Service Manual Viking™ XL mobile lift • 3EN350401 Rev. 11 7 www.liko.com lwww.hill-rom.com
1.4 Technical Data Viking™ XL mobile li, 2040043
Maximum load: 300 kg (660 lbs)
Material: Aluminium
Weight: Total: 39,9 kg (88.0 Ibs)
Heaviest part: 23,6 kg (52.0 Ibs)
Wheels: Front: 100 mm (4 in) twin wheels.
Rear: 125 mm (5 in) twin wheels.
Turning diameter: 1570 mm (61.8 in)
Emergency
lowering device:
Mechanical and electrical
Liing interval: 1370 mm (54.0 in)
Liing speed
(no load)
23 mm/s and 17 mm/s
(0.9 in/s and 0.7 in/s)
Sound level: 51 dB(A)
Protecon class: IP X4
Operang Forces
of Controls:
Handcontrol: 5 N
Electrical data: 24 V
Intermient
power:
Int. Op 10/90, acve operaon max 2 min. Only
10% of a given length of me may be acve, but
no more than 2 min.
Baeries: Lead-acid gel, valve-regulated baery
24 V 2.9 Ah Prod. No. 2006106.
Weight: 2,8 kg (6.2 Ibs.)
Li-ION baery
25.6 V, 3.3 Ah Prod. No. 2006110.
Weight: 1,4 kg (3.1 Ibs.)
Baery charger: Internal charger, 100-240 V AC, 50-60 Hz, max.
400 mA.
Li motor: Permanent magnet motor with mechanical
safety mechanism.
Viking XL: 24 V, 9.0 A
Motor for base-
width adjustment:
Permanent magnet motor
24 V, 5,5 A
Surrounding
funconal
environment:
Temperature:
+10°C to +40°C, (50° F to 104° F)
Humidity: 20% to 80% at 30°C non-condensing,
Air pressure: 800HPa to 1060HPa, Altude: max.
2000m.
The device is intended for use indoors
Type B, in accordance with the electrical shock
protecon class.
Class II equipment.
5
Viking™ L, XL mobile lis • 7EN137108 Rev. 1 www.hill-rom.com I www.liko.com
Dimensions
Note: The measurements are based on the li being equipped with standard sling bar. When changing to other liing accessories, check that the li sll
achieves desired liing height.
** Reference measurement according to Standard EN ISO 10535:2006.
Assembly
Before assembly, make sure you have the following parts: • Locking handles, pair (2 pair Viking XL)
• Baery
• Instrucon guide, charger cable,
charger connector cable.
• Li mast with li arm, li motor incl. cable, sling bar and
control box with handcontrol
• Base with motor for base-width adjustment, incl. cable
Top view
Lateral view
B
B1
FF1
Amax.
B2
Amin.
**600 mm (24 in)
E
M
N
O
Max. reach posion of li arm
B3
Lmin
Lmax
C
D2
D
Max. reach posion of li arm
D1
**700 mm (28 in)
1. Lock both rear wheels.
Place the li mast in
the foot of the base.
2. Use the pair of locking handles to secure the
li mast in the base.
Note! (XL) 2 pair, start in the lower hole and
use "Pull-out-rotate" to secure li mast, see
illustraon.
Viking L Measurements: mm / in
Viking XL Measurements: mm / in
Amax Amin BB1 B2 B3** CDD2** E F F1 Lmax Lmin MNO
max min max min
2030 1390 1360 900 630 560 1070 695 950 625 230 1060 140 43 1790 460 500 280 550
80.0 54.7 53.5 35.4 24.8 22.0 42.1 27.4 37.4 24.6 9.1 41.7 5.5 1.7 70.5 18.1 19.7 11.0 21.7
Amax. Amin. BB1 B2 B3** CDD2** E F F1 Lmax Lmin MNO
max. min. max. min.
2110 1500 1420 960 600 560 1130 815 1015 605 145 1010 140 40 1910 540 557 365 665
83.1 59.1 55.9 37.8 23.6 22.0 44.5 32.1 40.0 23.8 5.7 39.8 5.5 1.6 75.2 21.3 21.9 14.4 26.2
Pull-out-rotate XL L
3. Aer securing the mast use
"Pull-out-rotate" to set the
locking handles in a downwards
posion, see illustraon.
5
Viking™ L, XL mobile lis • 7EN137108 Rev. 1 www.hill-rom.com I www.liko.com
Dimensions
Note: The measurements are based on the li being equipped with standard sling bar. When changing to other liing accessories, check that the li sll
achieves desired liing height.
** Reference measurement according to Standard EN ISO 10535:2006.
Assembly
Before assembly, make sure you have the following parts: • Locking handles, pair (2 pair Viking XL)
• Baery
• Instrucon guide, charger cable,
charger connector cable.
• Li mast with li arm, li motor incl. cable, sling bar and
control box with handcontrol
• Base with motor for base-width adjustment, incl. cable
Top view
Lateral view
B
B1
FF1
Amax.
B2
Amin.
**600 mm (24 in)
E
M
N
O
Max. reach posion of li arm
B3
Lmin
Lmax
C
D2
D
Max. reach posion of li arm
D1
**700 mm (28 in)
1. Lock both rear wheels.
Place the li mast in
the foot of the base.
2. Use the pair of locking handles to secure the
li mast in the base.
Note! (XL) 2 pair, start in the lower hole and
use "Pull-out-rotate" to secure li mast, see
illustraon.
Viking L Measurements: mm / in
Viking XL Measurements: mm / in
Amax Amin BB1 B2 B3** CDD2** E F F1 Lmax Lmin MNO
max min max min
2030 1390 1360 900 630 560 1070 695 950 625 230 1060 140 43 1790 460 500 280 550
80.0 54.7 53.5 35.4 24.8 22.0 42.1 27.4 37.4 24.6 9.1 41.7 5.5 1.7 70.5 18.1 19.7 11.0 21.7
Amax. Amin. BB1 B2 B3** CDD2** E F F1 Lmax Lmin MNO
max. min. max. min.
2110 1500 1420 960 600 560 1130 815 1015 605 145 1010 140 40 1910 540 557 365 665
83.1 59.1 55.9 37.8 23.6 22.0 44.5 32.1 40.0 23.8 5.7 39.8 5.5 1.6 75.2 21.3 21.9 14.4 26.2
Pull-out-rotate XL L
3. Aer securing the mast use
"Pull-out-rotate" to set the
locking handles in a downwards
posion, see illustraon.