Hameg HM404-2.02 User manual

®
Instruments
HANDBUCH•MANUAL•MANUEL
Oscilloscope
HM404-2.02
ENGLISH

MANUAL•HANDBUCH•MANUEL

3
Subject to change without notice
Table of contents
St.210601-Hüb/tke
General information regarding the CE marking ............ 4
General Information ........................................................ 6
Symbols ......................................................................... 6
Use of tilt handle ............................................................ 6
Safety ............................................................................. 6
Intended purpose and operating conditions ................. 6
EMC ............................................................................... 7
Warranty......................................................................... 7
Maintenance .................................................................. 7
Protective Switch-Off .................................................... 7
Power supply ................................................................. 7
Type of signal voltage ..................................................... 8
Amplitude Measurements ............................................. 8
Total value of input voltage............................................ 9
Time Measurements ..................................................... 9
Connection of Test Signal ............................................ 10
Controls and Readout .................................................... 11
Menu ................................................................................ 20
First Time Operation ...................................................... 20
Trace Rotation TR ........................................................ 21
Probe compensation and use ...................................... 21
Adjustment at 1kHz ..................................................... 21
Adjustment at 1MHz ................................................... 21
Operating modes of the vertical
amplifiers in Yt mode. ................................................. 22
X-Y Operation ............................................................... 22
Phase comparison with Lissajous figures .................. 23
Phase difference measurement
in DUAL mode (Yt) ....................................................... 23
Phase difference measurement in DUAL mode ........ 23
Measurement of an amplitude modulation ................ 23
Triggering and time base .............................................. 24
Automatic Peak (value) -Triggering.............................. 24
Normal Triggering ......................................................... 24
SLOPE ................................................................... 25
Trigger coupling ............................................................ 25
Triggering of video signals ........................................... 25
Line triggering (~) ........................................................ 26
Alternate triggering ...................................................... 26
External triggering ........................................................ 26
Trigger indicator ”TR” .................................................. 27
HOLD OFF-time adjustment ....................................... 27
Delay / After Delay Triggering ..................................... 27
AUTO SET ....................................................................... 29
Mean Value Display ....................................................... 29
Component Tester .......................................................... 30
General ........................................................................ 30
Using the Component Tester ...................................... 30
Test Procedure ............................................................. 30
Test Pattern Displays ................................................... 30
Testing Resistors ......................................................... 30
Testing Capacitors and Inductors ................................ 30
Testing Semiconductors ............................................. 30
Testing Diodes ............................................................. 30
Testing Transistors ...................................................... 31
In-Circuit Tests ............................................................. 31
Adjustments.................................................................... 31
RS232 Interface - Remote Control ............................... 32
Safety ........................................................................... 32
Operation ..................................................................... 32
RS-232 Cable ............................................................... 32
RS-232 protocol ........................................................... 32
Baud-Rate Setting ........................................................ 32
Data Communication ................................................... 32
Front Panel HM404-2 ..................................................... 33
Oscilloscope
HM404-2.02

Subject to change without notice
4
General information regarding the CE marking
HAMEG instruments fulfill the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual generic- and
product standards. In cases where different limit values are applicable, HAMEG applies the severer standard. For emission the limits for
residential, commercial and light industry are applied. Regarding the immunity (susceptibility) the limits for industrial environment have been
used.
The measuring- and data lines of the instrument have much influence on emmission and immunity and therefore on meeting the acceptance
limits. For different applications the lines and/or cables used may be different. For measurement operation the following hints and conditions
regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) sufficiently screened cables must
be used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline must be less than 3
meters and not be used outside buildings. If an interface has several connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L from HAMEG
are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction in the
manual for a shorter length, signal lines must be less than 3 meters and not be used outside buildings.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators double
screened cables (RG223/U, RG214/U) must be used.
3. Influence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fields, even with careful setup of the measuring equipment an influence of
such signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the instruments
specifications may result from such conditions in individual cases.
4. RF immunity of oscilloscopes.
4.1 Electromagnetic RF field
The influence of electric and magnetic RF fields may become visible (e.g. RF superimposed), if the field intensity is high. In most cases the
coupling into the oscilloscope takes place via the device under test, mains/line supply, test leads, control cables and/or radiation. The device
under test as well as the oscilloscope may be effected by such fields.
Although the interior of the oscilloscope is screened by the cabinet, direct radiation can occur via the CRT gap. As the bandwidth of each
amplifier stage is higher than the total –3dB bandwidth of the oscilloscope, the influence RF fields of even higher frequencies may be noticeable.
4.2 Electrical fast transients / electrostatic discharge
Electrical fast transient signals (burst) may be coupled into the oscilloscope directly via the mains/line supply, or indirectly via test leads and/or
control cables. Due to the high trigger and input sensitivity of the oscilloscopes, such normally high signals may effect the trigger unit and/or
may become visible on the CRT, which is unavoidable. These effects can also be caused by direct or indirect electrostatic discharge.
HAMEG GmbH
KONFORMITÄTSERKLÄRUNG
DECLARATION OF CONFORMITY
DECLARATION DE CONFORMITE
Instruments
Herstellers HAMEG GmbH
Manufacturer Kelsterbacherstraße 15-19
Fabricant D - 60528 Frankfurt
Bezeichnung / Product name / Designation:
Oszilloskop/Oscilloscope/Oscilloscope
Typ / Type / Type: HM404-2
mit / with / avec: -
Optionen / Options / Options: -
mit den folgenden Bestimmungen / with applicable regulations / avec les
directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
Angewendete harmonisierte Normen / Harmonized standards applied / Normes
harmonisées utilisées
Sicherheit / Safety / Sécurité
EN 61010-1: 1993 / IEC (CEI) 1010-1: 1990 A 1: 1992 / VDE 0411: 1994
EN 61010-1/A2: 1995 / IEC 1010-1/A2: 1995 / VDE 0411 Teil 1/A1: 1996-05
Überspannungskategorie / Overvoltage category / Catégorie de surtension: II
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility
Compatibilité électromagnétique
EN 61326-1/A1
Störaussendung / Radiation / Emission: Tabelle / table / tableau 4, Klasse / Class
/ Classe B.
Störfestigkeit / Immunity / Imunitee: Tabelle / table / tableau A1.
EN 61000-3-2/A14
Oberschwingungsströme / Harmonic current emissions / Émissions de courant
harmonique:
Klasse / Class / Classe D.
EN 61000-3-3
Spannungsschwankungen u. Flicker / Voltage fluctuations and flicker /
Fluctuations de tension et du flicker.
Datum /Date /Date Unterschrift / Signature /Signatur
15.01.2001
E. Baumgartner
Technical Manager /Directeur Technique
®

5
Subject to change without notice
Accessories supplied: Line Cord, Operators Manual on CD-ROM, 2 Probes1:1/ 10:1
40MHz Analog-Oscilloscope HM404
-2
Autoset, Save / Recall, Readout / Cursor and RS-232 Interface
2 Channels, DC - 40MHz, 1mV - 20V/div., Component Tester
Triggering: DC to 100 MHz; Automatic Peak to Peak; ≤≤
≤≤
≤0,5div.
Time Base: 0.5 s/div. to 10 ns/div.; with Delay & 2nd Trigger.
Specifications
Vertical Deflection
Operating modes: Channel I or CH II separate,
Channel I and II: alternate or chopped
Chopper Frequency: approx. 0.5MHz
Sum or Difference: from Channel I and CH. II
Invert: CH II
XY-Mode: via CH I(X) and CH I(Y)
Frequency range: 2x DC to 40MHz (-3dB)
Overshoot: ≤1%. Risetime: <8.75ns
Deflection coefficient: 14 calibrated positions
from 1mV/div to 20V/div in 1-2-5 sequence,
with variable 2.5:1 up to 50V/div.
Accuracy in calibrated positions
1mV/div to 2mV/div: ±5% (DC - 10MHz (-3dB))
5mV/div to 20V/div: ±3%
Input impedance: 1MΩII 18pF
Input coupling: DC - AC - GD (ground)
Input voltage: max. 400V (DC + peak AC)
Triggering
Automatic (peak to peak): ≥0.5div.
Range: ≥20Hz-100MHz
Normal with level control:DC-100MHz≤ (0.5div.)
Indicator for trigger action: LED
Slope: positive or negative
Sources: CH I or II, line, external
ALT. Triggering: CHI / CHII (≥0.8div.)
Coupling: AC (10Hz-100MHz)
DC (0-100MHz)
HF (50kHz - 100MHz)
LF (0 - 1.5kHz)
2nd triggering: normal with level control
External: ≥0.3Vpp (0 – 100MHz)
Active TV Sync. Separator: field & line, +/ –
Horizontal Deflection
Time coefficients: 22 calibrated steps from
0.5s/div. - 50ns/div. (±3%) in 1-2-5 sequence
Variable 2.5:1 up to 1.25s/div.(uncal.)
X-MAG.x10: up to 10ns/div. ±5%
Delay: approx. 140ms - 200ns, variable
Hold-off time: variable to approx. 10:1
Bandwidth X-amplifier: 0-3MHz (-3dB)
Input X-amplifier:
via Channel I, Sensitivity see CH I
X-Y-phase shift: <3° below 120kHz.
Operation / Control
Manual (front panel switches)
Autoset (automatic parameter selection)
Save/Recall:
9user defined parameter settings
RS232: interface for remote control via a PC
Readout: Display of parameter settings
Cursor measurement: ∆V, ∆t or ∆1/t (frequ.)
Component Tester
Test voltage: approx. 7Vrms (open circuit)
Test current: max. 7mArms (short circuit)
Test frequency: approx.50Hz
One test lead is grounded (Safety Earth)
General Information
CRT: Screen (8x10cm) internal graticule
Acceleration voltage: approx 2000V
Trace rotation: adjustable on front panel
Z Input: (Intens. modulation), max. +5V (TTL)
Calibrator:
0,2V ±1%, ≈ 1kHz/1MHz (t
r
<4ns)
Line voltage: 100-240V AC ±10%, 50/60Hz
Power consumption:approx. 34 Watt at 50Hz
Min./Max. ambient temperature: 0°C...+40°C
Protective system: Safety class I (IEC1010-1)
Weight: approx. 5.6kg. Color: techno-brown
Cabinet: W 285, H 125, D 380 mm
Subject to change without notice. 06/98
Signals of 50 and 100 MHz, alternate mode,
display of cursors and frequency values.
TV burst signal in delay mode with 2. trigger.
The excellent user interface characteristics of the new HM404-2 oscilloscope are
comparable with high tech scopes. Supported by two microprocessors any front
panel input is executed in a fraction of a second. A selftest procedure checks all
relevant parameters of the device; the test results will be displayed on screen within
ten seconds after power on. Supported by an on screen menu adjustments can be
performed without opening the scope.
It is recommended to use the Autoset function if signals of lower complexity shall
be displayed. The scope’s logic circuitry performs all relevant parameter settings
automatically to optimize the presentation of the signal(s). Of course, any parameter
may be modified manually as required. Front panel settings (measurement parameters)
and selected features are alphanumerically displayed on the screen. The cursor
functions enable the user to analyze a signal while watching the numeric readout for
voltage difference, time difference, or frequency values. Another feature is the sto-
rage capability for nine complete parameter settings, which may be stored and recalled
randomly by pushing the according front panel key.
Because of its high performance characteristics of the broad band signal amplifiers
and its excellent trigger bandwidth the scope is capable to display 100 MHz signals.
A delayed time base combined with a second trigger circuit makes the HM404-2 an
ideal instrument for high-resolution analysis of expanded, asynchronous signals.
Furthermore, the built in component tester and the 1kHz/1MHz calibrator are standard
equipment for this class of HAMEG scopes.
The instrument may be remotely controlled by any personal computer via its built-
in serial interface. A CD-ROM supplied with the scope, contains the instrument
commands and programming examples.

Subject to change without notice
6
General Information
be followed by the user to ensure safe operation and to retain
the oscilloscope in a safe condition.
The case, chassis and all measuring terminals are connected
to the protective earth contact of the appliance inlet. The
instrument operates according to Safety Class I (three-
conductor power cord with protective earthing conductor and
a plug with earthing contact).
The mains/line plug shall only be inserted in a socket outlet
provided with a protective earth contact. The protective
action must not be negated by the use of an extension cord
without a protective conductor.
The mains/line plug must be inserted before connec-
tions are made to measuring circuits.
The grounded accessible metal parts (case, sockets, jacks)
and the mains/line supply contacts (line/live, neutral) of the
instrument have been tested against insulation breakdown
with 2200V DC.
Under certain conditions, 50Hz or 60Hz hum voltages can
occur in the measuring circuit due to the interconnection with
other mains/line powered equipment or instruments. This can
be avoided by using an isolation transformer (Safety Class II)
between the mains/line outlet and the power plug of the
device being investigated. Most cathode-ray tubes develop X-
rays. However, the dose equivalent rate falls far below the
maximum permissible value of 36pA/kg (0.5mR/h).
Whenever it is likely that protection has been impaired, the
instrument shall be made inoperative and be secured against
any unintended operation. The protection is likely to be
impaired if, for example, the instrument
• shows visible damage,
• fails to perform the intended measurements,
• has been subjected to prolonged storage under unfavorable
conditions (e.g. in the open or in moist environments),
• has been subject to severe transport stress (e.g. in poor
packaging).
Intended purpose and operating conditions
This instrument must be used only by qualified experts who
are aware of the risks of electrical measurement. The ins-
trument is specified for operation in industry, light industry,
commercial and residential environments.
Due to safety reasons the instrument must only be connected
to a properly installed power outlet, containing a protective
earth conductor. The protective earth connection must not be
broken. The power plug must be inserted in the power outlet
while any connection is made to the test device.
The instrument has been designed for indoor use. The
permissible ambient temperature range during operation is
+10°C (+50°F) ... +40°C (+104°F). It may occasionally be
subjected to temperatures between +10°C (+50°F) and -10°C
(+14°F) without degrading its safety. The permissible ambient
temperature range for storage or transportation is -40°C (-
40°F) ... +70°C (+158°F). The maximum operating altitude is
up to 2200m (non-operating 15000m). The maximum relative
humidity is up to 80%.
If condensed water exists in the instrument it should be
acclimatized before switching on. In some cases (e.g.
extremely cold oscilloscope) two hours should be allowed
General Information
This oscilloscope is easy to operate. The logical arrangement
of the controls allows anyone to quickly become familiar with
the operation of the instrument, however, experienced users
are also advised to read through these instructions so that all
functions are understood.
Immediately after unpacking, the instrument should be
checked for mechanical damage and loose parts in the interior.
If there is transport damage, the supplier must be informed
immediately. The instrument must then not be put into
operation.
Symbols
ATTENTION - refer to manual
Danger - High voltage
Protective ground (earth) terminal
Use of tilt handle
To view the screen from the best angle, there are three
different positions (C, D, E) for setting up the instrument. If
the instrument is set down on the floor after being carried,
the handle automatically remains in the upright carrying
position (A). In order to place the instrument onto a horizontal
surface, the handle should be turned to the upper side of the
oscilloscope (C). For the D position (10° inclination), the handle
should be turned to the opposite direction of the carrying
position until it locks in place automatically underneath the
instrument. For the E position (20° inclination), the handle
should be pulled to release it from the D position and swing
backwards until it locks once more. The handle may also be
set to a position for horizontal carrying by turning it to the
upper side to lock in the B position. At the same time, the
instrument must be lifted, because otherwise the handle will
jump back.
Safety
This instrument has been designed and tested in accordance
with IEC Publication 1010-1 (overvoltage category II, pollution
degree 2), Safety requirements for electrical equipment for
measurement, control, and laboratory use. The CENELEC
regulations EN 61010-1 correspond to this standard. It has
left the factory in a safe condition. This instruction manual
contains important information and warnings which have to

7
Subject to change without notice
on which the technical data are based. Purchase of the
HAMEG scope tester HZ60, which despite its low price is
highly suitable for tasks of this type, is very much
recommended. The exterior of the oscilloscope should be
cleaned regularly with a dusting brush. Dirt which is difficult
to remove on the casing and handle, the plastic and aluminum
parts, can be removed with a moistened cloth (99% water
+1% mild detergent). Spirit or washing benzene (petroleum
ether) can be used to remove greasy dirt. The screen may be
cleaned with water or washing benzene (but not with spirit
(alcohol) or solvents), it must then be wiped with a dry clean
lint-free cloth. Under no circumstances may the cleaning fluid
get into the instrument. The use of other cleaning agents can
attack the plastic and paint surfaces.
Protective Switch-Off
This instrument is equipped with a switch mode power
supply. It has both overvoltage and overload protection,
which will cause the switch mode supply to limit power
consumption to a minimum. In this case a ticking noise may
be heard.
Power supply
The oscilloscope operates on mains/line voltages between
100VAC and 240VAC. No means of switching to different
input voltages has therefore been provided.
The power input fuses are externally accessible. The fuseholder
is located above the 3-pole power connector. The power input
fuses are externally accessible, if the rubber connector is
removed. The fuseholder can be released by pressing its
plastic retainers with the aid of a small screwdriver. The
retainers are located on the right and left side of the holder and
must be pressed towards the center. The fuse(s) can then be
replaced and pressed in until locked on both sides.
Use of patched fuses or short-circuiting of the fuseholder is
not permissible; HAMEG assumes no liability whatsoever for
any damage caused as a result, and all warranty claims
become null and void.
Fuse type:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: time-lag (T).
Attention!
There is a fuse located inside the instrument within the
switch mode power supply:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: fast (F).
This fuse must not be replaced by the operator!
General Information
before the instrument is put into operation. The instrument
should be kept in a clean and dry room and must not be
operated in explosive, corrosive, dusty, or moist environments.
The oscilloscope can be operated in any position, but the
convection cooling must not be impaired. The ventilation
holes may not be covered. For continuous operation the
instrument should be used in the horizontal position, preferably
tilted upwards, resting on the tilt handle.
The specifications stating tolerances are only valid if
the instrument has warmed up for 30minutes at an
ambient temperature between +15°C (+59°F) and +30°C
(+86°F). Values without tolerances are typical for an
average instrument.
EMC
This instrument conforms to the European standards regarding
the electromagnetic compatibility. The applied standards are:
Generic immunity standard EN50082-2:1995 (for industrial
environment) Generic emission standard EN50081-1:1992 (
for residential, commercial and light industry environment).
This means that the instrument has been tested to the
highest standards.
Please note that under the influence of strong electromagnetic
fields, such signals may be superimposed on the measured
signals.
Under certain conditions this is unavoidable due to the
instrument’s high input sensitivity, high input impedance and
bandwidth. Shielded measuring cables, shielding and earthing
of the device under test may reduce or eliminate those
effects.
Warranty
HAMEG warrants to its Customers that the products it
manufactures and sells will be free from defects in materials
and workmanship for a period of 2 years. This warranty shall
not apply to any defect, failure or damage caused by improper
use or inadequate maintenance and care. HAMEG shall not be
obliged to provide service under this warranty to repair
damage resulting from attempts by personnel other than
HAMEG representatives to install, repair, service or modify
these products.
In order to obtain service under this warranty, Customers
must contact and notify the distributor who has sold the
product. Each instrument is subjected to a quality test with 10
hour burn-in before leaving the production. Practically all early
failures are detected by this method. In the case of shipments
by post, rail or carrier the original packing must be used.
Transport damages and damage due to gross negligence are
not covered by the guarantee.
In the case of a complaint, a label should be attached to the
housing of the instrument which describes briefly the faults
observed. If at the same time the name and telephone
number (dialing code and telephone or direct number or
department designation) is stated for possible queries, this
helps towards speeding up the processing of guarantee
claims.
Maintenance
Various important properties of the oscilloscope should be
carefully checked at certain intervals. Only in this way is it
largely certain that all signals are displayed with the accuracy

Subject to change without notice
8
Type of signal voltage
Type of signal voltage
The oscilloscope HM404-2 allows examination of DC voltages
and most repetitive signals in the frequency range up to at
least 40MHz (-3dB).
The vertical amplifiers have been designed for minimum
overshoot and therefore permit a true signal display.
The display of sinusoidal signals within the bandwidth limits
causes no problems, but an increasing error in measurement
due to gain reduction must be taken into account when
measuring high frequency signals. This error becomes
noticeable at approx. 14MHz. At approx. 18MHz the reduction
is approx. 10% and the real voltage value is 11% higher. The
gain reduction error can not be defined exactly as the -3dB
bandwidth of the amplifiers differ between 40MHz and
42MHz. For sinewave signals the -6dB limit is approx. 50MHz.
When examining square or pulse type waveforms, attention
must be paid to the harmonic content of such signals. The
repetition frequency (fundamental frequency) of the signal
must therefore be significantly smaller than the upper limit
frequency of the vertical amplifier.
Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content which can be
used for triggering. This is the case with bursts, for instance.
To obtain a well-triggered display in this case, the assistance
of the variable holdoff function or the delayed time base may
be required. Television video signals are relatively easy to
trigger using the built-in TV-Sync-Separator (TV).
For optional operation as a DC or AC voltage amplifier, each
vertical amplifier input is provided with a DC/AC switch. DC
coupling should only be used with a series-connected attenuator
probe or at very low frequencies or if the measurement of the
DC voltage content of the signal is absolutely necessary.
When displaying very low frequency pulses, the flat tops may
be sloping with AC coupling of the vertical amplifier (AC limit
frequency approx. 1.6 Hz for 3dB). In this case, DC operation
is preferred, provided the signal voltage is not superimposed on
a too high DC level. Otherwise a capacitor of adequate
capacitance must be connected to the input of the vertical
amplifier with DC coupling. This capacitor must have a sufficiently
high breakdown voltage rating. DC coupling is also
recommended for the display of logic and pulse signals,
especially if the pulse duty factor changes constantly. Otherwise
the display will move upwards or downwards at each change.
Pure direct voltages can only be measured with DC-coupling.
The input coupling is selectable by the AC/DC pushbutton.
The actual setting is displayed in the readout with the ”= ”
symbol for DC- and the ”~ ”symbol for AC coupling.
Amplitude Measurements
In general electrical engineering, alternating voltage data
normally refers to effective values (rms = root-mean-square
value). However, for signal magnitudes and voltage
designations in oscilloscope measurements, the peak-to-
peak voltage (Vpp) value is applied. The latter corresponds to
the real potential difference between the most positive and
most negative points of a signal waveform.
If a sinusoidal waveform, displayed on the oscilloscope screen,
is to be converted into an effective (rms) value, the resulting
peak-to-peak value must be divided by 2x√2 = 2.83. Conversely,
it should be observed that sinusoidal voltages indicated in
Vrms (Veff) have 2.83 times the potential difference in Vpp.
The relationship between the different voltage magnitudes
can be seen from the following figure.
Voltage values of a sine curve
Vrms = effective value; Vp= simple peak or crest value;
Vpp = peak-to-peak value; Vmom = momentary value.
The minimum signal voltage which must be applied to the Y
input for a trace of 1div height is 1mVpp (± 5%) when this
deflection coefficient is displayed on the screen (readout) and
the vernier is switched off (VAR-LED dark). However, smaller
signals than this may also be displayed. The deflection
coefficients are indicated in mV/div or V/div (peak-to-peak
value).
The magnitude of the applied voltage is ascertained by
multiplying the selected deflection coefficient by the vertical
display height in div. If an attenuator probe x10 is used, a
further multiplication by a factor of 10 is required to ascertain
the correct voltage value.
For exact amplitude measurements, the variable control
(VAR) must be set to its calibrated detent CAL position.
With the variable control activated the deflection sensitivity
can be reduced up to a ratio of 2.5 to 1 (please note ”controls
and readout”). Therefore any intermediate value is possible
within the 1-2-5 sequence of the attenuator(s).
With direct connection to the vertical input, signals up
to 400Vpp may be displayed (attenuator set to 20V/
div, variable control to 2.5:1).
With the designations
H = display height in div,
U = signal voltage in Vpp at the vertical input,
D = deflection coefficient in V/div at attenuator switch,
the required value can be calculated from the two given
quantities:
However, these three values are not freely selectable. They
have to be within the following limits (trigger threshold,
accuracy of reading):
H between 0.5 and 8div, if possible 3.2 to 8div,
U between 0.5mVpp and 160Vpp,
D between 1mV/div and 20V/div in 1-2-5 sequence.
Examples:
Set deflection coefficient D = 50mV/div 0.05V/div,

9
Subject to change without notice
Type of signal voltage
observed display height H = 4.6div,
required voltage U = 0.05x4.6 = 0.23Vpp.
Input voltage U = 5Vpp,
set deflection coefficient D = 1V/div,
required display height H = 5:1 = 5div.
Signal voltage U = 230Vrmsx2
√
2 = 651Vpp
(voltage > 160Vpp, with probe 10:1: U = 65.1Vpp),
desired display height H = min. 3.2div, max. 8div,
max. deflection coefficient D = 65.1:3.2 = 20.3V/div,
min. deflection coefficient D = 65.1:8 = 8.1V/div,
adjusted deflection coefficient D = 10V/div.
The previous examples are related to the CRT graticule
reading. The results can also be determined with the aid of the
DV cursor measurement (please note ”controls and
readout”).
The input voltage must not exceed 400V, independent from
the polarity.
If an AC voltage which is superimposed on a DC voltage is
applied, the maximum peak value of both voltages must not
exceed + or -400V. So for AC voltages with a mean value of
zero volt the maximum peak to peak value is 800Vpp.
If attenuator probes with higher limits are used, the
probes limits are valid only if the oscilloscope is set to
DC input coupling.
If DC voltages are applied under AC input coupling conditions
the oscilloscope maximum input voltage value remains 400V.
The attenuator consists of a resistor in the probe and the 1MΩ
input resistor of the oscilloscope, which are disabled by the
AC input coupling capacity when AC coupling is selected. This
also applies to DC voltages with superimposed AC voltages.
It also must be noted that due to the capacitive resistance of
the AC input coupling capacitor, the attenuation ratio depends
on the signal frequency. For sinewave signals with frequencies
higher than 40Hz this influence is negligible.
With the above listed exceptions HAMEG 10:1 probes can be
used for DC measurements up to 600V or AC voltages (with
a mean value of zero volt) of 1200Vpp. The 100:1 probe HZ53
allows for 1200V DC or 2400Vpp for AC.
It should be noted that its AC peak value is derated at higher
frequencies. If a normal x10 probe is used to measure high
voltages there is the risk that the compensation trimmer
bridging the attenuator series resistor will break down causing
damage to the input of the oscilloscope.
However, if for example only the residual ripple of a high
voltage is to be displayed on the oscilloscope, a normal x10
probe is sufficient. In this case, an appropriate high voltage
capacitor (approx. 22-68nF) must be connected in series with
the input tip of the probe.
With Y-POS. control (input coupling to GD) it is possible to use
a horizontal graticule line as reference line for ground potential
before the measurement. It can lie below or above the
horizontal central line according to whether positive and/or
negative deviations from the ground potential are to be
measured.
Total value of input voltage
The dotted line shows a voltage alternating at zero volt level. If
superimposed on a DC voltage, the addition of the positive peak
and the DC voltage results in the max. voltage (DC + ACpeak).
Time Measurements
As a rule, most signals to be displayed are periodically
repeating processes, also called periods. The number of
periods per second is the repetition frequency. Depending on
the time base setting (TIME/DIV.-knob) indicated by the
readout, one or several signal periods or only a part of a period
can be displayed. The time coefficients are stated in ms/div,
µs/div or ns/div. The following examples are related to the
CRT graticule reading. The results can also be determined
with the aid of the ∆T and 1/∆T cursor measurement (please
note ”controls and readout”).
The duration of a signal period or a part of it is determined by
multiplying the relevant time (horizontal distance in div) by the
(calibrated) time coefficient displayed in the readout.
Uncalibrated, the time base speed can be reduced until a
maximum factor of 2.5 is reached. Therefore any intermediate
value is possible within the 1-2-5 sequence.
With the designations
L = displayed wave length in div of one period,
T = time in seconds for one period,
F = recurrence frequency in Hz of the signal,
Tc = time coefficient in ms, µs or ns/div and the relation
F = 1/T, the following equations can be stated:
However, these four values are not freely selectable. They
have to be within the following limits:
L between 0.2 and 10div, if possible 4 to 10div,
T between 10ns and 5s,
F between 0.5Hz and 100MHz,
Tc between 100ns/div and 500ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and
Tc between 10ns/div and 50ms/div in 1-2-5 sequence (with X-
MAG. (x10) active).
Examples:
Displayed wavelength L = 7div,
set time coefficient Tc = 100ns/div,
required period T = 7x100x10-9 = 0.7µs

Subject to change without notice
10
Type of signal voltage
required rec. freq. F = 1:(0.7x10-6) = 1.428MHz.
Signal period T = 1s,
set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.
Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10-3) = 100Hz.
TV-Line frequency F = 15625Hz,
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15 625x10-5) = 6.4div.
Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x103) = 0.25ms/div,
min. time coefficient Tc = 1:(10x103) = 0.1ms/div,
set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(103x0.2x10-3) = 5div.
Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) button: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10-6) = 25MHz,
required period T = 1:(25x106) = 40ns.
If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(X-MAG. (x10) active). In this case, the time interval of interest
can be shifted to the screen center using the X-POS. control.
When investigating pulse or square waveforms, the critical
feature is the risetime of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly
influence the measuring accuracy, the risetime is generally
measured between 10% and 90% of the vertical pulse height.
For measurement, adjust the Y deflection coefficient using its
variable function (uncalibrated) together with the Y-POS.
control so that the pulse height is precisely aligned with the
0% and 100% lines of the internal graticule. The 10% and
90% points of the signal will now coincide with the 10% and
90% graticule lines. The risetime is given by the product of the
horizontal distance in div between these two coincident
points and the calibrated time coefficient setting. The fall time
of a pulse can also be measured by using this method.
The following figure shows correct positioning of the
oscilloscope trace for accurate risetime measurement.
With a time coefficient of 10ns/div (X x10 magnification
active), the example shown in the above figure results in a
total measured risetime of
ttot = 1.6div x 10ns/div = 16ns
When very fast risetimes are being measured, the risetimes
of the oscilloscope amplifier and of the attenuator probe has
to be deducted from the measured time value. The risetime
of the signal can be calculated using the following formula.
tr= √ttot2- tosc2- tp2
In this ttot is the total measured risetime, tosc is the risetime
of the oscilloscope amplifier (approx. 8.75ns), and tpthe
risetime of the probe (e.g. = 2ns). If ttot is greater than 100ns,
then ttot can be taken as the risetime of the pulse, and
calculation is unnecessary.
Calculation of the example in the figure above results in a
signal risetime:
tr = √162- 8.752- 22 = 13.25ns
The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to
measure in any display position and at any signal amplitude.
It is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness
and that the horizontal distance at 10% and 90% of the
amplitude is measured. If the edge shows rounding or
overshooting, the 100% should not be related to the peak
values but to the mean pulse heights. Breaks or peaks
(glitches) next to the edge are also not taken into account.
With very severe transient distortions, the rise and fall time
measurement has little meaning. For amplifiers with
approximately constant group delay (therefore good pulse
transmission performance) the following numerical relationship
between rise time tr (in ns) and bandwidth B (in MHz) applies:
Connection of Test Signal
In most cases briefly depressing the AUTO SET causes a
useful signal related instrument setting. The following
explanations refer to special applications and/or signals,
demanding a manual instrument setting. The description of
the controls is explained in the section ”controls and
readout”.
Caution:
When connecting unknown signals to the oscilloscope
input, always use automatic triggering and set the
input coupling switch to AC (readout). The attenuator
should initially be set to 20V/div.
Sometimes the trace will disappear after an input signal has
been applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160Vpp and
the deflection coefficient (VOLTS/DIV.) in calibrated condition,
an attenuator probe must be inserted before the vertical
input. If, after applying the signal, the trace is nearly blanked,
the period of the signal is probably substantially longer than
the set time deflection coefficient (TIME/DIV.). It should be
switched to an adequately larger time coefficient.
The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as
HZ32 or HZ34, or reduced through a x10 or x100 attenuator
probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to
approx. 50kHz). For higher frequencies, the signal source
must be of low impedance, i.e. matched to the characteristic
resistance of the cable (as a rule 50Ω). Especially when
transmitting square and pulse signals, a resistor equal to the
characteristic impedance of the cable must also be connected
across the cable directly at the Y-input of the oscilloscope.

11
Subject to change without notice
They should be as short and thick as possible. When the
attenuator probe is connected to a BNC-socket, a BNC-
adapter, should be used. In this way ground and matching
problems are eliminated. Hum or interference appearing in
the measuring circuit (especially when a small deflection
coefficient is used) is possibly caused by multiple grounding
because equalizing currents can flow in the shielding of the
test cables (voltage drop between the protective conductor
connections, caused by external equipment connected to the
mains/line, e.g. signal generators with interference protection
capacitors).
Controls and Readout
The following description assumes that the operating
mode ”COMPONENT TEST”is switched off.
All important measuring parameter settings are display-
ed in the screen Readout when the oscilloscope is on.
The LED indicators on the large front panel facilitate operation
and provide additional information. Electrical end positions of
controls are indicated by acoustic signal (beep).
All controls, except the power switch (POWER), the calibration
frequency pushbutton (CAL. 1kHz/1MHz), the FOCUScontrol
and the trace rotation control, are electronically set and
interrogated. Thus, all electronically set functions and their
current settings can be stored and also remotely controlled.
The large front panel is, as is usual with Hameg oscilloscopes,
marked with several fields.
The following controls and LED indicators are located
on the top, to the right of the screen, above the horizon-
tal line:
(1) POWER
Pushbutton and symbols for ON (I) and OFF (O).
After the oscilloscope is switched on, all LEDs are lit and
an automated instrument test is performed. During this
time the HAMEG logo and the software version are
displayed on the screen. After the internal test is
completed succesfully, the overlay is switched off and
the normal operation mode is present. Then the last used
settings become activated and one LED indicates the
ON condition.
(2) AUTO SET
Briefly depressing this pushbutton results in an automatic
instrument setting automatically selecting Yt mode. The
instrument is set to the last used Yt mode setting (CH I,
CH II or DUAL). SEARCH (SEA) and DELAY (DEL and
DTR) mode is automatically switched off. Please note
”AUTO SET”.
Automatic CURSOR supported voltage measurement:
If CURSORvoltage measurement is present, the CURSOR
lines are automatically set to the positive and negative
peak value of the signal. The accuracy of this function
decreases with higher frequencies and is also influenced
Controls and Readout
When using a 50Ωcable such as the HZ34, a 50Ωthrough
termination type HZ22 is available from HAMEG. When
transmitting square signals with short rise times, transient
phenomena on the edges and top of the signal may become
visible if the correct termination is not used. A terminating
resistance is sometimes recommended with sine signals as
well. Certain amplifiers, generators or their attenuators maintain
the nominal output voltage independent of frequency only if
their connection cable is terminated with the prescribed
resistance. Here it must be noted that the terminating resistor
HZ22 will only dissipate a maximum of 2Watts. This power is
reached with 10Vrms or at 28.3Vpp with sine signal. If a x10
or x100 attenuator probe is used, no termination is necessary.
In this case, the connecting cable is matched directly to the high
impedance input of the oscilloscope. When using attenuators
probes, even high internal impedance sources are only slightly
loaded (approx. 10MΩII 12pF or 100MΩII 5pF with HZ53).
Therefore, if the voltage loss due to the attenuation of the
probe can be compensated by a higher amplitude setting, the
probe should always be used. The series impedance of the
probe provides a certain amount of protection for the input of
the vertical amplifier. Because of their separate manufacture,
all attenuator probes are only partially compensated, therefore
accurate compensation must be performed on the oscilloscope
(see Probe compensation ).
Standard attenuator probes on the oscilloscope normally
reduce its bandwidth and increase the rise time. In all cases
where the oscilloscope bandwidth must be fully utilized (e.g.
for pulses with steep edges) we strongly advise using the
probes HZ51 (x10) HZ52 (x10 HF) and HZ54 (x1 and x10). This
can save the purchase of an oscilloscope with larger bandwidth.
The probes mentioned have a HF-calibration in addition to low
frequency calibration adjustment. Thus a group delay correction
to the upper limit frequency of the oscilloscope is possible
with the aid of an 1MHz calibrator, e.g. HZ60.
In fact the bandwidth and rise time of the oscilloscope are not
noticably changed with these probe types and the waveform
reproduction fidelity can even be improved because the probe
can be matched to the oscilloscopes individual pulse response.
If a x10 or x100 attenuator probe is used, DC input
coupling must always be used at voltages above 400V.
With AC coupling of low frequency signals, the attenu-
ation is no longer independent of frequency, pulses
can show pulse tilts. Direct voltages are suppressed
but load the oscilloscope input coupling capacitor
concerned. Its voltage rating is max. 400 V (DC + peak
AC). DC input coupling is therefore of quite special
importance with a x100 attenuation probe which usu-
ally has a voltage rating of max. 1200 V (DC + peak AC).
A capacitor of corresponding capacitance and voltage
rating may be connected in series with the attenuator
probe input for blocking DC voltage (e.g. for hum
voltage measurement).
With all attenuator probes, the maximum AC input voltage
must be derated with frequency usually above 20kHz.
Therefore the derating curve of the attenuator probe type
concerned must be taken into account.
The selection of the ground point on the test object is
important when displaying small signal voltages. It should
always be as close as possible to the measuring point. If this
is not done, serious signal distortion may result from spurious
currents through the ground leads or chassis parts. The
ground leads on attenuator probes are also particularly critical.

Subject to change without notice
12
can be used by the operator to save instrument settings
and to recall them. This relates to all settings with the
exception of FOCUS, TR (trace rotation) and the calibrator
frequency pushbutton.
SAVE:
Press the SAVE pushbutton briefly to start the save
procedure. The readout then indicates the letter ”S”
followed by a cipher between 1 and 9, indicating the
memory location. If the instrument settings stored in this
memory location must not be overwritten, briefly press
the SAVE or the RECALL pushbutton to select another
memory location. Each time the SAVE pushbutton is
briefly pressed the memory location cipher increases until
the location number 9 is reached. The RECALL
pushbutton function is similar but decreases the memory
location cipher until 1 is reached. Press and hold SAVE
for approx. 3 seconds to write the instruments settings
in the memory.
RECALL:
To recall a front panel setup, start that procedure by briefly
pressing the RECALL pushbutton. The readout then
indicates the letter ”R”and the memory location number.
If required, select a different memory location as
described above. Recall the settings by pressing and
holding the RECALL pushbutton for approx. 3 seconds.
Attention:
Make sure that the signal to be displayed is similar to
the one that was present when the settings were
stored. If the signal is different (frequency, amplitude)
to the one during storage then a distorted display may
result.
If the SAVE or the RECALL pushbutton was depressed
inadvertently, briefly press both pushbuttons at the same
time or wait approx. 10 seconds without pressing either
pushbutton to exit that function.
Switching the instrument off automatically stores the actual
settings in memory location 9, with the effect that different
settings previously stored in this location get lost. To prevent
this, RECALL 9 before switching the instrument off.
Attention!
Both pushbuttons have a second function if the instru-
ment is switched to menu operation. Please note
”MENU”.
The setting controls and LED’s for the Y amplifiers,
modes, triggering and time base are located under-
neath the sector of the front panel described before.
(8) Y-POS. I - Control knob.
The vertical trace position of channel I can be set with
this control knob. In ADD (addition) mode both (Y-POS. I
and Y-POS. II) control knobs are active. If the instrument
is set to XY mode this control knob is inactive and the
X-POS. knob must be used for a horizontal position shift.
DC voltage measurement:
If no signal is applied at the INPUT CHI (26), the vertical
trace position represents 0 Volt. This is the case if INPUT
CHI (26) or in addition (ADD) mode, both INPUT CHI (26)
and INPUT CHII (30), are set to GD (ground) and automatic
triggering (AT (10)) is present to make the trace visible.
The trace then can be set to vertical position which is
suited for the following DC voltage measurement.
Controls and Readout
by the signal‘s pulse duty factor.
In DUAL mode the CURSOR lines are related to the signal
which is used for internal triggering.
If the signal height is insufficient, the CURSOR lines do
not change.
(3) RM
The remote control mode can be switched on or off
(”RM”LED dark) via the RS232 interface. On condition
that the ”RM”LED is lit, all electronically selectable con-
trols on front panel are inactive. This state can be left by
depressing the AUTO SET pushbutton provided it was
not deactivated via the interface.
(4) INTENS - READOUT
Knob with associated pushbutton and LEDs.
This control knob is for adjusting the A trace and readout
intensity. Turning this knob clockwise increases and
turning it counterclockwise decreases the intensity.
The READOUT pushbutton below is for selecting the
function in two ways.
If the readout (RO) is not switched off, briefly pressing
the READOUT pushbutton switches over the INTENS
knob function indicated by a LED in the sequence:
Yt (time base) mode: A - RO - A
XY mode: A - RO - A.
Component Test: A - RO - A.
Pressing and holding the READOUT pushbutton switches
the readout on or off. In readout off condition the INTENS
knob function can consequently not be set to RO.
Switching the readout off, may be required if interference
is visible on the signal(s). Such interference may also
originate from the chopper generator if the instrument is
operated in chopped DUAL mode.
With the exception of the letters ”CT”all other READOUT
information is switched off in COMPONENT TEST mode.
All INTENS settings are stored after the instrument is
switched off.
The AUTOSET function switches the readout on. The
INTENS setting for each function is automatically set to
the mean value, if less intensity was previously selected.
(5) TR
The trace rotation control can be adjusted with a small
screwdriver (please note ”trace rotation TR”)
(6) FOCUS
This control knob effects both the trace and the readout
sharpness.
(7) SAVE / RECALL
The instrument contains 9 non volatile memories. These

13
Subject to change without notice
Controls and Readout
After switching GD (ground) off and selecting DC input
coupling, a DC signal applied at the input changes the
trace position in vertical direction. The DC voltage then
can be determined by taking the deflection coefficient,
the probe factor and the trace position change in respect
to the previous 0 Volt position into account.
”0 Volt”Symbol:
The determination of the ”0 Volt”position is not necessary
if the readout is switched on and the software setting
”DC Ref. = ON”is selected in the ”SETUP”submenu
”Miscellaneous”. Then the ”⊥⊥
⊥⊥
⊥”symbol to the left of the
screen‘s vertical center line always indicates the ”0 Volt”
trace position in CHI and DUAL mode.
The ”0 Volt”position symbol (⊥⊥
⊥⊥
⊥)will not be displayed in
XY and ADD (addition) mode.
(9) Y-POS. II - Control knob.
The vertical trace position of channel II can be set with
this control knob. In ADD (addition) mode both (Y-POS. I
and Y-POS. II) control knobs are active.
DC voltage measurement:
If no signal is applied at the INPUT CHII (30), the vertical
trace position represents 0 Volt. This is the case if INPUT
CHII (30) or in addition (ADD) mode, both INPUT CHI (26)
and INPUT CHII (30), are set to GD (ground) and automatic
triggering (AT (10)) is present to make the trace visible.
The trace then can be set to vertical position which is
suited for the following DC voltage measurement.
After switching GD (ground) off and selecting DC input
coupling, a DC signal applied at the input changes the
trace position in vertical direction. The DC voltage then
can be determined by taking the deflection coefficient,
the probe factor and the trace position change in respect
to the previous 0 Volt position into account.
”0 Volt”Symbol:
The determination of the ”0 Volt”position is not necessary
if the readout is switched on and the software setting
”DC Ref. = ON”is selected in the ”SETUP”submenu
”Miscellaneous”. Then the ”⊥⊥
⊥⊥
⊥”symbol to the right of
the screen‘s vertical center line always indicates the ”0
Volt”trace position in CHI and DUAL mode.
The ”0 Volt”position symbol (⊥⊥
⊥⊥
⊥) will not be displayed in
XY and ADD (addition) mode.
(10) NM - AT -
Pushbutton with a double function and associated NM-
LED.
NM - AT selection:
Press and hold the pushbutton to switch over from
automatic (peak value) to normal triggering (NM-LED
above the pushbutton lit) and vice versa. If the LED is
dark, automatic (peak value) triggering is selected.
Whether the peak value detection in automatic trigger
mode is automatically activated or not, depends on the
trigger coupling setting (TRIG.MODE). The way the
trigger point symbol in the readout responds on different
LEVEL control knob settings indicates the situation:
1. If the trigger symbol can not be shifted in the vertical
direction when a signal is not applied or the signal
height is not sufficient, the peak value detection is
active.
2. Under the condition that the trigger point symbol
cannot be shifted in such a way that it leaves the signal
display on the screen, the peak value detection is
active.
3. The peak value detection is switched off if the trigger
point can be set outside the maximum peak values of
the signal, thus causing an untriggered signal display.
Slope selection:
Briefly pressing this pushbutton selects which slope of
the signal is used for triggering the time base generator.
Each time this pushbutton is briefly pressed, the slope
direction switches from falling edge to rising edge and
vice versa.
The current setting is displayed in the readout by a slope
symbol. The last setting in undelayed time base mode is
stored and still active if triggered DELAY (DTR) time base
mode is selected. This allows for a different slope setting
for the triggered DELAY (DTR) time base mode.
(11)TR - Trigger indicator LED.
The TR LED is lit in Yt mode if the triggering conditions
are met. Whether the LED flashes or is lit constantly
depends on the frequency of the trigger signal.
(12)LEVEL - Control knob.
Turning the LEVEL knob causes a different trigger point
setting (voltage). The trigger unit starts the time base
when the edge of a trigger signal crosses the trigger
point. In most Yt modes the trigger point is displayed in
the readout by the symbol on the left vertical graticule
line. If the trigger point symbol would overwrite other
readout information or would be invisible when being
set above or below the screen, the symbol changes
and an arrow indicates in which vertical direction the
trigger point has left the screen.
The trigger point symbol is automatically switched off
in those modes where there is no direct relation
between the trigger signal and the displayed signal. The
last setting in undelayed time base mode is stored and
still active if triggered DELAY (DTR) time base mode is
selected.
This allows for a different level setting for the triggered
DELAY (DTR) time base mode.
(13) X-POS. - Control knob.
This control knob enables an X position shift of the
signal(s) in Yt and XY mode. In combination with X
magnification x10 (Yt mode) this function makes it
possible to shift any part of the signal on the screen.

Subject to change without notice
14
Controls and Readout
(14) X-MAG. x10 - Pushbutton and LED.
Each time this pushbutton is pressed the x10 LED located
above is switched on or off. If the x10 LED is lit, the
signal display in all Yt (time base) modes is expanded 10
fold and consequently only a tenth part of the signal curve
is visible. The interesting part of the signal can be made
visible with aid of the X-POS. (13) control. As the X
expansion results in a higher time base speed (lower time
deflection coefficient), all time and frequency relevant
information in the readout is switched over.
Please note!
The expansion is 5 fold if the time base is set to 50ns/
div. Consequently the lowest time deflection coeffi-
cient is 10ns/div.
This pushbutton is not operative in XY mode.
(15) VOLTS/DIV.
This control knob for channel I has a double function.
The following description relates to the input attenuator
function (VAR-LED dark).
Turning the control knob clockwise increases the sensi-
tivity (decreases the deflection coefficient) in a 1-2-5
sequence and decreases the sensitivity (increases the
deflection coefficient) if turned in the opposite direction
(ccw.). The available range is from 1mV/div up to 20V/
div. The knob is automatically switched inactive if the
channel related to it is switched off, or if the input coupling
is set to GD (ground).
The deflection coefficients and additional information
regarding the active channel(s) are displayed in the
readout, e.g. ”Y1: deflection coefficient, input coup-
ling”. The ”:”symbolizes calibrated measuring conditions
and is replaced by the ”>”symbol in uncalibrated con-
ditions.
(16) CH I - VAR. - Pushbutton with several functions.
CH I mode:
Briefly pressing the CHI button sets the instrument to
channel I (Mono CH I) mode. The deflection coefficient
displayed in the readout indicates the current conditions
(”Y1...”). If neither external nor line (mains) triggering
was active, the internal trigger source automatically
switches over to channel I (TRIG.-LED (18) CHI lits).
The last function setting of the VOLTS/DIV (15) knob
remains unchanged.
All channel I related controls are active if the input (26)
is not set to GD (28).
VAR.:
Pressing and holding this pushbutton selects the VOLTS/
DIV. (15) control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
After switching the VAR-LED (15) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV. (15)
control knob counter clockwise reduces the signal height
and the deflection coefficient becomes uncalibrated.
The readout then displays e.g. ”Y1>...”indicating the
uncalibrated condition instead of ”Y1:...”. Pressing and
holding the CHI pushbutton again switches the LED off,
sets the deflection coefficient into calibrated condition
and activates the attenuator function. The previous
vernier setting will not be stored.
The CHI pushbutton can also be pressed simultaneously
with the DUAL(17) button. Please note item (17).
(17) DUAL - XY - Pushbutton with multiple functions.
DUAL mode:
Briefly pressing this button switches over to DUAL
mode. Both deflection coefficients are then displayed.
The previous trigger setting stays as it was, but can be
changed.
All controls related to both channels are active, if the
inputs (26) and (30) are not set to GD (28) (32).
Whether alternated or chopped channel switching is
present depends on the actual time base setting, and is
displayed in the readout.
ALT:
displayed in the readout, indicates alternate channel
switching. After each time base sweep the instrument
internally switches over from channel I to channel II and
vice versa. This channel switching mode is automatically
selected if any time coefficient from 200µs/div to 50ns/
div is active.
CHP:
indicates chopper mode, whereby the channel switching
occurs constantly between channel I and II during each
sweep. This channel switching mode occurs when any
time base setting between 500ms/div and 500µs/div has
been chosen.
The actual channel switching can be changed to the
opposite mode by briefly pressing both CHI (16) and
DUAL (17) simultaneously. If afterwards the time
coefficient is changed, the channel switching is
automatically set to the time coefficient related mode.
ADD mode:
Addition mode can be selected by briefly pressing the
DUAL (17) and CHII (20) buttons simultaneously.
Whether the algebraic sum (addition) or the difference
(subtraction) of both input signals is displayed, depends
on the phase relationship and the INV (32) setting. As a
result both signals are displayed as one signal. For correct
measurements the deflection coefficients for both
channels must be equal.

15
Subject to change without notice
Controls and Readout
Please note ”Operating modes of the vertical amplifiers
in Yt mode”.
The readout indicates this mode by a ”+”sign located
between both channel deflection coefficients. While the
trigger mode is not affected, the trigger point symbol is
switched off.
The Y-position of the signal can be influenced by both Y-
POS controls (8) and (9).
XY mode:
This mode can be switched on or off by pressing and
holding the DUAL button (17).
In XY mode the deflection coefficients are displayed as
”Y...”for channel I and ”X...”for channel II, followed by
”XY”. Except the cursor lines which may be active, all
other readout information including the trigger point
symbol are switched off. In addition to all trigger and
time base related controls, the Y-POS. II (9) knob and
INV (32) button are deactivated. For X position alteration,
the X-POS. (13) knob can be used.
(18) TRIG.
Pushbutton with double function for trigger source
selection and associated LEDs.
The button and the LEDs are deactivated if line (mains)
triggering is selected or XY operation is chosen.
With the aid of this button, the trigger source can be
chosen. There are three trigger sources available:
channel I,channel II (both designated as internal trigger
sources) and the TRIG. EXT. (33) input for external trig-
gering.
The availability of the internal sources depends on the
actual channel mode. The actual setting is indicated by
the associated LED(s).
Briefly pressing the button switches over in the following
sequence:
I - II - EXT - I in DUAL and ADD (addition) mode,
I - EXT - I if mono channel I is present,
II - EXT - II under mono channel II conditions.
Each condition is indicated by the associated LED. The
trigger point symbol is switched off in external trigger
condition.
ALT:
Pressing and holding the button selects alternate
triggering in DUAL mode. Under these conditions both
I and II LEDs are lit. As alternate triggering requires
alternate channel operation, alternate channel switching
is set automatically. A change of the time coefficient then
has no affect regarding the channel switching mode. In
addition to the deflection coefficients display, ”ALT”is
displayed by the readout instead of ”CHP”.
In alternate trigger mode the trigger point symbol is
switched off.
Alternate triggering is not available or automatically
switched off under the following conditions:
ADD (addition) mode,
search (SEA),
delayed (DEL and DTR) time base mode,
TVL,TVF and line (mains) trigger coupling.
(19) VOLTS/DIV.
This control knob for channel II has a double function.
The following description relates to the input attenuator
function (VAR LED dark).
Turning the control knob clockwise increases the
sensitivity (decreases the deflection coefficient) in a 1-2-
5 sequence and decreases the sensitivity (increases the
deflection coefficient) if turned in the opposite direction
(ccw.). The available range is from 1mV/div up to 20V/
div. The knob is automatically switched inactive if the
channel related to it is switched off, or if the input coupling
is set to GD (ground).
The deflection coefficients and additional information
regarding the active channels are displayed in the readout,
e.g. ”Y2: deflection coefficient, input coupling”. The
”:”symbolizes calibrated measuring conditions and is
replaced by the ”>”symbol in uncalibrated conditions.
(20) CH II - VAR. - Pushbutton with several functions.
CH II mode:
Briefly pressing the button sets the instrument to
channel II (Mono CH II) mode. The deflection coefficient
displayed in the readout indicates the current conditions
(”Y2...”). If neither external nor line (mains) triggering
was active, the internal trigger source automatically
switches over to channel II (TRIG.-LED (18) CHII lits).
The last function setting of the VOLTS/DIV (19) knob
remains unchanged.
All channel related controls are active if the input (30) is
not set to GD (32).
VAR.:
Pressing and holding this pushbutton selects the VOLTS/
DIV. (19) control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
After switching the VAR-LED (19) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV. (19)
control knob counter clockwise reduces the signal height
and the deflection coefficient becomes uncalibrated.
The readout then displays ”Y2>...”indicating the uncali-
brated condition instead of ”Y2:...”. Pressing and holding
the CHII pushbutton again switches the LED off, sets
the deflection coefficient into calibrated condition and
activates the attenuator function. The previous vernier

Subject to change without notice
16
Controls and Readout
setting will not be stored.
The CHII pushbutton can also be pressed simultaneously
with the DUAL (17) button. Please note item (17).
(21) TRIG. MODE - Pushbuttons and indicator LEDs.
Pressing the upper or lower button selects the trigger
coupling. The actual setting is indicated by a TRIG.-LED
(18).
Each time the lower TRIG. MODE pushbutton is pressed
the trigger coupling changes in the sequence:
AC DC content suppressed,
DC peak value detection inactive,
HF high-pass filter cuts off frequencies below
approx. 50kHz, trigger point symbol switched off
LF low-pass filter cuts off frequencies above
approx. 1.5kHz,
TVL TV signal, line pulse triggering,
trigger point symbol switched off,
TVF TV signal, frame pulse triggering,
trigger point symbol switched off.
~line/mains triggering, trigger point symbol
and TRIG. LED (18) are switched off.
In some trigger modes such as alternate triggering, some
trigger coupling modes are automatically disabled and
can not be selected.
(22) DEL.POS. - HO
rotary knob with two functions and related HO LED.
The DEL.POS. rotary knob functions as a Hold off time
control, when the time base is not working in the SEA.
(SEARCH) or in DEL. (DELAY) mode.
The HO-LED is not lit when the hold off time is set to
minimum. The HO LED lights up and the hold off time
increases as the knob is rotated clockwise. A signal
sounds on reaching the maximum hold off time. Similarly
in the opposite direction until minimum hold off time is
reached (HO LED extinguishes).
The hold off time is automatically set to minimum when
the time base is changed. (For the application of hold off
time setting see the paragraph with the same heading).
A delay time between the trigger event and the start of
the trace can be set with the DEL.POS. control in the
time base modes SEA. (SEARCH) or DEL.(DELAY).See
SEA./DEL.-ON/OFF (24).
(23)TIME/DIV.
The time base is set with this rotary knob in the TIME/
DIV. field, and the setting is displayed at the top left in
the Readout (e.g. ”T:10µs”).
This knob acts as the time base switch when the VAR
LED above it is not lit. Then, the time deflection coefficient
can be set in a 1-2-5 sequence and the time base is
calibrated. Rotating anticlockwise increases the deflection
coefficient and rotating clockwise decreases the
deflection. The control acts as a vernier (fine adjustment)
when the VAR LED is lit.
The following description refers to the function as a time
base switch.
Time deflection coefficients between 500ms/div. and
50ns/div. in a 1-2-5 sequence can be selected without
x10 X magnification.
Time delay between 100ms and 100ns can be selected
in the ”SEA”(SEARCH) mode.
Time deflection coefficient range in the ”DEL”(DELAY)
mode extends from 20ms/div. to 50ns/div.
(24) SEA./DEL. - ON/OFF pushbutton.
This button is used to switch between delayed and
undelayed time base. The delayed time base operation
enables a magnified display in X direction which is
otherwise only possible with a second time base.
Pressing and holding the button switches to SEA.
(SEARCH) mode, when currently neither ”SEA”
(SEARCH) nor ”DEL”(DELAY) operation is effective.
Afterwards, it can be switched between SEA. and DEL.
by briefly pressing the button.
These operating modes are indicated in the Readout at
the right of the trigger slope indication thus:
In case of SEARCH ,”SEA”will be displayed;
In untriggered DELAY mode, ”DEL”(DEL.) and
in triggered DELAY mode, ”DTR”(DEL.TRIG.).
None of these will appear in the Readout in undelayed
time base operation.
When ”SEA”,”DEL”or ”DTR”mode is effective,
pressing and holding the button switches over to
undelayed time base.
The following description assumes that, with x10 X-
MAG. switched off, the trace starts at left edge of the
graticule.
In SEA. (SEARCH) mode, the hold off time is auto-
matically set to minimum and for the first few divisions
the trace is blanked. The trace will then be unblanked.
The point at which the trace starts can be varied with
DEL.POS (fine adjustment) from about 2 to 7 divisions.
The blanked section serves as a guide to the delay time.
The delay time is based on the current time deflection
coefficient setting and can also be coarsely set with the
TIME/DIV control (range: 20ms to 100ns).
Pressing the button briefly switches over from ”SEA”
to DEL”(DELAY) mode. Now the trace starts at the left

17
Subject to change without notice
Controls and Readout
edge without blanking. The start of the section, which
was previously unblanked in the ”SEA”(SEARCH) mode,
now lies at the left edge. The display can be expanded in
the X direction by rotating the TIME/DIV control
clockwise and thus decreasing the time deflection
coefficient. If a part of the signal of interest goes beyond
the right edge, it can be brought within the screen and
made visible by DEL.POS knob. Increasing the time
deflection coefficient beyond that used in the ”SEA”
(SEARCH) mode is not provided as it is practically
meaningless.
In the untriggered ”DEL”(DELAY) mode, a trigger event
does not start the trace at once but only starts the delay
time. After the delay time has elapsed the trace is started.
In triggered DELAY mode (DTR), to start the sweep, a
signal suitable for triggering must appear after the delay
time. The trace will be started if the instrument settings
(e.g. LEVEL setting) enable a triggering. See DEL.TRIG.
(31).
(25)DEL.TRIG. - VAR. -pushbutton with two functions.
DEL.TRIG. function
In the case of untriggered ”DEL”(DELAY) mode, briefly
pressing the button switches over to ”DTR”(triggered
DELAY mode). Thereby, the previously active settings,
Automatic/normal triggering (10), trigger LEVEL (12),
trigger slope (10) and trigger coupling (21) will be sto-
red.
It will be automatically switched to normal triggering and
DC trigger coupling in ”DTR”mode. Subsequently the
trigger LEVEL setting and the trigger slope should be so
adjusted that the signal for delayed trigger can trigger
the time base. Without triggering the screen will remain
blank.
Briefly pressing the button again switches back to
(untriggered) DEL. Operation.
VAR. function
Pressing and holding the button changes the function
of the TIME/DIV. knob.
The TIME/DIV. knob can function as a time deflection
coefficient switch (1-2-5 sequence) or as a time vernier
(fine adjustment). The current function is indicated by
the VAR-LED. The TIME/DIV. knob functions as a vernier
when the VAR-LED is switched on, but the time base
setting remains calibrated until the (vernier) knob is
operated. The readout now indicates ”T>...”instead of
”T:...”.Rotating further anticlockwise increases the time
deflection coefficient (uncalibrated) until the maximum
is reached indicated by a beep. Rotating the knob
clockwise has the opposite effect. Now, the vernier is
again in the calibrated position and the symbol ”>”will
be replaced by symbol ”:”.
The function of the knob can be switched back to normal
(calibrated) time base, pressing and holding the button.
Underneath the front panel sector described above, the
BNC sockets and four pushbuttons are located.
(26) INPUT CH I (X) - BNC socket.
This BNC socket is the signal input for channel I. The
outer (ground) connection is galvanically connected to
the instrument ground and consequently to the safety
earth contact of the line/mains plug.
In XY mode, signals at this input are used for the X
deflection.
(27) AC / DC - Pushbutton with two functions.
Input coupling:
Briefly pressing this pushbutton switches over from AC
(~ symbol) to DC (=symbol) input coupling and vice versa.
The setting is displayed in the readout with the deflection
coefficient.
Probe factor:
Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel I displayed in the readout,
between 1:1 and 10:1. In condition 10:1 the probe factor
is thus indicated by a probe symbol displayed by the
readout in front the channel information (e.g. ”probe
symbol”, Y1...). In the case of cursor voltage measure-
ment, the probe factor is automatically included.
Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.
(28)GD - Pushbutton.
Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is
displayed in the readout as an earth (ground) symbol
instead of the deflection coefficient and the ~ (AC) or =
(DC) symbol.
The GD setting disables the input signal, the AC/DC (27)
pushbutton and the VOLTS/DIV (15) knob. In automatic
trigger mode the undeflected trace is visible representing
the 0 Volt trace position.
(29)Ground socket - 4mm banana socket galvanically con-
nected to safety earth. This socket can be used as refe-
rence potential connection for DC and low frequency signal
measurement purposes and in COMPONENT TEST mode.

Subject to change without notice
18
Controls and Readout
(30) INPUT CH II - BNC socket.
This BNC socket is the signal input for channel II. The
outer (ground) connection is galvanically connected to
the instrument ground and consequently to the safety
earth contact of the line/mains plug.
In XY mode, signals at this input are used for the Y
deflection.
(31) AC / DC - Pushbutton with two functions.
Input coupling:
Briefly pressing this pushbutton switches over from AC
(~ symbol) to DC (= symbol) input coupling and vice versa.
The setting is displayed in the readout with the deflection
coefficient.
Probe factor:
Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel II displayed in the
readout, between 1:1 and 10:1. In condition 10:1 the
probe factor is thus indicated by a probe symbol displayed
by the readout in front the channel information (e.g.
”probe symbol”, Y2...). In the case of cursor voltage
measurement, the probe factor is automatically included.
Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.
(32) GD - INV. - Pushbutton with two functions.
GD:
Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is
displayed in the readout as an earth (ground) symbol
instead of the deflection coefficient and the ~ (AC) or =
(DC) symbol.
The GD setting disables the input signal, the AC/DC (31)
pushbutton and the VOLTS/DIV (20) knob. In automatic
trigger mode the undeflected trace is visible representing
the 0 Volt trace position.
The ”0 Volt”position is indicated by a ”” symbol in the
readout. Please note Y-POS. II (9).
INV.:
Pressing and holding this pushbutton switches the
channel II invert function on or off. The invert ”on”
condition is indicated by the readout with a horizontal
bar above ”Y2”(Yt mode). The invert function causes
the signal display of channel II to be inverted by 180°.
(33) TRIG. EXT. / INPUT (Z) - BNC socket with two functions.
The outer (ground) connection is galvanically connected
to the instrument ground and consequently to the safety
earth contact of the line/mains plug. The input impedance
is approx. 1MΩII 20pF.
TRIG. EXT:
This BNC socket is the external trigger signal input, if
external triggering is selected. Briefly pressing the TRIG.
(18) pushbutton, until the TRIG. ”EXT”-LED (18) is lit,
switches the external trigger input active.
The trigger coupling depends on the TRIG. MODE (21)
setting.
Z- Input:
If neither COMPONENT TEST nor external trigger
coupling (TRIG. EXT.) is chosen, the socket is operative
as a Z (trace intensity modulation) input.
High TTL level (positive logic) effects blanking, low level
gives unblanking. No higher voltages than +5 Volt are
permitted.
Below the CRT there are the controls for the readout,
the component tester and the squarewave calibrator
with their outputs.
Please note:
The following description of the cursor related controls
assumes that the readout is visible and the component
tester is switched off.
(34) MENU –Pushbutton
Pressing and holding the pushbutton activates the display
of the MAIN MENU. It contains the submenus SETUP
and CALIBRATE.
Once a menu is displayed, the following pushbuttons are
of importance:
1. SAVE and RECALL (7) pushbutton.
Briefly pressing selects the submenu or an item within
the submenu.
2. SAVE (7) pushbutton SET function.
Pressing and holding the SAVE (7) pushbutton calls (sets)
the menu or the previously selected item. In those cases
where the item is marked with ON / OFF the setting
changes from ON to OFF or vice versa.
If “PRESS SECURITY KEY!”is displayed, this function is
available only for HAMEG authorized service workshops.
Press AUTOSET (3) to switch back to the menu.
3. AUTOSET (3) pushbutton.
Each time the AUTOSET pushbutton is pressed the menu
is switched back one step until MAIN MENU is displayed.
Then pressing the AUTOSET pushbutton again switches
the menu operation off and AUTOSET is set to the normal
function.
(35) ON/OFF - CHI/II - 1/∆∆
∆∆
∆t
This pushbutton has several functions.
ON/OFF:
Pressing and holding the pushbutton switches both cursor
lines on or off. As the cursor lines are part of the readout,
they are visible only if the readout is switched on. If the
cursor lines are switched off and MEAN VALUE ON is
activated (Menu: SETUP > MISCELLANEOUS) the
instrument is prepared for DC Mean Value display. Please

19
Subject to change without notice
Controls and Readout
note section "Mean Value Display".
CHI/II:
This function is required and available only in DUAL and
XY mode in combination with ∆∆
∆∆
∆V (37) measurement.
The probe factor setting (27) (31) is automatically taken
into account.
Briefly pressing the button selects between the deflection
coefficients of channel I and channel II. The measured
result is displayed by the readout with ”∆∆
∆∆
∆V1...”or
”∆∆
∆∆
∆V2...”. The settings of the cursors must relate to the
signal of the selected channel.
In XY mode the instrument is automatically set to ∆∆
∆∆
∆V
measurement. In this mode two signals are normally
applied causing an X and a Y deflection. The deflection
coefficient selected for each channel may be different,
thus as in DUAL mode the ∆∆
∆∆
∆V cursor measurement
requires a channel selection. Under channel I (Y signal)
measuring condition the cursor lines are displayed as
horizontal lines and the readout displays ”∆∆
∆∆
∆VY...”. Briefly
pressing the pushbutton changes to channel II (X signal)
voltage measurement. Then the cursor lines are displayed
as vertical lines and the readout indicates ”∆∆
∆∆
∆VX...”.
In CH I or CH II mono mode, only one deflection
coefficient is present and there is no requirement to
select between different deflection coefficients.
Consequently the pushbutton is inactivated in
combination with ∆∆
∆∆
∆V measurements.
1/∆∆
∆∆
∆t:
Briefly pressing the button selects between time and
frequency measurement, if the ∆∆
∆∆
∆V function is not
present.
In calibrated time base condition the readout displays
”∆∆
∆∆
∆t:...”if time measurement is chosen. After briefly
pressing the pushbutton ”f:...”(frequency) is displayed.
If the time base is uncalibrated the readout displays
”∆∆
∆∆
∆t>...”or ”f<...”.
Attention:
Time and frequency measurements are not possible in
XY operation, since in this mode the time base is
switched off.
(36) TRK - CURSOR-Track function.
Briefly pressing simultaneously both buttons ON/OFF -
CHI/II - 1/∆∆
∆∆
∆t (35) and I/II - ∆∆
∆∆
∆V/∆∆
∆∆
∆t (37) switches over
from single cursor line operation to track mode and vice
versa.
In TRK (track) mode both cursor lines are indicated as
active (both dotted lines uninterrupted).
(37) I/II - ∆∆
∆∆
∆V / ∆∆
∆∆
∆t - Pushbutton with two functions.
I/II:
Briefly pressing this button changes the active (control-
lable) cursor in the sequence I - II - I, if TRK (track) mode
is not active. The active cursor is indicated by a conti-
nuously dotted line. An interrupted dotted line indicates
the inactive cursor.
∆∆
∆∆
∆V / ∆∆
∆∆
∆t:
Pressing and holding this pushbutton changes from voltage
to time (or frequency) measurement and vice versa. In
XY mode the instrument is automatically set to ∆V.
∆∆
∆∆
∆V:
Please note!
Inall DV (voltage) measurement conditions, the division
ratio of the probe(s) must be taken into account. The
voltage value displayed in the readout must be multi-
plied by 100 if e.g. a x100 (100:1) probe is used. In case
of x10 (10:1) probes, the probe factor can be automa-
tically included (see item (27) and (31)).
item 1: Time base mode (CH I or CH II single channel
mode, DUAL and ADD).
In the voltage measurement condition, the cursors are
displayed as horizontal lines and the result is displayed in
the readout.
Single channel mode (channel I or channel II).
The ∆∆
∆∆
∆V measuring result is automatically related to the
deflection coefficient of the active channel. The readout
displays ”∆∆
∆∆
∆V1...”or ”∆∆
∆∆
∆V2...”.
Dual mode.
The CURSOR lines must be set on the channel I or
channel II signal. As the deflection coefficients may be
different, it will be required to select between the
deflection coefficients of channel I and II. Please note
item CHI/II (35).
Addition (ADD) mode.
In ADD (addition) mode normally two input signals are
displayed as one signal (sum or difference). As the result
can only be determined if both (calibrated) deflection
coefficients are equal, the CHI/II (35) selection function
is deactivated. In that case the readout indicates ”∆∆
∆∆
∆V...”
without any additional channel information. Different
deflection coefficient settings or uncalibrated deflection
coefficient(s) are indicated in the readout as ”Y1<>Y2”.
item 2: XY mode.
In XY mode the instrument is automatically set to ∆∆
∆∆
∆V
measurement. The deflection coefficient selected for
each channel may be different, thus as in DUAL mode
the ∆∆
∆∆
∆V cursor measurement requires a channel selection.
Under channel I (Y signal) measuring condition the cursor
lines are displayed as horizontal lines and the readout
displays ”∆∆
∆∆
∆VY...”. Briefly pressing the CHI/II (35)
pushbutton selects channel II (X signal) measuring. Then
the cursor lines are displayed as vertical lines and the
readout indicates ”∆∆
∆∆
∆VX...”.
∆∆
∆∆
∆t:
In time or frequency measurement condition two vertical
cursor lines are displayed. The measurement result is
indicated as ”∆∆
∆∆
∆t...”(time measurement) or ”f...”
(frequency measurement). Please note item (36) 1/
∆∆
∆∆
∆
t.
NOTE:
For frequency measurement, the distance between
the cursors must equal exactly one signal period.
In XY mode the time base is switched off. Consequently
time or frequency measurement is disabled.

Subject to change without notice
20
Menu First Time Operation
(38) CURSOR - Center biased lever.
The active cursor line (in track mode: both lines) can be
shifted in the required direction, until the graticule limits
are reached. The directions are marked on the front panel
and depend on the selected measurement (∆∆
∆∆
∆V, ∆∆
∆∆
∆tor 1/
∆∆
∆∆
∆t= f).
Which of two shift speeds is used, depends on how far
the CURSOR knob is pressed. With slight knob pressure
the cursor line(s) moves slowly. If the knob is pressed to
the full extent the cursor moves fast. If the knob is
released it automatically reverts to the center position
and the cursor shift stops.
(39) CAL. - Pushbutton and concentric socket.
A squarewave signal of 0.2Vpp ±1% is available from
the socket for probe adjustment purposes.
The signal frequency depends on the pushbutton setting.
If the pushbutton is released it is approx. 1kHz and can
be switched over (depressed) to approx. 1MHz. The pulse
duty factor may deviate from 1:1 and the frequency is
not calibrated.
(40) CT - Pushbutton and banana jack. Pressing the push-
button switches the instrument over from oscilloscope
to component test mode and vice versa.
In component test mode, all controls are deactivated with
the exception of the CT button, AUTO SET (2) and
INTENS (4). All LEDs except ”A”or ”RO”(associated
with the INTENS knob) are dark. The readout displays
only ”CT”in this condition.
One test lead is connected to the CT socket. The second
test lead uses the ground socket (29). Please note
”Component Tester”.
The maximum test voltage is approx. 20Vpp under open
circuit conditions, while the max. test current under short
circuit condition is approx. 20mApp.
Menu
The instrument software contains several menus. The controls
regarding the menus are described under item MENU (34) in
section “Controls and Readout”.
The following menus, submenus and items within the
submenus are available:
1. MAIN MENU
1.1 CALIBRATE
Information regarding this can be found in the section ”Service
Instructions”item ”Adjustments”.
1.2 SETUP
This menu allows changes to the default settings regarding
the instrument behavior during operation. The SETUP menu
contains the submenus ”MISCELLANEOUS”and
”FACTORY”.
1.2.1 MISCELLANEOUS contains:
1.2.1.1 CONTROL BEEP ON/OFF
In OFF condition the acoustic signals actuated by
the control limits are switched off.
Note:
The default setting is ON. If different conditions are
required the setting must be performed each time
after switching the oscilloscope on.
1.2.1.2 ERROR BEEP ON/OFF
Acoustic signals indicating faulty control operation
are suppressed in OFF condition.
Note:
The default setting is ON. If different conditions are
required the setting must be performed each time
after switching the oscilloscope on.
1.2.1.3 QUICK START ON/OFF
In condition ON the HAMEG logo and the menus
will not be displayed after switching the instrument
on. Then the instrument is quickly ready for operation.
1.2.1.4 TRIG SYMBOL ON/OFF
In most of the Yt (time base) modes the readout
displays a trigger point symbol which will not be
displayed in condition OFF.
1.2.1.5 DC REF ON/OFF
If ON is selected and Yt (time base) mode is present,
the readout displays a ground symbol (⊥). This symbol
eases the evaluation and determination of DC
contents by indicating the 0 Volt reference position.
1.2.1.6 INPUT Z ON/OFF
In ON condition the TRIG.EXT. socket serves as an
unblanking input.
1.2.1.7 MEAN VALUE ON/OFF
The mean value display is activated in the readout if
ON is selected and cursor line measurement is
switched off. For further information please note
section “Mean Value Display”.
1.2.2 FACTORY
Attention!
Submenus are available only for HAMEG authorized
service workshops
First Time Operation
The following text assumes that the ”SAFETY”section of
this manual has been read carefully and understood.
Each time before the instrument is put into operation check
that the oscilloscope is connected to protective earth. For that
reason the power cable must be connected to the oscilloscope
and the power outlet. Then the test lead(s) must be connected
to the oscilloscope input(s). Check that the device under test
is switched off and connect the test lead(s) to the test
point(s). Then switch on the instrument and afterwards the
device under test.
The oscilloscope is switched on by depressing the red POWER
pushbutton. After a few seconds the HAMEG logo and the
instrument software release is displayed on the screen. As
long as the HAMEG logo is visible different internal checks are
Table of contents
Other Hameg Test Equipment manuals

Hameg
Hameg HM1507-3.02 User manual

Hameg
Hameg HM1008-2 User manual

Hameg
Hameg HM1000 User manual

Hameg
Hameg HM 203 User manual

Hameg
Hameg HZ65-3 User manual

Hameg
Hameg HM 1004-3 Series User manual

Hameg
Hameg HM 203-6 User manual

Hameg
Hameg HM 103-2 User manual

Hameg
Hameg HM 103 User manual

Hameg
Hameg HM205-2 User manual

Hameg
Hameg HMO Series User manual

Hameg
Hameg HM 307 User manual

Hameg
Hameg HM1004 User manual

Hameg
Hameg HMO Series User manual

Hameg
Hameg HMO3000 Series Instruction manual

Hameg
Hameg HM1500-2 User manual

Hameg
Hameg HM6050-2 User manual

Hameg
Hameg HM303-6 User manual

Hameg
Hameg HM1507-3 User manual

Hameg
Hameg HM 205 User manual