AEMC GROUNDFLEX 6474 User manual

ENGLISH User Manual
GROUNDFLEX®ADAPTER 6474
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated using
standards and instruments traceable to international
standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
NIST traceable calibration is only possible when
the Model 6474 is connected to the Ground Tester
Model 6472.
To obtain NIST calibration (for a nominal charge), the
instrument would need to be returned to our facility,
along with your Ground Tester Model 6472.
The recommended calibration interval is 12 months
and begins on the date of receipt by the customer.
Serial #: ________________________________
Catalog #: ______________________________
Model #: 6474
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

2
GroundFlex®Adapter Model 6474
Table of Contents
1. INTRODUCTION............................................................................... 3
1.1 International Electrical Symbols...........................................................4
1.2 DenitionofMeasurementCategories.................................................4
1.3 ReceivingYourShipment.....................................................................4
1.4 OrderingInformation ............................................................................5
1.4.1 AccessoriesandReplacementParts......................................5
2. GROUNDFLEX®ADAPTER................................................................ 6
2.1 ControlFeatures ..................................................................................6
2.2 GroundFlex®Sensor ............................................................................7
2.2.1 CalibrationofGroundFlex®Sensors .......................................8
2.3 MeasurementsinAUTOMode...........................................................10
2.3.1 PreparationoftheGroundFlex®Adapter...............................10
2.3.2 MakingaMeasurement(ResistanceofLegs) ......................12
2.3.3 OtherMeasurements ............................................................14
2.4 MANUALModeandSWEEPMode ...................................................14
3. SPECIFICATIONS........................................................................... 16
3.1 Electrical.............................................................................................16
3.2 Mechanical.........................................................................................18
3.3 Environmental ....................................................................................18
3.4 Safety.................................................................................................18
4. MAINTENANCE ............................................................................. 19
4.1 Maintenance.....................................................................................19
4.2 Cleaning...........................................................................................19
LimitedWarranty ................................................................................................21
WarrantyRepairs................................................................................................21
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
3
CHAPTER 1
INTRODUCTION
WARNING
Thesesafetywarningsareprovidedtoensurethesafetyofpersonnel.
Please read and complywiththeseprecautions:
• Thisinstrumentisprotectedfromaccidentalvoltagesofnotmorethan
50VwithrespecttoearthinmeasurementCATIV.Theguaranteedlevel
ofprotectionofthisequipmentmaybecompromisedifusedinamanner
notspeciedbythemanufacturer.
• Safetyistheresponsibilityoftheoperator.
• Allmetalobjectsorwiresconnectedtotheelectricalsystemshouldbe
assumedtobelethaluntiltested.Groundingsystemsarenoexception.
• Never exceed the maximum rated voltage and current, and the mea-
surementcategory.
• Neverexceedtheprotectionlimits,andalwayscomplywiththecondi-
tionsandplaceofuse,indicatedinthespecications.
• Donotusetheinstrumentoritsaccessoriesiftheyappeardamaged.
• Useaccessories thathave overvoltagecategory andservicevoltages
greaterthanorequaltothoseoftheinstrument(CATIV50V).Useonly
accessoriesthatcomplywithsafetystandards(IEC61010-2-031&32).
• Weartheappropriateprotectivegear(insulatingbootsandgloves).
• CheckthatnoterminalisconnectedandtheswitchissettoOFFbefore
openingthedevice.
• Useonlythechargingunitsuppliedwiththeinstrumenttorechargethe
battery.
• Troubleshootingandmetrologicalvericationproceduresmustonlybe
performedbyqualied,approvedpersonnel,orthefactory.
• NOTE: Thepotentialsonthevariousrodsusedforanearthmeasurement
maybedierentifanearbyelectricalinstallationisdefectiveorcertain
weather conditions prevail (thunderstorms). It is up to the operator
to decide whether to continue or postpone measurements in these
situations.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

4
GroundFlex®Adapter Model 6474
1.1 International Electrical Symbols
Signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING that the operator must
refer to the user manual for instructions before operating the instrument. In
this manual, the symbol preceding instructions indicates that if the instructions
are not followed, bodily injury, installation/sample and/or product damage may
result.
Risk of electric shock. The voltage at the parts marked with this symbol may be
dangerous.
Ground/Earth symbol
In conformity with WEEE 2002/96/EC
1.2 Definition of Measurement Categories
CAT II: Formeasurementsperformedoncircuitsdirectlyconnectedtotheelectrical
distributionsystem.Examplesaremeasurementsonhouseholdappliancesor
portabletools.
CAT III:
Formeasurementsperformedinthebuildinginstallationatthedistributionlevel
suchasonhardwiredequipmentinxedinstallationandcircuitbreakers.
CAT IV: Formeasurementsperformedattheprimaryelectricalsupply(<1000V)such
asonprimaryovercurrentprotectiondevices,ripplecontrolunits,ormeters.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistentwiththe
packinglist.Notifyyourdistributorofanymissingitems.Iftheequipmentappears
tobedamaged,leaclaimimmediatelywiththecarrierandnotifyyourdistributor
atonce,givingadetaileddescriptionofanydamage.Savethedamagedpacking
containertosubstantiateyourclaim.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
5
1.4 Ordering Information
NOTE: The GroundFlex®Adapter Model 6474 only operates in combination
with the Ground Tester Model 6472 Kit - 500ft.
GroundFlex®Adapter Model 6474.................................................Cat. #2136.01
Includes the GroundFlex®Adapter Model 6474, four GroundFlex®sensors (5m/16ft) with
twelve color-coded rings, carrying case for meter, connection lead, two extension leads on H
reel (black/green) with color-coded alligator clips, six BNC extension leads, calibration loop,
three C-clamps, and a user manual.
GroundFlex®Field Kit (Model 6472 and 6474) .............................Cat. #2136.03
Includes the GroundFlex®Adapter Model 6474, Ground Resistance Tester Model 6472,
carrying bag for meter, carrying bag for kit, hard carrying case, optical USB cable, power
adapter 110/240V with power cord 115V US, two 500ft color-coded leads on spools (red/
blue), two 100ft color-coded leads (hand-tied, green/black), one 30ft lead (green), four
T-shaped auxiliary ground electrodes, set of ve spaded lugs, one 100ft AEMC®tape
measure, rechargeable NiMH batteries, and a USB stick with DataView®software, ground
tester workbook and user manual.
1.4.1 Accessories and Replacement Parts
Bag#6-CarryingBagforGroundKits............................................. Cat. #2119.82
TapeMeasure-AEMC®(100ft) .......................................................Cat. #2130.60
Setoftwo,14.5"T-shapedAuxiliaryGroundElectrodes..................Cat. #2135.39
OpticalUSBCable............................................................................Cat. #2135.41
Replacement-OneGroundFlex®(5m /16ft)Sensor.........................Cat. #2135.73
Replacement-ConnectionLeadforModel6472/6474....................Cat. #2135.75
Replacement-OneBNC(15m/50ft)ExtensionLead......................Cat. #2135.76
Replacement-Setof12,GroundFlex®Rings
(red,yellow,brown,orange) .............................................................Cat. #2135.77
Replacement-GreenExtensionLead .............................................Cat. #2135.78
Replacement-BlackExtensionLead...............................................Cat. #2135.79
Replacement-Setof3,C-clamps....................................................Cat. #2135.80
Replacement-CalibrationLoop.......................................................Cat. #2135.82
Replacement-CarryingCase ..........................................................Cat. #2135.83
Replacement-SafetyAlligatorClip(black) ......................................Cat. #2140.53
Replacement-SafetyAlligatorClip(green) .....................................Cat. #2140.69
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

6
GroundFlex®Adapter Model 6474
CHAPTER 2
GROUNDFLEX®ADAPTER
2.1 Control Features
50V CAT IV
1-2-3-4
1-2-3
1-2
1
2
3
4
For Use with the
Ground Tester Model 6472
GroundFlex
®
Sensor
SENSOR TURNS
SENSITIVITY
To Model 6472
Overload
INPUT
1
2
3
4
GroundFlex®Adapter
MODEL 6474
2
4
1
6
5
3
1. Inputs1through4forGroundFlex®sensors
2. ConnectorforconnectionleadtotheGroundTesterModel6472
3. OVERLOAD indicator
4. SENSORTURNSselector:1,2,3or4
5. SENSITIVITYselector:x10,x1orx1/10
6. INPUTSELECTION:1,2,3,4,1-2,1-3or1-4
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
7
2.2 GroundFlex®Sensor
TheGroundFlex®sensorsareplacedaroundaconductor(towerlegorpole)and
closedusingascrewconnector,whichispartofthesensor.Thisclosedloopisthen
usedtomakeacontact-freemeasurementofthecurrentowingintheconductor
(leg)itiswrappedaround.
Thesesensorsarecurrentmeasurementclampsbutareabletoencloseobjects
thataremuchlarger.
GroundFlex®sensorsareavailableinlengthsofuptoapprox24ft(8m)andcan
beplacedaroundobjects(suchastowerlegs)uptoinapprox8ftdiameter(2.7m).
50V CAT IV
1-2-3-4
1-2-3
1-2
GroundFlex
®
Sensor
To Model 6472
Overload
INPUT
1
2
3
4
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

8
GroundFlex®Adapter Model 6474
2.2.1 Calibration of GroundFlex®Sensors
Calibrationisnecessaryforanewsensor,onethatisnolongeridentiedwhena
sensorhasbeenreplaced,ormorethan2lengthsofextensioncableisused.The
calibrationcoecientsarethenstoredintheModel6474.
NOTE: Do not move or touch the GroundFlex®sensor or the calibration loop
during the calibration. This could cause errors.
Startbyidentifyingeachsensorusingtheidenticationringsthataresuppliedwith
theModel6474:(1)Brown (2)Red (3)Orange (4)Yellow
• ConnectthecalibrationloopbetweenterminalsHandEoftheModel6472.
• ConnecttheModel6472and6474togetherusingtheconnectioncable.
• ConnecttheGroundFlex®sensorwiththeBrownringtoInput1oftheModel
6474,theonewiththeRedringtoInput2,theonewiththeOrangeringtoInput
3,andtheonewiththeYellowringtoInput4.
• CoiltheGroundFlex®sensorfourtimesaroundinthecalibrationloopandclose
itusingitsscrewconnector.Placetheconnectorasfaraspossiblefromthe
calibration device.
6472 6474
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
9
1. Settheswitchofthe6472toSET-UP.
2. SettheSENSOR TURNSswitchto4.
3. SettheSENSITIVITYswitchtox1.
4. SettheINPUT SELECTIONswitchto1.
1-2-3-4
1-2-3
1-2
1
2
3
4
SENSOR TURNS
SENSITIVITY INPUT
5. PresstheHz/OPTIONSbutton5times.
ThedevicestartsbycomparingthecurrentIIntinjectedbythedeviceintothe
calibrationlooptothecurrentIAmeasuredbytheGroundFlex®sensor.Ifthey
aredierent,the"CAL"symbolblinkstorecommendacalibration.
x 5
4
H
E
S
x
11
AUTO
mA
m A
NOTE: To avoid measurement errors, do not touch the GroundFlex®sensor or
the calibration loop during the calibration.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

10
GroundFlex®Adapter Model 6474
6. Tostartthecalibration,press2nd + START.
7. ThedevicecalculatesandstoresacalibrationcoecientfortheGroundFlex®
sensorconnectedtochannel1.
4
H
E
S
x
11
AUTO
mA
m A
Thisoperationmustberepeatedforchannels2,3and4;theINPUT SELECTION
switchmustbesettothecorrespondingpositioneachtime.
Thecalibrationvaluesarenowstoredintomemory.
Aftercalibration,theGroundFlex®sensorsmustalwaysbeconnectedtotheirre-
spectivechannels.
NOTE: Always connect the same sensor to the same input.
2.3 Measurements in AUTO Mode
2.3.1 Preparation of the GroundFlex®Adapter
TheGroundFlex®AdapterModel6474isdesignedtobeusedonlywiththeModel
6472device.Thetwoinstrumentsmustbeconnectedusingthespecialconnection
cablesupplied.TheModel6474hasnoON/OFFbutton,andispoweredonlyby
theModel6472,viathisconnectioncable.
TheModel6474isintendedprimarilytomeasurethecurrentthatowsthrough
ahigh-voltagetowertotheground.Youcanwrapupto4legsofthetowerseach
withaGroundFlex®sensorandmeasurethecurrentowingtoearththrougheach
ofthelegsorthroughseverallegs.
1. SettherotaryfunctionswitchtoGroundFlex®ontheModel6472.
WhenyouconnecttheGroundFlex®AdapterModel6474tothe6472,aself-
testwillbegin.Duringthistest,theOVERLOADchecklightonthe6474islit.
2. Connecttherequirednumberofsensorstochannels1,2,3,and/or4.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
11
NOTE: • The sensors have arrows to indicate a direction.
• All sensors should face the same direction.
(clockwise or counter-clockwise around the tower legs)
• All sensors should have the same number of wraps.
• Sensors should encircle both leg AND ground system.
6474
To the Model 6472
3. OntheModel6474,settheINPUT selectionswitchaccordingly.Youcanmea-
surethecurrentowingthroughoneGroundFlex®sensor(1, 2,3, or4),or
throughseveralsensors(1,2or1,2,3),orthroughallofthesensors(1,2,3,
4).
4. Select how many times the sensor is wrapped around the conductor to be
measured(from1to4turns)usingtheSENSOR TURNSselectionswitch.
NOTE: To increase measurement sensitivity, we recommend when possible,
wrapping a maximum number of turns (up to 4) around the structure.
5. SettheSENSITIVITYswitchtothedesiredsensitivity:x1/10,x1,orx10.The
choiceofsensitivitydependsontheexpectedcurrentvalue.Asaprecaution,
alwaysstartwiththelowestsensitivity,x1/10,andincreaseitasneeded.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

12
GroundFlex®Adapter Model 6474
2.3.2 Making a Measurement (Resistance of Legs)
1. PlaceauxiliaryelectrodesH (Z) and S (Y)onoppositesidesofthetower,as
farawayaspossible(100-150ft)and,ifpossible,perpendiculartothehigh-
voltageline(depthisnotimportant).Thiswilleliminateinterferencewiththe
measurementbystrayvoltages or currents induced under the high-voltage
line.
NOTE: The H (Z) and S (Y) electrodes can be placed in the same direction
if necessary. In this case, the S (Y) electrode should be placed at 62% the
distance of H (Z).
2. In order to avoid electromagnetic interference, it is best to unwind the full
lengthofeachcablefromthereel,tokeepthecablesasfarapartaspossible
ontheground,takingcarenottoformloops,andtoavoidplacingthecables
nearorparalleltometallicconductors(cables,rails,fences,etc.).Eectsof
interferencearemeasuredinmilli-ohms.
3. Connecttheseauxiliaryelectrodes to the terminals H (Z) and S (Y) onthe
Model6472'sfrontpanel.
4. Important: TheterminalsES (Xv) and E (X)mustbeconnectedtoaconduct-
ingmetallicpartofthetowerlocatedabovetheGroundFlex®sensors.This
ensuresmeasurementofthecurrentowingdownwardandtowardstheearth
(otherwise,youwouldmeasurethecurrentowingupward,fromtheearthto
thetopofthetower).
Thiswilltelltheconditionoftheoverheadgroundconductor.
5. Connect the required number of GroundFlex® sensors to channels 1, 2, 3,
and/or4(thechannelsforwhichtheyarecalibrated)andplacetheGround-
Flex®sensorsaroundthelegsofthetower.
NOTE: The direction of coiling of the sensors around tower legs have no
effect on the measurement, but all of the GroundFlex®sensors must be
coiled in the same direction and have the same orientation (check arrow
direction on the sensor) and they must all have the same number of turns.
Sensors should encircle both leg AND ground system.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
13
E
S
H
RS
RH
ES
1
2
3
4
6474
6472
Overhead earth wire
High-voltage
lines
4 GroundFlex®sensors
connected with the
same orientation
WARNING: If the GroundFlex®Adapter Model 6474 is connected to the earth
tester, the earth potential connected to the terminals E (X) and ES (Xv) will
also be present at the BNC connectors of the GroundFlex®sensors and at the
connecting cable between the adapter and tester. Whenever any doubt exists
about this earth potential, the user should proceed to a voltage measurement
using the S (Y) and ES (Xv) terminals of the tester.
1. Setthe3switches(Inputs,TurnsandSensitivity)oftheGroundFlex®Adapter
accordingtotheconnectionmade.
2. StartthemeasurementbypressingtheSTART/STOPbutton.
1
AUTO
k
Ω
mA
V
H
S
ES
E
S
x
12
Todisplaythemeasurementparameters,pressDISPLAY several times.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

14
GroundFlex®Adapter Model 6474
Thedevicedisplaysthefollowingparameters:
RSEL,US-ESanditsfrequency,ISEL,R-Act(RPASS),U-Act(US-ESanditsfrequency,
UH-Eanditsfrequency),I-Act(ISELanditsfrequency).
TomeasuretheresistancesofelectrodesH (Z) and S (Y),oriftheresistanceof
theelectrodesistoohigh,startthemeasurementbyalongpress(approx2sec)of
theSTART/STOPbutton(youwillheara2ndbeep).
RH,RS,UH-E,IH-Ewillbedisplayed.
2.3.3 Other Measurements
AsexplainedinStep4onthepreviouspage,terminalsES (Xv) and E (X)mustbe
connectedtoametallicpartofthetowerabovetheGroundFlex®sensor in order to
measurethecurrentowingISELdownwardthroughthesensortowardstheearth
tomeasuregroundresistanceoftheleg.
IfconnectionsES (Xv) and E (X)areplacedbelowtheGroundFlex®sensor,the
lattermeasuresthecurrentowingtowardsthetopofthetowerviatheoverhead
groundconductoratitstop,andyoucaninthiswayevaluatethequality of this
overhead ground conductor.
YoucanevenmakethismeasurementwithoutmovingthecontactsofES (Xv) and
E (X) onthetowerbysimplypassingthe2connectingcablesthroughtheGround-
Flex®sensor.ThecurrentinconnectingcableE(X) andthecurrentthroughtheleg
ofthetowerthencanceleachotherout.
Youcanalsomeasureany leakage currents(I-Act)owingthrougheachlegof
thetowerinnormaloperationofthehigh-voltageline.Ifyounddierentleakage
currentsineachleg,buttheycanceloutwhenaddedtogether,youcanconclude
thatthesecurrents,inducedbytherotatingeldofthehigh-voltagelines,indicate
an earthing faultofthetowervia the overhead ground wire at the tip of the
mast.
2.4 MANUAL Mode and SWEEP Mode
Use the Hz/OPTIONS button and ► button to change to MANUAL mode. The
followingparameterscanbemodied:
• 128Hzblinks Changethetestfrequency
• Testvoltageblinks Switchbetween16and32V
Youcanalsoperformafrequencysweepandasmoothing.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
15
Inthesweepmode,themeasurementscanbemadeatfrequenciesupto5kHz.
Onasingletower,frequenciesintheorderof5kHzdonotalwaysentailinductive
eects,butthe"currentloops"formedbytheadjacenttowersandthereturnchan-
nelthroughthegroundmayresultininductancesthatcanbedetectedonlyathigh
frequencies.
Anequivalentdiagramofthecomponentsformingthiscircuitisasfollows:
Point of injection
of the current
Overhead
earth wire
LLL
Rpylon
Next pylon
to the left
Pylon
measured
Next pylon
to the right
A measurement in sweep mode (SWEEP) of an installation like this having 10
towersgivesthefollowingresultsassuminganearthresistanceRpylonof10Ωfor
eachtower and an inductanceL of550µHfor the linkviathe overhead ground
conductoratthetipofthetowers:
Impedance of the towers
0
2
4
6
8
10
12
14
16
0100020003000 4000 5000
F (Hz)
Z (ohm)
MeasurementofatowerearthimpedancewithGroundFlex®
sensors
Conventional4polemeasurement
Measurementintothetopofatower(iterativenetworkimpedance)
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

16
GroundFlex®Adapter Model 6474
CHAPTER 3
SPECIFICATIONS
3.1 Electrical
Measuring Method:
Voltage/CurrentmeasurementwitharectangularACsignal
Short Circuit Current: > 200mAAC
Noise Suppression: >80dBatfrequenciesdieringby20%ormorefromthe
testfrequency
Max. Overload: 250Vrms
Max. Value for RH& RS: 100kΩ
Measuring Time: Short push on START: approx.7sforrstvalueofREat
128Hz,then3measurementspers.
Long push on START:approx.15sforrstvalueofREat
128Hz,then3measurementspers.
Pylon Measurement with GroundFlex®
Measurement
Range
0.067 to
9.999Ω
10.00 to
99.99Ω
100.0 to
999.9Ω
1.000 to
9.999kΩ
10.00 to
99.99kΩ
Resolution 0.001Ω0.01Ω0.1Ω1Ω10Ω
Test Voltage 16 or 32V user selectable
Frequency From 41 to 128Hz selectable
Resistance Measurement
Frequency From 41 to 5078Hz selectable
Accuracy ±5% of Reading + 1ct @ 128Hz
Theintrinsicerrorspecied refers to reference conditions with atestvoltageof
32V,testfrequencyof128Hz,RH,andRS=1kΩ,noexternalvoltage.
TheoperatingerrorformeasurementsofRH,RSand REisthesameasthatspecied
for4-Poleearthresistancemeasurement.
TheoperatingerrorofACresistancemeasurementscanbelessthanthatspecied
forvoltageorcurrentbecausefrequencycharacteristicsofthevoltagechannelare
matchedtothoseofthecurrentchannel.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

GroundFlex®Adapter Model 6474
17
For test frequencies between 41 and 5087Hz, between 1 and 4 turns of the
GroundFlex® sensor, and a selection of 1 to 4 measurement channels, the
operatingaccuraciesarethefollowing:
SENSITIVITY and minimum ISEL minimal Operating accuracy for RS-ES
S x 1/10 ISEL > 10mA ± (10% + 4cts)
S x 1 ISEL > 5mA ± (5% + 4cts)
S x 10 ISEL > 5mA ± (5% + 4cts)
5mA > ISEL > 0.5mA ± (15% + 10cts)
Current Measurement with GroundFlex®
Measurement Range 0.0 to
99.9mA (1)
100 to
999mA
1.00 to
9.99A
10.0 to
99.9A
Resolution 0.1mA (1) 1mA 0.01A 0.1A
Frequency 16 to 400Hz
Accuracy <5% of Reading @ 50/60Hz
(1): Valid only in the SENSITIVITY x 10 setting
ThecurrentmeasureddependsonthenumberofturnsoftheGroundFlex®sensor
aroundtheconductortobemeasured.(e.g.if the GroundFlex®sensor is wrapped
4 times around a conductor carrying a current of 1A, the input signal will be the
same as if the GroundFlex®sensor were wrapped once around a conductor car-
rying a current of 4A. The operating error is therefore indicated for an input signal
current in A.tr (Ampere.turns).
TheminimumcurrentmeasureddependsonthesettingoftheSENSITIVITYswitch
asindicatedbythetablebelow:
SENSITIVITY IMIN (A.tr) Points additional error (pt-er)
x 10 0.01 5
x 1 0.04 2
x 1/10 0.16 2
Accuracy
Current (A * turn) 16 to 49Hz 50 to 99Hz 100 to 400Hz
IMIN to 0.399 ± (20% + pt-er) ± (5% + pt-er) ± (3% + pt-er)
0.4 to 39.9 ± (10% + 2ct) ± (3% + 2ct) ± (3% + 2ct)
40 to 99.9 ± (10% + 2ct) ± (3% + 2ct) ± (20% + 2ct)
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

18
GroundFlex®Adapter Model 6474
3.2 Mechanical
Dimensions:10.7x9.76x5.12"(272,248x130mm)
Weight: 7lbs(3.2kg)approx
Case Material:ULV0Polypropylene
Terminals: 4mmrecessedbananajacks
Case Protection: EN60529-IP53(coverclosed)
Drop Test: Per EN 61010-1
Vibration Test: Per EN 61557-1
3.3 Environmental
Operating Temperature: 32°to113°F(0°to45°C);0to90%RH
Specied Operating Temperature(1):0°to95°F(0°to35°C);0to75%RH
Storage Temperature:-40°to158°F(-40°to70°C);0to90%RH
Altitude: <3000m
(1): This range corresponds to the one defined by standard EN 61557, for which an operating error including the
quantities of influence is defined. When the device is used outside this range, 1.5%/10°C and 1.5% between
75 and 90% RH must be added to the operating error.
3.4 Safety
Electromagnetic Compatibility
ThisinstrumentsatisestheEMCandLVDdirectivesrequiredfortheCEmarking
andproductstandardIEC61326-1(Ed.97)+A1(Ed.98).
• Immunityinindustrialenvironment
• Emissions in residential environment.
ElectricalsafetyaccordingtoEN61010-1(Ed.2of2001)
MeasurementaccordingtoEN61557(Ed.2of2007)parts1,4and5.
CATIV,50V
PollutionDegree2
*Specications are subject to change without notice
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412
Other manuals for GROUNDFLEX 6474
1
Table of contents
Other AEMC Test Equipment manuals

AEMC
AEMC 6416 Installation and operating instructions

AEMC
AEMC MN213 User manual

AEMC
AEMC FlexProbe User manual

AEMC
AEMC OX III Series User manual

AEMC
AEMC 6472 Service manual

AEMC
AEMC MN210 User manual

AEMC
AEMC 6471 User manual

AEMC
AEMC CM605 User manual

AEMC
AEMC 6400 series User manual

AEMC
AEMC 6417 User manual
Popular Test Equipment manuals by other brands

Fluke
Fluke 975 AirMeter Application note

ASSOCIATED EQUIPMENT CORPORATION
ASSOCIATED EQUIPMENT CORPORATION 6036B-24 Operator's manual

ATEQ
ATEQ CDP60 user manual

Besantek
Besantek BST-VM04 instruction manual

Haefely Technology
Haefely Technology AXOS5 user manual

PCE Health and Fitness
PCE Health and Fitness HT 210 manual