1

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
2 System Overview
536611JK.CDR
Figure 1
Device Comment
Modality The modality has to support the DICOM print protocol
DRYSTAR 5302 Dry printer for 14x17", 14x14", 14x11", 10x12", 8x10" DT2
blue base or clear base films
NOTE:
Remark to an installation in a 'HIPAA conform environment':
In Figure 1 two additional, optional modules are not listed:
1) Audit logging server: This is a computer in the networking allowing to store all
communication between the different devices in the network.
2) NTP (Network Time Protocol) Server: This is a computer providing time
synchronization. It is useful for audit logging, so that all messages of all
devices have the same time reference.
Both devices are only required in a 'HIPAA' environment, where high data security is
required.
For more information refer to section 6.3.
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 7
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3 Scope of Delivery and Accessories
3.1 Scope of delivery
NOTE:
The printer is delivered with all parts, which are required to install it in an existing
Ethernet network.
Device name: DRYSTAR 5302
Type: 5366/100
ABC Code EQOZV
Scope of delivery
x1 Basic DRYSTAR 5302 including:
- 2 input trays adjustable for all film sizes
- 1 output tray
x1 Compact Flash card (stored in backup CF card slot in input tray)
x2 power cables, 3 m long (Europe / US)
x1 UTP network cable, 5 m long (2 x RJ45 connection) incl. extension piece
x1 Reference manual (English)
x3 User Manuals (English; Chinese simplified, Spanish)
x1 CD with all user manuals and Plug&Play installation manuals in all languages*
x1 packing list
x1 Plug and Play Installation manual (English)
x25 films 14x17" DT 2B for test purposes
x25 films 8x10" DT 2B for test purposes
x2 sets film type labels
x1 Chinese warning sticker
x1 CE declaration of conformity
x1 China ROHS declaration
x1 Installation Report
*available languages: Chinese simplified, Chinese traditional, Czech, Danish, Dutch, English, Estonian,
Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Latvian, Lithuanian, Norwegian,
Polish, Portuguese, Russian, Slovakian, Slovenian, Spanish, Swedish, Thai, Turkish, Vietnamese
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 8
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
IMPORTANT:
The printer is delivered with two 3 m power cables. See Figure 9.
xFor US: NEMA 5-15P with IEC C13 plug (female)
xFor Europe: CEE7/7 with IEC C13 plug (female)
Organize a power cable for your country locally.
See section 8 for “Mains connection in the installation room” specifications.
3.2 Options
Item Order Number Comment
Mobile Kit EX2DV To be used in earthquake areas and in mobile
usage. May not be used in military vessels.
Figure 2
NOTE:
REMOTE SERVICE:
In general it is possible to diagnose and repair the printer to a certain extent remotely.
To enable remote service of the printer remote access has first to be organized.
Remote Access is established via the Agfa own Secure Remote Service System
(SRSS).
For more information to the SRSS refer to MEDNET – GSO Library - Technical
Services - Service Delivery - SRSS
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 9
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
4 Machine dimensions
Closed machine:
72,8cm
28.7"
67,6 cm
26.6"
71.5cm
28.2"
536611jo.cdr
Figure 3
Opened machine:
536611jp.cdr
76 cm
29,9"
116cm
45,7"
536611jn.cdr
Figure 4 Figure 5
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 10
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
5 Minimum free space for operation and maintenance
Space required for normal us
e
536611JL.cdr
Space required for servicing and installation
540mm
23.6"
600 mm
23.6"
600 mm
23.6"
600m
m
23.6"600mm
23.6"
800mm
31.5"
Figure 6
NOTE:
In case the minimum free space for installation and repair is not available, the printer
can be carried at a different place. The printer weight is 90 kg (198 lb). It can be
carried by three people.
Caution:
The device is heavy (90 kg;198 lb). Risk of injuries.
xAt least three people have to lift the device from one position to the other.
xUse proper foot and hand protection when lifting the device.
xThe table might break and cause injuries or damage to the device. The device
weight may be too heavy for the table.
xCommunicate to the customer, that the table must be able to carry more
than 90 kg (198 lb)
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 11
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
6.5 Example for System Integration
IP Address:192.9.200.155
Service
Host
Ultrasound
Printer
Digitizer:
CR 75.0
ADC QS
536311ad.cdr
IP address:
Hostname:
Subnet mask:
Router IP address:
Printer AE Title:
Domain name:
192.9.200.201
ds5302
255.255.255.0
-
DS5302
-
Nickname:
Modality Brand:
Modality Type:
Calling AE-Title:
Digitizer
AGFA
CR QS
QS
Ultrasound
ACUSON
ASPEN
US_ASPEN
Nickname:
Modality Brand:
Modality Type:
Calling AE-Title:
Figure 8
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 19
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
7 Ambient conditions
7.1 During operation
Ambient temperature: min.:
max.:
+15°C (59°F)
+30°C (86°F)
Relative humidity:
(non condensing)
min.:
max.:
20%
75%
7.2 During transport and storage
NOTE:
xClimate conditions for storage must be in accordance with
EN60721-3-1-class 1K4
xClimate conditions for transport must be in accordance with
EN60721-3-2(1993)-class 2K4
Ambient temperature
during storage:
min.:
max.:
- 25°C (-13°F)
+ 55°C (131°F)
Relative humidity
during storage:
min.:
max.:
10%
95%
Absolute humidity
during storage:
min.:
max.:
0.1 g/m³
35 g/m³
Rate of temperature
change during storage
max.: 1 °C/min.
Air pressure during
storage
Min.
max.:
70 kPa
106 kPa
Ambient temperature
during transport:
min.:
max.:
- 40°C (-40°F)
+ 70°C (158°F)
Relative humidity
during transport:
min.:
max.:
10%
95%
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 20
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
8 Electrical connections
Electrical installations have to be done in accordance
with national regulations. Examples:
VDE (Germany) Electrical installations in the
installation room must be in compliance with the
regulations IEC 364, VDE 0100, and VDE 0107.
Installation guidelines
UL (US) Electrical installations in the installation
room must be in compliance with the regulations:
“National Electrical Code" (NEC) (NFPA70).
IMPORTANT:
A ground fault interrupter (GFI) (IN= 30 mA) must be installed at a suitable point in
the circuit (in compliance with VDE 664).
VDE Double earthed pin outlet in compliance with
DIN 49441 and with CEE 7 standard cover V II.
Mains connection in the
installation room
UL Earthed contact outlet for the NEMA 5-15 P plug
Euro-
Version:
CEE7/7 - H05VV - F3G - 3 x 2.5 mm² with
IEC C13 plug (female); cable length 3 m
Mains cable
(part of the delivery)
US/CAN-
Version:
NEMA 5-15P - SJT, 3 x AWG 18 with
IEC C13 plug (female); Cable length 3 m
CEE7/7
IEC C13
IEC C13
NEMA 5-15P
Europe
US
5366_chap11_001.cdr
Figure 9
Network cable
(part of the delivery)
UTP; AWG24; Length: 5 m
Main switch Upon machine installation, it must be ensured that either
the mains connector or an all-pole circuit breaker for the
installation on site is located close to the machine and
easily accessible.
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 21
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
8.1 Electrical connections at the printer
536611JM.cdr
Network connection.
RJ45 UTP for 100BaseT network,
according to IEC 601
Service Interface.
RS232 Interface male
DTE 9 pin sub D (nullmodem)
Equipotential plug
Ground plug.
6 mm metric thread
100 - 240 V AC
power supply
Figure 10
9 Specifications
9.1 Type overview
Printer Type
DRYSTAR 5302 5366/100
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 22
12-2007 5366/100 Agfa Company Confidential

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 23
12-2007 5366/100 Agfa Company Confidential
9.2 Functional data
Film sizes 14x17", 14x14", 14x11", 10x12", 8x10"
Throughput (films/hour) 8x10": 140 14x17": 75
Access time1st film approx. (seconds) 77
Printable area(pixels) Refer to 9.2.1
resolution(DPI) 320 (79.4 µm Pixel size)
Bits per pixel / number of grey steps 12 / 4096
Daylight film loading yes
Films per film magazine 100 films
Number of film magazines 2
Image recording type Direct thermography
Densitometer built in yes
Automatic density control yes
Hard disk capacity t73 GB
RAM Memory Plugged on Themis: 512 MB
Warm-up time Approx. 7 min. (cold printer)
9.2.1 Pixel Matrices
The DRYSTAR 5302 has two film supply trays that can be configured to 14x17",
14x14", 14x11", 10x14" or 8x10" film size. If trim is used around the images (trim is a
small, white frame around each image on the film), the diagnostic area is 6 pixels
smaller at each edge of the film.
DRYSTAR 5302 Diagnostic Printable Area in Pixels
No Annotations Annotations
Film Size
Width Height Width Height
8x10" 2376 3070 2376 2705
10x12" 3070 3653 3070 3288
14x11" 3348 4358 2983 4358
14x14" 4358 4303 4358 3938
14x17" 4358 5232 4358 4867
Table 1: Pixel Matrices

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 24
12-2007 5366/100 Agfa Company Confidential
9.3 Electrical data
Rated voltage [~V] 100-240 AC (autosensing)
Frequency [Hz] 50/60
Power consumption [W] 70 standby
250 printing average
530 peak
9.4 Climatic data
Heat emission [kJ/h] 252 standby
900 printing average
1908 maximum
9.5 Sound emission data
Noise level: in standby:
in operation:
53 dB(A) maximum
62 dB(A) maximum
9.6 Interface data
Image / control data interface DICOM 3.0 Print SCP
Maximum number of simultaneous
DICOM associations
10
9.7 Service data
InfoCounter yes
Error diagnostics built in yes
Module exchange concept yes
Remote/preventive service capable yes

DD+DIS148.05E Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 25
12-2007 5366/100 Agfa Company Confidential
9.8 Dimensions and weight
9.8.1 Dimensions, packed machine
Length / depth (cm / inch) 90 / 35.4
Width (cm / inch) 87 / 34.2
Height (cm / inch) 69 / 27.2
Height incl. Pallet (cm / inch) 84 / 33.1
9.8.2 Dimensions, closed machine
Width (cm / inch) 72,8 / 28.6
Length / depth (cm / inch) 71,5 / 28.2
Height with output tray (cm / inch) 67,6 / 26.6
Height without output tray (cm / inch) 53,6 / 21.1
9.8.3 Dimensions, opened machine
Width (cm / inch) 72,8 / 28.6
Length / depth (cm / inch)
with film trays opened
116 / 45.7
Height (cm / inch)
with top cover opened
76 / 29.9
9.8.4 Weight
Printer alone (kg / lb) 90 / 198
Printer with packing incl. palette (kg / lb) 120,5 / 265,6
Other manuals for Drystar 5302
3
Table of contents
Other AGFA Medical Equipment manuals

AGFA
AGFA Drystar 4500 Plug & Play User manual

AGFA
AGFA WS-Motorized-T-001 User manual

AGFA
AGFA CR 30-X User manual

AGFA
AGFA D2RS User manual

AGFA
AGFA DR 100s User manual

AGFA
AGFA Classic E.O.S. Technical manual

AGFA
AGFA DR 14s User manual

AGFA
AGFA DR 600 User manual

AGFA
AGFA DX-D User manual

AGFA
AGFA DRY 3.4 User manual