AEA-2
AEA600F
AC-DC Power Supplies Medical Type
AE A 600 F -OO -O
Ordering information
R
1 2 3 4 5 6
Please refer to derating curve, because the rated load current depends on cooling method that is convection cooling or forced air.
*Make sure necessary tests will be carried out on your end equipment with the power supply installed in accordance with any required EMC/EMI regulations.
MODEL AEA600F-24 AEA600F-36 AEA600F-48
MAX OUTPUT WATTAGE[W] 600 601.2 600
DC OUTPUT (forced air)
ACIN 100V
24V 20.0 (Peak 42.0) A 36V 13.4 (Peak 28.0) A 48V 10.0 (Peak 21.0) A
ACIN 230V
24V 25.0 (Peak 52.5) A 36V 16.7 (Peak 35.0) A 48V 12.5 (Peak 26.3) A
SPECIFICATIONS
MODEL AEA600F-24 AEA600F-36 AEA600F-48
INPUT
VOLTAGE[V] AC85 - 264 1f(Output derating is required at AC85V - 170V. See “Derating”)
CURRENT[A]
ACIN 100V
5.7typ (Io=20A) 5.7typ (Io=13.4A) 5.7typ (Io=10A)
ACIN 230V
2.9typ (Io=25A) 2.9typ (Io=16.7A) 2.9typ (Io=12.5A)
FREQUENCY[Hz] 50/60 (45 - 66)
EFFICIENCY[%]
ACIN 100V
92.0%typ (Io=20A) 92.0%typ (Io=13.4A) 92.0%typ (Io=10A)
ACIN 230V
94.5%typ (Io=25A) 95.0%typ (Io=16.7A) 95.0%typ (Io=12.5A)
POWER FACTOR
ACIN 100V
0.98typ (Io=20A) 0.98typ (Io=13.4A) 0.98typ (Io=10A)
ACIN 230V
0.95typ (Io=25A) 0.95typ (Io=16.7A) 0.95typ (Io=12.5A)
INRUSH CURRENT[A]
*2
ACIN 100V
20/40typ (Io=20A) 20/40typ (Io=13.4A) 20/40typ (Io=10A)
ACIN 230V
40/40typ (Io=25A) 40/40typ (Io=16.7A) 40/40typ (Io=12.5A)
LEAKAGE CURRENT[mA] 0.3max (ACIN 240V 60Hz, Io=100%, According to IEC60601-1)
OUTPUT
VOLTAGE[V] 24 36 48
CURRENT (convection) [A]
ACIN 230V
17.5 11.7 8.8
CURRENT (forced air) [A]
ACIN 230V
25.0 16.7 12.5
PEAK CURRENT[A]
ACIN 230V
52.5 35.0 26.3
LINE REGULATION[mV] 96max 144max 192max
LOAD REGULATION[mV] 150max 240max 300max
RIPPLE[mVp-p]
*3
0 to +50C
120max 200max 200max
-20 to 0C
200max 300max 350max
RIPPLE NOISE[mVp-p]
*3
0 to +50C
150max 230max 250max
-20 to 0C
230max 350max 500max
TEMPERATURE REGULATION[mV]
0 to +50C
240max 360max 480max
DRIFT[mV]
*
4
96max 144max 192max
START-UP[ms] 750typ (ACIN 100V, Io=100%)
HOLD-UP[ms] 20typ (ACIN 230V, Io=100%)
OUTPUT VOLTAGE ADJUSTMENT RANGE[V]
21.6 to 26.4 32.4 to 39.6 43.2 to 52.8
OUTPUT VOLTAGE SETTING[V]
23.5 to 24.5 35.0 to 37.0 47.0 to 49.0
PROTECTION
CIRCUIT AND
OTHERS
OVERCURRENT PROTECTION
Works over 101% of peak current and recovers automatically
*
5
OVERVOLTAGE PROTECTION[V]
30 to 33.6 45 to 50.4 60 to 69.6
ALARM Optional (Input voltage alarm : PR, Output voltage alarm : PG)
REMOTE ON/OFF Optional
AUX1 Optional (12V1A)
AUX2 Optional (5V1A)
ISOLATION
INPUT-OUTPUT-PR-PG-RC-AUX *
6
AC4,000V 1minute, Cutoff current = 10mA, DC500V 50MΩmin (At Room Temperature) 2MOPP
INPUT-FG AC2,000V 1minute, Cutoff current = 10mA, DC500V 50MΩmin (At Room Temperature) 1MOPP
OUTPUT-PR-PG-RC-AUX-FG
*
6
AC1,500V 1minute, Cutoff current = 10mA, DC500V 50MΩmin (At Room Temperature) 1MOPP
OUTPUT-AUX1-PR-PG-RC-AUX2
*
6
AC100V 1minute, Cutoff current = 25mA, DC100V 10MΩmin (At Room Temperature)
ENVIRONMENT
OPERATING TEMP.,HUMID.AND ALTITUDE
-20 to +70C, 20 - 90%RH (Non condensing), 5,000m (16,500feet) max
STORAGE TEMP.,HUMID.AND ALTITUDE
-20 to +75C, 20 - 90%RH (Non condensing), 9,000m (30,000feet) max
VIBRATION 10 - 55Hz, 19.6m/s2(2G), 3minutes period, 60minutes each along X, Y and Z axis
IMPACT 196.1m/s2(20G), 11ms, once each X, Y and Z axis
SAFETY AND
NOISE
REGULATIONS
AGENCY APPROVALS
UL62368-1,ANSI/AAMI ES 60601-1, C-UL (equivalent to CAN/CSA-C22.2 No.62368-1,CAN/CSA-C22.2 No.60601-1)
EN62368-1,EN60601-1 3rd EN62477-1 (OVCIII) UL508 (Optional), Complies with IEC60601-1-2 4th Ed.
CONDUCTED NOISE Complies with FCC Part15 classB, VCCI-B, CISPR32-B, EN55011-B, EN55032-B
HARMONIC ATTENUATOR
*
7
Complies with IEC61000-3-2 (Class A)
OTHERS CASE SIZE/WEIGHT 41×127×186mm [1.61×5×7.32 inches] (W×H×D) (without terminal block) / 1.0kg max
COOLING METHOD Convection/Forced air
*1The listed options may affect the published standard specifications.
Please contact us for detailed product specification
*2
The current of input surge to a built-in EMI/EMS Filter (0.2ms or less) is excluded.
*3
Measured by 20MHz oscilloscope or Ripple-Noise meter (equivalent to KEISOKUGIKEN:RM104).
Please refer to the instruction manual 1.8.
*4Drift is the change in DC output for an eight hours period after a half-hour warm-up at 25C.
*5
The output is shut down when the overcurrent protection continues.
*6
Applicable when AUX and remote control (optional) is added.
*7
Please contact us about another class.
*
Sound noise may be generated by power supply in case of pulse load.
1Series name
2Single output
3Output wattage
4Universal input
5Output voltage
6Optional
*1
C : with Coating
N :
with cover
T : Vertical terminal block
J : Connector type
R3 :
with Subfeatures
(5V1A AUX,12V1A AUX
Remote ON/OFF
, Alarm)
T5 : UL508
P5 :
shutdown type
overcurrent protection
For option details, refer to instruction
manual 6.1.
* Use of an EMI/EMC filter is recommended
when a power supply is connected with several
devices so that additional filtering is necessary.
* Make sure that your final application will meet
the required EMC standard by measuring the
EMI level of the power supply used together
with an EMI/EMC filter.
High voltage pulse noise type : EAP series
Low leakage current type : EAM series
Example recommended EMI/EMC filter
EAC-20-472
2MOPP
May 25, 2021