Operating and adjusting the Patient Hoist must
be done only by persons who are knowledgeable
and trained on how to properly handle the Pa-
tient Lifter. This requires the ability to use the
unit with patients.
Patient Hoist used in wet rooms may be exposed
to humidity moisture, chlorine and heat, which
can lead to oxidation that appears as a white
coating.
When functional faults or damage of any type
occurs, then the Patient Hoist may not be used.
Danger of breakage!
The maximum allowable load of 125 kg may
not be exceeded because no safe operation
can be assured.
Whenever using a sling, be sure that it is secure-
ly hung to the carrier of the Patient Hoist. Do not
use any sling that shows signs of damage.
Do not place any objects between the main col-
umn and the drive axis. No perpendicular loads
may be applied to the drive axis.
The patient must not hold onto the drive axis.
No weight loads may be hung onto the holder for
an extended time period.
Danger of injury!
The Patient Hoist may not be operated while it is
being charged up.
The Patient Hoist may be used only inside on
even, slip-free surfaces. Using the Patient Hoist
on slopes, grades or inclinations must be avoid-
ed due to the increased risk.
Use caution when travelling in tight spaces.
Paths used by the Patient Hoist must be cleared
before they are travelled on.
Sudden movements and pushing, as well as
patient-swinging must be avoided because of an
increased risk of an accident.
Never leave the patient unsupervised when us-
ing the Patient Hoist! An unforeseen movement
or action by the patient may lead to a possible
risk of injury.
In the event of an emergency lowering of the
patient, use extreme caution to avoid putting the
patient in danger. In difficult situations, seek the
assistance of another person.
Check for a secure connection between the car-
rier and the lifter arm.
Danger! Function hazard!
It must be assured that no humidity reaches
inside the electrical systems. (Observe protec-
tion class!)
Do not open up the control / drive / battery unit or
hand switch systems!
If the drive doesn’t work (drive mustn’t start up
while it’s blocked) it’s not allowed to activate the
patient hoist.
Do not charge up the rechargeable battery in wet
rooms (e.g.: bathrooms, shower stalls).
Notice electromagnetic troubles caused by mo-
bile phones or medical equipment for example.
7. Product specifications
7.1 Delivery scope
The product has been carefully inspected in the
factory for freedom of defects and for complete-
ness. After having received it, carefully check over
the product for possible transport damages and for
completeness.
1 Patient Hoist “Arnold 125”
1 control unit
1 rechargeable battery
1 recharging cable
1 manual switch with hanging grip
1 User Manual
7.2. Technical specifications
We reserve the right to make technical develop-
ments and improvements with respect to the de-
sign shown in the User Manual.