manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Dräger
  6. •
  7. Respiratory Product
  8. •
  9. Dräger X-plore 9300 User manual

Dräger X-plore 9300 User manual

X-plore 9300
Airline belt manifold Instructions for Use
i
3363102 (A3-D-P)
1 For your safety
1.1 General safety statements
●Before using this product, carefully read the Instructions for Use.
●Strictly follow the Instructions for Use. The user must fully understand
and strictly observe the instructions. Use the product only for the
purposes specified in the Intended Use section of this document.
●Do not dispose of the Instructions for Use. Ensure that they are
retained and appropriately used by the product user.
●Only fully trained and competent users are permitted to use this
product.
●Comply with all local and national rules and regulations associated
with this product.
●Only trained and competent personnel are permitted to inspect, repair
and service the product. Dräger recommend a Dräger service contract
for all maintenance activities and that all repairs are carried out by
Dräger.
●Properly trained service personnel must inspect and service this
product as detailed in the Maintenance section of this document.
●Use only genuine Dräger spare parts and accessories, or the proper
functioning of the product may be impaired.
●Do not use a faulty or incomplete product, and do not modify the
product.
●Notify Dräger in the event of any component fault or failure.
1.2 Definitions of alert icons
Alert icons are used in this document to provide and highlight text that
requires a greater awareness by the user. A definition of the meaning of
each icon is as follows:
WARNING
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided,
could result in physical injury or damage to the product or
environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the product.
2 Description
2.1 Product overview
The Dräger X-plore 9300 (Fig 1) is belt mounted manifold that is used to
connect an external breathing air supply (an airline) to an item of headgear
worn by the user. The manifold has an input connector (5) to connect the
airline and has an output connector (4) to connect the headgear using a
corrugated hose (3) with a quick coupling on either end. The manifold has
a silencer which is a nylon mesh disc pressed into the top of the output
connector to limit the air noise in the headgear.
During use, a pressure reducer inside the manifold supplies a continuous
flow of air from the airline to the headgear. This air flow is within the safe
limits required at the headgear. The X-plore 9300 is available with or
without an adjustable flow control valve (2). When fitted, the flow control
valve allows the wearer to adjust the air flow within preset limits.
The whistle (1) is a low input pressure warning device. The whistle sounds
during use to alert the wearer if the breathing air supply falls below the
pressure required to provide adequate breathing air.
The X-plore 9300 is used with Dräger X-plore 8000 headgear including
hoods and protective helmets. The compatible Dräger hoods are classified
as Class 3B and the Dräger helmets are classified as Class 2B according
to EN 14594:2005. For a full description of the headgear see the
Instructions for Use supplied with the product.
Radio-frequency identification (RFID) is an option available for this product
to allow electronic asset management and tracking using a radio-
frequency reader. When fitted, the passive RFID tag is positioned inside
the manifold casing.
2.1.1 Breathing air supply
The breathing air supply (airline supply) is an external source of breathing
air from an air cylinder or compressor arrangement capable of supplying
the specified air quality, pressure and flow rates (see Section 3.1 for
details).
2.2 Intended use
The X-plore 9300 is intended for use with approved Dräger compressed
airline breathing equipment to provide respiratory protection against the
effects of harmful substances. The equipment is intended to be used only
for airline applications.
The compressed airline breathing equipment and other accessories used
with this product must be certified Dräger components, assembled in an
approved configuration; otherwise the operation of the device may be
impaired. Contact Dräger for further information.
Use in potentially explosive atmospheres
The X-plore 9300 is type tested as suitable for use in potentially explosive
atmospheres. Electronic sub-assemblies are ATEX certified. The
combinations are suitable for use in hazardous areas up to and including
zone 0 and zone 20. The product temperature class is T6 unless a passive
RFID tag is fitted in which case the temperature class is T4. The
combinations can be used in atmospheres containing gases of the gas
explosion group IIB.
The air supply must be located:
●In zone 1 if flammable gasses of class IIA are present.
●In zone 2 if flammable gasses of class IIA, IIB, or IIC are present.
There is no restriction for use in areas containing flammable dust.
2.3 Limitations on use
This product is not approved for use in CBRN (chemical, biological,
radiological and nuclear) applications.
2.4 Approvals
The European standards, guidelines, and directives according to which
this product is approved are specified in the declaration of conformity (see
declaration of conformity or www.draeger.com/product-certificates).
In addition the product is approved according to the following regulations:
●AS/NZS 1716:2012.
PED Important note: The X-plore 9300 is designed for use with airline
breathing-air supply systems only and is classified as safe and suitable for
use under the PED regulations – category SEP (Sound Engineering
Practice). The equipment cannot be certified to PED regulations.
2.5 Marking and symbols
The product type is indicated on the label on the manifold casing, and the
product serial number is marked on the body of the manifold.
Marking on the air supply hose indicates that the hose is heat resistant (H)
and/or antistatic (S).
3 Use
3.1 Breathing air supply
WARNING
The quality of the air supply shall meet the requirements for
breathing air according to EN 12021 in EU countries, or AS/
NZS 1715:2009 in Australia and New Zealand. Do not use oxygen
or oxygen-enriched air. The moisture content of breathing air
should be controlled within the specified limits in the standard to
avoid freezing the equipment.
Carry out a risk assessment of the workplace to ensure that it is
not possible to connect to any air supply other than breathable air
(e.g. Nitrox).
Adequate protection may not be provided by the equipment in
certain highly toxic atmospheres.
The user must ensure that the capacity of the air supply system is
sufficient for every user connected to it. The user must also ensure
that they can retreat safely if the supply from the airline fails.
Use a breathing air cylinder arrangement with a pressure reducer, or a
fixed or mobile compressor with a suitable filtration and air heating or
cooling system. Ensure that the air supply meets the air quality
requirements, and complies with the following pressure and flow
requirements:
●Input pressure 2.8 to 10 bar
●Input flow 200 to 1200 litres/minute (to provide the full range
of flow on versions with a flow control valve, a
minimum of 350 litres/minute is required)
The maximum number of users that can be connected simultaneously to
Dräger breathing air supply equipment is specified in the supply equipment
Instructions for Use.
It must be possible to move the air supply around a hazardous area if using
a breathing air cylinder arrangement. Conductive floors are mandatory in
zone 1 and zone 21 if flammable dusts with a minimum ignition energy of
less than 10 mJ are present.
3.1.1 Compressed air supply hose
●Ensure that the airline coupling is compatible with the product input
coupling. The standard input coupling is a Rectus 96 Series or CEJN-
type male quick coupling (other couplings are available from Dräger).
●The maximum working pressure of the Dräger compressed air supply
hose is 30 bar.
●Ensure that supply hoses with a total length greater than 50 m have an
appropriate earth connection.
EU countries
When using Dräger approved hose: a maximum of 100 m of hose may be
used, and no more than four individual hoses (a maximum of five hose
connections) are permitted in the airline.
Australia and New Zealand
Approved hoses must be used for the AS/NZS approval standard to be
valid. Hoses from 5 to 30 m or coupled to 90 m may be used.
3.1.2 Compressed air supply hose pressure drop
Using extension hoses causes a pressure drop between the breathing air
supply and the manifold. The pressure from the breathing air supply must
therefore be increased accordingly to compensate for any pressure drop
due to the length of hose used. The table below shows the pressure
required at the breathing air supply to ensure that the minimum pressure
of 2.8 bar is available at the manifold input connector.
This table is only applicable when using Dräger approved hose. When
using hose from any other manufacturer, the minimum input pressure of
2.8 bar must be supplied at the manifold input connector.
3.2 Preparation for use
3.2.1 Preparing the product
The manifold can be worn on the left or right side. To swap sides, loosen
the belt, remove the backplate if fitted and rotate the manifold (Fig 2).
When reassembling, ensure that the lip of the whistle cover sits under the
rib in the manifold bracket (Fig 3). Ensure that the position of the manifold
on the waist allows full head movement without pulling the hose, and that
the hose will not present a snag hazard.
1. Carry out a visual inspection (see Section 5.2).
2. Put on the belt and adjust as necessary. Ensure that the X-plore 9300
is fitted tightly against the body.
3. Connect the headgear hose to the output coupling and to the headgear
(Fig 4).
4. Connect the airline supply to the input coupling (Fig 5). The whistle will
sound briefly as the pressure increases.
5. Put on the headgear (see the Instructions for Use supplied with the
hood or helmet).
6. Ensure that the pneumatic hoses are not kinked or compressed by
clothing or equipment.
7. Take several deep breaths to ensure that sufficient air is available and
then breathe normally.
!
!
i
i
Hose length in
metres
Pressure in
bar
Hose length in
metres
Pressure in
bar
1–10 3.1 51–60 4.6
11–20 3.4 61–70 4.9
21–30 3.7 71–80 5.2
31–40 4.0 81–90 5.5
41–50 4.3 91–100 5.8
!
4439
4440
4528
4
4442 4429
6
MAXIMUM
MINIMUM
4446
1
12
3
4
5
3
2
5
X-plore 9300
Airline belt manifold Instructions for Use
i
3363102 (A3-D-P)
3363102
© Dräger Safety UK Limited
Edition 03 – August 2018 (Edition 01 – August 2014)
Subject to alteration
Draeger Safety UK Limited
Ullswater Close Tel +44 1670 352 891
Blyth, NE24 4RG Fax +44 1670 356 266
United Kingdom www.draeger.com
3.3 During use
NOTICE
Where hearing protection or sound attenuation communications
equipment is used, consider the reduction in audibility of the
whistle.
Flow control valve: Set the control knob to the minimum flow position at
the start of the task (Fig 6 shows the minimum and maximum positions).
Use the knob to adjust the air flow as necessary to suit your work intensity.
Whistle: If the whistle sounds, immediately cease the work activity and
leave the hazardous area by the shortest and safest route. Do not
disconnect the airline supply from the manifold. Remove the headgear as
soon as it is safe to do so.
3.3.1 In potentially explosive atmospheres
WARNING
Do not use or store the equipment near processes that produce a
charge.
Do not use the equipment unless conductive pars of the equipment
have an appropriate earth connection via the wearer.
Do not use materials in combination with the equipment that
influence the properties of the product.
3.4 After use
WARNING
Do not remove the equipment until you are in a non-hazardous and
safe breathing environment.
1. Remove the headgear (see the Instructions for Use supplied with the
headgear).
2. Disconnect the airline supply from the input coupling.
3. Disconnect the headgear hose from the output coupling and the
headgear inlet.
4. Open the waist belt buckle and remove the equipment.
5. Carry out the after use maintenance tasks in the maintenance table
(see Section 5.1).
4 Troubleshooting
The table shows fault diagnosis and repair information for product users.
Further troubleshooting information is available in Instructions for Use
supplied with associated equipment (e.g. the headgear or air supply
equipment).
Contact service personnel or Dräger if the symptom remains after the
remedy actions have been attempted.
5 Maintenance
5.1 Maintenance table
Dräger recommend that regular inspection, testing and servicing of the
breathing equipment is carried out in accordance with the table below. The
table applies also to out-of-use (stored) equipment. Record all
maintenance in the equipment log book. See also the maintenance
information for all equipment used (hood, headgear, etc.). Additional
inspection and testing may be required in the country of use to ensure
compliance with national regulations.
5.2 Visual inspection
Carry out a visual inspection, checking the full breathing equipment
including all component parts and accessories. Check that the equipment
is clean and undamaged, paying particular attention to pneumatic
components, hoses and connectors. Typical signs of damage that may
affect the operation of the breathing equipment include impact, abrasion,
cutting, corrosion and discoloration. Report damage to service personnel
and do not use the equipment until faults are rectified.
5.3 Cleaning and disinfecting
CAUTION
Do not exceed 60 °C for drying, and remove components from the
drying facility immediately when dry. Drying time in a heated dryer
must not exceed 30 minutes.
Do not immerse pneumatic or electronic components in cleaning
solutions or water. Contact Dräger for information about additional
cleaning methods for heavily soiled parts if required.
If water is trapped and then freezes inside the pneumatic system
of the breathing equipment, operation will be impaired. Prevent
any liquid from entering, and thoroughly dry the breathing
equipment after cleaning to prevent this from occurring.
Always clean Dräger X-plore equipment in a well ventilated room
or outdoors. There may be deposits of harmful substances in parts
of the equipment.
NOTICE
The X-plore 9300 has a leather belt when it is used with welding
headgear. Do not wash the leather belt using the cleaning agents
in this section – use a suitable leather cleaning product when
required.
For information about suitable cleaning and disinfecting agents
and their specifications refer to document 9100081 on
www.draeger.com/IFU.
See also the Instructions for Use for the hood or protective helmet, and
other associated equipment.
●Use only clean lint-free cloths
1. Clean the breathing apparatus manually using a cloth moistened with
cleaning solution to remove excess dirt.
○Disassemble and clean the parts separately if required. Fig 7
shows the parts disassembled and Fig 8 shows the belt routing
through the buckle during re-assembly to ensure that the belt can
be tightened.
2. Apply disinfecting solution to all internal and external surfaces.
3. Rinse all components thoroughly with clean water to remove all
cleaning and disinfecting agents.
4. Dry all components using a dry cloth, in a heated dryer or in air.
5. Contact service personnel or Dräger if disassembly of pneumatic or
electronic components is required.
6 Storage
Ensure that the storage environment is dry, free from dust and dirt, and
does not subject the equipment to wear or damage due to abrasion. Do not
store the equipment in direct sunlight. Route hoses in such a way that the
bend radius is not too acute and the hose is not stretched, compressed or
twisted.
See the storage temperature ranges in the technical data (Section 8).
7 Disposal
The product life is 10 years from the first use. When required, dispose of
the product in accordance with national or local regulations for waste
disposal.
8 Technical data
Weight (manifold and belt) < 0.5 kg
Belt size (standard belt) 38 mm x 750–1450 mm
Belt size (large belt) 38 mm x 750–1950 mm
Whistle volume >90 dB(A)
Input pressure 2.8 to 10 bar
Input flow 200 to 1200 litres/minute (to provide the full
range of flow on versions with a flow control
valve, a minimum of 350 litres/minute is
required)
Operating temperature -10 to +60 °C
Storage temperature -20 to +40 °C
Symptom Fault Remedy
Air leak Loose or dirty
connector
Disconnect, clean and
reconnect couplings and
retest
Faulty hose or
component
Substitute user
replaceable items and
retest
Whistle sounding or
insufficient air flow to
the wearer
Air supply restricted Remove any kinks or
restrictions from the
hoses
Faulty hose or
component
Substitute user
replaceable items and
retest
Input filter blocked Contact service personnel
or Dräger
Inoperative or poor
sounding whistle
Dirty whistle or
whistle cover
Clean the whistle flute
and cover as necessary
Faulty activation
mechanism
Contact service personnel
or Dräger
Excessive air noise Faulty or missing
silencer in the
manifold output
coupling
Check that the silencer is
fitted and undamaged.
Contact service personnel
or Dräger if replacement
is required
Hood over-inflates or
the screen fogs up
Blocked exhalation
strip
See the hood Instructions
for Use
i
i
!
!
Task Before
use
After
use
Every
year
Visual inspection (see Note 1 and
Section 5.2)
Notes
O Dräger recommendations
1 Clean the equipment if it is dirty. If it the equipment has been exposed
to contaminants, disinfect any components that come into direct and
prolonged contact with the skin.
OO
Functional test (see Note 2)
2 This maintenance task may only be carried out by Dräger or trained
service personnel. Details of the test are contained in the technical
manual which is issued to service personnel that have attended a
relevant Dräger maintenance course.
O
!
i
i
4433
8
7
'
'
4424

Other Dräger Respiratory Product manuals

Dräger PAS Micro CFH User manual

Dräger

Dräger PAS Micro CFH User manual

Dräger X-plore 6570 User manual

Dräger

Dräger X-plore 6570 User manual

Dräger X-plore Bayonett User manual

Dräger

Dräger X-plore Bayonett User manual

Dräger FPS-COM 7000 User manual

Dräger

Dräger FPS-COM 7000 User manual

Dräger X-plore 5500 User manual

Dräger

Dräger X-plore 5500 User manual

Dräger PARAT 3100 User manual

Dräger

Dräger PARAT 3100 User manual

Dräger PAS Micro Series User manual

Dräger

Dräger PAS Micro Series User manual

Dräger RBD 5000 User manual

Dräger

Dräger RBD 5000 User manual

Dräger Panorama Nova S - PE/ESA User manual

Dräger

Dräger Panorama Nova S - PE/ESA User manual

Dräger CDR 4500 User manual

Dräger

Dräger CDR 4500 User manual

Dräger Panorama Nova RP User manual

Dräger

Dräger Panorama Nova RP User manual

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger PAS Series User manual

Dräger

Dräger PAS Series User manual

Dräger UCF FireCore User manual

Dräger

Dräger UCF FireCore User manual

Dräger Panorama Nova S - ESA User manual

Dräger

Dräger Panorama Nova S - ESA User manual

Dräger Panorama Nova S User manual

Dräger

Dräger Panorama Nova S User manual

Dräger X-plore 4000 User manual

Dräger

Dräger X-plore 4000 User manual

Dräger PSS Series User manual

Dräger

Dräger PSS Series User manual

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger X-plore 8000 Series User manual

Dräger

Dräger X-plore 8000 Series User manual

Dräger X-plore 6000 User manual

Dräger

Dräger X-plore 6000 User manual

Dräger Panorama BelAir GC-RA User manual

Dräger

Dräger Panorama BelAir GC-RA User manual

Dräger saver CF Series User manual

Dräger

Dräger saver CF Series User manual

Dräger X-plore 8000 User manual

Dräger

Dräger X-plore 8000 User manual

Popular Respiratory Product manuals by other brands

ESAB EPR-X1 User instructions

ESAB

ESAB EPR-X1 User instructions

Dromex 6151 quick start guide

Dromex

Dromex 6151 quick start guide

ResMed AcuCare F1-0 user guide

ResMed

ResMed AcuCare F1-0 user guide

3M N95 Tips for use

3M

3M N95 Tips for use

Pari Lonhala Magnair Instructions for use

Pari

Pari Lonhala Magnair Instructions for use

Bullard PA1 user manual

Bullard

Bullard PA1 user manual

Wyndmere Classic instruction manual

Wyndmere

Wyndmere Classic instruction manual

Dpi Sekur SFERA Use and maintenance handbook

Dpi Sekur

Dpi Sekur SFERA Use and maintenance handbook

COMPASS HEALTH NEB-SOCCER user manual

COMPASS HEALTH

COMPASS HEALTH NEB-SOCCER user manual

MARTINDALE MAGNUM M23/1000 User instructions

MARTINDALE

MARTINDALE MAGNUM M23/1000 User instructions

Pari LC Sprint Instructions for use

Pari

Pari LC Sprint Instructions for use

G.Fox SAFE082A manual

G.Fox

G.Fox SAFE082A manual

Moldex 7000 Series instructions

Moldex

Moldex 7000 Series instructions

RPB PX4 AIR PAPR instruction manual

RPB

RPB PX4 AIR PAPR instruction manual

Medel Pro instruction manual

Medel

Medel Pro instruction manual

Omron NE-C900 instruction manual

Omron

Omron NE-C900 instruction manual

NIOSH SRS1010HDX quick start guide

NIOSH

NIOSH SRS1010HDX quick start guide

Lumiscope LUMINEB II 5710 Operation instructions

Lumiscope

Lumiscope LUMINEB II 5710 Operation instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.