KPS DMM4500PV User manual

IT
EN
User Manual / MANUAL DEL USUARIO /
Manuel d'utilisation /
Benutzerhandbuch / Manuale Utente
DMM4500PV
All New Designed Bluetooth DMM
Multímetro digital bluetooth con nuevo diseño
Tout nouveau design DMM Bluetooth
Ganz NEU gestaltetes Bluetooth DMM
DMM Bluetooth di Progettazione COMPLETMENTE NUOVA
LIMITED
WARRANTY
YEARS

EN
DMM4500PV
!Safety Information
Understand and follow operating instructions carefully. Use the
meter only as .
WARNING
!
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may
be impaired.
Always use proper terminals, switch position, and range for
measurements.
To reduce the risk of fire or electric shock, do not use this product
around explosive gas or in damp locations.
Verify the Meter operation by measuring a known voltage.
If in doubt, have the Meter serviced.
Do not apply more than the rated voltage, as marked on
Meter, between terminals or between any terminal and earth
ground.
To avoid false readings that can lead to electric shock and injury,
replace the battery as soon as low battery indicator blinks.
Avoid working alone so assistance can be rendered.
Do not use the Tester if the Tester is not operating properly or if it
is wet.
Individual protective device must be used if hazardous live parts
in the installation where the measurement is to be carried out
could be accessible.
Disconnect the test leads from the test points before changing
the position of the function rotary switch.
Never connect a source of voltage when the function rotary
switch is not in voltage position.
When using test leads or probes, keep your fingers behind the
finger guards.
Use caution with voltages above 30 Vac rms, 42 Vac peak, or 60
Vdc. These voltages pose a shock hazard.
Remove test lead from Meter before opening the battery door
or Meter case.
DO NOT USE the test leads when the internal white insulation
layer is exposed.
DO NOT USE the test leads above maximum ratings of CAT.
environment, voltage and current, that are indicated on the
probe and the probe tip guard cap.
DO NOT USE the test leads without the probe tip guard cap in
CAT III and CAT IV environments.
Probe assemblies to be used for MAINS measurements shall
be RATED as appropriate for MEASUREMENT CATEGORY
III or IV according to IEC 61010-031 and shall have a voltage
RATING of at least the voltage of the circuit to be measured.
Only replace the blown fuse with the proper rating as specified
in this manual.
01

EN
DMM4500PV
Symbols as marked on the Meter and Instruction
manual
Do not attempt a current measurement when the open voltage
is above the fuse protection rating. Suspected open voltage
can be checked with voltage function.
Never attempt a voltage measurement with the test lead inserted
into the A input terminal.
Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, or
capacitance.
!
Risk of electric shock
See instruction manual
DC measurement
AC measurement
Wireless transmission Both direct and
alternating current
Equipment protected by double or reinforced insulation
Low battery
Fuse
Earth
Application around and removal from
hazardous live conductors is permitted
Do not discard this product or throw away.
Attention! Magnets might affect the correct functioning
of cardiac pacemakers and implanted defibrillators.
As a user of such medical devices, keep a sufficient
distance to the magnet.
Conforms to EU directives
Unsafe Voltage
To alert you to the presence of a potentially hazardous voltage,
when the Tester detects a voltage ≧30 V or a voltage overload
(OL) in V, mV, PV. The symbol is displayed.
Maintenance
Do not attempt to repair this Meter. It contains no user
serviceable parts. Repair or servicing should only be performed
by qualified personnel.
02

EN
DMM4500PV
Cleaning
Periodically wipe the case with a dry cloth and detergent.
Do not use abrasives or solvents.
Introduction
The Meter Description
Front Panel Illustration
1. Volt Seek Light
2. 6,000 count digital display
3. Push-buttons.
4. Rotary switch for turn the Power On / Off and select the function.
5. Input Terminal for Multi-function.
6. Common (Ground reference) Input Terminal.
7. Input Terminal for A.
A COM
Volt Seek
V
Volt Seek Light
LCD Display
Button
Rotary Switch
V,ȍ, ,°C Input (or “+” )
Common Input
A Input
03

EN
DMM4500PV
Measuring Voltage
Making Basic Measurements
Preparation and Caution Before Measurement
Observe the rules of Warnings and Cautions.
! ! !
!CAUTION
When connecting the test leads to the DUT (Device Under Test)
connect the common test leads before connecting the live test
leads ; when removing the test leads, remove the live test leads
before removing the common test leads.
Dial the switch to select the measuring function.
04

EN
DMM4500PV
Measuring Voltage
Dial the switch and press the Function button to select the
measuring function.
!CAUTION
This function is only available with the dedicated PV test probe.
Always select correct DC / AC mode to perform high voltage
measurement. This meter will flash symbol and the correct
mode symbol (AC / DC) if the input voltage is different and
dangerous.
05

EN
DMM4500PV
Measuring Current
Dial the switch and press the Function button to select the
measuring function.
Dial the switch and press the Function button to select the
measuring function.
Disconnect
Measuring Capacitance / Diode
06

EN
DMM4500PV
Measuring Continuity / Resistance
Dial the switch and press the Function button to select the
measuring function.
Dial the switch and press the Function button to select the
measuring function. (°C / °F)
Measuring Temperature °C / °F
Red Black
07

EN
DMM4500PV
Volt Seek
Dial the switch to select the measuring function.
Press the Function button to change the function on the same
switch position.
Using the Function
Do not hold across this
line when sensing the
hazardous voltage.
Volt
Seek
!Warning
The Volt Seek LED indicates the electric field. If the Volt Seek
LED is not on, voltage could still be present.
Switch
Position
Function
08

EN
DMM4500PV
Range Button
The MAX/MIN mode records the min and max input values.
When the input goes below the recorded min value or above the
recorded max value, the meter records the new value. Press
Hold button to pause the recording.
MIN/MAX
Press Range Button >2 Sec to Enter Auto Range Mode
3
Auto Range Mode
1
Manual Range Mode
2
1
Switch range
Press Button > 2 sec
to exit the MIXMAX mode
09

EN
DMM4500PV
Smart Hold
Press the Relative button to enable/disenable this function.
Press the HOLD button over 2 sec to turn on/off Backlight.
The meter will beep continuously and the display will flash if the
measured signal is larger than the display reading by 50 counts.
(However, it can not detect across the AC and DC Voltage /
Current).
5HODWLYHǻ
Backlight
HOLD
10

EN
DMM4500PV
Auto Power Off
Press the function button and turn the meter on. Then press the
function button to select the time. The time can be 5 minutes,
10 minutes, 20 minutes, and disabled (OFF).
The High Frequency Rejection mode equip a low pass filter in
the AC measurements. The cut-off frequency (-3dB point) of low
pass filter is 800Hz.
Wakeup the meter by dialing the switch or pressing any button.
Time Setting of Auto Power Off
High Frequency Rejection (HFR)
After Specified Time
Without Operation
11

EN
DMM4500PV
The hazardous voltage may be present even if the LCD reading
is very low in HFR mode. Verify the voltage again without HFR
mode.
!Warning
To turn on the meter after keeping HOLD button down.
Testing LCD Monitor
12

EN
DMM4500PV
Function of LOG Button
123
Pressing button while powering-up to select the mode –
Data Logger mode, Manual Saving mode and Clear memory.
The meter can store up to 4000 data in memory.
Press button for more than 2 seconds to activate Data
log-ger mode. The meter will enter Time interval setting mode.
Press button again to select time interval. The interval can
be 1 second, 5 seconds, 10 seconds, 30 seconds, 60 seconds.
Data Logger
Start to record auto
matically after 2 sec
without operaction.
LOG icon flashes while logging
Time interval setting mode
!CAUTION
All stored data will be cleared next startup. Download the stored
data by App first if needed.
13

EN
DMM4500PV
Manual Saving Mode
!CAUTION
All stored data are saved until switching to data logger mode or
executing the clear function.
To save a reading
Show stored reading
amount for 2 seconds
Flash
The meter uses Bluetooth low energy (BLE) V4.0 wireless
technology to transfer the real-time reading and the stored data.
The open-air communication range is up to 10m.
Download “KPS Link” App via the following QR Code.
Turn on Bluetooth function of the meter and open “KPS Link” to
connect the DMM. The Bluetooth icon of the meter
will freeze on LCD after the connection establishes successfully.
14

EN
DMM4500PV
Turn on the meter after keeping the button down.
Default Temperature Units Setting
Default Temperature Units Setting
OR
15

EN
DMM4500PV
Replace the battery as soon as the low battery indicator
appears, to avoid false reading.
Refer to the following figure to replace the batteries
Specifications
Low Battery and Battery Replacement
test leads from Meter before opening the battery cover or Meter
case.
!Warning
General Specifications
Display : 6000 counts.
Overrange Indication : ”OL” or “-OL”
Measure : Samples 3 times per second .
Dimensions (W x H x D) : 74mm x 156mm x 44mm
Weight : 250g (including battery)
Fuse Specification: Fast Action AC/DC 11A, 1000V, IR 30kA
Batteries Life : 300 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage will flash.
Power Requirement : AA 1.5V x 2 batteries
Operating Temperature : -10 ~10°C
10°C ~ 30°C (≦80% RH),
30°C ~ 40°C (≦75% RH),
40°C ~ 50°C (≦45%RH)
Storage Temperature :
-20°C to 60°C , 0 to 80% R.H. (batteries not fitted)
Altitude : 6561.7 ft (2000m)
16

EN
DMM4500PV
CAT Application field
Safety : EN 61010-1, EN 61010-2-033 for CAT III 1000V, CAT IV
600V, EN 61326-1
Drop Protection : 4 feet drop to hardwood on concrete floor
Vibration : Random Vibration per MIL-PRF-28800F Class 2
Pollution degree : 2
Indoor Use
The building installation.
The circuits directly connected to Low-voltage installation.
The source of the Low-voltage installation.
Ⅱ
Ⅲ
Ⅳ
Electrical Specifications
Accuracy is given as ± (% of reading + counts of least significant
digit) at 23°C ± 5°C, with relative humidity Less than 80% R.H.,
and is specified for 1 year after calibration.
(1) Temperature coefficient
0.1 x (Specified accuracy) / °C, < 18°C, > 28°C
(2) AC Function
ACV and ACA specifications are ac coupled, true R.M.S.
The crest factor may be up to 3.0 as 4000 counts.
Accuracy is unspecified of Square Wave.
For non-sinusoidal waveforms, Additional Accuracy by
Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
Max. Crest Factor of Input Signal:
3.0 @ 3000 counts
2.0 @ 4500 counts
1.5 @ 6000 counts
Frequency Response is specified for sine waveform.
LCD displays 0 counts when the reading < 20 counts.
(3) DC mV
Range OL Reading Resolution Accuracy
600.0mV 660.0mV 0.1mV ± (0.5% + 5D)
Input Impedance :0ȍ
Overload Protection : AC/DC 1000V
17

EN
DMM4500PV
(4) DC Voltage
Range OL Reading Resolution Accuracy
6.000V 6.600V 0.001V
60.00V 66.00V 0.01V
600.0V 660.0V 0.1V
1000V 1100V 1V
±(0.5% + 2D)
Input Impedance :0ȍ
Overload Protection : AC/DC 1000V
(6) PV DC Voltage
Range OL Reading Resolution Accuracy
600.0V 660.0V 0.1V
2000V 2200V 1V
±(2.0% + 5D)
Input Impedance :0ȍ
Overload Protection : AC/DC 1000V
(5) AC Voltage
Range OL Reading Resolution Accuracy
600.0mV 660.0mV 0.1mV
60.00V 66.00V 0.01V
6.000V 6.600V 0.001V
600.0V 660.0V 0.1V
1000V 1100V 1V
±(1.0% + 3D)
±(1.0% + 5D)
Input Impedance : 0ȍOHVVWKDQS)
Frequency Response : 45 ~ 500Hz (Sine Wave)
Overload Protection : AC/DC 1000V
18

EN
DMM4500PV
(7) PV AC Voltage
Range OL Reading Resolution Accuracy
600.0V 660.0V 0.1V
1500V 1600V 1V
±(2.0% + 5D)
Frequency Response :5 ~ 500Hz (Sine Wave)
Input Impedance : 0ȍ
Overload Protection : AC/DC 1000V
(8) DC Current
Range OL Reading Resolution Accuracy
6.000A 6.600A 0.001A
10.00A 20.00A 0.01A
±(1.0% + 3D)
Maximum measurement time :
>5A for max.3 minutes with at least 20 minutes rest time.
>10A for max.30 seconds with at least 10 minutes rest time.
Overload Protection : Fuse AC/DC 11A
(9) AC Current
Range OL Reading Resolution Accuracy
6.000A 6.600A 0.001A
10.00A 20.00A 0.01A
±(1.5% + 3D)
Maximum measurement time :
>5A for max.3 minutes with at least 20 minutes rest time.
>10A for max.30 seconds with at least 10 minutes rest time.
Frequency Response : 45 ~ 500Hz (Sine Wave)
Overload Protection : Fuse AC/DC 11A
19