ... +15 -1 ... +24
Overload safety 10 17 35 35 80 50
psi Measuring range -15 inHg ... 0 -30 inHg ... 0 -30 inHg ... +15
Overload safety 72.5 72.5 145
Measuring range -30 inHg ... +30 -30 inHg ... +60 -30 inHg ... +100
Overload safety 240 240 500
Measuring range -30 inHg ... +160 -30 inHg ... +200 -30 inHg ... +300
Overload safety 1,160 1,160 1,160
10.Specications

EN
20 WIKA operating instructions pressure transmitter, model S-10
1604457.16 09/2017 EN/DE/FR/ES
Specications
Vacuum tightness Yes
Output signal
Output signal see product label
Load 4 ... 20 mA (2-wire) ≤ (power supply - 10 V) / 0.02 A
20 ... 4 mA (2-wire)
0 ... 20 mA (3-wire) ≤ (power supply - 3 V) / 0.02 A
DC 0 ... 10 V (3-wire) > max. output signal / 1 mA
DC 0 ... 5 V (3-wire)
DC 1 ... 5 V (3-wire)
DC 0.5 ... 4.5 V ratiometric (3-wire)
Voltage supply
Power supply see product label
The power supply for the pressure transmitter must be made via an energy-limited electri-
cal circuit in accordance with section 9.3 of UL/EN/IEC 61010-1, or an LPS per UL/EN/IEC
60950-1, or class 2 in accordance with UL1310/UL1585 (NEC or CEC). The voltage supply
must be suitable for operation above 2,000 m should the pressure transmitter be used at this
altitude.
Accuracyspecications
Non-linearity (per IEC 61298-2) ≤ ±0.2 % of span BFSL
Non-repeatability ≤ ±0.1 % of span
Accuracy at reference conditions see product label
Adjustability of zero point and span ±5 %
10.Specications
Other manuals for S-10
1
Table of contents
Languages:
Other WIKA Transmitter manuals
Popular Transmitter manuals by other brands

ASCON TECNOLOGIC S.r.l.
ASCON TECNOLOGIC S.r.l. ZTT-61/GD user guide

Honeywell
Honeywell Midas S2 Quick reference guide

Inovonics
Inovonics Frequency Agile FA210 installation manual

Inovonics
Inovonics EchoStream EN1233S/D installation instructions

ARMANO
ARMANO PTPi operating instructions

NAPCO
NAPCO GEM-TRANSLP installation instructions

Oppermann Regelgeräte
Oppermann Regelgeräte OPP-SENS P Series manual

tau
tau BUG Series Operation and Adjustments

Aplisens
Aplisens LI-24ALW Safety user manual

ABB
ABB Endura ACA592 operating instructions

Savant
Savant SmartLink HDBaseT HTX-SLN501 Quick reference guide

DynaMetric
DynaMetric TMP-646 instructions