NIMBUS®3 & NIMBUS®3 PROFESSIONAL
Quick Reference Guide
Power Switch & Alarm Reset - Switches the mains power to the pump on and off.
When the pump detects an alarm condition, it can be cancelled by switching the pump off and then back on.
The Wait indicator is illuminated when the mattress is being inated. The indicator will remain illuminated until the
mattress has been fully inated. This may take up to 15 minutes.
Alarms - During an alarm condition the triangular alarm symbol will ash, an audible alarm will sound and one of the
following indicators will be illuminated: Low Pressure, High Pressure, Pump Fault, Power (Fail).
The Low Pressure indicator is illuminated whenever the pump detects low pressure within the mattress. An audible
alarm will sound unless cancelled by the Mute button. The indicator will extinguish once normal pressure is reached.
Make sure the CPR is closed, the tubeset is connected properly and the mattress is set to Normal.
Press the Static button to select Static mode. An audible tone will sound when the button is pressed. The indicator on
the Static button will only illuminate when the pump is in Static mode. Press the button again to return to Dynamic
mode.
Pump Fault - Internal pump malfunction. Call Service Engineer.
The Power (Fail) alarm indicator will illuminate when a mains power failure has been detected. An audible alarm will
sound. The alarm will continue until the mains power is resumed or the pump is switched off using the Power switch
on the pump. If the power failure is prolonged, switch to Transport mode and disconnect the tubing.
Examples of the Power (Fail) alarm are mains lead removal and power failure.
The mattress cell pressure can be manually adjusted for patient comfort using the rotary Comfort Control dial.
The High Pressure indicator is illuminated whenever the pump detects high pressure within the mattress. An audible
alarm will sound unless cancelled by the Mute button. The indicator will extinguish once normal pressure is reached.
Make sure the tubeset is not kinked and the Auto-Matt® sensor pad is at and not kinked.
The Service indicator will illuminate and remain on after a set number of running hours indicating the pump is ready
for a service. The pump will continue to function normally while the Service indicator is illuminated.
During an alarm condition the audible alarm can be muted by pressing the Mute button. The indicator on the Mute
button will illuminate to show the audible alarm has been silenced. After approximately 15 minutes the alarm mute
will be cancelled, and the audible alarm will sound.
This Quick Reference Guide has been produced to assist Healthcare Professionals and other users
with only the functional aspects of the system. The guide does not replace the general safety warnings,
cautions and recommendations provided in the product instructions for use, part number 151996EN.
Refer to the product instructions for use before using this product.
Note: The Auto-Matt sensor consists of a exible air sensor pad which is located underneath the mattress cells.
Contraindication: Do not use Nimbus 3 and Nimbus 3 Professional systems for patients with unstable spinal fractures.
Caution: If patients have other unstable fractures, or conditions which may be complicated by a soft or moving surface,
advice should be sought from an appropriate clinician before use.
Note: For more information on putting the patient in the Prone position on the Nimbus 3 Professional mattress refer to the
Instructions for Use document, section “Patient Positioning Guidance for the Nimbus 3 Professional Mattress”.
Indications:
• The Nimbus 3 and Nimbus 3 Professional mattresses are designed for patients weighing up to 250 kg (550 lb).
• The Nimbus 3 and Nimbus 3 Professional systems are indicated for the prevention and/or management of all
categories1of pressure ulcer, when combined with an individualised, comprehensive pressure ulcer protocol.
1. NPUAP/EPUAP International Pressure Ulcer Guideline, 2009.
Installation
Nimbus 3 & Nimbus 3 Professional Mattress Replacement (MR)
Place the MR directly onto the bed base, with the tubeset and
CPR located at the foot end of the bed.
Attach the MR to the bed frame using the hook
and loop securing straps.
To complete mattress installation
If the pump is to be hung from
the end of the bed, attach the
bed bracket to the bed frame.
Alternatively the pump can be
placed underneath the bed, either
upright or lying on its back.
Quick Start
When the Nimbus 3 pump is switched on using the Power switch,
the pump will run a self diagnostic check for approximately 3 seconds
during which all indicators will be illuminated.
The Wait and Low Pressure indicators will remain illuminated
while the mattress inates. Allow approximately 15 minutes for the
mattress to inate fully.
Transport Mode
To transport a patient using the Nimbus 3 mattress, turn the
Transport control at the foot end of the mattress clockwise to
Transport. Switch off the pump and disconnect the tubeset.
In this mode the mattress will support the patient for up to
12 hours.
To resume operation, reconnect the pump, turn the transport
control counterclockwise to Normal and switch on.
ArjoHuntleigh AB
Verkstadsvägen 5
241 38 Eslöv
SWEDEN
www.arjohuntleigh.com
® and ™ are trademarks of ArjoHuntleigh
As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.
151912EN_02: 10/2012
CPR Control
IN THE EVENT OF CARDIAC ARREST
In the event of a patient suffering cardiac arrest and
CPR needing to be administered:
To Activate the CPR
1. Lift the red handle on the control panel at foot end
of the mattress.
SAFETY SIDES
Regarding the use of safety sides, always refer to your local
trust policy AND the product instructions for use.
If the bed has divided sections for independent
elevation of a patient’s head and/or knees, attach the
mattress to the movable parts of the bed frame only.
Zip the protective cover over the mattress. Make sure that the
ArjoHuntleigh logo is uppermost and at the foot end of the mattress.
Check that the transport control is set to Normal and the CPR
control is closed and locked in position.
Make sure that the mains power cable and tubeset are positioned
to avoid causing a trip or other hazard, and are clear of
moving bed mechanisms or other possible entrapment areas.
The CPR control must be visible and accessible at all times.
Make sure the tubeset is securely connected to the pump and
the mattress.
Connect the mains power cable to a suitable mains power outlet.
2. Turn the handle counterclockwise.
3. Pull the handle away from the panel.
To Reset the CPR
1. Turn the grey triangular seal clockwise and push
onto the connectors.
2. Turn the red handle clockwise.
3. Fold the handle at to lock in position.
Vent Valve Control (Nimbus 3 Professional Only)
Each of the 16 cells in the leg and
torso section of the mattress can be
individually deated using the Vent
Valve Controls. To deate a cell, turn
the Vent Valve 180° clockwise, to point
downwards.
Caution: Do not deate adjacent cells or more than 3 cells
at any one time.
Head Section Deate (Nimbus 3 Professional Only)
To deate the three cells at the head end of the bed, turn the
Prole Control Valve to the Head Section Deate position.
4. The grey triangular seal will rotate and the air will
rapidly exhaust from the mattress.
To Connect
Tubeset
To Disconnect
Tubeset
1
2
1
2
Once the mattress is fully inated, the Wait and Low Pressure
indicators will extinguish. Place a bed sheet over the mattress and
tuck in loosely.
Once the patient is on the mattress, the pump will automatically
sense and adjust the pressure in the cells to support the patient.
Warning: Do not place the patient on the mattress until it is
fully inated and normal operating pressure has
been reached.
Transport
Mode
Normal
Transport
Normal
Mode
Normal
Transport
Closed Open
Head Section Deate
Normal
CPR
1
3
NORMAL
TRANSPORT
2
FAST DEFLATE
CPR
1
3
NORMAL
TRANSPORT
2
FAST DEFLATE
1
3
NORMAL
TRANSPORT
2
FAST DEFLATE
CPR
Some older Nimbus 3 Systems may have the old style tubeset. If
this is the case DO NOT t as above. Instead see below.
1
2
1
2
To Connect
Tubeset
To Disconnect
Tubeset

Other manuals for Arjohuntleigh Nimbus 3 Professional

2

This manual suits for next models

1

Other Getinge Medical Equipment manuals

Getinge Arjohuntleigh Parker User manual

Getinge

Getinge Arjohuntleigh Parker User manual

Getinge Volista User manual

Getinge

Getinge Volista User manual

Getinge ArjoHuntleigh Flowtron ACS900 User manual

Getinge

Getinge ArjoHuntleigh Flowtron ACS900 User manual

Getinge Arjohuntleigh Akron Bobath 4 User manual

Getinge

Getinge Arjohuntleigh Akron Bobath 4 User manual

Getinge Flow-i User manual

Getinge

Getinge Flow-i User manual

Getinge CS100 User manual

Getinge

Getinge CS100 User manual

Getinge Blue 80/80 User manual

Getinge

Getinge Blue 80/80 User manual

Getinge Arjohuntleigh Maxi Sky 600 ECS User manual

Getinge

Getinge Arjohuntleigh Maxi Sky 600 ECS User manual

Getinge Arjohuntleigh Citadel Plus User manual

Getinge

Getinge Arjohuntleigh Citadel Plus User manual

Getinge Maquet FLOW-i 4.2 User manual

Getinge

Getinge Maquet FLOW-i 4.2 User manual

Getinge Arjohuntleigh Nimbus 3 Professional User manual

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional User manual

Getinge MAQUET HLX 3000 User manual

Getinge

Getinge MAQUET HLX 3000 User manual

Getinge 9900 Series User manual

Getinge

Getinge 9900 Series User manual

Getinge LUCEA 10 User manual

Getinge

Getinge LUCEA 10 User manual

Getinge MAQUET CARDIOHELP User manual

Getinge

Getinge MAQUET CARDIOHELP User manual

Getinge Arjohuntleigh Maxi Air User manual

Getinge

Getinge Arjohuntleigh Maxi Air User manual

Getinge Lucea 50 User manual

Getinge

Getinge Lucea 50 User manual

Getinge Arjohuntleigh ALPHA RESPONSE User manual

Getinge

Getinge Arjohuntleigh ALPHA RESPONSE User manual

Getinge Maquet Atrium OCEAN User manual

Getinge

Getinge Maquet Atrium OCEAN User manual

Getinge Maquet PowerLED II User manual

Getinge

Getinge Maquet PowerLED II User manual

Getinge Arjohuntleigh Akron Streamline 2 Section User manual

Getinge

Getinge Arjohuntleigh Akron Streamline 2 Section User manual

Getinge ARJOHUNTLEIGH AURA User manual

Getinge

Getinge ARJOHUNTLEIGH AURA User manual

Getinge Volista User manual

Getinge

Getinge Volista User manual

Getinge Arjohuntleigh NIMBUS II User manual

Getinge

Getinge Arjohuntleigh NIMBUS II User manual

Popular Medical Equipment manuals by other brands

Optec 5000 Getting to know

Optec

Optec 5000 Getting to know

Gima OXY-4 Use and maintenance book

Gima

Gima OXY-4 Use and maintenance book

Medifa 81220 user manual

Medifa

Medifa 81220 user manual

IngMar Medical ASL 5000 quick guide

IngMar Medical

IngMar Medical ASL 5000 quick guide

Gate Bure Standard EL 56-312T Instructions for use

Gate

Gate Bure Standard EL 56-312T Instructions for use

Alpinion Medical Systems E CUBE 8 quick guide

Alpinion Medical Systems

Alpinion Medical Systems E CUBE 8 quick guide

Thomashilfen A chassis Series user manual

Thomashilfen

Thomashilfen A chassis Series user manual

Nasco Healthcare 2271 instruction manual

Nasco Healthcare

Nasco Healthcare 2271 instruction manual

Cristofoli vitale 21 manual

Cristofoli

Cristofoli vitale 21 manual

Nellcor OxiMax N-550 Service manual

Nellcor

Nellcor OxiMax N-550 Service manual

HEINE mini-c Cliplamp instructions

HEINE

HEINE mini-c Cliplamp instructions

laerdal V-Vac manual

laerdal

laerdal V-Vac manual

Merit Medical ONE Snare Endovascular Microsnare System Instructions for use

Merit Medical

Merit Medical ONE Snare Endovascular Microsnare System Instructions for use

Arrow MAC manual

Arrow

Arrow MAC manual

COOK Medical Duette DT-6-5F Quick reference guide

COOK Medical

COOK Medical Duette DT-6-5F Quick reference guide

Huntleigh Smartsigns Compact SC1000 Instructions for use

Huntleigh

Huntleigh Smartsigns Compact SC1000 Instructions for use

TransLite Veinlite LED quick start guide

TransLite

TransLite Veinlite LED quick start guide

Emerson Albr BDS-256XL Specification sheet

Emerson

Emerson Albr BDS-256XL Specification sheet

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.