
10
English
Additional 
work
0 - 120 s After the liquid level falls 
below the set value “STOP 
Level”, the pump will work 
for a period of time speci-
edhere.
Wait time 
comm.
1 - 30 min No communication time. 
Time after which the 
device enters automatic 
operation when no com-
munication is present.
24h-5s work enabled After 24 hours standby 
the pump is automatically 
turned on for 5 seconds in 
order to lubricate bearings.
disabled
Sound alarm enabled Failure is reported by an 
internal sound signalling 
device.
disabled
Pulsing alarm enabled The failure report relay is 
turned on every second. It 
can be used for supplying 
powertoanexternalsignal
lamp.
disabled In the case of failure the 
relay is switched on perma-
nently.
Therm.protect1 enabled Set to “disabled” if thermal 
protection contacts have 
not been connected to 
terminals 20, 21.
disabled
Work waterless allowed The pump can be start-
ed even if the liquid level 
in the tank is below the 
value at which the pump is 
turned off.
disallowed The pump cannot be start-
ed if the liquid level in the 
tank is below the value at 
which the pump is turned 
off. This parameter should 
be set in case of hazard of 
explosion.
Phase control enabled The phase monitoring 
system protects the pump 
against non-uniform volt-
age or incorrect order of 
phases, and detects con-
tactor failure. In the case of 
a single-phase pump, set 
to “disabled”.
disabled
Language polish, english, german, czech
2.6. Additional information
2.6.1. Setting level values
The MR-Lift prevents the setting of unacceptable 
levels values, e.g. the value of the level at which 
the pump is turned off higher than the respec-
tive level of turning it on. Minimum difference 
between level settings is 3 cm.
2.6.2. Setting current values
The MR-Lift prevents the setting of unacceptable 
current values, e.g. minimum current value higher 
thanmaximumcurrentvalue.
2.6.3. LCD contrast setting
The default contrast setting is optimum, and 
there is usually no need to change it. However, 
if such a necessity occurs, the following action 
shouldbetaken:
1. Turn off the device.
2. Press and hold the keys “MANUAL OPERATION” 
and “AUTOMATIC OPERATION” at the same time.
3. Turnonthedevice–themessage“Contrast”will
be displayed.
4. Set required value with navigation keys.
5. Release the pressed keys “MANUAL OPERATION”
and “AUTOMATIC OPERATION”.
2.6.4. Calibration of internal level sensor
After long use the indications of the internal level 
sensor may stop being accurate. In such a case, 
even if there is no pressure, indication of level 
may be other than zero. In order to calibrate the 
sensorproceedasfollows:
1. Turn off the device.
2. Press and hold the keys “MANUAL OPERATION”
and “AUTOMATIC OPERATION” at the same time.
3. Turnonthedevice–themessage“Contrast”will
be displayed.
4. Press the acceptance key “OK” - the message 
“Internal sensor calibrated” will be displayed.
5. Release the pressed keys “MANUAL OPERATION” 
and “AUTOMATIC OPERATION”.
It is recommended to calibrate the sensor at each 
technical review.
NOTE!
Before you start to calibrate the level sensor, 
raise the bell over the water surface.
2.6.5. Settings lock
To avoid making changes to the settings by unau-
thorized persons, the MR-Lift is equipped with a 
lock function. 
If the settings lock is on, when you try to change 
settings, the display shows the message “Settings 
locked”. 
To activate the settings lock, press and hold the 
“OK” key and press “MANUAL OPERATION” 
key(thedisplayshowsthemessage„Settings
locked”).Todisablethelockmustrepeatthe
aboveprocedure(thedisplayshowsthemessage
„Settingsunlocked”).
Locking and unlocking can be activated only in 
the home screen mode.
2.6.6. Wireless communication
MR-Lift operation is controlled by means of a 
radio communication system in ISM 868MHz 
standard. The system consists of controlling soft-