EMC Table
Guidance and manufacturer’s declaraon – electromagnec emissions
The mobile li is intended for use in the electromagnec environment specied below. The customer or the user of the
mobile li should assure that it is used in such an environment.
”Essenal performance according to the manufacturer: The hoist shall not move unintenonally while being submied to
disturbances.”
Emissions test Compliance Electromagnec environment - guidance
RF emissions
CISPR 11
Group 1 The mobile li uses RF energy only for its internal funcon.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The mobile li is suitable for use in all establishments inclu-
ding domesc establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage uctuaons/ icker emis-
sions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaraon – electromagnec immunity
The mobile li is intended for use in the electromagnec environment specied below. The customer or the user of the
mobile li should assure that it is used in such an environment.
”Essenal performance according to the manufacturer: The hoist shall not move unintenonally while being submied to
disturbances.”
Immunity test IEC 60601 test level Compliance level Electromagnec
environment - guidance
Electrostac
discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 15 kV air
+/- 8 kV contact
+/- 15 kV air
+/- 8 kV contact
+/- 15 kV air
Floors should be wood, concrete or
ceramic le. If oors are covered
with synthec material, the relave
humidity should be at least 30 %.
Electrical fast
transient / Burst
IEC 61000-4-4
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 2 kV for power supply
lines
n/a. for input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+/- 1 kV dierenal mode
+/- 2 kV common mode
+/- 1 kV dierenal mode
n/a. for common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short inter-
rupons and voltage va-
riaons on power supply
input lines
IEC 61000-4-11
0 % UT for 0,5 cycle, at 0, 45,
90, 135, 180, 225, 270 and
315 degrees
0 % UT for 1 cycle, at 0
degrees
70 % UT for 25 cycles at 50
Hz and 30 cycles at 60 Hz, at
0 degrees
0 % UT for 250 cycles at 50
Hz and 300 cycles at 60 Hz.
0 % UT for 0,5 cycle, at 0,
45, 90, 135, 180, 225, 270
and 315 degrees
0 % UT for 1 cycle, at 0
degrees
70 % UT for 25 cycles at 50
Hz and 30 cycles at 60 Hz,
at 0 degrees
0 % UT for 250 cycles at 50
Hz and 300 cycles at 60 Hz.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
[Equipment or System] requires
connued operaon during power
mains interrupons, it is recom-
mended that the [Equipment or Sys-
tem] be powered from an uninter-
rupble power supply or baery.
Power frequency (50/60
Hz) magnec eld
IEC 61000-4-8
30 A/m Complies Power frequency magnec elds
should be at levels characterisc of
a typical locaon in a typical com-
mercial or hospital environment
NOTE UT is the a.c. mains voltage prior to applicaon of the test level.