2
Aesculap®
S4Cervical – Navigated Instruments
Aesculap®
S4Cervical – N avigated Instruments
Legend
ANavigated tapper ∅3.5 mm (FW655R)
BReference star unit (pre-calibrated) ?(Brainlab Art. no. 55830-20A)
CHandle (FW165R with ratchet or FW067R without ratchet)
DNavigated screwdriver for polyaxial screw (FW656R)
EReference star unit ML (calibration with ICM4) (Brainlab
Art. no. 55830-25A)
FNavigated drill guide ∅3.5 mm, (FW664R)
GNavigated drill guide ∅4.0 mm, (FW665R)
HAesculap star unit navigation attachment (FW652R)
INavigated guide sleeve ∅4.0 mm for smooth-shaft screws (FW658R)
JReduction sleeve ∅13 mm for Brainlab instrument calibration matrix
version 4 (FW657R)
KNavigated guide sleeve ∅3.5 mm/4.0 mm (FW660R)
LC1/C2 inner drill sleeve ∅4 mm (FJ985R)
MC1/C2 inner drill sleeve ∅3.5 mm (FW661R)
NC1/C2 drill ∅2.4 mm (FW662SU)
OC1/C2 tapper ∅3.5 mm (FW663R)
PCentering tool for navigated drill guide ∅3.5 mm FW664R (FW688R)
QCentering tool for navigated drill guide ∅4.0 mm FW665R (FW689R)
RDrill ∅2.4 mm for FW664R (FW666SU)
SDrill ∅2.9 mm for FW665R (FW667SU)
Symbols on product and packages
Applicable to
►For item-specific instructions for use and information on material
compatibility, see also the Aesculap Extranet at
https://extranet.bbraun.com
Intended use
The S4Cervical system (S4C system) is used for posterior occipital cervical
and thoracic stabilization and fusion. Indications and contraindications
are specified in the instructions for use for the implants (TA011796). The
instruments listed in the key form part of this system. They are used for
the individual adaptation, positioning and insertion of S4C implants in
patients. For safe usage, follow instructions for use for S4C System Instru-
ments (TA011984) and Operating Techniques (O34202/O83002).
The instruments listed in the key may only be used with the Brainlab nav-
igation system. For safe handling prior to the operation, read Spine &
Trauma user manual for Brainlab instruments and the corresponding soft-
ware manual for the Brainlab spine application used.
Combination specifications
Note
Aesculap and Brainlab accept absolutely no responsibility if instruments,
centering tools or drills other than those named below are used with the
corresponding drill guides and guide sleeves.
■Only combine S4C navigated drill guide Fwith:
–S
4C centering tool Pfor S4C drill guide F
–S
4C drill R∅2.4 mm for ∅3.5 mm screws
■Only combine S4C navigated drill guide Gwith:
–S
4C centering tool Qfor drill guide G
–S
4C drill S∅2.9 mm for ∅4.0 mm screws
■Only combine S4C navigated guide sleeve ∅4.0 mm Ifor smoothshank
screws with:
– Smooth-shaft screw bone awl (FW085R)
– Drill for ∅2.9 mm smooth-shaft screws (FW086SU)
– Smooth-shaft (Shank) screw tapper (Tap) (FW087R)
– Polyaxial screwdriver (FW070R/FW131R)
– Navigated screwdriver for polyaxial screw (FW656R)
– Apfelbaum ball head screwdriver, short (FJ968R)
■Only combine S4C navigated guide sleeve ∅3.5 mm/4.0 mm Kwith:
– Apfelbaum C1/C2 obturator (FJ983R)
– Apfelbaum trocar (FJ984R)
– Favored Angle screw drill ∅2.9 mm for ∅4mm screws
(FW088SU)
– Favored Angle tapper (Tap)∅4 mm (FW089R)
– C1/C2 inner drill sleeve ∅4 mm L(FJ985R)
–S
4C Favored Angle screwdriver (FW069R/FW132R)
– C1/C2 drill ∅2.4 mm for ∅3.5 mm screws (FW662SU) N
– C1/C2 tapper (Tap) ∅3.5 mm (FW663R) O
– C1/C2 inner drill sleeve ∅3.5 mm (FW661R) M
– Apfelbaum ball head screwdriver (FJ988R)
Note
K-wires generally may not be used with the S4C system.
Sterilization using irradiation
Not for reuse in intended applications as defined by the
manufacturer
Use by
Caution, general warning symbol
Caution, see documentation supplied with the product
Date of manufacture