Hameg HM1507-3 User manual

3
Subject to change without notice
Table of contents
Oscilloscope
HM1507-3.02
St.250900-Hüb/tke
General information regarding the CE marking .......... 4
Specifications ................................................................... 5
General Information ........................................................ 6
Symbols ........................................................................ 6
Use of tilt handle .......................................................... 6
Safety ............................................................................ 6
EMC .............................................................................. 7
Warranty ....................................................................... 7
Maintenance ................................................................. 7
Protective Switch-Off ................................................... 7
Power supply ................................................................ 7
Type of signal voltage ..................................................... 8
Amplitude Measurements ............................................ 8
Total value of input voltage .......................................... 9
Time Measurements .................................................... 9
Connection of Test Signal........................................... 10
Controls and Readout .................................................... 11
First Time Operation ...................................................... 27
Trace Rotation TR ....................................................... 27
Probe compensation and use .................................... 27
Adjustment at 1kHz .................................................... 28
Adjustment at 1MHz .................................................. 28
Operating modes of the vertical
amplifiers in Yt mode ................................................. 28
X-Y Operation .............................................................. 29
Phase comparison with Lissajous figures ................. 29
Phase difference measurement
in DUAL mode (Yt) ...................................................... 29
Phase difference measurement in DUAL mode ....... 30
Measurement of an amplitude modulation ............... 30
Triggering and time base .............................................. 30
Automatic Peak (value) -Triggering ............................ 31
Normal Triggering ....................................................... 31
- Slope .................................................................... 31
Trigger coupling ........................................................... 31
Triggering of video signals .......................................... 32
Line / Mains triggering (~) .......................................... 32
Alternate triggering ..................................................... 32
External triggering ...................................................... 33
HOLD OFF-time adjustment ...................................... 33
B time base (2nd time base) / .................................... 33
Triggering after Delay ................................................. 33
AUTO SET ....................................................................... 34
Component Tester (analog mode) ............................... 34
General ........................................................................ 34
Using the Component Tester ..................................... 35
Test Procedure ............................................................ 35
Test Pattern Displays .................................................. 35
Testing Resistors ........................................................ 35
Testing Capacitors and Inductors ............................... 35
Testing Semiconductors ............................................. 35
Testing Diodes ............................................................ 35
Testing Transistors ...................................................... 36
In-Circuit Tests ............................................................ 36
Storage mode ................................................................. 36
Signal recording modes .............................................. 37
Vertical resolution ....................................................... 37
Horizontal resolution ................................................... 37
Maximum signal frequency in storage mode ............ 37
Alias signal display ...................................................... 38
Test Instructions ............................................................. 38
General ........................................................................ 38
Cathode Ray Tube:
Brightness and Focus,
Linearity, Raster Distortion ......................................... 38
Astigmatism Check .................................................... 38
Symmetry and Drift of the Vertical Amplifier ............ 38
Calibration of the Vertical Amplifier ............................ 38
Transmission Performance ......................................... 39
of the Vertical Amplifier .............................................. 39
Operating Modes: CH.I/II, DUAL, ADD,
CHOP., INVERT and X-Y Operation ............................ 39
Triggering Checks ....................................................... 39
Time base .................................................................... 40
Hold Off time .............................................................. 40
Component Tester ...................................................... 40
Trace Alignment .......................................................... 40
Adjustments ................................................................ 40
RS232 Interface - Remote Control ............................... 40
Safety .......................................................................... 40
Operation .................................................................... 40
Baud-Rate Setting ....................................................... 40
Data Communication .................................................. 40
Front control HM1507-3 ................................................ 41

Subject to change without notice
4
KONFORMITÄTSERKLÄRUNG
DECLARATION OF CONFORMITY
DECLARATION DE CONFORMITE
®
Instruments
Herstellers HAMEG GmbH
Manufacturer Kelsterbacherstraße 15-19
Fabricant D - 60528 Frankfurt
Bezeichnung / Product name / Designation:
Oszilloskop/Oscilloscope/Oscilloscope
Typ / Type / Type: HM1507-3
mit / with / avec: -
Optionen / Options / Options: HO79-6
mit den folgenden Bestimmungen / with applicable regulations / avec les
directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
Angewendete harmonisierte Normen / Harmonized standards applied / Normes
harmonisées utilisées
Sicherheit / Safety / Sécurité
EN 61010-1: 1993 / IEC (CEI) 1010-1: 1990 A 1: 1992 / VDE 0411: 1994
EN 61010-1/A2: 1995 / IEC 1010-1/A2: 1995 / VDE 0411 Teil 1/A1: 1996-05
Überspannungskategorie / Overvoltage category / Catégorie de surtension: II
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility
Compatibilité électromagnétique
EN 50082-2: 1995 / VDE 0839 T82-2
ENV 50140: 1993 / IEC (CEI) 1004-4-3: 1995 / VDE 0847 T3
ENV 50141: 1993 / IEC (CEI) 1000-4-6 / VDE 0843 / 6
EN 61000-4-2: 1995 / IEC (CEI) 1000-4-2: 1995 / VDE 0847 T4-2
Prüfschärfe / Level / Niveau = 2
EN 61000-4-4: 1995 / IEC (CEI) 1000-4-4: 1995 / VDE 0847 T4-4:
Prüfschärfe / Level / Niveau = 3
EN 50081-1: 1992 / EN 55011: 1991 / CISPR11: 1991 / VDE0875 T11: 1992
Gruppe / group / groupe = 1, Klasse / Class / Classe = B
Datum /Date /Date Unterschrift / Signature /Signatur
23.04.1999
G. Hübenett
QMB
General information regarding the CE marking
HAMEG instruments fulfill the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual
generic- and product standards. In cases where different limit values are applicable, HAMEG applies the severer standard. For
emission the limits for residential, commercial and light industry are applied. Regarding the immunity (susceptibility) the limits
for industrial environment have been used.
The measuring- and data lines of the instrument have much influence on emmission and immunity and therefore on meeting the
acceptance limits. For different applications the lines and/or cables used may be different. For measurement operation the
following hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) sufficiently screened
cables must be used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a
dataline must be less than 3 meters long. If an interface has several connectors only one connector must have a connection to
a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L
from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without
instruction in the manual for a shorter length, signal lines must be less than 3 meters long. Signal lines must screened (coaxial
cable - RG58/U). A proper ground connection is required. In combination with signal generators double screened cables (RG223/
U, RG214/U) must be used.
3. Influence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fields, even with careful setup of the measuring equipment an
influence of such signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding
the instruments specifications may result from such conditions in individual cases.
4. RF immunity of oscilloscopes.
4.1 Electromagnetic RF field
The influence of electric and magnetic RF fields may become visible (e.g. RF superimposed), if the field intensity is high. In most
cases the coupling into the oscilloscope takes place via the device under test, mains/line supply, test leads, control cables and/
or radiation. The device under test as well as the oscilloscope may be effected by such fields.
Although the interior of the oscilloscope is screened by the cabinet, direct radiation can occur via the CRT gap. As the bandwidth
of each amplifier stage is higher than the total –3dB bandwidth of the oscilloscope, the influence RF fields of even higher
frequencies may be noticeable.
4.2 Electrical fast transients / electrostatic discharge
Electrical fast transient signals (burst) may be coupled into the oscilloscope directly via the mains/line supply, or indirectly via
test leads and/or control cables. Due to the high trigger and input sensitivity of the oscilloscopes, such normally high signals
may effect the trigger unit and/or may become visible on the CRT, which is unavoidable. These effects can also be caused by
direct or indirect electrostatic discharge.
HAMEG GmbH

5
Subject to change without notice
The 150 MHz (200MS/s) Analog-/Digital-Oscilloscope HM1507-3
Autoset
Auto Cursor
Readout / Cursor
Save / Recall
2 Reference Memories
Dual Time Base
Component Tester
1kHz/1MHz Calibrator
RS232 Interface
Digital:
■■
■■
■Refresh, Single, Roll-, Envelope-, Average-,XY-Mode
■■
■■
■Max. Sampling Rate 200MS/s, Storage 2x2048x8 bit
■■
■■
■Time Base A: 100s - 50ns/div., B: 20ms - 50ns/div.
■■
■■
■Pre Trigger 25-50-75-100%, Post Trigger 25-50-75%
■■
■■
■Screen Refresh 180/s, Dot Join (linear)
Analog:
■■
■■
■2 x DC to 150MHz, 2 x 1mV-50V/div
■■
■■
■Time Base A with Trig. DC to 250MHz
■■
■■
■Time Base B with 2ndTrig. to 250MHz
■■
■■
■Trig. DC to 250MHz, TV Sync. Separator
■■
■■
■1kHz/1MHz Calibrator, CRT with 14kV
Specifications
Vertical Deflection
Operating modes: Channel I or II separate
both Channels (alternated or chopped)
Chopper frequency: approx. 0.5MHz)
Sum or Difference: from CH I and CH II
Invert: CH I and CH II
XY-Mode: via channel I (Y) and channel II(X)
Frequency range: DC to 150MHz (-3dB)
Rise time: <2.3ns
Overshoot: ≤1%
Deflection coefficient: 14 calibrated positions
from 1mV/div to 20V/div in 1-2-5 sequence,
variable 2.5:1 to min. 50V/div.
Accuracy in calibrated positions
1mV/div – 2mV/div: ±5%(DC-10MHz(-3dB))
5mV/div – 20V/div: ±3%
Input impedance: 1MΩII 15pF
Input coupling: DC-AC-GD (ground)
Input voltage: max. 400V (DC + peak AC)
Delay line: approx. 70ns
Triggering
Automatic (peak to peak):20Hz-250MHz (≥0.5div.)
Normal with level control:DC-250MHz (≥0.5div.)
Indicator for trigger action: LED
Slope: positive or negative
Sources: Channel I or II, line and external
ALT. Triggering: CH I/CH II (≥0.8div.)
Coupling: AC (10 – 250MHz)
DC (0 – 250MHz)
HF (50kHz – 250MHz)
LF (0 – 1.5kHz)
NR (Noise reject)0 - 50MHz (≥0.8div.)
Triggering time base B:normal with level control
and slope selection (0 – 250 MHz)
External: ≥0.3Vpp (0 – 250MHz)
Active TV Sync. Separator: field & line, + / –
Horizontal Deflection
Analog Time Base:
Accuracy in calibr. position 3%; 1-2-5 sequence
A: 0.5s-50ns/div.
B: 20ms-50ns/div.
Operating modes: A or B, alternate A/B
Variable: 2.5:1 up to 1.25s/div.
X-MAG. x10 (±5%) max. 5ns/div.
Holdoff time: variable to approx. 10:1
Bandwidth X-amplifier: 0 – 3MHz (-3dB)
X-Y phase shift: <3° below 220kHz
Digital Time Base:
Accuracy: 3%; 1-2-5 sequence
A: 100s-0.1µs/div.
Peak detect: 100s – 5µs/div.
B: 20ms-0.1µs/div.
Peak detect: 20ms – 5µs/div.
Operating modes: A or B, alternate A/B
X-MAG. x10 (±5%): 10ns/div.
Bandwidth X-Amplifier: 0 – 20MHz (-3dB)
X-Y phase shift: <3° below 20MHz
Input X-amplifier: via Channel II
Sensitivity: see CH II
Digital Storage
Operating modes: Refresh, Roll, Single, XY
Peak Detect, Average (2 to 512), Envelope
Dot Join function: automatically
Acquisition (real time)
8 bit flash A/D max. 200MS/s
Peak detect: 5ns
Display refresh rate: max. 180/s
Memory & display: 2k x 8bit per channel
Reference memory: 2 waveforms 2k x 8bit
Saved in: (EEPROM).
Resolution (samples/div.): X 200/div.
Y25 /div.
XY 25 x 25/div.
Pre-/Post Trigger: 25,50,75,100, -25,-50,-75%
Operation / Control
Manual: front panel switches
Auto Set: signal related automatic
parameter selection
Save & Recall:9 user defined parameter settings
Readout & Cursor (analog/digital)
Display of parameter settings and other functions
on the screen. Trigger point indication.
Cursor measurement of ∆U, ∆t or 1/∆t (frequency),
separate or in tracking mode.
Readout intensity: separately adjustable.
Interface
PC remote control: built in RS232 interface
Option: HO79-6 Multifunction-Interface
IEEE-Bus, RS232, and Centronics
Output formats (HO79-6):PCL, Post Script
HPGL, EPSON
Opto interface HZ70
Component Tester
Test voltage: max. 7Vrms (o/c).
Test current: max. 7mArms (s/c)
Test frequency: approx.50Hz
One test lead is grounded (Safety Earth)
General Information
CRT: D14-375GY, 8x10cm internal graticule
Acceleration voltage: approx. 14kV
Trace rotation: adjustable on front panel
Calibrator: 0.2V ±1%, ≈1kHz/1MHz (tr <4ns)
Line voltage: 100-240V AC ±10%, 50/60Hz
Power consumption: approx. 47 Watt at 50Hz
Min./Max. ambient temperature: 0°C...+40°C
Protective system: Safety class I (IEC1010-1)
Weight: approx. 6.5kg (12.4lbs)
Color: techno-brown
Cabinet: W 285, H125, D380 mm
Lockable tilt handle 7/00
Accessories supplied: Operators Manual, 4 Disks, Line Cord, 2 Probes 10:1

Subject to change without notice
6
General Information
This oscilloscope is easy to operate. The logical arrangement
of the controls allows anyone to quickly become familiar with
the operation of the instrument, however, experienced users
are also advised to read through these instructions so that all
functions are understood.
Immediately after unpacking, the instrument should be
checked for mechanical damage and loose parts in the interior.
If there is transport damage, the supplier must be informed
immediately. The instrument must then not be put into
operation.
Symbols
ATTENTION - refer to manual
Danger - High voltage
Protective ground (earth) terminal
Use of tilt handle
To view the screen from the best angle, there are three
different positions (C, D, E) for setting up the instrument. If
the instrument is set down on the floor after being carried,
the handle automatically remains in the upright carrying
position (A). In order to place the instrument onto a horizontal
surface, the handle should be turned to the upper side of the
oscilloscope (C). For the D position (10° inclination), the handle
should be turned to the opposite direction of the carrying
position until it locks in place automatically underneath the
instrument. For the E position (20° inclination), the handle
should be pulled to release it from the D position and swing
backwards until it locks once more. The handle may also be
set to a position for horizontal carrying by turning it to the
upper side to lock in the B position. At the same time, the
instrument must be lifted, because otherwise the handle will
jump back.
Safety
This instrument has been designed and tested in accordance
with IEC Publication 1010-1 (overvoltage category II, pollution
degree 2), Safety requirements for electrical equipment for
measurement, control, and laboratory use. The CENELEC
regulations EN 61010-1 correspond to this standard. It has
left the factory in a safe condition. This instruction manual
contains important information and warnings which have to
be followed by the user to ensure safe operation and to retain
the oscilloscope in a safe condition.
The case, chassis and all measuring terminals are
connected to the protective earth contact of the
appliance inlet. The instrument operates according
to Safety Class I (three-conductor power cord with
protective earthing conductor and a plug with
earthing contact).
The mains/line plug shall only be inserted in a socket outlet
provided with a protective earth contact. The protective action
must not be negated by the use of an extension cord without
a protective conductor.
The mains/line plug must be inserted before connec-
tions are made to measuring circuits.
The grounded accessible metal parts (case, sockets, jacks)
and the mains/line supply contacts (line/live, neutral) of the
instrument have been tested against insulation breakdown
with 2200V DC.
Under certain conditions, 50Hz or 60Hz hum voltages can
occur in the measuring circuit due to the interconnection with
other mains/line powered equipment or instruments. This
can be avoided by using an isolation transformer (Safety Class
II) between the mains/line outlet and the power plug of the
device being investigated.
Most cathode-ray tubes develop X-rays.
However, the dose equivalent rate falls far below the
maximum permissible value of 36pA/kg (0.5mR/h).
Whenever it is likely that protection has been impaired, the
instrument shall be made inoperative and be secured against
any unintended operation. The protection is likely to be
impaired if, for example, the instrument
• shows visible damage,
• fails to perform the intended measurements,
• has been subjected to prolonged storage under unfavorable
conditions (e.g. in the open or in moist environments),
• has been subject to severe transport stress (e.g. in poor
packaging).
Intended purpose and operating conditions
This instrument must be used only by qualified experts who
are aware of the risks of electrical measurement.
The instrument is specified for operation in industry, light
industry, commercial and residential environments.
Due to safety reasons the instrument must only be connected
to a properly installed power outlet, containing a protective
earth conductor. The protective earth connection must not
be broken. The power plug must be inserted in the power
outlet while any connection is made to the test device.
The instrument has been designed for indoor use. The
permissible ambient temperature range during operation is
+10°C (+50°F) ... +40°C (+104°F). It may occasionally be
subjected to temperatures between +10°C (+50°F) and -10°C
(+14°F) without degrading its safety. The permissible ambient
temperature range for storage or transportation is -40°C (-
0°F) ... +70°C (+158°F). The maximum operating altitude is
up to 2200m (non-operating 15000m). The maximum relative
humidity is up to 80%.
If condensed water exists in the instrument it should be
acclimatized before switching on. In some cases (e.g.
extremely cold oscilloscope) two hours should be allowed
before the instrument is put into operation. The instrument
General Information

7
Subject to change without notice
on which the technical data are based. The test methods
described in the test plan of this manual can be performed
without great expenditure on measuring instruments.
However, purchase of the HAMEG scope tester HZ60, which
despite its low price is highly suitable for tasks of this type,
is very much recommended. The exterior of the oscilloscope
should be cleaned regularly with a dusting brush. Dirt which
is difficult to remove on the casing and handle, the plastic
and aluminum parts, can be removed with a moistened cloth
(99% water +1% mild detergent). Spirit or washing benzene
(petroleum ether) can be used to remove greasy dirt. The
screen may be cleaned with water or washing benzene (but
not with spirit (alcohol) or solvents), it must then be wiped
with a dry clean lint-free cloth. Under no circumstances may
the cleaning fluid get into the instrument. The use of other
cleaning agents can attack the plastic and paint surfaces.
Protective Switch-Off
This instrument is equipped with a switch mode power supply.
It has both overvoltage and overload protection, which will
cause the switch mode supply to limit power consumption
to a minimum. In this case a ticking noise may be heard.
Power supply
The oscilloscope operates on mains/line voltages between
100VAC and 240VAC. No means of switching to different
input voltages has therefore been provided.
The power input fuses are externally accessible. The fuse
holder is located above the 3-pole power connector. The
power input fuses are externally accessible, if the rubber
connector is removed. The fuse holder can be released by
pressing its plastic retainers with the aid of a small screw-
driver. The retainers are located on the right and left side of
the holder and must be pressed towards the center. The
fuse(s) can then be replaced and pressed in until locked on
both sides.
Use of patched fuses or short-circuiting of the fuse holder is
not permissible; HAMEG assumes no liability whatsoever for
any damage caused as a result, and all warranty claims
become null and void.
Fuse type:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: time-lag (T).
Attention!
There is a fuse located inside the instrument within
the switch mode power supply:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: fast (F).
This fuse must not be replaced by the operator!
should be kept in a clean and dry room and must not be
operated in explosive, corrosive, dusty, or moist environ-
ments. The oscilloscope can be operated in any position, but
the convection cooling must not be impaired. The ventilation
holes may not be covered. For continuous operation the
instrument should be used in the horizontal position,
preferably tilted upwards, resting on the tilt handle.
The specifications stating tolerances are only valid if the
instrument has warmed up for 30minutes at an ambient
temperature between +15°C (+59°F) and +30°C (+86°F).
Values without tolerances are typical for an average instru-
ment.
EMC
This instrument conforms to the European standards regar-
ding the electromagnetic compatibility. The applied standards
are: Generic immunity standard EN50082-2:1995 (for indus-
trial environment) Generic emission standard EN50081-1:1992
(for residential, commercial and light industry environment).
This means that the instrument has been tested to the highest
standards.
Please note that under the influence of strong electro-
magnetic fields, such signals may be superimposed on the
measured signals.
Under certain conditions this is unavoidable due to the
instrument’s high input sensitivity, high input impedance and
bandwidth. Shielded measuring cables, shielding and earthing
of the device under test may reduce or eliminate those
effects.
Warranty
HAMEG warrants to its Customers that the products it
manufactures and sells will be free from defects in materials
and workmanship for a period of 2 years. This warranty
shall not apply to any defect, failure or damage caused by
improper use or inadequate maintenance and care. HAMEG
shall not be obliged to provide service under this warranty to
repair damage resulting from attempts by personnel other
than HAMEG representatives to install, repair, service or
modify these products.
In order to obtain service under this warranty, Customers
must contact and notify the distributor who has sold the
product. Each instrument is subjected to a quality test with
10 hour burn-in before leaving the production. Practically all
early failures are detected by this method. In the case of
shipments by post, rail or carrier it is recommended that the
original packing is carefully preserved. Transport damages
and damage due to gross negligence are not covered by the
warranty.
In the case of a complaint, a label should be attached to the
housing of the instrument which describes briefly the faults
observed. If at the same time the name and telephone
number (dialing code and telephone or direct number or
department designation) is stated for possible queries, this
helps towards speeding up the processing of warranty claims.
Maintenance
Various important properties of the oscilloscope should be
carefully checked at certain intervals. Only in this way is it
largely certain that all signals are displayed with the accuracy
General Information

Subject to change without notice
8
Type of signal voltage
The following description of the HM1507-2 relates to the
analog-oscilloscope mode. Please note “Storage Opera-
tion”.
The oscilloscope HM1507-2 allows examination of DC vol-
tages and most repetitive signals in the frequency range up
to at least 150MHz (-3dB).
The vertical amplifiers have been designed for minimum
overshoot and therefore permit a true signal display.
The display of sinusoidal signals within the bandwidth limits
causes no problems, but an increasing error in measurement
due to gain reduction must be taken into account when
measuring high frequency signals. This error becomes
noticeable at approx. 70MHz. At approx. 110MHz the redu-
ction is approx. 10% and the real voltage value is 11% higher.
The gain reduction error can not be defined exactly as the -
3dB bandwidth of the amplifiers differ between 150MHz
and 170MHz.
For sine wave signals the -6dB limit is approx. 220MHz.
When examining square or pulse type waveforms, attention
must be paid to the harmonic content of such signals. The
repetition frequency (fundamental frequency) of the signal
must therefore be significantly smaller than the upper limit
frequency of the vertical amplifier.
Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content which can be
used for triggering. This is the case with bursts, for instance.
To obtain a well-triggered display in this case, the assistance
of the variable hold off function or the second time base may
be required. Television video signals are relatively easy to
trigger using the built-in TV-Sync-Separator (TV).
For optional operation as a DC or AC voltage amplifier, each
vertical amplifier input is provided with a DC/AC switch. DC
coupling should only be used with a series-connected atte-
nuator probe or at very low frequencies or if the measure-
ment of the DC voltage content of the signal is absolutely
necessary.
When displaying very low frequency pulses, the flat tops may
be sloping with AC coupling of the vertical amplifier (AC limit
frequency approx. 1.6 Hz for 3dB). In this case, DC operation
is preferred, provided the signal voltage is not superimposed
on a too high DC level. Otherwise a capacitor of adequate
capacitance must be connected to the input of the vertical
amplifier with DC coupling. This capacitor must have a
sufficiently high breakdown voltage rating. DC coupling is
also recommended for the display of logic and pulse signals,
especially if the pulse duty factor changes constantly.
Otherwise the display will move upwards or downwards at
each change. Pure direct voltages can only be measured with
DC coupling.
The input coupling is selectable by the AC/DC pushbutton.
The actual setting is displayed in the readout with the “ = “
symbol for DC- and the “ ~ “ symbol for AC coupling.
Amplitude Measurements
In general electrical engineering, alternating voltage data
normally refers to effective values (rms = root-mean-square
value). However, for signal magnitudes and voltage desi-
gnations in oscilloscope measurements, the peak-to-peak
voltage (Vpp) value is applied. The latter corresponds to the
real potential difference between the most positive and most
negative points of a signal waveform. If a sinusoidal
waveform, displayed on the oscilloscope screen, is to be
converted into an effective (rms) value, the resulting peak-
to-peak value must be divided by 2x√2 = 2.83. Conversely, it
should be observed that sinusoidal voltages indicated in Vrms
(Veff) have 2.83 times the potential difference in Vpp.
The relationship between the different voltage magnitudes
can be seen from the following figure.
Voltage values of a sine curve
Vrms = effective value; Vp= simple peak or crest value;
Vpp = peak-to-peak value; Vmom = momentary value.
The minimum signal voltage which must be applied to the Y
input for a trace of 1div height is 1mVpp (± 5%) when this
deflection coefficient is displayed on the screen (readout)
and the vernier is switched off (VAR-LED dark). However,
smaller signals than this may also be displayed. The deflection
coefficients are indicated in mV/div or V/div (peak-to-peak
value).
The magnitude of the applied voltage is ascertained by
multiplying the selected deflection coefficient by the vertical
display height in div. If an attenuator probe x10 is used, a
further multiplication by a factor of 10 is required to ascertain
the correct voltage value.
For exact amplitude measurements, the variable control (VAR)
must be set to its calibrated detent CAL position.
With the variable control activated the deflection sensitivity
can be reduced up to a ratio of 2.5 to 1 (please note “controls
and readout”). Therefore any intermediate value is possible
within the 1-2-5 sequence of the attenuator(s).
With direct connection to the vertical input, signals
up to 400Vpp may be displayed (attenuator set to
20V/div, variable control to 2.5:1).
With the designations
H= display height in div,
U= signal voltage in Vpp at the vertical input,
D= deflection coefficient in V/div at attenuator switch,
the required value can be calculated from the two given
quantities:
However, these three values are not freely selectable.
They have to be within the following limits (trigger threshold,
accuracy of reading):
Hbetween 0.5 and 8div, if possible 3.2 to 8div,
Ubetween 0.5mVpp and 160Vpp,
Dbetween 1mV/div and 20V/div in 1-2-5 sequence.
Type of signal voltage

9
Subject to change without notice
Examples:
Set deflection coefficient D= 50mV/div 0.05V/div,
observed display height H= 4.6div,
required voltage U= 0.05x4.6 = 0.23Vpp.
Input voltage U = 5Vpp,
set deflection coefficient D= 1V/div,
required display height H= 5:1 = 5div.
Signal voltage U= 230Vrmsx 2
?√
2 = 651Vpp
(voltage > 160Vpp, with probe 10:1: U= 65.1Vpp),
desired display height H= min. 3.2div, max. 8div,
max. deflection coefficient D= 65.1:3.2 = 20.3V/div,
min. deflection coefficient D= 65.1:8 = 8.1V/div,
adjusted deflection coefficient D= 10V/div.
The previous examples are related to the CRT graticule
reading. The results can also be determined with the aid of
the ∆V cursor measurement (please note “controls and
readout”).
The input voltage must not exceed 400V, independent
from the polarity.
If an AC voltage which is superimposed on a DC voltage is
applied, the maximum peak value of both voltages must not
exceed + or -400V. So for AC voltages with a mean value of
zero volt the maximum peak to peak value is 800Vpp.
If attenuator probes with higher limits are used, the
probes limits are valid only if the oscilloscope is set
to DC input coupling.
If DC voltages are applied under AC input coupling conditions
the oscilloscope maximum input voltage value remains 400V.
The attenuator consists of a resistor in the probe and the
1MΩinput resistor of the oscilloscope, which are disabled
by the AC input coupling capacity when AC coupling is
selected. This also applies to DC voltages with superimposed
AC voltages. It also must be noted that due to the capacitive
resistance of the AC input coupling capacitor, the attenuation
ratio depends on the signal frequency. For sine wave signals
with frequencies higher than 40Hz this influence is negligible.
With the above listed exceptions HAMEG 10:1 probes can
be used for DC measurements up to 600V or AC voltages
(with a mean value of zero volt) of 1200Vpp. The 100:1 probe
HZ53 allows for 1200V DC or 2400Vpp for AC.
It should be noted that its AC peak value is derated at higher
frequencies. If a normal x10 probe is used to measure high
voltages there is the risk that the compensation trimmer
bridging the attenuator series resistor will break down causing
damage to the input of the oscilloscope. However, if for
example only the residual ripple of a high voltage is to be
displayed on the oscilloscope, a normal x10 probe is sufficient.
In this case, an appropriate high voltage capacitor (approx.
22-68nF) must be connected in series with the input tip of
the probe.
With Y-POS. control (input coupling to GD) it is possible to
use a horizontal graticule line as reference line for ground
potential before the measurement. It can lie below or above
the horizontal central line according to whether positive and/
or negative deviations from the ground potential are to be
measured.
Total value of input voltage
The dotted line shows a voltage alternating at zero volt level. If
superimposed on a DC voltage, the addition of the positive peak
and the DC voltage results in the max. voltage (DC + ACpeak).
Time Measurements
As a rule, most signals to be displayed are periodically repea-
ting processes, also called periods. The number of periods
per second is the repetition frequency. Depending on the
time base setting (TIME/DIV.-knob) indicated by the readout,
one or several signal periods or only a part of a period can be
displayed. The time coefficients are stated in ms/div, µs/
div or ns/div. The following examples are related to the CRT
graticule reading. The results can also be determined with
the aid of the ∆t and 1/∆t cursor measurement (please note
“controls and readout”).
The duration of a signal period or a part of it is determined by
multiplying the relevant time (horizontal distance in div) by
the (calibrated) time coefficient displayed in the readout .
Uncalibrated, the time base speed can be reduced until a
maximum factor of 2.5 is reached. Therefore any intermediate
value is possible within the 1-2-5 sequence.
With the designations
L= displayed wave length in div of one period,
T= time in seconds for one period,
F= recurrence frequency in Hz of the signal,
Tc = time coefficient in ms, µs or ns/div and the relation
F = 1/T, the following equations can be stated:
However, these four values are not freely selectable. They
have to be within the following limits:
Lbetween 0.2 and 10div, if possible 4 to 10div,
Tbetween 5ns and 5s,
Fbetween 0.5Hz and 100MHz,
Tc between 50ns/div and 500ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and
Tc between 5ns/div and 50ms/div in 1-2-5 sequence
(with X-MAG. (x10) active).
Examples:
Displayed wavelength L = 7div,
set time coefficient Tc = 100ns/div,
required period T = 7x100x10-9 = 0.7µs
required rec. freq. F = 1:(0.7x10-6) = 1.428MHz.
Signal period T = 1s,
set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.
Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10-3) = 100Hz.
TV-line frequency F = 15625Hz,
Type of signal voltage

Subject to change without notice
10
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15625x10-5) = 6.4div.
Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x103) = 0.25ms/div,
min. time coefficient Tc = 1:(10x103) = 0.1ms/div,
set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(103x0.2x10-3) = 5div.
Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) pushbutton: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10-6) = 25MHz,
required period T = 1:(25x106) = 40ns.
If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be
applied (X-MAG. (x10) active). In this case, the time interval
of interest can be shifted to the screen center using the X-
POS. control.
When investigating pulse or square waveforms, the critical
feature is the rise time of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly
influence the measuring accuracy, the rise time is generally
measured between 10% and 90% of the vertical pulse height.
For measurement, adjust the Y deflection coefficient using
its variable function (uncalibrated) together with the Y-POS.
control so that the pulse height is precisely aligned with the
0% and 100% lines of the internal graticule. The 10% and
90% points of the signal will now coincide with the 10% and
90% graticule lines. The rise time is given by the product of
the horizontal distance in div between these two coincident
points and the calibrated time coefficient setting. The fall
time of a pulse can also be measured by using this method.
The following figure shows correct positioning of the
oscilloscope trace for accurate rise time measurement.
With a time coefficient of 5ns/div (X x10 magnification active),
the example shown in the above figure results in a total
measured rise time of
ttot = 1.6div x 5ns/div : 10 = 8ns
When very fast rise times are being measured, the rise times
of the oscilloscope amplifier and of the attenuator probe has
to be deducted from the measured time value. The rise time
of the signal can be calculated using the following formula.
In this ttot is the total measured rise time, tosc is the rise time
of the oscilloscope amplifier (approx. 2.3ns), and tpthe rise
time of the probe (e.g. = 2ns). If ttot is greater than 34ns,
then ttot can be taken as the rise time of the pulse, and calcu-
lation is unnecessary.
Calculation of the example in the figure above results in a
signal rise time
The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to
measure in any display position and at any signal amplitude.
It is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness
and that the horizontal distance at 10% and 90% of the
amplitude is measured. If the edge shows rounding or over-
shooting, the 100% should not be related to the peak values
but to the mean pulse heights. Breaks or peaks (glitches)
next to the edge are also not taken into account. With very
severe transient distortions, the rise and fall time measure-
ment has little meaning. For amplifiers with approximately
constant group delay (therefore good pulse transmission
performance) the following numerical relationship between
rise time tr (in ns) and bandwidth B (in MHz) applies:
Connection of Test Signal
In most cases briefly depressing the AUTO SET causes a
useful signal related instrument setting. The following
explanations refer to special applications and/or signals,
demanding a manual instrument setting. The description of
the controls is explained in the section “controls and read-
out”.
Caution:
When connecting unknown signals to the oscil-
loscope input, always use automatic triggering and
set the input coupling switch to AC (readout). The
attenuator should initially be set to 20V/div.
Sometimes the trace will disappear after an input signal has
been applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160Vpp and
the deflection coefficient (VOLTS/DIV.) in calibrated
condition, an attenuator probe must be inserted before the
vertical input. If, after applying the signal, the trace is nearly
blanked, the period of the signal is probably substantially
longer than the set time deflection coefficient (TIME/DIV.).
It should be switched to an adequately larger time coefficient.
The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as
HZ32 or HZ34, or reduced through a x10 or x100 attenuator
probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to
approx. 50kHz). For higher frequencies, the signal source
must be of low impedance, i.e. matched to the characteristic
resistance of the cable (as a rule 50Ω). Especially when
transmitting square and pulse signals, a resistor equal to the
characteristic impedance of the cable must also be connected
across the cable directly at the Y-input of the oscilloscope.
When using a 50Ωcable such as the HZ34, a 50Ωthrough
termination type HZ22 is available from HAMEG. When
transmitting square signals with short rise times, transient
phenomena on the edges and top of the signal may become
visible if the correct termination is not used. A terminating
resistance is sometimes recommended with sine signals as
well. Certain amplifiers, generators or their attenuators
maintain the nominal output voltage independent of frequency
only if their connection cable is terminated with the prescribed
resistance. Here it must be noted that the terminating resistor
Type of signal voltage

11
Subject to change without notice
HZ22 will only dissipate a maximum of 2Watts. This power
is reached with 10Vrms or at 28.3Vpp with sine signal. If a
x10 or x100 attenuator probe is used, no termination is
necessary. In this case, the connecting cable is matched
directly to the high impedance input of the oscilloscope. When
using attenuators probes, even high internal impedance
sources are only slightly loaded (approx. 10MΩII 12pF or
100MΩII 5pF with HZ53). Therefore, if the voltage loss due
to the attenuation of the probe can be compensated by a
higher amplitude setting, the probe should always be used.
The series impedance of the probe provides a certain amount
of protection for the input of the vertical amplifier. Because
of their separate manufacture, all attenuator probes are only
partially compensated, therefore accurate compensation must
be performed on the oscilloscope (see Probe compensation).
Standard attenuator probes on the oscilloscope normally
reduce its bandwidth and increase the rise time. In all cases
where the oscilloscope bandwidth must be fully utilized (e.g.
for pulses with steep edges) we strongly advise using the
probes HZ51 (x10) HZ52 (x10 HF) and HZ54 (x1 and x10).
This can save the purchase of an oscilloscope with larger
bandwidth.
The probes mentioned have a HF-calibration in addition to
low frequency calibration adjustment. Thus a group delay
correction to the upper limit frequency of the oscilloscope is
possible with the aid of an 1MHz calibrator, e.g. HZ60.
In fact the bandwidth and rise time of the oscilloscope are
not noticeably changed with these probe types and the
waveform reproduction fidelity can even be improved because
the probe can be matched to the oscilloscopes individual pulse
response.
If a x10 or x100 attenuator probe is used, DC input
coupling must always be used at voltages above
400V. With AC coupling of low frequency signals, the
attenuation is no longer independent of frequency,
pulses can show pulse tilts. Direct voltages are
suppressed but load the oscilloscope input coupling
capacitor concerned. Its voltage rating is max. 400 V
(DC + peak AC). DC input coupling is therefore of quite
special importance with a x100 attenuation probe
which usually has a voltage rating of max. 1200 V
(DC + peak AC). A capacitor of corresponding
capacitance and voltage rating may be connected in
series with the attenuator probe input for blocking
DC voltage (e.g. for hum voltage measurement).
With all attenuator probes, the maximum AC input voltage
must be derated with frequency usually above 20kHz.
Therefore the derating curve of the attenuator probe type
concerned must be taken into account.
The selection of the ground point on the test object is
important when displaying small signal voltages. It should
always be as close as possible to the measuring point. If this
is not done, serious signal distortion may result from spurious
currents through the ground leads or chassis parts. The
ground leads on attenuator probes are also particularly critical.
They should be as short and thick as possible. When the
attenuator probe is connected to a BNC-socket, a BNC-adapter,
should be used. In this way ground and matching problems
are eliminated. Hum or interference appearing in the measuring
circuit (especially when a small deflection coefficient is used)
is possibly caused by multiple grounding because equalizing
currents can flow in the shielding of the test cables (voltage
drop between the protective conductor connections, caused
by external equipment connected to the mains/line, e.g. signal
generators with interference protection capacitors).
Controls and Readout
The following description assumes that the instrument is not
set to “COMPONENT TESTER”mode.
If the instrument is switched on, all important settings are
displayed in the readout. The LED’s located on the front panel
assist operation and indicate additional information. Incorrect
operation and the electrical end positions of control knobs
are indicated by a warning beep.
Except for the power pushbutton (POWER), the calibrator
frequency pushbutton (CAL. 1kHz/1MHz), the focus control
(FOCUS) and the trace rotation control (TR) all other controls
are electronically selected. All other functions and their settings
can therefore be remote controlled and stored. Some controls
are only operative in storage mode or have different functions
in analog operation. See “STORAGE MODE ONLY”.
The front panel is subdivided into sections.
On the top, immediately to the right of the CRT screen,
the following controls and LED indicators are placed:
(1) POWER - Pushbutton and symbols for ON (I) and OFF
(O).
After the oscilloscope is switched on, all LEDs lit and an
automated instrument test is performed. During this
time the HAMEG logo and the software version are
displayed on the screen. After the internal test is
completed successfully, the overlay is switched off and
the normal operation mode is present. Then the last
used settings become activated and one LED indicates
the ON condition.
Some mode functions can be modified (SETUP) and/or
automated adjustment procedures (CALIBRATE) can be
called if the “MAIN MENU”is present. To enter this
menu the AUTO SET pushbutton must be pressed
constantly when the HAMEG logo is displayed until
“MAIN MENU”becomes visible. For further information
please note “MENU”.
(2) AUTO SET - Briefly depressing this pushbutton results
in an automatic signal related instrument setting (please
note “AUTO SET”), if the signal frequency and height
are suited for automatic triggering (AT). In Yt mode the
actual channel operating conditions (CH I, CH II or DUAL)
remain unchanged, whereas the time base is auto-
matically set to Atime base mode.
In case of XY or CT (Component Tester) operation, the
instrument is set to the last used Yt mode setting.
Automatic CURSOR supported voltage measurement
If CURSOR voltage measurement is present, the
CURSOR lines are automatically set to the positive and
negative peak value of the signal. The accuracy of this
function depends on the signal frequency and is also
influenced by the signal‘s pulse duty factor. If the signal
height is insufficient, the CURSOR lines do not change.
In DUAL mode the CURSOR lines are related to the
signal which is used for internal triggering.
Controls and Readout

Subject to change without notice
12
STORAGE MODE ONLY
Additionally, AUTOSET automatically selects refresh
mode (RFR) when SINGLE (SGL) or ROLL (ROL)
function is in operation.
Automatic CURSOR supported measurement
In contrast to analog mode, AUTO SET also causes an
automatic CURSOR line setting if time or frequency
measurement has been selected and at least one signal
period is displayed. Neither the signal frequency nor the
pulse duty factor have an effect on the accuracy when
CURSOR voltage measurement is chosen.
(3) RM - The remote control mode can be switched on or
off via the RS232 interface. In the latter case the “RM”
LED is lit and the electronically selectable controls on
front panel are inactive. This state can be left by
depressing the AUTO SET pushbutton provided it was
not inactivated via the interface.
STORAGE MODE ONLY
The RM LED is lit during signal transfer via the built in
RS232 interface. At this time the controls are inactive.
(4) INTENS - Knob with associated pushbutton and LEDs.
This control knob is for adjusting both the trace and
readout intensity. Turning this knob clockwise increases
and turning it counterclockwise decreases the intensity
of the selected function (A, RO resp. B).
The READ OUT pushbutton below is for selecting the
function in two ways.
Depending on the actual time base mode and the readout
(RO) not switched off, briefly pressing the READ OUT
pushbutton switches over the INTENS knob function
indicated by a LED in the sequences:
A - RO - A in condition A time base,
A - RO - B - A if alternate time base mode is present,
B - RO - B in condition B time base,
A - RO - B in XY mode and
A - RO - A in Component Tester (CT) mode.
Pressing and holding the READ OUT pushbutton swit-
ches the readout on or off. In readout off condition the
INTENS knob function can consequently not be set to
RO. Briefly pressing the pushbutton causes an error tone
if only A or B time base mode are present. If alternate
time base mode is used the switching sequence is A - B
- A.
Switching the readout off, may be required if interference
distortions are visible on the signal(s). Such distortions
may also originate from the chopper generator if the
instrument is operated in chopped DUAL mode.
In XY mode only A (for the signal) and RO can be
selected unless the readout is switched off. Then just
the A-LED is lit.
The readout is automatically switched off in COMPO-
NENT TEST mode and no other LED on the front panel
is lit except A.
All INTENS settings are stored after the instrument is
switched off.
The AUTO SET function switches the readout on and
selects A time base mode (A-LED lit). The INTENS
setting for each function is automatically set to the mean
value, if less intensity was previously selected.
(5) TR - The trace rotation control can be adjusted with a
small screwdriver (please note “trace rotation TR”)
(6) FOCUS - This control knob effects both the trace and
the readout sharpness.
(7) STOR. ON / HOLD - Pushbutton with two functions.
STOR. ON
Pressing and holding the pushbutton switches from
analog (Yt or XY) to storage mode and vice versa. If CT
(Component Tester) mode is present (only available in
analog mode), it must be switched off first to enable
switching over to storage mode.
The oscilloscope is in analog mode if none of the LED’s
associated with the STOR.MODE (9) pushbuttons are
lit and a pre- or post trigger value (PT...%) is not indicated
by the readout. Pressing and holding the STOR. ON
pushbutton switches over to the digital mode, but
without changing the channel operating mode (CH I, CH
II, DUAL, ADD and XY).
The actual signal capture mode is indicated by one of
the STOR. MODE-LED‘s (RFR - ENV - AVM - ROL) and
in addition displayed by the readout. In digital XY mode
the RFR-LED is lit and the readout indicates XY.
If digital SINGLE event (SGL) capture mode is selected,
all STOR. MODE-LED‘s are dark, but the readout displays
the pre- or post trigger value (PT...%).
Attention:
The time base ranges are different between ana-
log and storage mode operation depending on
the operating mode!
In ALTernate and B time base mode the B time
coefficient can never be set to a larger value than
the actual A time coefficient. The following
information excludes the X magnifier factor.
Analog mode:
A time base from 500ms/div to 50ns/div.
B time base from 20ms/div to 50ns/div.
Storage mode:
A time base from 100s/div to 100ns/div,
B time base from 20ms/div to 100ns/div,
This results in the following behavior when switched
from analog to digital mode and vice versa:
1.If in analog mode, the time base has been selected
between 200ns/div and 50ns/div, then on switching
to digital mode the lowest available time coefficient
will be automatically selected, i.e. 100ns/div. If now
one switches back to analog mode without having
made any time base changes in the digital mode, then
the last time base selected in the analog mode is again
active (e.g. 50ns/div).
If on the other hand, the time base is changed after
switching over to digital mode (e.g. to 2µs/div). Then,
Controls and Readout

13
Subject to change without notice
when switched back to analog mode, the time base
in analog mode will be set to the value selected in the
digital mode (e.g. 2µs/div).
2.If a time base between 100s/div and 1s/div has been
set in the digital mode and the mode is switched to
analog, then the time base in analog mode is
automatically set to 500ms/div. The rest is as described
before.
The X-MAG x10 setting remains unchanged when
switched from analog to digital mode and vice versa.
STORAGE MODE ONLY
If by pressing and holding the STOR. ON / HOLD
pushbutton, the mode is switched to digital, then one
of the associated LED’s lights up. Which one it is,
depends on the last selected digital operation.
Exception
Switching over from analog SINGLE mode to di-
gital mode sets the instrument automatically to
digital SINGLE mode.
For additional information regarding the digital mode,
see section STORAGE OPERATION.
HOLD
STORAGE MODE ONLY
Briefly pressing the STOR. ON / HOLD pushbutton
switches over between protected and unprotected mode
of the current memory contents.
The current contents of the memory are protected
against overwriting when HLD (HOLD) instead of
channel information (e.g. Y1... ) is displayed in the
readout. This prevents a change in the Yt mode setting,
but it is possible to select between DUAL (Yt) and XY
display by pressing the DUAL (23) pushbutton if one of
these modes was selected before activating HOLD.
If HOLD is switched off, one can observe how the
existing memory contents are successively overwritten
by new data especially with slow time base settings
and refresh mode. Protecting the memory contents in
the middle of a data acquisition process can result in an
irregularity at the junction of old (right) and new data
(left). This can be avoided by recording in single shot
mode (SGL), even though the input signal is repetitive.
At the end of a sweep, one can use HOLD to protect
the contents against being overwritten by an uninten-
tional actuation of RESET (RES).
The signal in each of the current memory can be shifted
in the vertical direction (+/- 4div) with the corresponding
Y-POS rotary knob when HOLD is operative.
The original trace position will be lost when shifted
vertically, but this can be found again. To this end the Y-
POS knob in question must be rotated quickly. Once
the original position is reached, the trace does not shift
anymore although the knob is rotated further. Simul-
taneously a signal tone sounds. To shift the trace verti-
cally again it will be required to stop rotating the knob
for at least about 2 seconds.
Attention!
The dynamic range limits of the A/D converter may
become visible if a Y -position shift is performed after
storage. This can affect those signal parts which were
originally above or below the screen.
(8) PTR / PK Det - Pushbutton with two functions.
Neither function is available in analog mode.
PTR
Briefly pressing selects the PRE- and POST-Trigger value.
The PRE TRIGGER function is used to capture signals
that occur prior to a trigger event, making the prehistory
visible. In contrast to this function, the POST TRIGGER
is used to capture signals occurring after the trigger
event, which could not be captured in “0%”pre trigger
condition. Due to the dependence on trigger events,
neither function is available in the trigger independent
modes XY and ROLL.
The actual PRE- or POST TRIGGER value is displayed
by the readout and changes each time the PTR
pushbutton is pressed briefly, in the following sequence:
PT0%, PT25%, PT50%, PT75%, PT100%, PT-75%, PT-
50%, PT-25% and back to PT0%.
The values refer to the X-axis (graticule) of the screen
display (10% = 1div).
The following description assumes that the X magnifier
(x10) is inactive and the signal display starts on the
leftmost vertical graticule line. It is also assumed that a
trigger mode (source, coupling) is chosen, in which the
trigger point symbol is displayed. In contrast to analog
mode, using pre trigger the trigger point symbol can be
shifted in X-direction.
PRE TRIGGER
0% PRE TRIGGER (readout ”PT0%”) means that the
signal display starts with the trigger event. The trigger
point symbol indicates this position. If the X-POS. control
is not in center position, an arrow pointing to the left
may be displayed. Then the X-POS. (19) control must
be turned clockwise until the arrow is no longer visible.
25% PRE TRIGGER (readout ”PT25%”) is achieved after
pressing the PTR pushbutton once. The signal display
starts with 25% prehistory and the trigger point symbol
is shifted 2.5 divisions to the right.
Each time the PTR pushbutton is pressed the PRE
TRIGGER value increases by 25% until 100% is reached.
If in 100% condition an arrow symbol is displayed in
addition to the trigger point symbol, the X-POS. control
should be turned ccw. to make the trigger point visible
on the screen.
The duration of the prehistory is determined by multi-
plying the time coefficient by the pre trigger value (in
divisions). E.g. 20ms/div x 7,5 div (= 75% pre trigger) =
150ms.
POST TRIGGER
In POST TRIGGER condition the trigger point is always
to the left of the screen and therefore not visible. The
trigger point symbol then only indicates the LEVEL
Controls and Readout

Subject to change without notice
14
setting. An additional arrow symbol which points to the
left is displayed to indicate post trigger operation. In
POST TRIGGER condition the arrow symbol does not
indicate a wrong X-POS. setting. A minus sign (-) placed
in front of the percentage value, is displayed by the
readout for POST TRIGGER mode indication.
Proceeding from 100% pre trigger, the instrument swit-
ches over to 75% POST TRIGGER (”PT-75%”) after
the PTR pushbutton is pressed. Then the trigger point
is 7.5 div to the left of the trace start on the screen.
This means that the signal capture starts 7.5 x time
deflection coefficient after the trigger event occurred.
Every time the PTR pushbutton is pressed the POST
TRIGGER value changes in 25% steps until PTR-25% is
active. When the PTR pushbutton is pressed again, both
post and pre trigger are switched off and the readout
indicates ”PT0%”.
Attention!
In time base settings from 100s/div to 50ms/div the
pre- or post trigger is automatically switched off
(”PT0%) if refresh (RFR), envelope (ENV) or average
(AVM) mode is active. This is to avoid excessive wai-
ting times.
If the pre- or post trigger function is required in combi-
nation with those time coefficients, SINGLE (SGL) mode
operation must be used.
PK Det
Pressing and holding switches the peak value detection
(“PK Det”= peak detect) on or off. This function is
available only with deflection coefficients from 100s/div
to 5µs/div in combination with REFRESH, ENVELOPE,
ROLL or SINGLE modes. “PK Det”will be disabled
automatically if AVERAGE mode is active or a time
coefficient from 2µs/div to 100ns/div is chosen.
The “PK Det”function is indicated by the time coefficient
display in the readout. Switching “PK Det”on, changes
from e.g. “A:20ms”to “P:20ms”and consequently in
B time base mode from “B:100µs”to “P:100µs”.In
alternate (A and B) time base mode, the “PK Det”
function only affects the A time base and the readout
displays e.g. “P:20ms”and “B:100µs”.
In “PK Det”operation the sampling rate is always 40MS/
s and the signal will be sampled every 25ns. The advan-
tage of this sampling method is as follows:
Without “PK Det”and a time coefficient of 100s/div,
the signal is sampled every 0.5 seconds (2 Samples/
second) and stored at a new address. A signal amplitude
change with a duration of e.g. 30ns appearing 0.2 se-
conds after the last sampling procedure will not be cap-
tured. In combination with “PK Det”the sampling
interval is reduced to 25ns and then the samples will be
evaluated and the most deviating value captured within
0.5s after the last storage procedure, will be stored at
the next address.
(9) STOR. MODE - Pushbuttons with associated LEDs.
These functions are not available in analog mode.
If digital SINGLE (SGL) mode has not been chosen,
one of the associated LEDs is lit. The signal capture and
display mode can be selected by pressing one of the
pushbuttons. The mode setting is indicated by one of
the LEDs (RFR, ENV, AVM and ROL) and also displayed
by the readout. The only exception is in XY storage
mode. Then the RFR-LED is lit and the readout displays
XY. No other signal capture and display mode can be
chosen in XY mode.
The desired Yt signal capture mode can be selected by
pressing the upper or lower STOR. MODE pushbutton.
The following description presumes that HOLD (HLD)
is not activated and the trigger conditions are met.
(9) RFR - stands for refresh operation. In this mode, as in
analog mode, periodically repeating signals can be
captured and displayed.
The signal acquisition is started by triggering the digital
time base. Then the previously captured and displayed
signal will be overwritten with the current signal. This
will be displayed until the digital time base is triggered
again. This is in contrast to analog operation where the
screen remains blank when the time base is not
triggered.
In refresh mode, the signal acquisition can be effected
with pre- or post triggering when a time base between
20ms/div and 100ns/div is selected. The pre triggering
or post triggering will be automatically switched off
(PT0%), with larger time coefficients (100s/div to 50ms/
div) in order to avoid excessive waiting times. If it is
required to measure with pre- or post trigger in this time
base range, one should select single shot (SINGLE =
SGL).
In XY digital mode the RFR-LED lights. It indicates a
continuous, trigger independent signal acquisition. The
trigger circuit is switched off.
(9) ENV - is the abbreviation for ENVELOPE operation.
In this mode the minimum and maximum values of the
signal during several signal acquisitions will be deter-
mined and displayed. Except for this display, the ENVE-
LOPE operation is identical to the refresh operation.
Changes in the signal are easier to measure and are
more visible in ENVELOPE operation. This is valid not
only for amplitude changes but also for frequency
variations (Jitter).
The ENVELOPE evaluation begins anew when the
SINGLE (10) pushbutton is pressed briefly, to actuate
the RESET (RES) function.
Attention!
The pre- or post trigger will be automatically switched
off (PT0%) in the time base range from 100s/div to
50ms/div.
(9) AVM - indicates AVERAGE (mean value) mode.
This operation is effective when the AVM-LED lights up
and the readout displays AV... .
In this case also several signal acquisition scans are
required; hence, it is similar to Refresh operation. The
signal is averaged over the several acquisitions so that
Controls and Readout

15
Subject to change without notice
amplitude variations ( e.g. noise) and frequency variations
(Jitter) are minimized or eliminated in the display.
The accuracy of the mean value evaluation increases as
the number of signal acquisition scans used for
evaluation is increased. One can select the number
between 2 and 512. The selected setting is displayed in
the readout. Of course, with increasing accuracy the
time required for this also increases.
To select a different value briefly press both STOR.
MODE pushbuttons simultaneously. The AV... display
in the readout flashes indicating the setting mode. Now,
the value can be changed by briefly pressing the upper
or lower STOR. MODE pushbutton. The setting mode
can be exited by again briefly pressing the two
pushbuttons simultaneously. The setting mode will also
be switched off automatically if none of the two
pushbuttons is actuated during about 10 seconds.
The averaging begins anew after briefly pressing the
SINGLE (10) pushbutton (RESET function).
Attention!
The pre- or post trigger will be automatically switched
off (PT0%) in the time base range from 100s/div to
50ms/div.
(9) ROL - indicates ROLL mode.
In ROLL mode the ROL-LED is lit and the readout
displays ”ROL”.
In this mode, the memory contents and thus also the
signal display, are continuously updated. Because signal
capture is untriggered, no idle states arise while waiting
for a new trigger event to start signal capture. With each
signal sampling the new value is shown on the right-
hand edge of the screen, while the previously captured
data are shifted to the left. The leftmost value is shifted
out of the memory and lost.
The recording can be stopped at any time by selecting
the HOLD (7) function.
ROLL mode can only be used with time coefficients
from 100s/div to 50ms/div, as lower time coefficients
(faster time base speeds) are impractical.
If the time base is set to values between 20ms/div and
100ns/div and ROLL mode is selected, the time base
will be automatically set to 50ms/div. The time deflection
coefficient set previously before switching to ROLL
mode will be internally stored (e.g. 20ms/div). If ROLL
mode has been selected inadvertently and the TIME/
DIV. knob has not been changed, the time base will be
automatically set to the internally stored coefficient when
switching from ROLL to AVERAGE mode.
(10) SINGLE - Pushbutton with two functions and associated
LEDs.
SINGLE
Pressing and holding this pushbutton switches the
SINGLE event capturing mode on or off. SINGLE mode
is indicated by the associated SGL-LED.
SINGLE mode is available in digital as well as in analog
mode and remains unchanged when switching over from
analog to digital mode or vice versa. The main purpose
of SINGLE is the capture of one time events, but it can
also be used in combination with repetitive signals.
SINGLE mode automatically selects A time base mode
and normal triggering (NM-LED lit). Otherwise the
automatic trigger (AT) would occur without an input
(trigger) signal.
If the trigger circuit is activated by RESET, one time
base sweep (analog mode) or one complete data
acquisition (digital mode) is performed after a suited
signal caused triggering. Switching over to SINGLE in
analog mode interrupts the time base sweep and blanks
the beam.
Storage mode only
Selecting SINGLE stops the current data acquisition. Until
a new data acquisition is started, the memory content
is displayed continuously.
In combination with Yt (time base) mode, SINGLE is
indicated by the readout. It displays the actual pre- or
post trigger value and ”SGL”instead of ”RFR”, ”ENV”,
”AV...”or ”ROL”. If XY mode is active the A time
deflection coefficient is replaced by the sampling rate
(e.g. 100MS/s) display in the readout and additionally
the trigger regarding information is switched off.
Attention!
If SINGLE mode is present in combination with
DUAL mode, the minimum time coefficient is 2µs/
div instead of 100ns/div. Similarly if X-MAG. x10 is
operative, then 200ns/div replaces 10ns/div.
ANALOG MODE ONLY
Selecting SINGLE mode switches the current sweep
off and blanks the screen.
SINGLE mode is indicated by the lit SGL-LED and the
readout displaying SGL next to the / (SLOPE) symbol.
RESET (RES)
Briefly pressing the SINGLE pushbutton causes a RESET
activating the trigger circuit if Yt mode is active. The
result depends on the current signal capture mode.
STORAGE MODE ONLY
a)
In combination with SINGLE, briefly pressing the
SINGLE pushbutton activates the RESET function. Then
both LEDs (SGL and RES) are lit. Whether the RES-
LED flashes once or is lit constantly, depends on:
1. the presence or absence of a trigger signal,
2. the selected time coefficient (time base) and
3. the pre- or post trigger setting.
After the RESET function is switched on, the signal
acquisition will be effective at once if the HOLD function
is not active. If the pre trigger function is active, the
prehistory must elapse before the trigger event becomes
effective. The signal capture terminates with the trigger
event only with 100% pre trigger setting.
With all other pre trigger and post trigger settings, the
signal acquisition is not complete when the trigger occurs
Controls and Readout

Subject to change without notice
16
and will only be terminated later. After termination the
RES-LED extinguishes but the signal display remains.
Briefly pressing the SINGLE pushbutton (RESET
function) again restarts a new single event capture which
then overwrites the previously recorded display.
Single events recorded in DUAL mode can also be
displayed in the XY mode when switched over to XY
operation.
XY mode
Briefly pressing the SINGLE pushbutton (RESET function)
causes one complete trigger independent signal
recording. The RES-LED extinguishes after completion.
Thereafter it is possible to switch over to DUAL mode
to display the signals in Yt (time base) mode.
Attention!
If time coefficients between 100s/div and 50ms/
div are present the signal acquisition becomes
visible at once as a ROLL display, but the signal
acquisition has nothing to do with ROLL mode.
b)
Briefly pressing the SINGLE pushbutton (RESET
function) is also effective if (instead of SINGLE)
ENVELOPE or AVERAGE mode is selected.
In both modes the evaluation / averaging begins anew.
ANALOG MODE ONLY
Capturing single events can also be carried out in analog
Yt (time base) mode (e.g. photographing).
Briefly pressing the SINGLE pushbutton activates the
RES-LED in SINGLE mode. The next trigger event then
unblanks the beam and causes one time base sweep.
Only in chopped DUAL mode can both channels be
displayed during one time base sweep.
XY mode is not available in combination with SINGLE
operation.
(11) REFERENCE - Reference memory pushbutton with two
associated LEDs.
The instrument contains two non volatile signal data
memories. Signal(s) stored in these memories can be
displayed separately or together in addition to the current
signal(s). The REFERENCE memory content will not be
erased by switching the instrument off.
Display
If neither the associated “I”nor the “II”LED is lit, no
reference signal is displayed. Briefly pressing the
REFERENCE pushbutton switches LED “I”on. Then in
addition to the current signal the reference Imemory
content is displayed. The switching sequence is:
dark –I–II - Iand II - dark.
The LED(s) indicate the memory content(s) which are
displayed. The display of the current signal(s) is not
affected by the reference display.
In XY mode the switching sequence is:
dark –Iand II –dark.
Overwrite
To overwrite the reference memory content with current
signal(s) the following procedure is required:
Determine the reference memory(ies) by briefly pressing
the REFERENCE pushbutton. Then press and hold the
pushbutton until a beep confirms reference memory
acquisition. Before starting the procedure, the current
signal can, but must not be protected by the HOLD
function.
As the reference signal is then equal to the current signal,
it is usually not visible at once. If neither HOLD nor
SINGLE is active, the current signal can be shifted by
the Y-POS. control(s) to make the reference signal visible.
If both REFERENCE Iand II LEDs lit and DUAL or XY
mode is active, both reference memories can be
overwritten at the same time. The signal from channel
I is stored in reference memory I and consequently
channel II in reference memory II. The relation between
channel and reference memory is present only during
DUAL or XY operation. If only REFERENCE I is active
and consequently displayed, only the current signal of
channel I can overwrite it.
In single channel mode (CHI or CHII) the current signal
can be stored in each of both reference memories. This
enables to create two reference curves with different Y
positions and to display them in condition REFERENCE
Iand II LEDs lit. The current signal then may not exceed
these limits. To avoid overwriting both reference
memories with the same signal, pressing and holding
the REFERENCE pushbutton only causes a warning beep
if single channel mode is present.
(12)SAVE / RECALL –Pushbuttons.
The instrument contains 9 non volatile memories. These
can be used by the operator to save instrument settings
and to recall them. This relates to all controls which are
electronically selected.
SAVE
Press the SAVE pushbutton briefly to start the save
procedure. The readout then indicates the letter “S”
followed by a cipher between 1 and 9, indicating the
memory location. If the instrument settings stored in
this memory location shall not be overwritten, briefly
press the SAVE or the RECALL pushbutton to select
another memory location. Each time the SAVE
pushbutton is briefly pressed the memory location cipher
increases until the location number 9 is reached. The
RECALL pushbutton function is similar but decreases
the memory location cipher until 1 is reached. Press and
hold SAVE for approx. 3 seconds to write the
instruments settings in the memory and additionally
switch the readout information (i.e. “S8”) off.
RECALL
To recall a front panel setup, start that procedure by
briefly pressing the RECALL pushbutton. The readout
then indicates the letter “R”and the memory location
number. If required, select a different memory location
as described above. Recall the settings by pressing and
holding the RECALL pushbutton for approx. 3 seconds.
If the SAVE or the RECALL pushbutton was depressed
inadvertently, briefly press both pushbuttons at the same
Controls and Readout

17
Subject to change without notice
time or wait approx. 10 seconds without pressing either
pushbutton to leave that function.
Attention:
Make sure that the signal to be displayed is similar
to the one that was present when the settings were
stored. If the signal is different (frequency,
amplitude) to the one during storage then a
distorted display may result.
If the SAVE or the RECALL pushbutton was depressed
inadvertently, briefly press both pushbuttons at the same
time or wait approx. 10 seconds without pressing either
pushbutton to exit that function.
Switching the instrument off results in an automatic SAVE
procedure of the present settings in memory location 9
and overwrites the data in that location. If the instrument
settings in memory location 9 are of importance, RECALL
9 before switching the instrument off.
The setting controls and LED’s for the Y amplifiers,
modes, triggering and time base are located
underneath the sector of the front panel described
before.
(13) TRS –Pushbutton and associated LED.
The instrument contains a trace separation function
which is required in the alternate time base mode to
separate the Btime base trace from the Atime base in
Y direction. Subsequently this function is only available
in alternate time base mode.
After the TRS pushbutton was pressed once the LED
related to that pushbutton is lit. Then the Y-POS. I control
knob is operative as vertical position control for the trace
of the Btime base. The maximum position shift is approx.
+/- 4 div. Without a change of the Y-POS. I control the
trace separation function is switched off automatically
after approx. 10 seconds. The trace separation function
can also be left by pressing the TRS pushbutton.
(14) Y-POS. I –Control knob with two functions.
The vertical trace position of channel I can be set with
this control knob. In ADD (addition) mode both (Y-POS.
I and Y-POS. II) control knobs are active. In alternate
time base mode, this control knob can be used to
separate the Btime base trace from the Atime base
trace. Please note TRS (13).
If automatic triggering (AT) is present and the input is
set to GD (34), the vertical trace position corresponds
with 0 Volt (reference) at the input and can be set to any
suitable position. In ADD mode these conditions apply
to both channels. After switching GD off and selecting
DC input coupling it is possible to determine the DC
content of a signal by comparing the actual Y position
with the previously determined 0 Volt Y position.
Y-POS. I Symbol
Provided that the readout is displayed and ”DC REFE-
RENCE = ON”is selected in the ”SETUP”submenu
”MISCELLANEOUS”the 0 Volt reference position is
indicated by a ground (⊥⊥
⊥⊥
⊥) symbol. For channel I this
symbol is displayed on the left of the vertical (graticule)
center line, if the Y-POS. I trace position is set within
the screen. This allows you to determine the 0 Volt
reference position at any time.
Attention!
In XY mode the 0 Volt reference symbol is automati-
cally switched off.
STORAGE MODE ONLY
The Y-POS. I control knob can also be used for shifting
the position of a signal stored with HOLD in vertical
direction. Additional information relating to this operation
is described under HOLD (7).
(15) Y-POS. II –Control knob with two functions.
The vertical trace position of channel II can be set with
this control knob. In ADD (addition) mode both (Y-POS.
I and Y-POS. II) control knobs are active. If the instrument
is set to analog XY mode this control knob is inactive
and the X-POS. knob must be used for a horizontal
position shift.
If automatic triggering (AT) is present and the input is
set to GD (38), the vertical trace position corresponds
with 0 Volt (reference) at the input and can be set to any
suitable position. In ADD mode these conditions apply
to both channels. After switching GD off and selecting
DC input coupling it is possible to determine the DC
content of a signal by comparing the actual Y position
with the previously determined 0 Volt Y position.
Y-POS. II Symbol
Provided that the readout is displayed and ”DC REFE-
RENCE = ON”is selected in the ”SETUP”submenu
”MISCELLANEOUS”the 0 Volt reference position is
indicated by a ground (⊥⊥
⊥⊥
⊥) symbol. For channel II this
symbol is displayed on the right of the vertical (graticule)
center line, if the Y-POS. II trace position is set within
the screen. This allows you to determine the 0 Volt
reference position at any time.
Attention!
In XY mode the 0 Volt reference symbol is automa-
tically switched off.
STORAGE MODE ONLY
In contrast to analog mode the Y-POS. II knob must be
used for X position shift in storage XY mode and the
X-POS. knob is disabled. The Y-POS. II control knob
can also be used for shifting the position of a signal
stored with HOLD. Additional information relating to this
operation is described under HOLD (7).
(16) NM - AT - (SLOPE)- Pushbutton with a double
function and associated NM-LED.
The following description assumes that Yt (time base)
mode has been chosen.
Controls and Readout

Subject to change without notice
18
NM - AT selection
Press and hold the pushbutton to switch over from auto-
matic to normal triggering (NM-LED above the push-
button lit) and vice versa. If the LED is dark, automatic
triggering is selected.
Whether the peak value detection in automatic trigger
mode is automatically activated or not, depends on the
trigger coupling setting (TRIG.MODE (27)). The way the
trigger point symbol in the readout responds on different
LEVEL control knob settings indicates the situation:
1.If the trigger symbol can not be shifted in the vertical
direction when no signal is applied or the signal height
is not sufficient, the peak value detection is active.
2.Under the condition that the trigger point symbol
cannot be shifted in such a way that it leaves the signal
display on the screen, the peak value detection is
active.
3.The peak value detection is switched off if the trigger
point can be set outside the maximum peak values of
the signal, thus causing an untriggered signal display.
Slope selection
Briefly pressing this pushbutton selects which slope of
the signal is used for triggering the time base generator.
Each time this pushbutton is briefly pressed, the slope
direction switches from falling edge to rising edge and
vice versa. The current setting is displayed in the readout
by a slope symbol.
AT/NM and (SLOPE)
The current time base mode determines the pushbutton
function:
a) A time base mode: Both functions affect the A time
base only.
b)alternate time base mode, with Bin free run condition:
Both functions affect the Atime base only.
c) alternate time base mode, with Bin triggered (DEL.-
TRIG.) condition: Btime base SLOPE selection only.
d)Btime base in free run condition: Both functions affect
the - not displayed - A time base.
e)Btime base in triggered (DEL.TRIG.) condition: Btime
base SLOPE selection only.
Function settings which are not accessible in the current
operating conditions remain unchanged.
Attention!
In digital mode, the operating conditions described
under item c) are not available.
(17) TR - Trigger indicator LED
The TR LED is lit in Yt (time base) mode if the triggering
conditions are met. Whether the LED flashes or is lit
constantly depends on the frequency of the trigger
signal.
(18) LEVEL –Control knob
Turning the control knob causes a different trigger point
setting (voltage). The trigger unit starts the time base
when the edge of a trigger signal (voltage) crosses the
trigger point. In most Yt modes the trigger point is
displayed in the readout by the symbol on the left vertical
graticule line. If the trigger point symbol would overwrite
other readout information or would be invisible when
being set above or below the screen, the symbol
changes and an arrow indicates in which vertical direction
the trigger point has left the screen.
The trigger point symbol is automatically switched off
in those modes where there is no direct relation between
the trigger signal and the displayed signal. The last set-
ting in Atime base mode is stored and still active if the
alternate (A and B) or Btime base are selected.
This allows for a different level setting for the B time
base if the DEL. TRIG. function is active. Under this
condition the letter “B”is added to the trigger point
symbol.
STORAGE MODE ONLY
In storage mode the trigger point symbol also indicates
the post or pre trigger condition by a horizontal position
shift. Please note “PTR”(8).
(19) X-POS. –Control knob
This control knob enables an X position shift of the
signal(s) in Yt and analog XY mode. In combination with
X magnification x10 this function makes it possible to
shift any part of the (Yt) signal on the screen.
STORAGE MODE ONLY
In XY mode the X-POS. knob is inoperative. The Y-POS.
II (15) knob must be used for X position shift.
(20) X-MAG. x10 –Pushbutton and LED
Each time this pushbutton is pressed the x10 LED
located above is switched on or off. If the x10 LED is lit,
the signal display in all Yt and time base modes is
expanded 10 fold and consequently only a tenth part of
the signal curve is visible. The interesting part of the
signal can be made visible with aid of the X-POS. control.
As the X expansion results in a higher time base speed
(lower time deflection coefficient), all time and frequency
relevant information in the readout is switched over.
Please note that in alternate time base mode the
intensified sector may become invisible due to the X
position setting.
This pushbutton is not operative in XY mode.
(21) VOLTS/DIV. –Control knob.
This control knob for channel I has a double function.
The following description relates to the input attenuator
function (VAR LED dark).
Turning the control knob clockwise increases the
sensitivity in a 1-2-5 sequence and decreases it if turned
Controls and Readout

19
Subject to change without notice
in the opposite direction (ccw.). The available range is
from 1mV/div up to 20V/div. The knob is automatically
switched inactive if the channel related to it is switched
off, or if the input coupling is set to GD (ground).
The deflection coefficients and additional information
regarding the active channels are displayed in the read-
out, i.e. “Y1: deflection coefficient, input coupling”.
The “: “symbolizes calibrated measuring conditions
and is replaced by the “> “symbol in uncalibrated
conditions.
In DUAL, ADD (addition) and XY mode both deflection
coefficients etc. are displayed. Instead of “Y1:...”and
“Y2:...”the readout displays “Y:...”and “X:...”in XY mode.
(22) CH I
VAR. - Pushbutton with several functions
CHI
Briefly pressing the pushbutton sets the instrument to
channel I (Mono CH I) mode. The deflection coefficient
displayed in the readout indicates the current conditions
(“Y1...”). If neither external nor line (mains) triggering
was active, the internal trigger source automatically
switches over to channel I (“TR:Y1...”). The last function
setting of the VOLTS/DIV (21) knob remains unchanged.
All channel related controls are active if INPUT CHI (32)
is not set to GD (34).
VAR.
Pressing and holding this pushbutton selects the VOLTS/
DIV. (21) control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
After switching the VAR-LED (21) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV. (21)
control knob counter clockwise reduces the signal height
and the deflection coefficient becomes uncalibrated. The
readout then displays i.e. “Y1>...”indicating the uncali-
brated condition instead of “Y1:...”. Pressing and holding
the CHI pushbutton again switches the LED off, sets
the deflection coefficient into calibrated condition and
activates the attenuator function. The previous vernier
setting will not be stored.
The CHI pushbutton can also be pressed simultaneously
with the DUAL(23) pushbutton.Please note item (23).
(23) DUAL - Pushbutton with multiple functions
DUAL
Briefly pressing this pushbutton switches over to DUAL
mode. Then both deflection coefficients are displayed.
The previous trigger setting stays as it was, but can be
changed.
All controls related to both channels are active, if the
inputs (32) and (36) are not set to GD (34) (38).
Whether alternated or chopped channel switching is
present in analog mode depends on the actual time base
setting, and is displayed in the readout.
ALT (analog mode only)
displayed in the readout, indicates alternate channel
switching. After each time base sweep the instrument
internally switches over from channel I to channel II and
vice versa. This channel switching mode is automatically
selected if any time coefficient from 200µs/div to 50ns/
div is active.
CHP (analog mode only)
indicates chopper mode, whereby the channel switching
occurs constantly between channel I and II during each
sweep. This channel switching mode occurs when any
time base setting between 500ms/div to 500µs/div has
been chosen.
The actual channel switching can be changed to the
opposite mode by briefly pressing both CHI (22) and
DUAL (23) simultaneously. If afterwards the time coef-
ficient is changed, the channel switching is automatically
set to the time coefficient related mode.
STORAGE MODE ONLY
In DUAL storage mode the signal capture is performed
by both A/D converters. As there is no requirement to
select between ALT and CHP, the selection is switched
off. The readout then displays the current signal capture
and display mode.
ADD
(addition) mode can be selected by briefly pressing the
DUAL (23) and CHII (26) pushbuttons simultaneously.
Whether the algebraic sum (addition) or the difference
(subtraction) of both input signals is displayed, depends
on the phase relationship and the INV (34) (38) setting(s).
As a result both signals are displayed as one signal. For
correct measurements the deflection coefficients for
both channels must be equal.
Please note “Operating modes of the vertical ampli-
fiers in Yt mode”.
In ADD mode the readout indicates a “+”sign located
between both channel deflection coefficients. The trigger
point symbol is switched off.
The Y -position of the signal can be influenced by both
Y-POS controls (14) and (15).
XY (analog mode)
mode can be switched on or off by pressing and holding
the DUAL pushbutton (23).
In XY mode the deflection coefficients are displayed as
“Y...”for channel I and “X...”) for channel II, followed
by “XY”. Except the cursor lines which may be active
and cause the display of the voltage measurement, all
other readout information including the trigger point
symbol are switched off. In addition to all trigger and
time base related controls, the Y-POS. II (15) knob and
INV (38) pushbutton are deactivated. For X position
alteration, the X-POS. (19) knob can be used.
Controls and Readout

Subject to change without notice
20
STORAGE MODE ONLY
In XY storage mode the readout indicates “XY”and
the RFR LED (9) is lit. No other STOR. MODE can be
chosen. Instead of the time deflection coefficient, the
readout displays the sampling rate (e.g. “100MS/s”)
which can be selected by the TIME/DIV. knob (29).
In contrast to analog mode, the INV. (38) pushbutton is
active and the X signal can be inverted. Also the Y-POS.
II (15) knob must be used for X position shift instead of
the X-POS. (19) control. If HOLD is active, the current
XY signal can be moved in X (Y-POS. II) and Y (Y-POS.
I) direction.
(24) TRIG.
ALT- Pushbutton with double function and associated
LEDs.
The pushbutton and the LEDs are deactivated in XY
mode and if line (mains) triggering is selected.
TRIG.
With the aid of this pushbutton, the trigger source can
be chosen. There are three trigger sources available:
channel I, channel II (both designated as internal trigger
sources) and the TRIG. EXT. (39) input for external
triggering. The availability of the internal sources depends
on the actual channel mode. The actual setting is
indicated by the associated LED(s) and the readout.
Briefly pressing the pushbutton switches over in the
following sequence:
I - II - EXT - I in DUAL mode,
I - EXT - I if mono channel I is present,
II - EXT - II under mono channel II conditions.
Each condition is indicated by the associated LED and
displayed by the readout (“TR:Y1...”, “TR:Y2...”and
“TR:EXT...”). The trigger point symbol is switched off
in external trigger condition.
ALT
Pressing and holding the pushbutton selects alternate
triggering in DUAL mode. Under these conditions both I
and II LEDs lit and the readout displays “TR:ALT...”. As
alternate triggering requires alternate channel operation
in analog mode, alternate channel switching is set
automatically. A change of the time base then has no
affect regarding the channel switching mode. In addition
to the deflection coefficients display, “ALT”is displayed
by the readout instead of “CHP”if analog mode is present.
In alternate trigger mode the trigger point symbol is
switched off. TVL,TVF trigger coupling and line (~)
triggering can not be selected.
Alternate triggering is not available or automatically
switched off under the following conditions:
SINGLE (SGL) mode
ADD (addition) mode,
alternate (A & B) time base mode,
B time base mode and
line (mains) trigger coupling.
STORAGE MODE ONLY
In ROLL mode the signal capture is untriggered. Con-
sequently alternate triggering is not available.
(25) VOLTS/DIV. –Control knob
This control knob for channel II has a double function.
The following description relates to the input attenuator
function (VAR LED dark).
Turning the control knob clockwise increases the
sensitivity in a 1-2-5 sequence and decreases it if turned
in the opposite direction (ccw.). The available range is
from 1mV/div up to 20V/div. The knob is automatically
switched inactive if the channel related to it is switched
off, or if the input coupling is set to GD (ground).
The deflection coefficients and additional information
regarding the active channels are displayed in the
readout, i.e. “Y2: deflection coefficient, input coup-
ling”. The “: “symbolizes calibrated measuring condi-
tions and is replaced by the “> “symbol in uncalibrated
conditions.
In DUAL,ADD (addition) and XY mode both deflection
coefficients etc. are displayed. Instead of “Y1:...”and
“Y2:...”the readout displays “Y:...”and “X:...”in XY
mode.
(26) CH II
VAR. - Pushbutton with several functions.
Channel mode
Briefly pressing the pushbutton sets the instrument to
channel II (Mono CH II) mode. The deflection coefficient
displayed in the readout indicates the current conditions
(“Y2...“). If neither external nor line (mains) triggering
was active, the internal trigger source automatically
switches over to channel II (“TR:Y2...“). The last function
setting of the VOLTS/DIV (25) knob remains unchanged.
All channel related controls are active if INPUT CHII (36)
is not set to GD (38).
VAR.
Pressing and holding this pushbutton selects the VOLTS/
DIV. (25) control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
After switching the VAR-LED (25) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV. (25)
control knob counter clockwise reduces the signal height
and the deflection coefficient becomes uncalibrated. The
readout then displays i.e. “Y2>...”indicating the
uncalibrated condition instead of “Y2:...”. Pressing and
holding the CHII pushbutton again switches the LED
off, sets the deflection coefficient into calibrated
condition and activates the attenuator function. The
previous vernier setting will not be stored.
The CHII pushbutton can also be pressed simultaneously
with the DUAL (23) pushbutton. Please note item (23).
Controls and Readout

21
Subject to change without notice
(27) TRIG. MODE - Pushbutton with double function and
indicator LEDs.
Pressing the upper or lower pushbutton selects the
trigger coupling. The actual setting is indicated by a
LED and by the readout (“TR: source, slope, AC”).
Each time the lower TRIG. MODE pushbutton is pressed
the trigger coupling changes in the sequence:
AC (DC content suppressed),
DC (peak value detection inactive),
HF (high-pass filter cuts off frequencies below
approx. 50kHz)
trigger point symbol switched off,
NR (high frequency noise rejected),
LF (low-pass filter cuts off frequencies above
approx. 1.5kHz),
TVL (TV signal, line pulse triggering)
trigger point symbol switched off,
TVF (TV signal, frame pulse triggering)
trigger point symbol switched off.
~(line/mains triggering) trigger point symbol and
TRIG. LEDs (24) switched off.
The readout displays “TR:~”. The ~symbol can
be changed by 180°by using the trigger slope
function (NM/AT - (16)).
Please note:
In delay trigger mode (B time base) the instrument
is automatically set to DC trigger coupling. This set-
ting can not be changed with the TRIG. MODE push-
buttons.
In some trigger modes such as alternate triggering, some
trigger coupling modes are automatically disabled and
can not be selected.
This trigger mode is present if the trigger coupling information
in The trigger point symbol is inactive in line/mains trigger
mode as there is no direct amplitude relationship between
the trigger voltage and the signal voltage.
(28) DEL.POS.
HO –LED - Control knob with two functions and
associated LED.
The function of this control knob depends on the time
base mode.
HO (hold off)
In A time base mode, the control knob applies to the
hold off time setting. If the HO-LED associated with the
knob is dark, the hold off time is set to minimum.
Turning the control knob clockwise switches the LED
on and extends the hold off time until the maximum is
reached (please note “Hold Off-time adjustment”). The
hold off time is automatically set to minimum (LED dark),
if the A time base setting is changed. The last hold off
time setting is stored if alternate (A and B) or B time
base mode is selected.
DEL.POS.
In alternate (A and B) and B time base modes, the knob
controls the delay time setting. Under alternate time
base mode conditions, the delay time is visible on the A
trace, beginning at the trace start and ending at the start
of the intensified sector. In the free run condition (delay
trigger not active) of the Btime base, an approximate
value is displayed in the readout (“Dt:...”). This is an aid
to find the position of the intensified sector which may
be very small.
If only the B time base is being operated, the delay time
can be varied, but there is no intensified sector as the A
trace is not visible.
STORAGE MODE ONLY
The HOLD OFF function is not available in storage mode.
(29) TIME/DIV. - Control knob with two functions.
The following description applies to the time base switch
function (VAR LED dark).
ANALOG MODE
Turning the control knob clockwise reduces the deflec-
tion coefficient in a 1-2-5 sequence and increases it if
turned in the opposite direction (ccw.). The time coef-
ficient(s) is (are) displayed in the readout.
In A time base mode, time deflection coefficients
between 500ms/div and 50ns/div can be chosen in 1-2-
5 sequence, if the X-MAG. x10 function is not activated.
During alternate (A and B) and Btime base operation,
the control knob changes the B time base setting in 1-2-
5 sequence. The available deflection coefficient range
is from 20ms/div up to 50ns/div (without X-MAG. x10)
but the availability depends on the Atime base setting.
The internal control of the oscilloscope prevents the B
time deflection coefficient from becoming higher than
the Adeflection coefficient, as such an operation
condition would make no sense. If the Atime base
setting is 200µs/div the Btime base range from 20ms/
div up to 500µs/div is not available and the maximum
time deflection coefficient for Bwould be 200µs/div. In
the last named condition the change of the Atime base
from 200µs/div to 100µs/div switches the Btime base
also to 100µs/div. However the Btime base setting
remains unchanged if the Atime base is set to 500µs/
div.
As already mentioned under DUAL (23) the channel
switching depends on the time deflection coefficient
setting. In the time base ranges from 500ms/div to
500µs/div chopped (CHP) channel switching is auto-
matically selected, through which the switching occurs
constantly during the time base sweeps. Alternate (ALT)
channel switching is automatically used in all other time
deflection coefficient settings (200µs/div - 50ns/div). In
the latter case the active channel is switched off and
the previously inactive channel is switched on after the
completion of each time base sweep. To avoid inter-
ference in chopped mode, or to make both channels
appear simultaneously visible, the actual setting (ALT
Controls and Readout

Subject to change without notice
22
or CHP) displayed in the readout can be overwritten
and changed to the opposite mode. This is carried out
by simultaneously pressing and holding the CHI (22) and
the DUAL (23) pushbutton.
STORAGE MODE ONLY
The time deflection ranges are different in storage mode
and therefore must be considered (please note item
(7)). In DUAL mode a selection between alternate or
chopped signal capture is neither required nor
possible. Consequently the readout does not display
such information (ALT, CHP).
In XY storage mode the TIME/DIV. knob is still active.
It is used to determine the sampling rate, which is then
displayed by the readout.
(30) A/ALT-
B - Pushbutton with multiple function.
The instrument contains two time bases designated A
and B. With the aid of the Btime base, signal parts
displayed by the Atime base can be expanded in X
direction. The expansion ratio depends on the time
deflection coefficient ratio of both time bases (i.e. “A:
100µs”,“B:1µs”= 100). With higher expansion ratios
the Btime base trace intensity reduces if analog mode
is active.
SINGLE (SGL) event signal capture mode is enabled in
Atime base mode only.
A/ALT function:
Each time the A/ALT- B pushbutton is briefly pressed,
the time base mode changes in the sequence A–
alternate - A.
A
In Atime base mode the TIME/DIV. (29) control knob
is operative only for this time base. The readout then
only displays the Atime coefficient. The time base
settings for this condition are stored if the time base
mode is changed.
ALT
If alternate (ALT) time base mode is selected, the
TIME/DIV (29) knob only controls the Btime base.
The alternate time base mode is a sub function of the B
time base mode where both time base traces are
displayed. Consequently the readout displays both time
deflection coefficients (e.g. “A:100µs B:1µs”). Unlike
the former Atime base mode, an intensified sector is
also visible on the Atrace. This sector indicates the signal
part which is displayed by the Btime base. The
intensified segment can be shifted horizontally by the
DEL. POS. (28) control knob continuously (if the B time
base is operated in free run conditions). The difference
between the start of the Atime base trace and the
beginning of the intensified sector shows the delay time.
This information is also displayed in the readout as an
approximate value (e.g. “Dt:2.5ms”) related to the
calibrated Atime coefficient (uncalibrated i.e. “Dt>
2.5ms”). The width of the segment decreases when
the B time coefficient is set to a lower value (higher
time deflection speed).
Alternate time base mode causes the display to alter
between Aand Btime base after each sweep (analog
mode). In alternate DUAL mode the sequence is channel
I with A time base, channel I with B time base, channel
II with A time base and channel II with B time base.
For better reading, the vertical position of the Btrace
position can be shifted (please note “TRS”(13)).
B function:
Pressing and holding the pushbutton switches over from
Aor alternate time base mode to Btime base mode
and vice versa.
In Btime base mode the display of the Atrace(s), the
intensified sector(s) and the Atime coefficient display
in the readout are disabled. As the trace separation (TRS)
is no longer required under these circumstances, this
function is switched off too. Consequently only the B
time coefficient is displayed by the readout.
Briefly pressing the pushbutton switches Btime base
mode off and Atime base mode on. Pressing and holding
the pushbutton switches over from Bto alternate time
base mode.
STORAGE MODE ONLY
In the combination of storage mode and alternate or B
time base operation, the pre- and post trigger are
automatically switched off. The PTR (8) pushbutton is
deactivated and instead of a pre- or post trigger value
the readout displays the delay time.
ROLL mode is not available in alternate or Btime base
mode.
(31) DEL. TRIG.
VAR. - Pushbutton with double function.
DEL. TRIG. function:
ANALOG MODE ONLY
Each time the pushbutton is briefly pressed, the instrument
switches between free run (untriggered) and triggered B
time base, if alternate or Btime base mode is present.
DIGITAL MODE ONLY
In contrast to analog operation the Btime base can not
be triggered in alternate time base mode.
ANALOG and DIGITAL MODE
The actual setting is displayed in the readout. Instead of
the approximate delay time (“Dt:...”) in free run mode,
the readout displays “DTr: slope, DC (trigger coup-
ling)”in triggered delay mode. In this mode, the former
Atime base trigger settings regarding the trigger mode
(automatic or normal), -coupling, -slope and -level
settings are stored but still active.
With the activated delay trigger the instrument is
automatically set to normal (NM) trigger mode and DC
Controls and Readout
Other manuals for HM1507-3
1
Table of contents
Other Hameg Test Equipment manuals

Hameg
Hameg HM 2005 User manual

Hameg
Hameg HM303-6 User manual

Hameg
Hameg HM205-2 User manual

Hameg
Hameg HM303-6 User manual

Hameg
Hameg HMO3000 Series Instruction manual

Hameg
Hameg HM204-2 User manual

Hameg
Hameg HM1008-2 User manual

Hameg
Hameg HM1507-3 User manual

Hameg
Hameg HM1507-3.02 User manual

Hameg
Hameg HM 1004-3.01 User manual

Hameg
Hameg HMO3000 Series Installation guide

Hameg
Hameg HM1004 User manual

Hameg
Hameg HM504-2 User manual

Hameg
Hameg HM 1004-3 Series User manual

Hameg
Hameg HM 203 User manual

Hameg
Hameg HMO Series User manual

Hameg
Hameg HM 1007 User manual

Hameg
Hameg HM 203-6 User manual

Hameg
Hameg HM 203-7 User manual

Hameg
Hameg CombiScope HM2008 User manual