Before using the rst time make sure that:
• the lift is assembled in accordance with the assembly instructions
• the lifting accessories are properly attached to the lift
• the batteries have been charged for at least 6 hours
• you have read the instruction guides for the lift and lifting accessories
• personnel using the lift are trained of the correct operation and use of the lift.
Before lifting, always make sure that:
• the lifting accessories are not damaged
• the lifting accessory is selected appropriately in terms of type, size, material and design with regard to
the patient’s needs
• the lifting accessory is correctly and safely applied to the patient in order to avoid bodily injury
• the lifting accessory is correctly applied to the sling bar
• the sling bar latches are intact. Missing or damaged latches must always be replaced with new ones;
• the sit-to-stand vest’s/sling’s straps are properly connected to the sling bar hooks when the straps have been
fully extended but before the patient is lifted from the underlying surface.
• as a caregiver assure that the patient not are at risk of falling forward or to any side during lifting.
Safety Instructions
Safety Instructions.............................................................................................. 2
Denitions........................................................................................................... 3
Technical Data.................................................................................................... 3
Measurement ..................................................................................................... 4
Assembly............................................................................................................ 5
Operation............................................................................................................ 7
Charging the Batteries........................................................................................ 8
Maximum Load................................................................................................... 9
Recommended Lifting Accessories .................................................................... 9
Using Sabina to Assist Users into a Standing Position..................................... 10
Using Sabina in Passive Lifting Situations ....................................................... 13
Simple Troubleshooting.................................................................................... 15
Inspection and Maintenance ............................................................................ 16
Sabina II EM and Sabina II EE have been tested by accredited testing institutes, and comply with the requirements
for MDD Class 1 products (MDD 93/42/EEC).
Sabina II EM and Sabina II EE comply with the requirements of EN ISO 10535:2006, IEC 60601-1, EN 60601-1-2,
ANSI/AAMI ES60601-1 and CAN/CSA C22.2 no 60601-1.
Under no circumstances must the lift be modied. Please contact Liko/Hill-Rom for more information.
Particular care must be taken when using strong sources of potential disturbance, such as diathermy, etc, so that
cables are not positioned on or near the lift. If you have questions, please consult the responsible assistive-device
technician or the supplier.
The lift should not be used in areas where ammable mixtures may occur, for example in areas where ammable
goods are stored.
Never leave a patient unattended in Sabina!
NOTE!
This instruction guide contains important information for users of the product. All those who use the product should
review and completely understand the contents of the instruction guide. Remember to keep the instruction guide
in a place where it is always available to those using the product.
Table of Contents
After use, store the lift in forward direction towards a wall and out of reach of unauthorized!