Date of Issue: May. 22, 2017
Document No.: NIC-600-2240-02
2/9
1. Overview
The LP-WA vaporizes LPG/natural gas sampled into a high pressure cylinder. It is a
pretreatment unit for measurements by the WA-5A/F.
2. Configuration/specifications
1) Objects to be measured: LP gas, Natural gas ; Filling pressure 10MPa or lower
2) Dimension/Weight: 230W x 370D x 490H, 11kg
3) Power supply
Power voltage: Single phase 200 to 240V, 50/60Hz
Voltage fluctuation: Within ±5%
Capacity: 110VA (3 pole plug)
Fuse: 10A
Power quality: No surges, harmonics, or noise.
Distance from power outlet to the system: Within 2m. Grounding required.
4) Cylinder size: 1L Max.
5) Line filter: PEEK 2μm
6) Regulator: Heating temperature 90°C Max.
Parts exposed to the gas are coated with silica.
Valve seats are made of PCTFE.
3. Precautions for use
1) Since historical data of the previous measurement remains about 5% due to the
configuration of the system, a low-concentration gas measurement is impossible immediately
after a high-concentration gas measurement. In such a case, be sure to perform the blank
measurement and check the blank values before proceeding to the next measurement.
2) This system is not explosion proof. Be careful of any gas leak, particularly from the
connection between the cylinder and the heated regulator.
LPG
Cylindar
Heated regurator
WA-5F/A