Sunx FX-100 SERIES User manual

FX-100 SERIES
INDEX
..............................P. 1
.......................P. 1
..............................P. 1
...........P. 2
...P. 2
.......................................................P. 3
...P. 3
.................................P. 3
.......P. 3
................................................P. 4
.........................................P. 4
...P. 4
............................................P. 4
.................................................P. 5
....P. 5
............................................P. 7
...................P.10
.....................P.10
...P.11
..............P.12
.................................................P.13
....P.13
........................................P.18
.........................P.18
...P.19
.......................P.20
.......................................P.21
........P.22
.........P.22
......................................P.23
...............................................P.24
...............................P.25
...............................P.26
DIGITAL FIBER SENSOR
1Functional Description
1-1. Functional Description
1-2. Operating Portion
2 Diagram of Functions and Settings
2-1. Diagram of Functions and Settings
3 Others
3-1.
Precautions When Selecting Settings
3-2. Factory Settings
3-3. Error Display Indicator Readings
4RUN Mode
4-1. RUN Mode
4-2.
Threshold Value Fine Adjustment Function
4-3. Key Lock
5SET Mode
5-1. SET Mode Functions and Settings
5-2. Teaching
5-3. Output Operation Setting
5-4. Timer Operation Setting
5-5.
Light-emitting Amount Selection Setting
5-6. Emission Frequency Setting
6PRO Mode
6-1. PRO Mode Functions and Settings
6-2. Shift setting
6-3. External Input Setting
6-4.
Threshold Value Follow-up Cycle Setting
6-5. GETA Function Setting
6-6. ECO Setting
6-7. Digital Display Inversion Setting
6-8. Threshold Value Margin Setting
6-9. Setting Copy
6-10. Reset
7Quick Setting Function
8Code Setting Function
Operation Guide

1
Functional Description
1-1. Functional Description
1-2. Operating Portion
1
Three keys [ (MODE key), (ON key / setting value UP key) and (OFF key / setting value DOWN key)]
are used to carry out all settings.
In addition, the keys can be used in combination to carry out the following.
and and
MODE key
· Selection of setting items
· Confirmation of set contents
· RUN Mode
Press for 2 sec.: To SET mode
Press for 4 sec.: To PRO mode
<When using the GETA function (Refer to p.20)>
Press once (1 sec. or less):
The incident light intensity before
the GETA function is set is indicated
for approx. 2 sec.
· Other than RUN mode
Press for 2 sec. or more: To RUN mode
or Cancel
· Key lock setting / release
(Refer to p.4)
Press simultaneously for 2 sec.
· Quick setting function (Refer to p.25)
Press simultaneously for 2 sec.
· Code setting function (Refer to p.26)
Press simultaneously for 4 sec.
ON key / Setting value UP key
OFF key / Setting value DOWN key
· Selection of setting contents
· Threshold value settings in teaching mode
· Fine adjustments to the threshold value
(When pressed and held, the value will change more rapidly.)
Digital display (Red)
(Incident light intensity)
Digital display (Green)
(Threshold value)
Operation indicator (Orange)
(Lights up when the output is ON)
OFF Key /
Setting value DOWN key
MODE key
ON Key /
Setting value UP key

2
Diagram of Functions and Settings
2-1. Diagram of Functions and Settings
2
The amplifier features and settings can be generally classified into three main modes: the ‘RUN mode’ for normal
sensing operation, the ‘SET mode’ for basic settings, and the ‘PRO mode’ that contains more detailed settings.
This mode is used for carrying out basic settings.
SET Mode P. 5
Press for 4 sec.
during RUN mode.
Press for 2 sec.
during RUN mode.
Keep pressing until
‘ ’ is displayed. Keep pressing until
‘ ’ is displayed.
Press simultaneously
for 2 sec. during
RUN mode.
Press simultaneously
for 4 sec. during
RUN mode.
RUN Mode QUIC CODESET ModePRO Mode
Threshold value can be set in 2-level
teaching, limit teaching [,], or
full-auto teaching.
Teaching
Procedure: P.7lDetails: P.5
Light-ON or Dark-ON can be set.
Output Operation Setting
Procedure: P.10Details: P.5
Configures timer operation and timer
period.
Timer Operation Setting
Procedure: P.10Details: P.6
When fiber heads are used side-by-side,
the emission frequencies for each amplifier
can be set to different emission frequencies
in order to prevent interference.
Emission Frequency Setting
Procedure: P.12Details: P.6
Allows light reduction to be set for the
light-emitting amount. This is useful if
sensing is difficult because the
incident light intensity is saturated.
Light-emitting Amount Selection Setting
Procedure: P.11Details: P.6
This mode is used for carrying out detailed settings.
PRO Mode P. 1 3
Allows the shift amount for limit
teaching [,] and for threshold
value follow-up cycle setting to be set
within the range of 0 to 80 %.
Shift Setting
Procedure: P.18Details: P.13
External input can be selected from
emission halt, limit teaching [], limit
teaching [], full-auto teaching, and
ECO.
External Input Setting
Procedure: P.18Details: P.13l
Allows follow-up of the threshold value to
be carried out within a set range (shift
amount) in response to fluctuations in the
incident light intensity.
Threshold Value Follow-up Cycle Setting
Procedure: P.19Details: P.15
Allows the viewing orientation of the digital
display to be inverted in accordance with
the setting direction of the amplifier.
Digital Display Inversion Setting
Procedure: P.22Details: P.17
Sets the digital display to turn off in
order to reduce power consumption
(current consumption).
ECO Setting
Procedure: P.21Details: P.17
Allows the display value for the
incident light intensity to be offset by
the desired value (target value).
GETA
Function
Setting
Procedure: P.20Details: P.16
Margin for threshold value to the
present incident light intensity can be
checked.
Threshold Value Margin Setting
Procedure: P.22Details: P.17
The settings of the master side
amplifier can be copied to the slave
side amplifier.
Setting Copy
Procedure: P.23Details: P.17
Allows all settings to be returned to
the factory settings.
Reset
Procedure: P.24Details: P.17
This indicates normal sensing operation.
RUN Mode
Fine adjustment of
the threshold value
can be done when in
RUN mode.
Threshold value fine
adjustment function
P. 4
Allows various settings (output
operation settings, timer operation
settings, light-emitting amount selection
settings and emission frequency
settings) to be made in SET mode
simply by selecting a setting number.
Quick Setting Function
Procedure: P.25Details: P.25
Settings for output operation, timer
operation, light-emitting amount
selection, emission frequency,
ECO, external input, and shift
amount are possible by selecting
codes discretionary.
Code Setting Function
Procedure: P.26Details: P.26
The key lock function prevents
key operations so that the
conditions set in each setting
mode are not inadvertently
changed.
Key Lock Function
Procedure: P.4Details: P.4
Procedure: P.4
Details: P.4

3
3-2. Factory Settings
Factory settings for the FX-100 series are indicated below:
If the unit is reset using the ‘Reset’ from ‘PRO Mode’ on p.24, the resulting settings will be those indicated below:
Settings
Item
SET Mode
PRO Mode
Digital display
*When timer is not set,
this mode is not displayed.
Settings
Item
40
Dark-ON
Without timer
ON-delay timer : 10 ms
OFF-delay timer : 10 ms
OFF
0 (Response time: 250 !s or less)
1 (Response time: 2.5 ms or less)
Threshold value
Output operation setting
Timer operation setting
Timer period setting
Light-emitting amount selection setting
Shift amount 15 %
Emission halt
OFF
OFF
OFF
OFF
OFF
NO
NO
Shift setting
External input setting
Threshold value follow-up cycle setting
GETA function setting
ECO setting
Digital display inversion setting
Threshold value margin setting
Setting copy
Reset
Digital display
Factory Setting
Factory Setting
3-3. Error Display Indicator Readings
In case of errors, attempt the following measures:
Digital display Error description Measures
The load has short-circuited
and excess current is flowing.
Communication error.
Disconnection, connection
failure, etc.
EEPROM writing error Contact our office.
Turn off the power, then check the load.
Turn off the power, and then check the
wiring for using the setting copy function.
FX-101:
FX-102:
Emission frequency
setting
Others
3-1. Precautions When Selecting Settings
3
You can use the ‘key lock function’ to protect settings. (Refer to p.4)
Key lock function can be used to prevent the operator from accidentally changing the sensor settings.
Setting protection
If (MODE key) is pressed for 2 sec. while setting is in progress in a mode other than RUN mode, the mode
will switch to RUN mode. When this occurs, the settings which had been changed before (MODE key) was
pressed for 2 sec. will be confirmed.

4
RUN Mode
4-1. RUN Mode
4-2. Threshold Value Fine Adjustment Function
4
Digital display
Key lock set
Key lock released
4-3. Key Lock Function
When in RUN mode
When in RUN mode
;When turn ON the power, the Model No. is displayed in the
green digital display, and the emission frequency is displayed in
the red digital display. Then switch into RUN mode [digital
display (green: threshold value, red: incident light intensity.)]
;‘ ’ is displayed in the red digital display when emission halt
is selected in the external input setting and externally received
the signal.
;When ECO setting mode is ON, the digital display turns off in 20 sec. approx. in
RUN mode. To light up the digital display again, press any key for 2 sec. or more.
;Fine adjustment of the threshold value can be done when in RUN mode.
;The threshold value changes when (setting value UP key) or (setting value DOWN key) is pressed.
(when the key is pressed long, the value is accelerated.)
;The threshold value is stored after 3 sec.
;The key lock function prevents key operations so that the conditions set in each setting mode are not inadvertently changed.
;In the key lock condition, ‘ ’ is displayed when any key is pressed.
When in RUN mode
Automatic
Automatic Automatic
Press and simultaneously for 2 sec.
Press and simultaneously for 2 sec. or more.
(Keep pressing until ‘ ’ is displayed.)
After approx. 20 sec.
Blinks Turns off
Automatic Automatic
Caution
When the power is ON RUN mode
Displays blink alternately
When external input
Automatic
Emission
frequency
Threshold
value
Incident light
intensity
Threshold
value
Incident light
intensity
Model
No.
For the setting of external input, refer to ‘ PRO MODE 6-3. External input setting (P.18).’
For the setting of ECO, refer to ‘ PRO MODE 6-6. ECO setting (P.21).’
If ‘ ’ and ‘ ’ (incident light intensity) are blinking alternately
in the digital displays, check if there is a contact between the output
wire (black) and the external input wire (white).
6
6

5
SET Mode
5-1. SET Mode Functions and Settings
5
SET mode is the mode used for making basic settings. Settings for five functions can be configured.
: Teaching The ‘threshold value’ can be set by utilizing three kinds of teaching, whichever
‘2-level teaching’, ‘limit teaching [,]’ or ‘full-auto teaching’.
Make sure that detection may become unstable if less margin is applied in the
use environment when teaching.
Refer to p.7lfor setting procedure
: Output Operation Setting
This mode allows the selection of output operation from either L-ON (Light-ON),
or D-ON (Dark-ON).
When set to ‘L-ON’, the output will be ON if the incident light intensity becomes
greater than the ‘threshold value’.
When set to ‘D-ON’, the output will be ON if the incident light intensity becomes
less than the ‘threshold value’.
* The factory setting is ‘Dark-ON (D-ON)’.
Refer to p.10 for setting procedure
2-level teaching is a method of setting the threshold value by teaching the amplifier unit
two different status conditions - sensing object present and sensing object absent.
The ‘threshold value’ is normally set using this method.
P. 7
P. 8 l
2-level Teaching
Limit Teaching [, ]
Teaches only the status condition in which no sensing object is within sensing range
(status in which incident light intensity is stable). This method is used to set a ‘threshold
value’ for conducting sensing in the presence of a background, or when extremely
small objects are to be detected.
P. 9
Full-auto Teaching
This method is used to set the threshold value while the sensing objects are still
moving on the production line, without stopping the production line.
Sensing object
present
Sensing object
present
Sensing object
present
Sensing object
absent
Sensing object
absent
Sensing object
absent
Minute
object Background
Minute
object
Thru-beam type Reflective type Retroreflective type
Thru-beam type Reflective type Retroreflective type
Thru-beam type Reflective type Retroreflective type
Incident light intensity
Threshold value
04,000
ONOFF
OFFON
Output operation
L-ON
D-ON

6
If the incident light intensity is still saturated (4,000 or higher) even when the
light reduction setting is used, make the distance between the sensing
objects and the fiber head longer.
• For the FX-101:, the interference prevention function will not operate for
emission frequency 0 (factory setting).
• The response time will change when the emission frequency is changed.
: Timer Operation Setting
This sets timer operation and timer period.
Without timer, ON-delay timer, or OFF-delay timer can be set.
In case of setting ON-delay timer or OFF-delay timer in the timer operation
setting mode, timer period can be set.
Select timer period from 1 ms, 5 ms, 10 ms, 20 ms, 40 ms, 50 ms, 100 ms, 500
ms, and 1,000 ms.
* The factory setting is ‘Without timer’.
Refer to p.10 for setting procedure
: Light-emitting Amount Selection Setting
Allows light reduction to be set for the light-emitting amount.
When the setting is ON, the light-emitting amount is reduced.
Use this setting if sensing is difficult because the incident light intensity is
saturated (4,000 or higher).
* The factory setting is ‘OFF’.
* The factory setting is emission frequency 0 for the FX-101:and emission frequency
1 for the FX-102:.
Refer to p.11 for setting procedure
: Emission Frequency Setting
When fiber heads are used side-by-side, the emission frequencies for each
amplifier can be set to different frequencies in order to prevent interference.
For the FX-101:, this function is activated by setting the emission frequency to
1, 2 or 3.
For the FX-102:, this function is activated by setting the emission frequency to
1, 2, 3 or 4.
<Number of units for interference prevention>
• Combination of FX-101:only: Maximum 3 units
• Combination of FX-102:only: Maximum 4 units
• Combination of FX-101:and FX-102:: Maximum 4 units
The beam-emitting part and the operation indicator (orange) will blink while the
emission frequency is being set. (They light up steadily when emission frequency 0 is
set.) The blinking cycle will change depending on the emission frequency which is
selected. (Refer to the table below.)
Refer to p.12 for setting procedure
Emission
frequency
Note: The operation indicator (orange) will operate in the same way.
Combination for
interference prevention
function to be enabled
Light-emitting operation
during setting (Note)
Display
Response time
FX-101:
0
250 !s or less
Display
Response time
FX-102:
—— None Lights up
1
450 !s or less 2.5 ms or less Blinks extremely quickly
2
500 !s or less 2.8 ms or less
Blinks quickly
3
600 !s or less 3.2 ms or less
Blinks slowly
4––
5.0 ms or less
Blinks extremely slowly
,,
,,
,,
,,

7
5-2. Teaching
Make sure that detection may become unstable if less margin is applied in the use environment when teaching.
This is the method of setting the threshold value by teaching two levels, corresponding to
object present and object absent conditions. Normally, setting is done by this method.
The output operation setting of Light-ON or Dark-ON is reflected automatically.
2-level Teaching
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
2Press (ON key) in object present condition.
3
The first incident light intensity is set and indicated on the green digital display.
Cancellation of the setting for the first incident light intensity is
possible when (MODE key) is pressed.
5Press (OFF key) in object absent condition.
For output ON when in object present condition
The ‘threshold value’ can be set by utilizing three kinds of teaching, whichever ‘2-level teaching’, ‘limit teaching [,]’ or ‘full-auto teaching’.
Automatic
Automatic
Automatic
Automatic
Automatic
For output ON when in object absent condition
Press ON key in object
present condition.
Press OFF key in object
absent condition.
Automatic
Press ON key in object
absent condition.
Press OFF key in object
present condition.
Threshold value : 2,500
Margin : 25 %
Threshold value : 2,500
Margin : 25 %
Sensing object
present
Sensing object
present
Sensing object
present
Sensing object
present
Sensing object
present
Sensing object
present
Sensing object
absent
Sensing object
absent
Sensing object
absent
Sensing object
absent
Sensing object
absent
Sensing object
absent
Thru-beam type Reflective type
Retroreflective type
Thru-beam type Reflective type
Retroreflective type
Thru-beam type Reflective type
Retroreflective type
Thru-beam type Reflective type
Retroreflective type
6
The threshold value which is set between the first and the second
incident light intensity is indicated on the green digital display.
Margin (%) for the threshold value to the first or the second
incident light intensity is indicated on the red digital display. When
the margin is 200 % or more, ‘ ’ is displayed.
7Press the (MODE key) 5 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
4The present incident light intensity is indicated on the red
digital display with blinking.
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
2Press (ON key) in object absent condition.
3
The first incident light intensity is set and indicated on the green digital display.
Cancellation of the setting for the first incident light intensity is
possible when (MODE key) is pressed.
5Press (OFF key) in object present condition.
6
The threshold value which is set between the first and the second
incident light intensity is indicated on the green digital display.
Margin (%) for the threshold value to the first or the second
incident light intensity is indicated on the red digital display. When
the margin is 200 % or more, ‘ ’ is displayed.
7Press the (MODE key) 5 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
4The present incident light intensity is indicated on the red
digital display with blinking.
Press MODE key
for 2 sec.
Press MODE key
for 2 sec. or 5 times.
Press MODE key
for 2 sec.
Press MODE key
for 2 sec. or 5 times.

8
Limit teaching [] is carried out.
In case of limit teaching []
Threshold value
Incident light intensity
with object absent
This is the method of setting the threshold value by teaching only the object absent condition
(stable incident light condition.) This is used for detection in the presence of a background body or
for detection of minute objects.
Limit Teaching [,]
Automatic
Automatic
Automatic
Press OFF key
Press OFF key
When using thru-beam / retroreflective type fiber When using reflective fiber
Background
Reflective type
Thru-beam type
Retroreflective type
Incident light intensity
Shift amount
b
*Shift amount: The factory setting is ‘15 %’. For the shift amount, refer to ‘Shift Setting (P.13, P.18)’.
Thru-beam type
Retroreflective type
4Press the (MODE key) 5 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
3Press (OFF key) once more.
The incident light intensity at this time has no relationship to
the threshold value.
The set threshold value is indicated on the green digital display.
Margin (%) for the threshold value to the incident light
intensity is indicated on the red digital display. When the
margin is 200 % or more, ‘ ’ is displayed.
2Press (OFF key) in object absent condition.
Reference incident light intensity is set and indicated on the
green digital display. The present incident light intensity is
indicated on the red digital display with blinking.
Cancellation of the setting for the reference incident light
intensity is possible when (MODE key) is pressed.
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
<Limit teaching []operations>
Operation 1 :
Press (OFF key) in object
absent condition.
Reference incident light
intensity setting)
Operation 2 :
Press (OFF key) once
more.
Automatic threshold value
setting
<Limit teaching []operations>
Operation 1 :
Press (ON key) in object
absent condition.
Reference incident light
intensity setting
Operation 2 :
Press (ON key) once
more.
Automatic threshold value
setting
Limit teaching [] is carried out.
The reference incident light
intensity is set in operation 1.
In operation 2, the threshold
value is set to the reference
incident light intensity with the
shift amountbsubtracted.
Incident light intensity
The reference incident light
intensity is set in operation 1.
In operation 2, the threshold
value is set to the reference
incident light intensity with the
shift amountbsubtracted.
Shift amount
b
Threshold value
Incident light intensity
with object absent
Threshold value
Incident light intensity
with object absent
Incident light intensity
Shift amount
b
*
The factory setting is ‘15 %’.
Press MODE key
for 2 sec.
Press MODE key
for 2 sec. or 5 times.

9
Automatic
Automatic
Automatic
Background
Automatic
Automatic
Full-auto teaching is used when it is desired to set the threshold value without
stopping the assembly line, with the object in the moving condition.
Full-Auto Teaching
In case of limit teaching []
Reflective type
Threshold value
Incident light intensity
with object absent
Incident light intensity
Shift amount
b
*
The factory setting is ‘15 %’.
4Press the (MODE key) 5 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
3Press (ON key) once more.
The incident light intensity at this time has no relationship to
the threshold value.
The set threshold value is indicated on the green digital display.
Margin (%) for the threshold value to the incident light
intensity is indicated on the red digital display. When the
margin is 200 % or more, ‘ ’ is displayed.
2Press (ON key) in object absent condition.
Reference incident light intensity is set and indicated on the
green digital display. The present incident light intensity is
indicated on the red digital display with blinking.
Cancellation of the setting for the reference incident light
intensity is possible when (MODE key) is pressed.
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
4Press the (MODE key) 5 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
3
‘ ’ appears in the green digital display after 2 sec. approx.,
and starts sampling incident light intensity from that point.
The threshold value is set when (ON key) or (OFF
key) is releassed.
The set threshold value is indicated on the green digital display.
Margin (%) for the threshold value to the incident light
intensity is indicated on the red digital display. When the
margin is 200 % or more, ‘ ’ is displayed.
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
2Press (ON key) or (OFF key) long.
Thru-beam type
Reflective type
Retroreflective type
Press ON key
Press ON key
Press MODE key
for 2 sec.
Press MODE key
for 2 sec. or 5 times.
Press MODE key
for 2 sec.
Press MODE key
for 2 sec. or 5 times.
ON key or OFF key
Press long.
Release ON key
or OFF key.

10
5-3. Output Operation Setting Factory setting:
Factory setting:
Light-ON or Dark-ON can be set.
Automatic
(Dark-ON)
The current
output operation
(Light-ON)(Dark-ON)
5-4. Timer Operation Setting
This mode sets the timer operation and timer period. Without timer, ON-delay timer, or OFF-delay timer can be set.
In case of setting ON-delay timer or OFF-delay timer in the timer operation setting mode, timer period can be set.
Automatic
(Without timer)
The current
timer operation
Press or to select the output operation to be set.
Press or to select the timer operation to be set.
(Without timer) (ON-delay timer) (OFF-delay timer)
(ON-delay timer period)
(OFF-delay timer period)
(Without timer)
The current
timer operation
<Timer period setting>
Press or to select the
timer period from 1 ms, 5 ms,
10 ms, 20 ms, 40 ms, 50 ms,
100 ms, 500 ms or 1,000 ms.
(OFF)
The current light-emitting
amount setting
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
4
Press
(
MODE key) to confirm the setting.
*
After the setting is confirmed, operation will move to ‘timer operation setting’.
To continue with setting timer operation, carry out the procedure from step
3
of ‘5-4. Timer Operation Settings’ (p.10). In addition, to carry out any
other basic settings, press (MODE key) to select the setting item.
3Press (setting value UP key) or (setting value
DOWN key) to select the output operation to be set.
2Press the (MODE key) once to select ‘output operation
setting mode’.
5Press the (MODE key) 3 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
4Press (MODE key) to confirm the setting.
6Press (MODE key) to confirm the setting.
*After the setting is confirmed, operation will move to ‘light-emitting amount selection
setting’.To continue with setting light-emitting amount selection, carry out the procedure
from step
3
of ‘5-5. Light-emitting Amount Selection Settings’ (p.11). In addition, to
carry out any other basic settings, press (MODE key) to select the setting item.
3Press (setting value UP key) or (setting value
DOWN key) to select the timer operation to be set.
5Press (setting value UP key) or (setting value
DOWN key) to select the timer operation to be set.
<In case of setting ON-delay timer or OFF-delay timer>
The current timer period will be indicated.
2Press the (MODE key) 2 times to select ‘timer setting
mode’.
7Press the (MODE key) 2 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
Press MODE key
for 2 sec.
Press MODE key once.
Press MODE key
to confirm.
Press MODE key
for 2 sec.
Press MODE key
for 2 sec. or 3 times.
Press MODE key 2 times.
Press MODE key
to confirm.
Press MODE key
to confirm.
Press MODE key
to confirm.
Press MODE key
for 2 sec. or 2 times.

11
Factory setting:5-5. Light-emitting Amount Selection Setting
Allows light reduction to be set for the light-emitting amount.
Use this setting if sensing is difficult because the incident light intensity is saturated (4,000 or higher).
Automatic
(OFF)
The current light-emitting
amount selection setting
(Setting for reduced intensity: ON)
(Setting for reduced intensity: OFF)
Press or to select the light-emitting amount selection setting to be set.
The current emission
frequency setting
In case of FX-101:
In case of FX-102:
Emission frequency 0
Emission frequency 1
If the incident light intensity is still saturated (4,000 or higher) even when the light reduction setting ‘ ’ is
used, make the distance between the sensing objects and the fiber head longer.
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
3Press (setting value UP key) or (setting value DOWN
key), and then select ‘ ’ (setting for reduced
intensity: ON) to set light reduction.
2Press the (MODE key) 3 times* to select ‘light-emitting
amount selection setting’.
* Press 4 times when setting the ON-delay timer or the OFF-delay timer.
4Press (MODE key) to confirm the setting.
*After the setting is confirmed, operation will move to ‘emission
frequency setting’. To continue with setting emission frequency,
carry out the procedure from step 3of ‘5-6. Emission
Frequency Setting’ (p.12). In addition, to carry out any other
basic settings, press (MODE key) to select the setting item.
5Press the (MODE key) once or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
Press MODE key
for 2 sec. or 5 times.
Press MODE key
for 2 sec. or once.
Press MODE key 3 times.*
Press MODE key
to confirm.

12
Factory setting: FX-101:,FX-102:h
5-6.
Emission Frequency Setting
Automatic
The current emission
frequency setting
In case of FX-101:
In case of FX-102:
Emission frequency 0
Emission frequency 1
Press or to select the emission frequency to be set.
Emission
frequency
Combination for interference
prevention function to be enabled
Light-emitting operation
during setting
Note: The operation indicator (orange) will operate in the same way.
Display
Response time
FX-101:
0
250 !s
or less
450 !s
or less
500 !s
or less
600 !s
or less
2.5 ms
or less
2.8 ms
or less
3.2 ms
or less
5.0 ms
or less
Display
Response time
FX-102:
—— None Lights up
1
2Blinks quickly
3Blinks slowly
4––
,,
,,
,,
,,
Blinks extremely
quickly
Blinks extremely
slowly
• For the FX-101:, the interference prevention function will not operate for emission frequency 0 (factory setting).
• The response time will change when the emission frequency is changed.
When fiber heads are used side-by-side, the emission frequencies for each amplifier can be set to different
frequencies in order to prevent interference.
In case of FX-101:: Select emission frequency from
‘ ’, ‘ ’, ‘ ’, ‘ ’
In case of FX-102:: Select emission frequency from
‘ ’, ‘ ’, ‘ ’, ‘ ’
1Press the (MODE key) for 2 sec. to select ‘SET mode’.
2Press the (MODE key) 4 timesbto select the ‘emission
frequency setting’.
* Press 5 times when setting the ON-delay timer or the OFF-delay timer.
4Press (MODE key) to confirm the setting.
After the setting is confirmed, operation will return to ‘RUN
mode’ (normal sensing operation).
3Press (setting value UP key) or (setting value DOWN
key) to select the emission frequency to be set.
The operation indicator and the beam-emitting part blink while
setting emission frequency. (When emission frequency 0 is set,
they light up.) The blinking cycle depends on each emission
frequency. (Refer to the table at left.)
Press MODE key
for 2 sec.
Press MODE key 4 times.*
Press MODE key
to confirm.

13
PRO Mode
6-1. PRO Mode Functions and Settings
6
PRO mode is the mode used for making detailed settings. There are nine items that can be set.
: Shift Setting Shift amount can be selected from 0 to 80 % (1 % intervals) in the limit teaching
[,] and the threshold value follow-up cycle setting.
Select 0 % when it is desired to set the present incident light intensity as a
threshold value.
*The factory setting is ‘15 %’.
• Note that when set to ECO, key operations are disabled during external input.
• If ‘ ’ and ‘ ’ (incident light intensity) are blinking alternately in the digital displays,
check if there is a contact between the output wire (black) and the external input wire (white).
Refer to p.18 for setting procedure
: External Input Setting
External input can be selected from emission halt, limit teaching [], limit
teaching [], full-auto teaching, and ECO.
*The factory setting is ‘emission halt’.
Non-voltage contact or NPN open-collector transistor
High (8 V to V DC or open): Ineffective
Low [0 to 2 V DC (source current 0.5 mA or less)] : Effective
Refer to p.18 for setting procedure
[I/O circuits and external input conditions]
<NPN output type>
<PNP output type>
1
2
3
4
1
3
2
4
or
or
m1
Non-voltage contact or PNP open-collector transistor
High [
4 V to
V DC (sink current 0.5 to 3 mA)]: Effective
Low (0 to
0.6 V DC or open): Ineffective
m1
1
2
3
4
1
2
3
4
Terminal arrangement diagram
Terminal arrangement diagram
• External input
Input impedance: 10 kΩapprox.
• External input
Input impedance: 10 kΩapprox.
Sensor circuit
Users’ circuitInternal circuit
100 mA max.
D
Tr
(Blue) 0 V
(White) Extemal input
(Black) Output
Terminal No.
Color code of cable with connector
ZD
(Brown) V
12 to 24 V DC
10 %
m1
8 V
Sensor circuit
Users’ circuitInternal circuit
100 mA max.
D
Tr
(Blue) 0 V
(White) Extemal input
(Black) Output
Terminal No.
Color code of cable with connector
ZD
(Brown) V
12 to 24 V DC
10 %
m1
Load
Load

14
If full-auto teaching ( ) is selected
If limit teaching ( , ) is selected
If emission halt ( ) is selected
(Note 2)
External input High (NPN output type: Low)
Low (NPN output type: High)
Teaching in progress
Normal operation
(Note)
Limit teaching
External input High (NPN output type: Low)
Low (NPN output type: High)
Teaching in progress
Normal operation
Full-auto
teaching
Time chart
25 ms or more (Note 1)
20 ms
Sampling in progress
20 ms
If ECO ( ) is selected
External input High (NPN output type: Low)
Low (NPN output type: High)
For ECO
Normal operation
(Note)
ECO
25 ms or more
20 ms 20 ms
25 ms or more
20 ms
External input High (NPN output type: Low)
Low (NPN output type: High)
Emission halt
Emission
(Note)
Emission halt
25 ms or more
20 ms 20 ms
Note: Output operation will be undetermined only during the response time.
If the output signal is received by something such as a PLC, set the timer to a value
of 20 msfiber sensor response time or greater.
Example: For the FX-101:with emission frequency 0 (response time 250 !s or less)
Timer period: 20 ms 0.25 ms (250 !s) 20.25 ms
Note:
After teaching is complete, output operation will be undetermined only during the response time.
If the output signal is received by something such as a PLC, set the timer to the fiber
sensor response time or greater.
The threshold value will be set based on the incident light intensity at the instant when
teaching is verified.
Notes: 1) Move the sensing object past once during the time that the input signal is being input.
2)
After teaching is complete, output operation will be undetermined only during the response time.
If the output signal is received by something such as a PLC, set the timer to the fiber
sensor response time or greater.
Note: Output operation will be undetermined only during the response time.
If the output signal is received by something such as a PLC, set the timer to a value
of 20 ms fiber sensor response time or greater.
Example: For the FX-101:with emission frequency 0 (response time 250 !s or less)
Timer period: 20 ms 0.25 ms (250 !s) 20.25 ms

15
: Threshold Value Follow-up
Cycle Setting
Follow-up of the threshold value can be carried out within a set range (shift amount)
in response to fluctuations in the incident light intensity. The threshold value
follow-up operation functions by setting the time for refreshing of the threshold value.
At this time, the threshold value at the time of the refresh is indicated on the green
digital display.
However, the threshold value is not stored. (The threshold value returns to the
original value when the power is turned off.)
The threshold value follow-up cycle can be set to between 1 to 59 sec. (1-sec.
intervals), 1 to 9 min. (1-min. intervals) and 10 to 60 min. (10-min. intervals).
The incident light intensity is checked and the threshold value is reset each time
after the threshold value follow-up cycle which has been set.
However, the threshold value is not stored.
The shift amount for follow-up is the shift amount which has been set according
to ‘6-2. Shift setting’ (p.18).
Shift amount: Can be set to between 0 to 80 % (1 % intervals).
(The factory setting is ‘15 %’. )
Refer to p.19 for setting procedure
When threshold value follow-up operation is being carried out, the follow-up
operation stops when the incident light intensity drops to 300 or lower.
At this time, the threshold value blinks in the digital display (green).
[Usage conditions]
This is enabled when thru-beam type fibers or retroreflective type fibers are
being used and output operation is set to Dark-ON.
*If output operation is set to Light-ON, the threshold value is not refreshed.
<Graph showing long-term follow-up>
*The factory setting is ‘OFF’.
Output operation: Dark-ON Output operation: Dark-ON
Shift amount
Incident light
intensity
Incident light
intensity
Threshold
value
Threshold
value
Threshold value
follow-up cycle
Time
Time
Thru-beam type fiber Retroreflective type fiber
Incident light intensityIncident light intensity

16
: GETA Function Setting
The display value for the incident light intensity can be offset by the desired
value (target value).
The target value can be set to between 0 and 2,000 (in intervals of 100).
For example, if the incident light intensity is 1,500 and the target value is set to
2,000, then the value appearing in the digital display will be 2,000.
Refer to p.20 for setting procedure
* The factory setting is ‘GETA OFF’.
4,000
500
0500
1,000
0
[digits]
4,000
[digits]
Example 1: Setting an incident light intensity display of ‘ ’ to
display as ‘ ’
Normal
Incident light
intensity display
Setting possible
in intervals of 100
When using the GETA function
2000
1900
1800
•
•
•
300
200
100
0
Not
displayed
4,000
500
0
0
[digits]
3,500
Example 2: Setting an incident light intensity display of ‘ ’ to
display as ‘ ’ (0-adjust)
4,000
[digits]
Incident light intensityIncident light intensity
• If you press (MODE key) during RUN mode when using the GETA
function, the incident light intensity before the GETA function was used will
be indicated on the red digital display for 2 sec.
• If this function is used while the incident light intensity is saturated (4,000 or
higher), ‘ ’ will be indicated on the red digital display. The maximum
offset value is 4,000.
Because the original
value is 500, the
displays for the lower
limit value will be 500
and the upper limit
value will be 4,000.
Because the original
value is 0, the
displays for the lower
limit value will be 0
and the upper limit
value will be 4,000.

17
: Digital Display Inversion Setting
The viewing orientation of the digital display can be inverted in accordance with
the mounting direction of the amplifier.
*The factory setting is ‘Turn OFF’.
• Be sure to use the setting copy function between the identical models (for all
FX-101:units or all FX-102:units). This function cannot be used between
different models.
• Only one amplifier can be connected on slave side with a master side
amplifier for the setting copy function. If copying to more than one unit,
make the settings for one unit at a time.
• ‘Threshold value’, ‘output operation setting’, ‘timer operation setting’, ‘timer
period setting’, ‘light-emitting amount selection setting’, ‘shift setting’, ‘ECO
setting’, ‘digital display inversion setting’, and ‘threshold value margin
setting’ can be copied.
Refer to p.22 for setting procedure
: ECO Setting The digital display can be set to turn off in order to reduce power consumption
(current consumption). When the ECO setting has been set to ON, ‘ ’ will
blink if no keys have been pressed for approx. 20 sec. while in RUN mode, and
then the digital display will switch off.
To switch the digital display back on again, press and hold any one of the three
keys for 2 sec. or more.
*The factory setting is ‘ECO OFF’.
Refer to p.21 for setting procedure
Note that key operations will be invalid if the ECO setting has been set by means of external input.
;ECO OFF
720 mW or less
Current consumption 30 mA or less
at 24 V supply voltage
Power consumption
;ECO ON
;Turn OFF
: Threshold Value Margin Setting
Margin for threshold value to the present incident light intensity can be checked.
When there is no margin, it is possible to make the digital display blink. The
incident light intensity when the display starts blinking will be approximately half
of the shift amount. [For the shift amount, refer to ‘Shift setting (p.13, p.18)’.]
The blinking method for the digital display can be selected from blinking green
(threshold value blinks), blinking red (incident light intensity blinks) and blinking
green and red simultaneously (threshold value and incident light intensity blink
simultaneously).
*The factory setting is ‘OFF’.
The incident light intensity range
that will cause the digital display
to blink will be from approx. 900 to
approx. 1,100.
Refer to p.22 for setting procedure
: Setting Copy The settings of the master side amplifier can be copied to the slave side
amplifier.
The copying method used is data communication by means of connected wiring.
For details on the setting procedure and wiring, refer to p.23.
*The factory setting is ‘Do not copy settings (Copy NO)’.
Refer to p.23 for setting procedure
: Reset All settings can be returned to factory settings.
For details on the default values for factory settings, refer to p.3.
*The factory setting is ‘Do not reset (Reset NO)’.
Refer to p.24 for setting procedure
<If using a shift amount of 20 % and a threshold value of 1,000>
;Turn ON
Digital display
blinking range
1,100
approx.
Threshold
value
1,000
900
approx.
Incident light
intensity
600 mW or less
Current consumption 25 mA or less
at 24 V supply voltage
Power consumption

18
6-2. Shift Setting
Shift amount during limit teaching [,] and threshold value follow-up cycle setting can be selected from 0 to 80 % (1 % intervals).
Select 0 % when it is desired to set the present incident light intensity as a threshold value.
Automatic
The current
shift amount
The current
external input
The current threshold value
follow-up cycle setting
Automatic
6-3. External Input Setting
External input can be selected from emission halt, limit teaching [
], limit teaching [
], full-auto teaching, and ECO.
(Emission halt) (Limit teaching []) (Limit teaching [])
(ECO) (Full-auto teaching)
Factory setting:
Factory setting:
(15 %)
(Emission halt)
(15 %)
The current
external input (Emission halt)
Press or to set the shift amount to between 0 %
and 80 % (1 % intervals).
(OFF)
1Press the (MODE key) for 4 sec. to select ‘PRO mode’.
3Press (MODE key) to confirm the setting.
*After the setting is confirmed, operation will move to ‘external
input setting’. To continue with setting external input, carry out
the procedure from step 3of ‘6-3. External Input Setting’
(p.18). In addition, to carry out any other basic settings, press
(MODE key) to select the setting item.
2Press (setting value UP key) or (setting value
DOWN key) to select the shift amount to be set.
4Press the (MODE key) 8 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
5Press the (MODE key) 7 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
4
Press
(
MODE key) to confirm the setting.
*
After the setting is confirmed, operation will move to threshold value
‘follow-up cycle setting’. To continue with setting threshold value
follow-up cycle, carry out the procedure from step 3of ‘6-4. Threshold
Value Follow-up Cycle Setting’ (p.19). In addition, to carry out any
other basic settings, press (MODE key) to select the setting item.
3Press (setting value UP key) or (setting value
DOWN key) to select the external input to be set.
1Press the (MODE key) for 4 sec. to select ‘PRO mode’.
2
Press the (MODE key)
once to select ‘
external input setting
’.
Press or to select the external input to be set.
• Note that when set to ECO, key operations are disabled during external input.
• If ‘ ’ and ‘ ’ (incident light intensity) are blinking alternately in the digital displays, check if there is a contact
between the output wire (black) and the external input wire (white).
Press MODE key
for 4 sec.
Press MODE key
to confirm.
Press MODE key
for 2 sec. or 8 times.
Press MODE key
for 4 sec.
Press MODE key
for 2 sec. or 7 times.
Press MODE key once.
Press MODE key
to confirm.

19
Automatic
6-4. Threshold Value Follow-up Cycle Setting
(OFF)
(1 to 59 sec. <1-sec. intervals>)
(1 to 9 min. <1-min. intervals>)
(1 to 60 min. <10-min. intervals>)
Factory setting:
Follow-up of the threshold value can be carried out within a set range (shift amount) in response to fluctuations in
the incident light intensity. The threshold value follow-up operation functions by setting the time for refreshing of
the threshold value. However, the threshold value is not stored.
The shift amount for follow-up is the shift amount which has been set according to ‘6-2. Shift Setting’ (p.18).
When threshold value follow-up operation is being carried out, the follow-up operation stops when the incident light
intensity drops to 300 or lower.
At this time, the threshold value blinks in the digital display (green).
The current threshold value
follow-up cycle setting
The current
GETA function setting (OFF)
(OFF)
(15 %)
Press or to select the threshold value follow-up cycle to be set.
5Press the (MODE key) 6 times or keep it pressed for
2 sec. or more, the amplifier will return to ‘RUN mode’
(normal sensing operation).
3
Press (setting value UP key) or (setting value DOWN
key) to select the threshold value follow-up cycle to be set.
2
Press the (MODE key)
2 times to select ‘
Threshold value
follow-up cycle setting
’.
1Press the (MODE key) for 4 sec. to select ‘PRO mode’.
4Press (MODE key) to confirm the setting.
*After the setting is confirmed, operation will move to ‘GETA
function setting’. To continue with setting GETA function, carry
out the procedure from step 3of ‘6-5. GETA Function Setting’
(p.20). In addition, to carry out any other basic settings, press
(MODE key) to select the setting item.
Press MODE key
for 4 sec.
Press MODE key
for 2 sec. or 6 times.
Press MODE key 2 times.
Press MODE key
to confirm.
This manual suits for next models
2
Table of contents
Other Sunx Accessories manuals

Sunx
Sunx FX-301 User manual

Sunx
Sunx FX-D1 User manual

Sunx
Sunx FX-100-Z Series User manual

Sunx
Sunx EX-21A User manual

Sunx
Sunx DP4 Series User manual

Sunx
Sunx EX-L211-P User manual

Sunx
Sunx LM10 User manual

Sunx
Sunx DP-100 Series User manual

Sunx
Sunx LS Series User manual

Sunx
Sunx FM-200 Series User manual