Service Manual Golvo™ 7000/7007 Mobile Lift • 3EN400401 Rev.10
17 www.liko.com lwww.hill-rom.com
6.0 Periodic Inspecon Informaon, Golvo™ 7000/7007 mobile li
According to ISO 10535:2006 Annex B, Periodic Inspecon of medical device Liko™ Mobile Li should be
performed at least once a year. For Liko™ mobile lis Periodic Inspecon Liko Mobile lis (3EN371001) is
available in Agile and on the Partner Web.
Liko™ lis must be thoroughly inspected when the Service Light at the Control Unit illuminates (if applicable)
or at least once per year. Inspecon and service must be carried out by Hill-Rom authorized personnel.
Periodic Inspecon
Liko™ mobile lis
Liko™ sit-to-stand lis
3EN371001 Rev. 1
Valid for following lifts: Viking™, Golvo™, Uno™, Liko™ M220/230, LikoLight™, Sabina™, RollOn™.
Control Unit CBL
If Liko Service System (LSS) or Liko Diagnosc System (LDS) are available, data can be read
from lis equipped with Control Unit CBL. LSS store data and LDS show data, write down
the values in the boxes below.
TA A x T Ah
Control Unit Jumbo Care
From the display, write down the
values in the boxes below.
As
Total work, liing actuator
To track how the li has been used, write down the values from the control units (if applicable):
Expected life me:
Viking™ M mobile li 700 Ah
Viking L mobile li 910 Ah
Viking XL mobile li 1020 Ah
Golvo 8/88 mobile li 700 Ah
Expected life me:
Viking™ M mobile li 2 520 000 As
Li informaon Customer reference
Li: Agreement No:
Product Number: Name:
Serial Number: Address:
Producon year: Post Code
Inspecon sign o
Complete inspecon according to Instrucons for the check points (page 3-9), ll in page 2 and sign below.
Approval to use the mobile li Approved Not approved
If the mobile li has one or more inspecon points “Not approved”, the li must not be used.
To be aconed: Acons according to the “INSPECTION ITEMS NOT APPROVED” should be performed immediately. Aer performed acons
sign below. If anything is unclear or if you have quesons, please contact Hill-Rom or your local Hill-Rom representave.
Contact informaon is to be found at www.hill-rom.com.
Inspecon performed
by:
Date:
Next inspecon:
Inspecon performed in accordance with ISO 10535:2006 Annex B- Periodic inspecon
In this document, this warning symbol indicates that special care should be taken.
If instrucons are not followed there is a risk of serious injury.
Utgå från PI:
Viking M, L XL
kollat igenom punkterna i övriga:
Golvo 8/88
Golvo 7/77
Uno 200
Uno 100-102
Uno 102 EE
Liko M220/230
LL
Viking S/XS
Sabina
RollON
1(9)
Liko technical documentaon can be found on the Partnerweb:
hps://partner.liko.com - “My informaon” - “Documents”
If you don’t ha
ve access, send a mail for r
egistr
aon to: se.cus
tomer
[email protected]
Periodic Inspecon may only be performed by personnel authorized by Hill-Rom and using
original Liko™ spare parts.
Available Liko Periodic Inspecons:
• 3EN371001PeriodicInspeconLikoMobilelis
• 3EN111001PeriodicInspeconLikoOHRailsystem
• 3EN191001PeriodicInspeconOverheadLis
• 3EN301001 PI Freespan Freestand
• 3EN601001PeriodicInspeconLikoaccessories
Also available is Periodic Inspecon Slings. This inspecon should be performed by a person who is
suitable, has good experience and sucient knowledge about design, use and maintenance of the sling.
• 7EN160834GeneralPeriodicInspeconProtocolLikoSlingsandSabinaSupportVests
NOTE!
Periodic Inspecon shall be performed at least once per year.