manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Honeywell
  6. •
  7. Gas Detector
  8. •
  9. Honeywell GasAlertMicro 5 Series User manual

Honeywell GasAlertMicro 5 Series User manual

Intro
The Quick Reference Guide provides basic information to operate the GasAlertMicro 5, GasAlertMicro 5 PID, and
GasAlertMicro 5 IR gas detectors.
For complete operating instructions, refer to the GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR
User Manual provided on the Honeywell https://safety.honeywell.com web site.
The GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR gas detectors (“the detector”) are designed to
warn of hazardous gas levels above user-defined alarm setpoints.
The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
What’s in the box
• Detector complete with specified sensor(s), stainless steel alligator clip and concussion-proof housing
• Rechargeable battery pack or alkaline pack with three AA batteries
• Cradle charger and wall outlet charging adaptor (with rechargeable battery option)
• Sample probe (with motorized pump option)
• Screwdriver
• Calibration/test adaptor and hose
• Quick Reference Guide
Safety Info - Read First
Use the detector only as specified in the manual and quick reference guide, otherwise the protection provided by
the detector may be impaired.
The detector is a personal safety device. It is your responsibility to respond properly to the alarm. Read the
following Cautions before using the detector.
aCAUTION
• Caution: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand the user
manual completely before operating or servicing.
• Charge the detector before first-time use. Honeywell Analytics recommends the detector be charged after every workday.
• Charge the battery pack immediately when a low battery alarm occurs.
• Read and adhere to the battery cautions.
• Honeywell Analytics recommends that the combustible sensor be checked with a known concentration of calibration gas after any
exposure to contaminants/poisons such as sulfur compounds, silicon vapors, halogenated compounds, etc.
• Honeywell Analytics recommends to bump test the sensors before each day’s use to confirm their ability to respond to gas by exposing
the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated.
Calibrate if the readings are not within the specified limits.
• For an additional bump test caution relating to the European Regulations, refer to user manual.
• Calibratethedetectorbeforefirst-timeuseandthenonaregularschedule,dependingonuseandsensorexposuretopoisonsandcontaminants.
The sensors must be calibrated regularly and at least once every 180 days (6 months).
• Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
• Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International.
• The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in the % LEL range,
calibrate the sensor using the appropriate gas.
• Caution: High off-scale LEL readings may indicate an explosive concentration.
• Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons. Although certain organic
vapors(such asleaded gasoline and halogenated hydrocarbons)may temporarily inhibit sensor performance, inmost cases, the sensor
will recover after calibration.
• Foruseonlyin potentially explosive atmospheres whereoxygen concentrations do notexceed20.9%(v/v). Oxygen deficientatmospheres
(<10% v/v) may suppress some sensor outputs.
• Any rapid up scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond the upper scale limit,
which can be hazardous.
• Extended exposure of the GasAlertMicro 5, GasAlertMicro 5 PID, or GasAlertMicro 5 IR to certain concentrations of combustible gases
and air may stress the detector element that can seriously affect its performance. If an alarm occurs due to a high concentration of
combustible gases, recalibration should be performed or, if needed, the sensor replaced.
• TheBWpump module (M5-PUMP)iscertified forusewiththe GasAlertMicro5,GasAlertMicro5PID,andGasAlertMicro5 IRmodelsonly.
• Protect the PID sensor from exposure to silicone vapors.
• When calibrating Oand ClOsensors that are located in the Toxic 2 position of the detector, a single gas calibration cap must be used
to ensure accurate calibration.erratic reading may indicate a gas concentration beyond the upper scale limit, which can be hazardous.
• Extended exposure of the GasAlertMicro 5, GasAlertMicro 5 PID, or GasAlertMicro 5 IR to certain concentrations of combustible gases
and air may stress the detector element that can seriously affect its performance. If an alarm occurs due to a high concentration of
combustible gases, recalibration should be performed or, if needed, the sensor replaced.
• TheBWpump module (M5-PUMP)iscertified forusewiththe GasAlertMicro5,GasAlertMicro5PID,andGasAlertMicro5 IR modelsonly.
• Protect the PID sensor from exposure to silicone vapors.
• When calibrating Oand ClOsensors that are located in the Toxic 2 position of the detector, a single gas calibration cap must be used
to ensure accurate calibration.
• Replace the sensors only in a safe and non-hazardous area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
• Dispose of used lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste stream.
Spent batteries must be disposed of by a qualified recycler or hazardous materials handler.
• Keep lithium cells away from children.
• Beforeusingcommon products around sensors,referto Sensor PoisonsandContaminants onthe GasAlertMicro5/PID/IRUserManual.
• High concentrations of certain toxic gases, for example HS, may have an adverse effect on the LEL sensor. This effect, known as
inhibition, is usually temporary but in extreme circumstances can impair the sensitivity of the LEL sensor.
• After any gasexposurethat causes analarmin the toxicgassensors, theLEL sensor shouldbe verified witha bumptest, andrecalibrated
if necessary.
• Products may contain materials that are regulated for transportation under domestic and international dangerous goods regulations.
Return product in compliance with appropriate dangerous goods regulations. Contact freight carrier for further instructions.
aWARNING
• Substitution of components may impair Intrinsic Safety.
• The battery pack (M5-BAT08) is equipped with a lithium battery that may present a risk of fire or chemical burn hazard if misused. Do
not disassemble, heat above 212ºF (100ºC), or incinerate.
• Do not use any other lithium batteries that are not designed for use with the GasAlertMicro 5, GasAlertMicro5 PID, or GasAlertMicro5 IR
detectors. Use of any other cell can cause fire and/or explosion. To order and replace the M5-BAT07 or the M5IR-BAT08 lithium battery
pack, contact Honeywell Analytics.
• Lithium polymer cells exposed to heat at 266°F (130°C) for 10 minutes can cause fire and/or explosion.
aMISES EN GARDE
• Attention: Pourdes raisonsde sécurité, cetappareildoit être utiliséet entretenu pardupersonnel qualifié uniquement. Lisez attentivement
le manuel d’utilisation avant d’utiliser l’appareil ou d’en assurer l’entretien et assurez-vous d’en avoir bien compris les instructions.
• Chargez le détecteur avant sa première utilisation. Honeywell Analytics recommande de recharger le détecteur après chaque journée
d’utilisation.
• Chargez la batterie dès l’émission d’une alarme de batterie faible.
• Lisez et respectez les avertissements donnés à la section Remplacement des batteries et des piles dans le manuel d’utilisation.
• Honeywell Analytics recommande de contrôler le capteur de gaz combustibles à l’aide d’une concentration connue de gaz d’étalonnage
après toute exposition à des contaminants/poisons (composés de soufre, vapeurs de silicium, produits halogénés, etc.)
• Avantchaque utilisation quotidienne,HoneywellAnalytics recommande d’effectuerun testfonctionneldes capteurs afinde vérifier qu’ils
réagissentbienauxgazprésents, en exposant le détecteuràuneconcentration de gazsupérieureauxseuilsd’alarme. Vérifiezmanuellement
que les alarmes sonore et visuelle sont activées. Étalonnez l’appareil si les relevés ne sont pas conformes aux limites spécifiées.
• Pour une mise en garde supplémentaire sur le test fonctionnel en rapport avec la réglementation européenne.
• Étalonnez le détecteur avant sa première utilisation, puis de manière régulière, en fonction de l’utilisation et de l’exposition du capteur
aux poisons et autres contaminants. Les capteurs doivent être étalonnés régulièrement et au moins une fois tous les 180 jours (6 mois).
• Veillez à effectuer l’étalonnage dans une zone sûre, exempte de gaz dangereux, dans une atmosphère contenant 20,9 % d’oxygène.
• Seul l’élément de détection de gaz combustibles de cet appareil a fait l’objet d’une évaluation des performances homologuée par CSA
International.
• Le capteur de gaz combustibles est étalonné en usine au méthane, à une concentration de 50 % de la LIE. Si la surveillance porte sur
un autre gaz combustible dans la plage % LIE, étalonnez le capteur en utilisant le gaz approprié.
• Attention : Des valeurs LIE hors échelle élevées peuvent indiquer la présence d’une concentration explosive.
• Protégezlecapteur de gazcombustibles contre touteexpositionaux composésde plomb, auxsilicones et auxhydrocarbures chlorés. Bien
que certaines vapeurs organiques (comme l’essence au plomb ou les hydrocarbures halogénés) puissent neutraliser provisoirement les
performances du capteur, dans la plupart des cas, le capteur retrouvera son fonctionnement normal après étalonnage.
• Cetappareil estdestinéuniquement àune utilisation dansdes atmosphèrespotentiellement explosives, danslesquelles la concentration
d’oxygènene dépasse pas20,9 % (v/v).Les atmosphères appauvriesen oxygène(<10% v/v) peuventinhiber certainessortiesducapteur.
• Toute mesure en rapide augmentation suivie d’une diminution ou d’une mesure fantaisiste peut indiquer une concentration de gaz
au-delà de la limite d’échelle\ supérieure, risquant donc d’être dangereuse.
• L’expositionprolongéedu GasAlertMicro5, GasAlertMicro5 PID ou GasAlertMicro5 IR àcertainesconcentrations de gazcombustibles et
d’airpeutfortementsolliciter l’élémentdudétecteur et nuiregravementà ses performances.En cas d’alarmesuiteà une forteconcentration
de gaz combustible, il convient d’effectuer un réétalonnage, voire de remplacer le capteur si nécessaire.
• LemoduledepompeBW(M5-PUMP)est certifiépourles modèles GasAlertMicro5, GasAlertMicro5PID et GasAlertMicro5IR uniquement.
• Protégez le capteur à photo-ionisation (PID) des expositions aux vapeurs de silicium.
• Pour garantir un étalonnage précis des capteurs d’Oet de ClOplacés en position toxique 2 du détecteur, employez un seul capuchon
d’étalonnage monogaz.
• Remplacez les capteurs uniquement en zone sûre et non dangereuse, dans une atmosphère exempte de gaz dangereux et contenant
20,9 % d’oxygène.
• Mettez immédiatement au rebut les batteriesau lithiumusagées. Veillezà nejamais les démonter ou lesjeter au feu. Ne pas les mélanger
aux autres déchets solides. Les piles usagées doivent être éliminées par un centre de recyclage agréé ou un centre de traitement des
matières dangereuses.
• Conservez les batteries au lithium hors de portée des enfants.
• Avant d’utiliser des produits usuels à proximité des capteurs, reportez-vous à la section Poisons et contaminants des capteurs.
• Des concentrations élevées en certains gaz toxiques, tels que HS, peuvent avoir un effet néfaste sur le capteur LIE. Cet effet, appelé
inhibition, est généralement temporaire, mais peut, dans des conditions extrêmes, altérer la sensibilité du capteur LIE. Après toute
exposition à un gaz entraînant le déclenchement d’une alarme des capteurs de gaz toxiques, le capteur LIE devra subir un test
fonctionnel et être réétalonné si nécessaire.
• Lesproduits peuvent contenirdes matériauxqui sontréglementés pour le transporten vertudes règlementsnationaux etinternationaux
de marchandises dangereuses. Retourner le produit conformément à la réglementation sur les marchandises dangereuses appropriées.
Contactez transporteur pour plus d’instructions.
Le détecteur contient des piles alcalines ou une batterie au lithium polymère. Reportez-vous aux avertissements
suivants.
aec AVERTISSEMENT
• Le remplacement d’un composant de l’appareil peut compromettre la sécurité intrinsèque du détecteur.
• Cet appareil contient des piles alcalines. Ne pas les mélanger aux autres déchets solides. Les piles usagées doivent être éliminées par un
centre de recyclage agréé ou un centre de traitement des matières dangereuses.
• Cet appareil contient une batterie au lithium polymère. Mettez immédiatement au rebut les batteries au lithium usagées.
Veillez à ne jamais les démonter ou les jeter au feu. Ne pas les mélanger aux autres déchets solides.
Les piles usagées doivent être éliminées par un centre de recyclage agréé ou par un centre de traitement des matières dangereuses.
• La batterie (M5-BAT08) est constituée d’une pile au lithium qui pourrait présenter un risque d’incendie ou de brûlure chimique en cas
de mauvaise utilisation. Ne pas la démonter ni la chauffer au-delà de 100 °C (212 °F) ou l’incinérer.
• N’utilisez jamais d’autres piles au lithium avec les détecteurs GasAlertMicro 5, GasAlertMicro5 PID ou GasAlertMicro5 IR Toute autre
batterie pourrait provoquer un incendie et/ou une explosion. Pour commander ou remplacer la batterie au lithium M5-BAT07 ou M5IR-
BAT08, contactez Honeywell Analytics.
• Les batteries au lithium polymère exposées à une température supérieure à 130 °C (266 °F) pendant plus de 10 minutes peuvent
provoquer un incendie et/ou une explosion.
Parts of the GasAlertMicro 5 Series
Item Description
1 Liquid crystal display (LCD)
2Button
3◯Button
4 Audible alarms
5 Toxic 2 sensor
6 Toxic 1/PID sensor (GasAlertMicro 5 PID),
or Toxic 1/IR (CO) sensor (GasAlertMicro
5 IR)
7 Visual alarm indicators (LEDs)
8 LEL sensor
9 Oxygen sensor
10 🔼Up button
11 🔽Down button
12 Battery pack
13 Alligator clip
Activating/ Deactivating the detector
To activate the detector, press button.
To deactivate the detector, press and hold button until the countdown is complete.
For more information about other buttons actions, refer to the GasAlertMicro 5, GasAlertMicro 5 PID, and
GasAlertMicro 5 IR User Manual provided on the Honeywell https://safety.honeywell.com web site.
Display Elements
Item Description
1 Alarm condition
2 Automatically span sensor
3 Gas cylinder
4 Gas identifier bars
5 Battery life indicator
6 Passcode lock
7 Data transmission
8 Clock
9 Stealth mode
10 Optional pump indicator
11 Datalog card indicator (optional)
12 Alarm condition (low, high, TWA, STEL, or
multi) or view TWA, STEL and peak (MAX)
gas exposures
13 Automatically zero sensor
Bump Test
A bump test is the process of applying a small amount of test gas to
force the detector into alarm. A bump test should be performed regularly
to confirm the sensors are responding correctly to gas, and that the
audible, visual, and vibrator alarms activate during an alarm condition.
Calibrate if the readings are not within specified limits.
aCAUTION
Honeywell recommends to bump test the sensors before each day’s use
to confirm their ability to respond to gas by exposing the sensors to a gas
concentration that exceeds the alarm setpoints. Should the user wish to
comply with European Regulations, a bump test must be completed before
each day’s use. Refer to EN 60079-29-2.
Follow this procedure when Bump Daily is enabled:
1.Connect the calibration hose to the 0.5l/min regulator on the gas cylinder.
Note: If performing a bump test on a pump unit, connect the calibration hose directly to the
pump module.
2.Connectthe calibration hose tothe intake inlet onthe calibration cap.Arrowson the calibration
cap indicate the direction of gas flow.
3.Attach the calibration cap and tighten the knob.
4.Apply gas. Verify the visual, audible, and vibrator alarms activate.
5.Close the regulator and remove the calibration cap from the detector. The detector will temporarily remain in alarm until the gas clears
from the sensors.
6.Disconnect the calibration hose from the cap and regulator.
aIMPORTANT: Only use the calibration cap for calibrations and bump tests.
Calibration
Procedure Display Procedure Display
aCaution
Calibrate only in a safe area in an
atmosphere of 20.9% oxygen.
1.Activatethe detector.Toenter calibration,
pressandhold○ and🔽simultaneously.
The detector beeps and flashes to the
correspondingcountdown. The LCDthen
displays Starting calibration.
2. flashes while the detector
zeros all of the sensors (except CO) and
calibrates the oxygen sensor. If a sensor
fails to auto zero, that sensor will bypass
the span.
If calibrating a GasAlertMicro 5 or
GasAlertMicro 5 PID, proceed to step #4.
Step #3 for Micro 5 IR Only
3.The Zero-CO? screen displays. Press
○to zero the CO sensor, or press
to bypass.
If ○is pressed to zero the CO sensor,
the following screens display: Apply CO
zero gas now (nitrogen must be used to
zero the CO sensor).
4.The Auto-Zero COscreen displays, and
flashes.
5.The following three screens display:
- Apply span gas now to calibrate
- or press ○to select sensor(s)
- or press to skip calibration
If none of the buttons are pressed, proceed
to step #6.
If ○is pressed, proceed to step #5.
If is pressed, proceed to the end of
step #7.
6.Select which sensor to span. Press 🔼or
🔽to scroll to the required sensor and
then press ○to select. Sensors must be
spanned in the following order:
- Exotics (NH, ClO, O, and CO)
- Single gases
- Quad gases (HS, CO, O and LEL).
7.Attach the calibration cap and apply gas
at a flow rate of 500 ml/min.
Tank flashes while the detector
determines which gas is being applied.
After 30 seconds, flashes
and a countdown displays while the
detector completes the span.
8.When the span is complete, the following
three screens display:
- Calibration successful
- Press 🔼to apply a new cal gas
- Press 🔽to end span
Repeat steps #4-7 to calibrate the
remaining sensors. The LCD displays the
following options:
- Press ○to set the calibration due dates or
press to bypass.
9.Press🔼or 🔽to changethecalibration
duedate. Press ○to acceptthe valueand
proceed to the next due date. If a sensor
fails or does not span, the calibration due
date cannot be changed for that sensor.
The LCD displays the following options:
- Press ○ to change the calibration due
dates or press to bypass.
10. Press 🔼or 🔽to change the alarm
setpoint. Press ○to save the value
and proceed to the next setpoint.
Define the remaining setpoints.
The detector beeps twice when all of
the alarm setpoints have been defined
or bypassed.
10. When calibration is complete, Saving
calibration displays. Note: Only use the calibration cap and single gas calibration cap
during calibration and bump tests.
For additional information about performing calibrations and
bump tests, refer to the GasAlertMicro 5, GasAlertMicro 5 PID, and
GasAlertMicro 5 IR User Manual provided on the Honeywell https://
safety.honeywell.com web site.
GASALERTMICRO 5
SERIES
QUICK
REFERENCE
GUIDE
1, 2, 3, and 4 Gas Detector
Battery operating time
Toxic, O, and LEL sensor configuration: Three alkaline cells or one lithium battery pack at 20°C/68°F provides 20
hours operating runtime
Toxic, O, LEL, and PID sensor configuration: Three alkaline cells or one lithium battery pack at 20°C/68°F provides
15 hours operating runtime
Toxic, O, LEL, and CO sensor configuration: Three alkaline cells or one lithium battery pack at 20°C/68°F provides
15 hours operating runtime
Approved batteries
Approved batteries for GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR product:
• Alkaline (M5-BAT02): as per standards EN 60079-11, EN 60079-0, UL913, CSA C22.2 No. 157
• Lithium-ion polymer (M5-BAT07): as per standards EN 60079-11, EN 60079-0, UL913, CSA C22.2 No. 157
• Lithium-ion polymer (M5-BAT08): as per standards EN 60079-11, EN 60079-0, EN 60079-29-1, EN 50104, UL913, CSA C22.2 No.
157, EN 45544-1 and EN 45544-2.
Rechargeable battery (M5-BAT08):
Lithium polymer -20°C ≤ Ta ≤ +50°C. Temperature code: T4
Alkaline batteries:¹
Duracell MN1500 -20°C ≤ Ta ≤ +50°C. Temperature code: T4 (129.9°C)
Energizer E91VP -20°C ≤ Ta ≤ +50°C. Temperature code: T3C (135.3°C)
Battery charger: GasAlertMicro 5/PID/IR battery charger
First-time charge: Lithium 6 hours
Normal charge: Lithium 6 hours
Warranty: 2 years including sensors (1 year NH3 sensor and PID lamp)
Year of manufacture: The detector's year of manufacture is determined from the serial number. The second and third
number after the first letter determines the year of manufacture.
Example: H311-001000 = 2011 year of manufacture.
¹Standard release data. This product has extended characteristics. For specifications refer to EAC Ex certificate and Pattern Approval document.
Approvals
GasAlertMicro 5 and GasAlertMicro 5 PID (Zone 0): Approved by CSA to both U.S. and Canadian Standards
Standards: CAN/CSA C22.2 No. 157 and C22.2 152
ANSI/UL – 913 and ANSI/ISA – S12.13 Part 1
ABS Type Approved: VA-348169-X
CSA Class I, Division 1, Group A, B, C, and D. Class I, Zone 0, Group IIC.
ATEX CE 0539 II 1 G Ex da ia IIC T4 Ga
KEMA 06 ATEX 0206
EN 60079-0, EN 60079-1, EN 60079-11
IECEx Ex da ia IIC Ga IECEx CSA 06.0011X
IEC 60079-0, IEC 60079-1, IEC 60079-11
INMETRO Ex d ia IIC T4 Ga DNV 12.0138 X
KTL 12-KB4BO-0055X, 18-KB4BO-0672X
GasAlertMicro 5 IR (Zone 1): Approved by CSA to both U.S. and Canadian Standards
Standards: CAN/CSA C22.2 No. 157 and C22.2 152
ANSI/UL – 913 and ANSI/ISA – S12.13 Part 1
ABS Type Approved: VA-348169-X
CSA Class I, Division 1, Group A, B, C, and D. Class I, Zone 0, Group IIC.
ATEX CE 0539 II 2 G Ex d ia IIC Gb
KEMA 06 ATEX 0206
EN 60079-0, EN 60079-1, EN 60079-11,
IECEx Ex d ia IIC Gb IECEx CSA 06.0011X
IEC 60079-0, IEC 60079-1, IEC 60079-11
INMETRO Ex d ia IIC T4 Gb DNV 12.0138 X
KTL 18-KB4BO-0673X
For more information on Approvals and Conformity Declarations refer to https://safety.honeywell.com
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules and ICES-003 Canadian EMI requirements. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Information Required for European Regulations Compliance
Should the user wish to comply, the detector must be operated in the following manner.
Daily bump test: A bump test must be completed before each day’s use.
Condition of use: The detector may only be used with the concussion-proof boot (GA-BM5-1 if it is a diffusion unit, or
GA-BM5-2 if it is a pump unit).
Stealth mode: Stealth mode must be disabled.
Latching alarms: The latching alarm option must be enabled.
Warm up time: 60 seconds; ≤ 90 seconds for pump unit
Stabilization time for methane sensor: ≥ 120 seconds
Stabilization time for oxygen sensor: ≥ 120 seconds
Methane response time t: 10 seconds
Oxygen response time t: 15 seconds for oxygen deficiency
Oxygen response time t: 14 seconds for oxygen surplus
Changing measurement range from % LEL to % Vol.: If changing the measurement unit from % LEL to % Vol. or
from % Vol. to % LEL, a calibration must be completed and the alarm setpoints changed. For calibration information
refer to Calibration chart and for alarm setpoint information refer to the GasAlertMicro 5/PID/IR User Manual.
Effect of other toxic gases on the LEL sensor: High concentrations of certain toxic gases, for example HS, may
have an adverse effect on the LEL sensor. This effect, known as inhibition, is usually temporary but in extreme
circumstances can impair the sensitivity of the LEL sensor. After any gas exposure that causes an alarm in the toxic
gas sensors, the LEL sensor should be verified with a bump test, and recalibrated if necessary.
Approved batteries for GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR product: Lithium-ion
polymer (M5-BAT08): as per standards EN 60079-11, EN 60079-0, UL913, CSA C22.2 No. 157
General Use: For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20.9%
(v/v). Oxygen deficient atmospheres (<10% v/v) may suppress some sensor outputs.
Limited Warranty and Limitation Liability
Honeywell Analytics warrants the product to be free from defects in material and workmanship under normal use
and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to
the sale of new and unused products to the original buyer. Honeywell’s warranty obligation is limited, at Honeywell’s
option, to refund of the purchase price, repair or replacement of a defective product that is returned to a Honeywell
authorized service center within the warranty period. In no event shall Honeywell’s liability hereunder exceed the
purchase price actually paid by the buyer for the Product.
This warranty does not include:
a. fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use;
b.any product which in Honeywell’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of
operation, handling or use;
c. any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved
parts on the product; or the obligations set forth in this warranty are conditional on:
d.proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable
recommendations of Honeywell;
e. the buyer promptly notifying Honeywell of any defect and, if required, promptly making the product available for correction. No goods shall
be returned to Honeywell until receipt by the buyer of shipping instructions from Honeywell; and
f. the right of Honeywell to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish
that the product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF
ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. HONEYWELL SHALL NOT BE LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL, OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of
incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If
any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will
not affect the validity or enforceability of any other provision.
Contact
Europe, Middle East, Africa
Life Safety Distribution GmbH
Life Safety Distribution GmbH
Javastrasse 2
8604 Hegnau
Switzerland
Toll-Free 00800 333 222 44
Middle East +971 4 450 5800
Middle East +971 4 450 5852
(Portable Gas Detection)
gasdetection@honeywell.com
Technical Services
EMEA: HAexpert@honeywell.com
US: ha.us.service@honeywell.com
AP: ha.ap.service@honeywell.com
Americas
Honeywell Analytics Distribution Inc.
405 Barclay Boulevard
Lincolnshire, Illinois.
USA 60069
Tel: +1 847 955 8200
Toll free: +1 800 538 0363
detectgas@honeywell.com
Asia Pacific
Honeywell Analytics Asia Pacific
7F SangAm IT Tower,
434 Worldcup Buk-ro, Mapo-gu,
Seoul 03922, South Korea
Tel: +82 (0) 2 6909 0300
India Tel: +91 124 4752700
analytics.ap@honeywell.com
https://safety.honeywell.com
WEEE Directive and Battery Directive
This symbol indicates that the product must not be disposed of as general industrial or domestic waste.
This product should be disposed of through suitable WEEE disposal facilities.
For more information about disposal of this product, contact your local authority, distributor or the
manufacturer.
Factory Calibration Certificate
We are committed to providing quality product. This instrument has undergone rigorous testing throughout its
manufacture. This is the final report by the people that take pride in the products they build.
This instrument has been factory inspected, tested, and calibrated in accordance with the conditions and
requirements of our registered Quality System, Operating Standards, and Sales Agreements.
Place Factory Calibration
Certification here
User Options Menu
To access the user options, press and hold 🔼and 🔽simultaneously until the detector completes the countdown.
To scroll through the options press 🔼or 🔽. Press ○to select the option.
Exit: Exits the user options menu.
For more information about the User Options Menu, refer to the GasAlertMicro 5, GasAlertMicro 5 PID, and
GasAlertMicro 5 IR User Manual provided on the Honeywell https://safety.honeywell.com web site.
Maintenance
To maintain the detector in good operating condition, perform the following basic maintenance as required.
• Calibrate, bump test, and inspect the detector on a regular schedule.
• Maintain an operations log of all maintenance, bump tests, calibrations, and alarm events.
• Clean the exterior with a soft damp cloth. Do not use solvents, soaps, or polishes. Refer to Sensor Poisons and Contaminants table.
• Do not immerse the detector in liquids.
For more information about Maintenance, refer to the GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR
User Manual provided on the Honeywell https://safety.honeywell.com web site.
Replacing Battery Cells and Packs
To avoid personal injury and/or property damage, adhere to the following battery cautions:
aec WARNING
• This instrument contains alkaline batteries. Do not mix with the solid waste stream. Spent batteries must be disposed of by a qualified
recycler or hazardous materials handler.
• This instrument contains a lithium polymer battery. Dispose of lithium cells immediately. Do not disassemble and do not dispose of in
fire. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler.
• Replace the alkaline cells or rechargeable battery pack immediately when the detector emits a low battery alarm.
• Use only batteries that are recommended by Honeywell.
• Use only approved alkaline batteries that are properly installed in the battery pack. Refer to Specifications.
• To order lithium battery packs (M5-BAT08/M5-BAT07), contact Honeywell.
• Charge the batteries and battery packs using only a recommended Honeywell charger. Failure to adhere to this caution can lead to fire
and/or explosion.
• The detector must be deactivated to charge the battery pack.
• Do not calibrate the detector immediately after charging is complete.
• Boththe lithium batterypack and thealkaline battery packare user-changeable inhazardous locations, butthe alkaline battery cells inside
the pack can only be replaced in a safe area that is free of hazardous gas.
• The M5-BAT08 and M5-BAT07 battery packs are equipped with lithium batteries that can present a risk of fire or chemical burn hazard
if misused. Do not recharge, disassemble, heat above 212°F (100°C), or incinerate.
• Do not use any other lithium batteries with the GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR detectors. Use of any other
cell can cause fire and/or explosion.
• Lithium polymer cells exposed to heat at 266°F (130°C) for 10 minutes can cause fire and/or explosion.
• Keep lithium cells away from children.
For more information about replacing battery cells and packs, refer to the GasAlertMicro 5, GasAlertMicro 5 PID, and
GasAlertMicro 5 IR User Manual provided on the Honeywell https://safety.honeywell.com web site.
Specifications
Instrument dimensions: 14.5 x 7.4 x 3.8 cm (5.7 x 2.9 x 1.5 in.)
Weight: 370 g (13.1 oz.)
Operating temperature:
• VOC: -10°C to +40°C (14°F to +104°F)¹
• Other gases: -20°C to +50°C (-4°F to +122°F)
• Combustible gas sensor: Certified by CSA International to ±3%
• LEL accuracy from -10°C to +40°C (4°F to 104°F)
Storage temperature: -25°C to +60°C
Operating humidity:
• O: 0% to 90%¹ relative humidity (non-condensing)
• VOC: 0% to 95% relative humidity (non-condensing)
• Combustibles: 5% to 95% relative humidity (non-condensing)
• Cl: 10% to 95% relative humidity (non-condensing)
• HCN, ClO: 15% to 90%¹ relative humidity (non-condensing)
• Other gases: 15% to 90%¹ relative humidity (non-condensing)
Pressure: 95 to 110 kPa¹
Dust and moisture ingress: IP65/66¹
Alarm setpoints: May vary by region and are user-defined
Detection range:
• O: 0 - 30.0% vol. (0.1% vol. increments)¹
• CO: 0 - 999 ppm (1 ppm increments)
• CO (TwinTox sensor): 0 - 500 ppm (1 ppm increments)
• HS: 0 - 500 ppm (1 ppm increments)
• HS (TwinTox sensor): 0 - 500 ppm (1 ppm increments)
• Combustible (LEL): 0 - 100% LEL (1% LEL increments) or 0 - 5.0% v/v methane; certified by CSA International to C22.2 No. 152 and
ISA 12.13.01 within 0 - 60% or 3.0% v/v methane
• PH: 0 - 5.0 ppm (0.1 ppm increments)
• SO: 0 - 150 ppm (1 ppm increments)
• Cl: 0 - 50.0 ppm (0.1 ppm increments)
• NH: 0 - 100 ppm (1 ppm increments)
• NO: 0 - 99.9 ppm (0.1 ppm increments)
• HCN: 0 - 30.0 ppm (0.1 ppm increments)
• ClO: 0 - 1.00 ppm (1.00 ppm increments)
• O: 0 - 1.00 ppm (0.01 ppm increments)
• VOC: 0 - 1000 ppm (1.0 ppm increments)
• CO IR: 0 - 50,000 ppm (50 ppm increments) or 0-5.0% v/v CO
Sensors not certified for use with the GasAlertMicro 5 IR: ClO, HCN, NO, PH, Cl
Sensor type:
• HS/CO: Twin plug-in electrochemical cell
• Combustibles: Plug-in catalytic bead
• VOC: Photoionization detector (PID)
• CO: IR detector
• Other gases: Single plug-in electrochemical cell
O measuring principle: Capillary controlled concentration sensor
Alarm conditions: TWA alarm, STEL alarm, low alarm, high alarm, multi alarm, over limit (OL) alarm, sensor alarm,
pump alarm, MMC/SD fail alarm, low battery alarm, confidence beep, automatic deactivation alarm
Audible alarm: 95 dB at 0.3 m (1 ft.) variable pulsed dual beepers
Visual alarm: Dual red light-emitting diodes (LEDs)
Display: Alphanumeric liquid crystal display (LCD)
Backlight: Activates briefly during startup, when there is insufficient light to view the display (if enabled), and during
alarm conditions
Self-test: Initiated during activation
Calibration: Automatic zero and automatic span
Oxygen sensor: Automatic span during startup (if enabled)
User field options: Confidence beep, latching low and high alarms, passcode protection, enable/disable safe display
mode, combustible sensor measurement, sensor enable/disable, language selection, enable/disable automatic
O calibration, set span concentration values, set STEL calculation period, set TWA method, gas measurement
resolution, enable/disable automatic backlight, adjust clock/calendar, set datalogging rate (datalog models only),
CO sensor measurement
Datalog Models Approved for GasAlertMicro 5 and GasAlertMicro 5 PID Models: Delkin 128 MB SD card and 64
MB Unigen SD card
Datalog Models Approved for GasAlertMicro 5 IR Models: Delkin 128 MB MMC, Delkin 128 MB SD card,
Transcend 128 MB SD, and 64 MB Unigen SD card
¹Standard release data. This product has extended characteristics. For specifications refer to EAC Ex certificate and Pattern Approval document.
https://sps.honeywell.com
50105442-036 EN Rev.N
©2021 Oct 25, All rights reserved

Other Honeywell Gas Detector manuals

Honeywell BW Ultra User manual

Honeywell

Honeywell BW Ultra User manual

Honeywell BW Technologies GasAlertMicro 5 User manual

Honeywell

Honeywell BW Technologies GasAlertMicro 5 User manual

Honeywell BW Technologies GasAlertQuattro 1 User manual

Honeywell

Honeywell BW Technologies GasAlertQuattro 1 User manual

Honeywell SPM Flex User manual

Honeywell

Honeywell SPM Flex User manual

Honeywell Gamewell-FCI E3 Series User manual

Honeywell

Honeywell Gamewell-FCI E3 Series User manual

Honeywell SEARCHLINE EXCEL EDGE User manual

Honeywell

Honeywell SEARCHLINE EXCEL EDGE User manual

Honeywell SEARCHZONE SONIK User manual

Honeywell

Honeywell SEARCHZONE SONIK User manual

Honeywell BW MaxXT II User manual

Honeywell

Honeywell BW MaxXT II User manual

Honeywell HON COCON 13 User manual

Honeywell

Honeywell HON COCON 13 User manual

Honeywell BW Technologies GasAlertMicroClip Use and care manual

Honeywell

Honeywell BW Technologies GasAlertMicroClip Use and care manual

Honeywell SEARCHLINE EXCEL PLUS User manual

Honeywell

Honeywell SEARCHLINE EXCEL PLUS User manual

Honeywell Apex Release Note

Honeywell

Honeywell Apex Release Note

Honeywell BW Solo User manual

Honeywell

Honeywell BW Solo User manual

Honeywell MultiRAE User manual

Honeywell

Honeywell MultiRAE User manual

Honeywell Touchpoint Plus User manual

Honeywell

Honeywell Touchpoint Plus User manual

Honeywell BW GasAlertMicro 5 User manual

Honeywell

Honeywell BW GasAlertMicro 5 User manual

Honeywell Searchline Excel Open-Path Release Note

Honeywell

Honeywell Searchline Excel Open-Path Release Note

Honeywell BW RigRat User manual

Honeywell

Honeywell BW RigRat User manual

Honeywell BW Clip Real Time User manual

Honeywell

Honeywell BW Clip Real Time User manual

Honeywell Sensepoint XCL User manual

Honeywell

Honeywell Sensepoint XCL User manual

Honeywell BW MaxXT II User manual

Honeywell

Honeywell BW MaxXT II User manual

Honeywell MBeacon User manual

Honeywell

Honeywell MBeacon User manual

Honeywell Midas-M User manual

Honeywell

Honeywell Midas-M User manual

Honeywell BW Icon User manual

Honeywell

Honeywell BW Icon User manual

Popular Gas Detector manuals by other brands

Charmeg SafeGas DTL26 Installation and use instruction

Charmeg

Charmeg SafeGas DTL26 Installation and use instruction

BEINAT GS913- V.6 manual

BEINAT

BEINAT GS913- V.6 manual

Det-Tronics GTXSM26N4 instructions

Det-Tronics

Det-Tronics GTXSM26N4 instructions

Macurco TX-6-AM User instructions

Macurco

Macurco TX-6-AM User instructions

Alecto GA-10 manual

Alecto

Alecto GA-10 manual

Tecnocontrol SE126K user manual

Tecnocontrol

Tecnocontrol SE126K user manual

Det-Tronics Open Path Eclipse OPECL instructions

Det-Tronics

Det-Tronics Open Path Eclipse OPECL instructions

GasTech G-Shield Fire instruction manual

GasTech

GasTech G-Shield Fire instruction manual

Evikon E2608-CH4 user manual

Evikon

Evikon E2608-CH4 user manual

DOEATOOW CO-3 Operation manual

DOEATOOW

DOEATOOW CO-3 Operation manual

Macurco CD Series User instructions

Macurco

Macurco CD Series User instructions

BEINAT BXI32 Installation and user guide

BEINAT

BEINAT BXI32 Installation and user guide

PNI Safe House Smart Gas 300 WiFi user manual

PNI

PNI Safe House Smart Gas 300 WiFi user manual

EDS EG Series user manual

EDS

EDS EG Series user manual

Metrohm 853 CO2 Suppressor manual

Metrohm

Metrohm 853 CO2 Suppressor manual

Medem AGDS-2em v2 installation instructions

Medem

Medem AGDS-2em v2 installation instructions

GasDog GD300 Series instruction manual

GasDog

GasDog GD300 Series instruction manual

Tecnocontrol TS293E  Series user manual

Tecnocontrol

Tecnocontrol TS293E Series user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.