AGFA DR 100e User manual

DR 100e (Film/CR)
6012/001
6012/002
User Manual
0360B EN 20181015 0853

Contents
Legal Notice ..........................................................................4
Getting started .......................................................................5
Introduction to this Manual ................................................... 6
Scope of this Manual ..................................................7
About the safety notices in this document .................. 8
System Documentation ..............................................9
Disclaimer ................................................................10
Introduction ........................................................................ 11
Intended Use ........................................................... 12
Intended User ..........................................................12
Training .......................................................12
Configuration .......................................................... 13
Operation Controls .................................................. 15
Circuit breaker ............................................. 18
X-ray generator control panel .......................19
Manual Collimator .......................................20
Exposure button ...........................................21
Infrared remote control ................................22
Messages ................................................................. 23
Sound signals .............................................. 24
Status messages ........................................... 24
Warning messages ....................................... 25
Error messages .............................................27
Labels ...................................................................... 29
General ........................................................30
Type label .................................................... 33
Labeling of the collimator ............................ 34
Installation .............................................................. 35
Equipotential earth connection .................... 36
Electrostatic discharge ................................. 37
Radiation Protection ................................................38
Monitoring of Personnel .............................. 38
Cleaning and Disinfecting ........................................ 39
Cleaning ...................................................... 40
Disinfecting ................................................. 42
Disinfecting safety directions ....................... 43
Approved disinfectants ................................ 44
Maintenance ............................................................45
Preventive maintenance schedule ................ 45
Safety Directions ..................................................................47
Basic Workflow ....................................................................50
Moving the unit ....................................................... 51
Starting the device ...................................................54
Positioning .............................................................. 55
ii | DR 100e (Film/CR) | Contents
0360B EN 20181015 0853

Setting the exposure parameters on the X-ray generator
control panel ............................................................57
Executing the exposure ............................................59
Using the wired exposure button ..................60
Using the remote control exposure button ....61
Stopping the device ................................................. 62
Advanced Operation ............................................................ 63
Using a moving grid .................................................64
Using a DAP meter ................................................... 65
Printing the DAP values ........................................... 66
Replacing the roll of paper ...........................66
Modifying the exposure parameters of an anatomic
program .................................................................. 68
Hanging a leaded apron ...........................................70
Product Information ............................................................ 71
Options and Accessories ...........................................72
Compatibility ...........................................................72
Compliance ..............................................................72
General ........................................................73
Safety .......................................................... 73
Electromagnetic Compatibility .....................73
Laser Safety ................................................. 73
Equipment Classification ..........................................74
Product Complaints ................................................. 75
Environmental protection ........................................ 76
Technical Data .........................................................77
DR 100e .......................................................78
Anatomic program exposure parameters ......83
DAP meter ................................................... 86
Collimator ................................................... 87
X-ray generator and X-ray tube assembly ......88
DAP value printer ........................................ 92
Remarks for HF-emission and immunity ..................93
Enclosure Port Immunity to RF wireless
communication equipment .......................... 98
Precautions on EMC .....................................99
Cables, transducers and accessories ........... 100
DR 100e (Film/CR) | Contents | iii
0360B EN 20181015 0853

Legal Notice
0413
Agfa NV, Septestraat 27, B-2640 Mortsel - Belgium
For more information on Agfa products, please visit www.agfa.com.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium or
its affiliates. DR 100e and NX are trademarks of Agfa NV, Belgium or one of its
affiliates. All other trademarks are held by their respective owners and are
used in an editorial fashion with no intention of infringement.
Agfa NV makes no warranties or representation, expressed or implied, with
respect to the accuracy, completeness or usefulness of the information
contained in this document and specifically disclaims warranties of suitability
for any particular purpose. Products and services may not be available for
your local area. Please contact your local sales representative for availability
information. Agfa NV diligently strives to provide as accurate information as
possible, but shall not be responsible for any typographical error. Agfa NV
shall under no circumstances be liable for any damage arising from the use or
inability to use any information, apparatus, method or process disclosed in
this document. Agfa NV reserves the right to make changes to this document
without prior notice. The original version of this document is in English.
Copyright 2018 Agfa NV
All rights reserved.
Published by Agfa NV
B-2640 Mortsel - Belgium.
No part of this document may be reproduced, copied, adapted or transmitted
in any form or by any means without the written permission of Agfa NV
4 | DR 100e (Film/CR) | Legal Notice
0360B EN 20181015 0853

Scope of this Manual
This manual contains information for safe and effective operation of the DR
100e mobile X-ray system, further referred to as the device.
DR 100e (Film/CR) | Introduction to this Manual | 7
0360B EN 20181015 0853

About the safety notices in this document
The following samples show how warnings, cautions, instructions and notes
appear in this document. The text explains their intended use.
DANGER:
A danger safety notice indicates a hazardous situation of direct,
immediate danger for a potential serious injury to a user,
engineer, patient or any other person.
WARNING:
A warning safety notice indicates a hazardous situation which
can lead to a potential serious injury to a user, engineer, patient
or any other person.
CAUTION:
A caution safety notice indicates a hazardous situation which
can lead to a potential minor injury to a user, engineer, patient
or any other person.
An instruction is a direction which, if it is not followed, can
cause damage to the equipment described in this manual or any
other equipment or goods and can cause environmental
pollution.
A prohibition is a direction which, if it is not followed, can cause
damage to the equipment described in this manual or any other
equipment or goods and can cause environmental pollution.
Note: Notes provide advice and highlight unusual points. A note is
not intended as an instruction.
8 | DR 100e (Film/CR) | Introduction to this Manual
0360B EN 20181015 0853

System Documentation
The documentation shall be kept with the system for easy reference.
The most extensive configuration is described within this manual, including
the maximum number of options and accessories. Not every function, option
or accessory described may have been purchased or licensed on a particular
piece of equipment.
Technical documentation is available in the product service documentation
which is available from your local support organization.
The most recent version of this document is available on http://
www.agfahealthcare.com/global/en/library/index.jsp
DR 100e (Film/CR) | Introduction to this Manual | 9
0360B EN 20181015 0853

Disclaimer
Agfa assumes no liability for use of this document if any unauthorized
changes to the content or format have been made.
Every care has been taken to ensure the accuracy of the information in this
document. However, Agfa assumes no responsibility or liability for errors,
inaccuracies or omissions that may appear in this document. To improve
reliability, function or design Agfa reserves the right to change the product
without further notice. This manual is provided without warranty of any kind,
implied or expressed, including, but not limited to, the implied warranties of
merchantability and fitness for a particular purpose.
Note: In the United States, Federal law restricts this device to sale
by or on the order of a physician.
10 | DR 100e (Film/CR) | Introduction to this Manual
0360B EN 20181015 0853

Intended Use
The DR 100e system is a mobile X-ray imaging system used in hospitals, clinics
and medical practices by physicists, radiographers and radiologists to make
static X-ray radiographic images of the skeleton (including skull, spinal
column and extremities), chest, abdomen and other body parts on adult,
pediatric or neonatal patients.
Applications can be performed with the patient in the sitting, standing or
lying position.
This device is not intended for mammography applications.
Intended User
This manual has been written for trained users of Agfa products and trained
diagnostic X–Ray clinical personnel who have received proper training.
Users are those persons who actually handle the equipment and those who
have authority over the equipment.
Before attempting to work with this equipment, the user must read,
understand, note and strictly observe all warnings, cautions and safety
markings on the equipment.
Training
The user must have received adequate training on the safe and effective use of
the system before attempting to work with it. Training requirements may vary
from country to country. The user must make sure that training is received in
accordance with local laws or regulations that have the force of law. Your local
Agfa or dealer representative can provide further information on training.
The user must note the following information in the system documentation:
• Intended Use.
• Intended User.
• Safety Directions.
12 | DR 100e (Film/CR) | Introduction
0360B EN 20181015 0853

Configuration
DR 100e is a mobile X-Ray System to perform X-ray exposures on CR Cassettes
or on film.
Figure 1: DR 100e mobile X-Ray System with rotating column (type 6012/002)
DR 100e (Film/CR) | Introduction | 13
0360B EN 20181015 0853

Figure 2: DR 100e mobile X-Ray System with fixed column (type 6012/001)
14 | DR 100e (Film/CR) | Introduction
0360B EN 20181015 0853

Operation Controls
7
8
9
10
1
6
5
4
3
2
1. Lock of the arm movement during transport
2. X-ray generator control panel
3. Exposure button
4. Circuit breaker and equipotential node
5. Power supply cable reel
6. Rotating column
7. X-ray generator and X-ray tube assembly
8. Goniometer
9. Collimator
10. X-ray tube handle
DR 100e (Film/CR) | Introduction | 15
0360B EN 20181015 0853

2
4
6
5
3
1
1. X-ray tube rotation movement lock
2. Arm
3. Handlebar with brake lever
4. Remote control receiver (optional)
5. Cassette holder
6. Tilting pedal
The operation controls on the system with fixed column are similar, apart
from the brake that is operated by pedals and not at the handlebar.
1
1
1. Brake pedals
Topics:
•Circuit breaker
•X-ray generator control panel
•Manual Collimator
•Exposure button
16 | DR 100e (Film/CR) | Introduction
0360B EN 20181015 0853

Circuit breaker
Figure 3: System with rotating column
Figure 4: System with fixed column
The equipment is protected by a circuit breaker against excessive mains
fluctuations.
To restore operation of the equipment after activation of the circuit breaker,
return the switch to position "I".
18 | DR 100e (Film/CR) | Introduction
0360B EN 20181015 0853

X-ray generator control panel
1 2 3 4 5
6
7
891011
12
13
14
15
1. Mains power indicator light.
The indicator lights up when the device is connected to the mains and
powered.
2. Turn the device on.
3. Control panel display for device status, X-ray parameters and messages.
4. The indicator lights up when the device is ready for exposure.
5. The indicator lights up when exposure is ongoing.
6. Reset the alarm.
Return to the upper menu.
7. Function buttons for selecting a line on the display.
8. Modify the exposure settings for mAs.
9. Modify the exposure settings for kV.
10. Open the list of anatomical programs.
Scroll down.
11. Activate remote control.
The indicator lights up when the remote control is enabled.
12. Small focus and large focus.
The indicator light shows the active setting.
13. Moving grid.
The indicator lights up when the signal for the moving grid is enabled.
14. Collimator light.
After pressing the button, the light field indicating the collimated area and
the laser light indicating the center position remain lit for a few seconds
before automatically switching off.
15. Turn the device off.
DR 100e (Film/CR) | Introduction | 19
0360B EN 20181015 0853

Manual Collimator
1
2
3
4
5
6
1. Longitudinal collimation
2. Button to switch on the light field indicating the collimated area and the
laser light indicating the center position.
After pressing the button, they remain lit for a few seconds before
automatically switching off.
3. Rails to insert a DAP meter or a filter.
4. Measurement tape to measure the distance between the focal spot of X-ray
tube and the tabletop.
The measurement tape is at the rear side of the collimator.
5. Transversal collimation
6. Filter selection.
Another button to switch on the light field is available on the X-ray generator
control panel.
20 | DR 100e (Film/CR) | Introduction
0360B EN 20181015 0853
Other manuals for DR 100e
3
This manual suits for next models
2
Table of contents
Other AGFA Medical Equipment manuals
Popular Medical Equipment manuals by other brands

AQUARION
AQUARION AQ100 Installation, operation & maintenance manual

SUNNEX
SUNNEX TRI-STAR TS2035M Installation and service manual

Medin
Medin C-NAIL Information on use

AmpliVox
AmpliVox Otowave 102-C operating manual

BMC
BMC G2S A20 Quick operation manual

Somnetics
Somnetics Transcend Sleep Apnea Therapy user manual

GE
GE Datex-Ohmeda Aisys Carestation Reference manual

Kinetec
Kinetec Centura B&W user manual

YOSHIDA
YOSHIDA X-era Smart Operation manual

Netatmo
Netatmo June manual

Mangar
Mangar ELK User instructions and warranty

Beijing Choice Electronic Technology
Beijing Choice Electronic Technology MD300C1C Operator's manual