
Technical Service Hotline: 978-750-9876
3.9 Ending a Combustion Test
WARNING! HEAT HAZARD! Do Not Touch the probe after removing it from the
stack. Allow the probe to cool before handling (about 5 minutes).
When testing is complete, remove the probe from the stack and allow the analyzer
pump to run to purge the hose and analyzer with room air until normal ambient gas
readings are observed on the display. Turn OFF the analyzer by setting the OFF/ON
switch to its “O” position. When storing the analyzer, it is a good idea to empty the wa-
ter trap and dry it out. It is also suggested that you remove the hose and allow it to dry.
3.10 Using the Gas Sample Conditioner when Measuring NO
, the metal heat exchangers must be replaced with plastic
versions in the A 500’s condensation block. It is also highly recommended that a Wöh-
ler Peltier Cooler gas sample conditioner be attached to the probe assembly. The gas
sample conditioner removes water vapor from the gas sample preventing the formation
of water droplets inside the probe hose. If water droplets were to form inside the hose,
gas molecules in the gas sample would be removed by
the water resulting in inaccurate, lower-than-actual readings for these gases.
Refer to Appendix A for detailed information on how to install, operate, and maintain
the Gas Sampler Conditioner.
3.11 Status and Error Messages
The following is a list of status and error messages that appear in the top line of the
display while operating the analyzer.
**LOW BATTERY**: Battery life below 4%.
FUEL SELECTION: Select the fuel being burned by touching the name of the approp-
SYSTEM IN HOLD: The display is frozen at the values being displayed when the
CO:PROTECT ON: If the unit is exposed to CO levels in excess of the pre-set excess
level, the CO clean out pump will start. If the unit has no COhigh
sensor installed you must remove the probe from the stack and
wait for the CO level in the analyzer to reach zero. Then you must
press the „PUM2“ button on the bottom of the display screen. This
will shut-off the CO clean out pump and you can begin testing
again. If the unit has a COhigh sensor installed the CO clean
out pump will start when the pre-set excess CO level has been
reached, however, you can continue testing and the COhigh sen-
sor will give you a reading up to 32,000ppm. When the CO(low)