Ecleris PROCAM SD User manual

USER MANUAL
PROCAM SD
CCD CAMERA AND CONSOLE
ABOUT THIS MANUAL
Version: 2020.04.01
Year: 2020
Language: English
All the rights to this manual are reserved, especially reproduction, publication and translation rights. No portion of this manual may be
reproduced by no means (photocopy, microfilm or other methods) without previous written authorization from ECLERIS. Neither can it
be reproduced, multiplied or published through any electronic data processing form.
The information contained in this manual may be modified or broaden without previous notification, without it representing any
responsibility to ECLERIS.
ECLERIS USA.
11801 NW 100th Rd
Suite 17
Medley, FL 33178, U.S.A.
Phone: +1 786 483 7490
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº18, CP29006,
Málaga, Spain
Ph. +34951214054

USER MANUAL –PROCAM SD
Page 2of 18
INMANUUIN201 Rev.2020.04.01 / ENG
TABLE OF CONTENT
1. SYMBOLS................................................................................................................................3
2. SAFETY INFORMATION.........................................................................................................3
2.1 WARNINGS..........................................................................................................................3
3. DESCRIPTION.........................................................................................................................5
3.1 INTENDED USE...................................................................................................................5
3.2 INTENDED USER ................................................................................................................5
3.4 COMPONENTS....................................................................................................................6
3.4.1. CONSOLE FRONT PANEL ...............................................................................................6
3.4.2. CONSOLE REAR PANEL..................................................................................................6
3.4.3. PROCAM SD BASIC CAMERA .........................................................................................8
3.4.4. PROCAM SD FULL CAMERA ...........................................................................................8
4. INSTALLATION AND SET UP .................................................................................................8
4.1. INSTALLATION....................................................................................................................8
4.2. COUPLING TO ECLERIS EVERLUX LIGHT SOURCE........................................................9
4.3. REVISION..........................................................................................................................11
5. OPERATION & FEATURES...................................................................................................12
5.1. WHITE BALANCE..................................................................................................................12
5.2. WINDOW FUNCTION............................................................................................................12
5.3. ANTI-MOIRE FUNCTION.......................................................................................................13
6. MAINTENANCE.....................................................................................................................14
6.1 CLEANING.........................................................................................................................14
6.2 ELECTRICAL CONNECTIONS..........................................................................................14
6.3 CHANGING THE FUSE......................................................................................................14
7. TECHNICAL SPECIFICATIONS............................................................................................15
8. TROUBLESHOOTING...........................................................................................................16
9. TECHNICAL SERVICE..........................................................................................................17
10. WARRANTY.......................................................................................................................18

USER MANUAL –PROCAM SD
Page 3of 18
INMANUUIN201 Rev.2020.04.01 / ENG
1. SYMBOLS
Read the related documentation
CE Mark
Serial number
Manufacturer Information
(Name, Address)
Warning
EU Authorized
Representative
Alternative current
BF type
Protect from water.
Do not discard with the
household waste
This side up
Protective Earth (Ground)
Enviroment transport conditions
Equipotential bonding
Pressure Limits
Date of Manufacture
(MM/AAAA)
Humidity Limits
Product code
Camera connection
Fragile
White Balance
Video output
Anti-Moiré Function
Fuse
Window Function
USB Port
2. SAFETY INFORMATION
2.1 WARNINGS
It is important to read this manual carefully before using the device.
ECLERIS will not be held responsible for damage that can occur to the equipment due to not
following the instructions or not observing the following criteria:
Keep the equipment out of children’s reach.
Do not install the equipment next to heat sources, such as heaters.
Install the equipment on a firm horizontal surface, in a well-ventilated area.
Protect the equipment and accessories from direct solar light, dust, humidity and vibrations
or excessive banging.

USER MANUAL –PROCAM SD
Page 4of 18
INMANUUIN201 Rev.2020.04.01 / ENG
Equipment not appropriate to be used near or around flammable anesthetic mixture with air,
oxygen or nitrous oxide.
The equipment does not need sterilization.
Do not put any container with liquids on top of the device.
Main power cord and PROCAM SD cable are resistant to deformations and mechanical
tensions however do not damage the cables and avoid violent shocks to them.
Do not use the equipment during electrical storms.
Unplug the equipment when it is not being used.
This equipment has to be connected to ground through a ground conductor in the power
cable. This ground connection is essential for the safe use of the equipment.
Do not connect the equipment to devices that do not comply with IEC60601-1ed 3.0 or later.
The device is provided with a connectable power cable that assures proper electrical
isolation. If it needs to be replaced, the replacement should not be longer that the original
cable and it has to be certified under IEC 60245-1 or IEC 60227-1.
In case of interference with other electro medical or electric equipment, we recommend
increasing the distance between PROCAM SD and the other equipment to minimize these
effects.
Do not open the equipment. Maintenance and repairs must be done by ECLERIS or an
authorized company. ECLERIS will not be responsible for the consequences of repairs or
maintenance carried out by non-authorized personnel.
In case of any problem, disconnect the power cable.
The device must be used only by personnel informed of all this warnings.
This instruction manual must be kept together with the equipment for future reference.
Should the equipment be sold or donated, this manual must be enclosed.
Any use outside of the manufacturer specific intent use, will exclude ECLERIS of all liability
If by any of the following:
Inadequate use of the equipment.
Failure to use the instruction manual.
Bad maintenance and care of the equipment.
Repair or manipulation by unauthorized personnel.
…personal or material damage is produced to a third party, ECLERIS, the manufacturer and
sales agent of this equipment, will not assume responsibility for said damage.
The device should never be discarded with the household waste. Please check local
regulations for electronic components disposal or recycling. It is the user’s
responsibility to deliver the system to the proper location for disposal or recycle of
electrical and electronic devices.

USER MANUAL –PROCAM SD
Page 5of 18
INMANUUIN201 Rev.2020.04.01 / ENG
3. DESCRIPTION
3.1 INTENDED USE
The device is intended for use in all types of diagnoses and surgeries that require real-time
capture of high sharpness and contrast images, where stereoscopy is not an exclusively
conditioning technique.
3.2 INTENDED USER
The device is intended to be used by a physician, paramedic or health care professional.
3.3 DESCRIPTION & FUNCTIONING
PROCAM SD video camera is a system that combines one camera head and a power and video
processing console. The camera head has a powerful 1/3"CCD color sensor and integrated
controls, making the unit easy to operate.
PROCAM SD video camera delivers sharp images with vivid colors thanks to an evolved
automatic gain control and shutter.
The CCD chip, which is located in the front of the camera head, captures the image and converts
the light signals into electrical signals; these signals are processed and optimized to obtain a
sharp, high-contrast image.
The image is processed in the central control unit (CCU) and is provided as CVBS signal in the
BNC connector in the back panel to be either projected on the monitor or other devices such as
video printer, video recorder, etc.
The camera has several controls that allow the selection, by single buttons, of different functions
such as White Balance, Anti-Moiré (to minimize the edges of the fiber optic grid when using
flexible fiber endoscopes) and Windows, (to minimize the light saturation when using small
diameter endoscopes).
PROCAM SD power supply (console) was specifically designed to be used together with video
cameras from ECLERIS. This Central Control Unit (CCU) gives power to the video camera, which
processes the video. Up to two imaging devices can be connected. PROCAM console supports
NTSC and PAL ECLERIS video cameras interchangeably without any prior configuration. The
HC220/1 console has an USB interface for its connection to the PC capture system.
PROCAM SD power supply can be fixed to Everlux light source manufactured by ECLERIS. It
can be placed hanging together with the light source down the colposcope, or microscope
forearm from ECLERIS. It can also be attached under an Everlux light source on a desk, cart or
any flat surface. It can be used also as a standalone unit as it includes rubber slip supports.
This symbol will be used to indicate a warning.
Read the manual carefully before using the device.
Should you have any doubts about this manual’s content, or the safe operation or security of the
device, please contact your local distributor or the nearest Ecleris office.

USER MANUAL –PROCAM SD
Page 6of 18
INMANUUIN201 Rev.2020.04.01 / ENG
3.4 COMPONENTS
The system is composed by:
Main unit (console)
Video camera
Main power cable
Video cable with BNC connector
3.4.1. CONSOLE FRONT PANEL
1. ON/OFF button.
2. PROCAM SD Connector
3.4.2. CONSOLE REAR PANEL
HC112G2/1 console
1. Power connector.
2. Equal-potential Ground connection.
3. Composite Video Output (CVBS)
4. Fuse holders.
2
1
1
3
3
4
2

USER MANUAL –PROCAM SD
Page 7of 18
INMANUUIN201 Rev.2020.04.01 / ENG
HC220 console
References:
1. Power connector.
2. Equal-potential Ground connection.
3. S-Video output
4. Composite Video Output (CVBS)
5. Ecleris Endodigi connector (image and video capture) and other
peripheral devices*
* Standard output (normally-open) to remotely operate other image capture
devices such as Video Printers.
HC220/1 console
Referencias:
1. Power connector.
2. Equal-potential Ground connection.
3. Composite Video Output (CVBS)
4. USB port
1
3
2
4
1
2
3
4
5

USER MANUAL –PROCAM SD
Page 8of 18
INMANUUIN201 Rev.2020.04.01 / ENG
3.4.3. PROCAM SD BASIC CAMERA
1. White balance button.
2. White balance LED.
3.4.4. PROCAM SD FULL CAMERA
1. White balance button.
2. Anti-Moiré button.
3. Window button.
4. White balance LED indicator.
5. Anti-Moiré LED indicator.
6. Window LED indicator.
4. INSTALLATION AND SET UP
4.1. INSTALLATION
Connection and revision of this equipment must be done by technical personnel with
electrical installation knowledge.
The PROCAM SD includes the following components. Check and identify them before proceeding
with the installation.
Console
Videocamera
Main power cable
Video cable with BNC connector
1) Unpack the device carefully.
2) Place the PROCAM SD console on a solid surface.
1
2
1
2
3
4
5
6

USER MANUAL –PROCAM SD
Page 9of 18
INMANUUIN201 Rev.2020.04.01 / ENG
3) If the building structure has equipotential grounding, connect the equipment to ground
through the equipotential ground terminal located on the rear panel of the unit.
4) Connect the video camera ECLERIS to the connector in the front panel.
5) Plug the video output connector on the rear panel of the PROCAM SD console to the
video input of the video monitor.
6) Connect the optical device to the camera head with the C-Mount screw (Endoscope,
Microscope, Colposcope, etc).
7) With the equipment off, connect power cable to the electrical socket
Use video monitors certified under IEC60601-1 ed 3.0 or later.
Use endoscopes compliant with IEC60601-2-18.
4.2. COUPLING TO ECLERIS EVERLUX LIGHT SOURCE
PROCAM SD console (CCU) can be attached to the ECLERIS Everlux light source or be used as
a standalone unit.
To attach the PROCAM SD console to the ECLERIS Everlux light source, you will need a 2,5
millimeters Allen key. With this Allen key, unscrew the two screws at the bottom of the PROCAM
SD console as shown in the picture.
By doing so, two heads of screws will emerge from the top of the PROCAM SD console. These
screws shall be anchored to the bottom side of the ECLERIS Everlux light source.
The base of the ECLERIS Everlux light source has two holes where the head of the screws of the
PROCAM SD console shall be introduced in order to attach it to the ECLERIS Everlux light
source.

USER MANUAL –PROCAM SD
Page 10 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
Once the head of the anchor screws are in place into the holes of the ECLERIS Everlux light
source base, you need to slightly push back the PROCAM SD console in order to attach it
completely to the ECLERIS Everlux light source.
Finally, to ensure a strong and stable coupling, tighten the screws located in the base of the
PROCAM SD console.
Once the coupling is finished, the combined set ECLERIS Everlux light source and PROCAM SD
console is ready to be used by itself or it can be attached to an ECLERIS microscope forearm or
to an ECLERIS colposcope forearm through the ECLERIS Everlux light source (after removing
the handle on top of the Everlux unit).

USER MANUAL –PROCAM SD
Page 11 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
In order to avoid the use of two electrical power sockets to supply both devices and the use of
many cables, ECLERIS has specially designed an interlock male-female cable of 24 centimeters.
With this cable, the PROCAM SD console can be supplied by the ECLERIS Everlux light source
directly. This cable is an optional accessory and it has to be requested by the user.
This cable connects both units as it shown in the picture.
4.3. REVISION
Once the installation is complete and before turning on the PROCAM SD camera we recommend
to perform the following controls:
Verify that the electrical circuit is connected to ground and in perfect condition, and that
the connection to the equipment is done perfectly too.
Confirm that the power supply fuses are in the socket, and that it is suitable.
Make sure no water or liquid supplies, or gases or inflammable fumes that can harm the
equipment are in the environment surrounding it.
The equipment has to be connected to the electrical socket through the power cable
supplied by ECLERIS.
The electrical supply has to be done at the electrical socket, without power adapters.
The equipment must be used only in places where the electrical installation complies with
local regulations.
If any abnormality is detected before turning on the equipment, please contact the technical
service of your local distributor or your nearest ECLERIS office before using it.

USER MANUAL –PROCAM SD
Page 12 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
5. OPERATION & FEATURES
There are detailed procedures for turning on and verifying the equipment functions in this section.
We recommended following them after the installation.
Only medical, paramedical or health personnel should use this equipment.
1) Press the power button in front of the PROCAM SD console. The LEDs of the PROCAM
SD camera will blink for few seconds and then they will turn off. It indicates that the
PROCAM SD camera head is being powered.
2) Verify that the video monitor is ON and in Video mode through its CVBS input.
3) Adjust the picture on the monitor according to your preferences of Brightness, Contrast,
Saturation, Color, HUE, Gamma etc.
4) Proceed with the white balance (see 5.1. section).
5) Adjust the level of Window and Anti-Moiré functions according to your preferences (see
5.2. and 5.3. sections)
6) Once the procedure is over, turn the device off in order to maximize the system lifetime
and energy saving.
5.1. WHITE BALANCE
To obtain the white balance, focus on a completely bright and white image lit with the maximum
intensity and press the White Balance button for approximately 3 seconds. This white balance
configuration will be stored in the memory even if the camera is off. While the white balance
function is working, the light on the camera function panel will flicker.
The proper adjustment of the white balance function guarantees the reproduction of natural
colors.
Perform a new white balance every time the light source is changed.
Perform the White Balance with the same light source that you will use in the intervention.
5.2. WINDOW FUNCTION
With the Window Function activated, only the surface that contains image information is used to
control the camera sensitivity, this way the camera adjustment and the image contrast is
optimized even for very thin optics and short work distances.
Once the object to be observed is in the center of the image, as with very thin diameter
endoscopes in which the outer edge of the image is black, activate the window function pressing
the Window button. The object will now be seen with the adequate lighting on the screen and the
outer margins of the image will appear proportionally brighter.
Contrary to the standard measuring where, the average brightness is intergraded on the whole
image, when the Window function is active, only the central part of the image is evaluated for an
average brightness, not taking into account the external margins.
When the Window function is activated, a light around the button turns on.

USER MANUAL –PROCAM SD
Page 13 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
If the object to be observed is centered in the monitor screen, as when thin endoscopes
are being used (<10mm), and where the edge of the image displayed is black, select the
Window function by pressing the corresponding button.
Window function indicator:
OFF → Window Deactivated (Area of interest: External, Medium and Central)
Blinking → Window Level 1 (Area of interest: Medium and Central)
ON → Window Level 2 (Area of interest: Central)
5.3. ANTI-MOIRE FUNCTION
The PROCAM HD camera has 2 levels of Anti-Moiré digital filters to optimize the images
obtained, minimizing the grid affect originated in the visualization of images taken from flexible
endoscopes considerably.
Anti-Moiré function indicator:
OFF → Anti-Moiré Deactivated
Blinking → Anti-Moiré level 1
ON → Anti-Moiré level 2
To active the first level of filtration press the Anti-Moiré button on the camera once. When the first
level is activated the green light around the button flickers.
To activate the second level of filtration, press the Anti-Moiré button again. When the second level
is activated the green light around the button stays on.
Press the Anti-Moiré button one more time to deactivate this option.
OFF
Level 1
Level
2
OFF
Level 1
Level
2

USER MANUAL –PROCAM SD
Page 14 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
6. MAINTENANCE
The following procedures are recommended to preserve PROCAM SD in optimum working
conditions:
Before the maintenance of the PROCAM SD, disconnect the power cable from the electric
power supply.
Periodic cleaning of all external components.
Checking the connections.
Annual technical verification of the equipment.
For any type of maintenance work or revision contact your local distributor or nearest ECLERIS
office.
6.1 CLEANING
To clean the outer surfaces of the device use only cotton or a piece of cloth dampened in mild
detergent or soap. Use only adequate cleaning products prepared for electrical equipment,
following the product’s instructions.
While cleaning, be careful not to let water or humidity get into the equipment.
Never use flammable products, especially alcohol, when the unit is hot.
Do not turn the equipment on until one (1) hour after cleaning process.
We recommend cleaning the external surface of the equipment every 15 days.
Do not sterilize any parts of the device in autoclave unless they are marked as
autoclavable.
Do no immerse the device or its accessories.
The cleaning, disinfection and sterilization of accessories not provided by ECLERIS must
be performed following the instructions provided by their manufacturer.
6.2 ELECTRICAL CONNECTIONS
Permanently check the condition of all the electrical components.
Check that there are no:
Cables in bad conditions; cut or torn.
Bad connection of the cable with the socket.
Rusty pins in the connectors.
Disconnected equipotential ground connection (in places with equipotential ground)
6.3 CHANGING THE FUSE
To avoid downtime, we recommend having spare fuses available.
1) Disconnect the power cable from the electric network and from the device.
2) Next to the power cord connector there are two plastic knobs where the fuses are housed.
Manually turning these knobs in the direction indicated by the arrows, both handles can be
removed, one at a time, to replace damaged fuses.
3) Remove the damaged fuses and replace them with the spare fuses, make sure to use the
correct fuses.

USER MANUAL –PROCAM SD
Page 15 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
Only use fuses as shown at the rear side of the PROCAM SD console and certified
by IEC60127.
4) Screw again the knobs.
5) Connect the power cord.
7. TECHNICAL SPECIFICATIONS
POWER SUPPLY
100 –240V
Nº OF PHASES
1 (single phase)
FREQUENCY
50 –60Hz
CLASS
I
TYPE
BF
OPERATION MODE
Continuous
STERILIZATION METHOD
The device can not be sterilized
GRADE OF WATER PROTECTION
IPX0
CONSUMPTION
15 VA
FUSES
2 x T2AL250V
VIDEO OUTPUT
2 x CVBS
S-video and CVBS
CVBS and USB
(HC112G2/1)
(HC220)
(HC220/1)
THERMAL PROTECTION
Automatic shutdown of power supply in case
of overheating
OPERATING CONDITIONS
50 –104ºF
35 –75% relative humidity
700 –1060 hPa
STORAGE AND TRANSPORT CONDITIONS
-22 –122ºF
10 –95% relative humidity
500 –1060 hPa
This system is not designed to be used in the presence of inflammable anesthetic gases
mixture with air, oxygen, nitrous oxide, or environments with high oxygen content.
The device is provided with a connectable power cable that assures proper electrical
isolation.
Do not sterilize.

USER MANUAL –PROCAM SD
Page 16 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
8. TROUBLESHOOTING
The following table indicates the most common problems and their solutions. Please consult this
table if you have any inconvenience.
As PROCAM SD console is always used in combination with a video monitor and a videocamera,
faults will be analyzed considering the whole system.
PROBLEM
PROBABLE CAUSE
ACTION
The equipment
will not turn on.
It is not connected to the
electric power supply.
Connect the cable to the electric
power supply.
The power cable is damaged
or cut.
Replace the power cable.
The plug is in bad conditions.
Replace the plug.
The green pilot
light on the switch
turns on but there
is no.
Monitor off.
Turn the monitor on.
Monitor disconnected.
Plug the monitor to the PROCAM
SD console through the provided
cord.
Monitor in wrong video mode.
Set the monitor in CVBS video
input.
The camera is not connected.
Connect the camera connector to
the front panel of the console.
Video camera incorrectly
connected.
Check if the video camera
connector is correctly and
completely connected to the
frontal panel of the PROCAM SD
console.
Image is in black
and white.
Camera is defective.
Please contact to the nearest
Technical Service Office.
Incorrect screen
programming.
Check that the video standard
programmed in the screen
matches the PROCAM video
standard (NTSC or PAL).
Image without
real colors.
Incorrect White Balance.
Adjust White Balance.
Blurred Image.
Lack of focus adjustment of
the external optical system.
Perform the focus adjustment of
the optical system.
Anti-Moiré function is
If you don’t use flexible scope,

USER MANUAL –PROCAM SD
Page 17 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
activated.
deactivate Anti Moiré function.
Image is too
bright.
Misadjusted monitor.
Calibrate properly the image
levels of the monitor.
Misadjusted Window Function.
Change the levels of Window
Function to adjust the image.
Image is too dark.
Not enough light.
Increase the levels of intensity of
illumination of the light source.
Verify that the video monitor is
configured with adequate levels
of brightness.
Verify that the external optical
system of the video camera is
not dirty.
If the inconvenience cannot be solved, please contact your local distributor or the nearest
ECLERIS office and ask for technical service.
9. TECHNICAL SERVICE
PROCAMSD should be repaired and/or maintained exclusively by ECLERIS trained and
qualified personnel in authorized installations and only the parts stated as repairable by
ECLERIS can be repaired. ECLERIS will provide, under request, circuits, list of components,
calibration instructions or any other information necessary in order to assist the ECLERIS
authorized technical personnel.
We recommend using the following procedure to expedite repair service and reposition of
products under guarantee.
1) Contact the technical service department of your local distributor or nearest ECLERIS
office.
2) Provide detailed information on the problem.
3) If the problem cannot be solved through troubleshooting, it must be sent for repair.
The following information should be sent with the product:
Owners name
Owners address
Person to contact and a telephone number
Product serial number
Problem description
The shipment should be done in the original packaging to avoid damage to the equipment
during the transportation. ECLERIS will not be responsible for damage caused during the
transportation due to improper packaging.

USER MANUAL –PROCAM SD
Page 18 of 18
INMANUUIN201 Rev.2020.04.01 / ENG
10. WARRANTY
ECLERIS declares that the PROCAMSD has passed our Quality Control and no defects have
been found in its manufacture, therefore guarantees proper working order under the adequate
conditions that are indicated in this manual. The guarantee is for the term of one (1) year
starting on the purchase date.
ECLERIS shall not be responsible for damage caused by/to the PROCAM SD and its power
supply for inadequate use, neglect, incorrect cleaning or storage. Any repairs carried out by
anyone other than ECLERIS authorized personal will annul this warranty. This warranty is only
valid for the original buyer.
Regardless of the warranty state, all shipping costs to and from ECLERIS authorized
installations are responsibility of the client exclusively.
Table of contents
Other Ecleris Medical Equipment manuals

Ecleris
Ecleris STROBOLIGHT User manual

Ecleris
Ecleris SPIDER User manual

Ecleris
Ecleris CAVIX User manual

Ecleris
Ecleris FRAXFACE User manual

Ecleris
Ecleris PROCAM HD 3D User manual

Ecleris
Ecleris XENOLUX 180 User manual

Ecleris
Ecleris C100HD User manual

Ecleris
Ecleris VNG Plus User manual

Ecleris
Ecleris PORTALED User manual

Ecleris
Ecleris STROBOLED User manual

Ecleris
Ecleris PROCAM HD User manual

Ecleris
Ecleris MICROSTAR C100 User manual

Ecleris
Ecleris EVERLUX MINI-PRO LED111 User manual

Ecleris
Ecleris MINIVAC User manual

Ecleris
Ecleris EVERLUX User manual

Ecleris
Ecleris MICROSTAR C-100A5E User manual

Ecleris
Ecleris PROCAM HD 2D User manual

Ecleris
Ecleris MINIVAC+ User manual