Thermal mass flow sensor

Dear Customer,
Thank you for choosing our product.
Before you start up the device please read the operating instructions in
full and carefully observe them. The manufacturer cannot be held liable
for any damage which occurs as a result of non-observance or non-
compliance with this manual.
Should the device be tampered with in any manner other than a
procedure which is described and specified in the manual, the warranty
is canceled and the manufacturer is e empt from liability.
The device is destined e clusively for the described application.
SUTO offers no guarantee for the suitability for any other purpose.
SUTO is also not liable for consequential damage resulting from the
delivery, capability or use of this device.
2 S415

Table of Contents
1 Safety instructions......................................................................4
2 Registered trademarks.................................................................6
3 RF e posure information and statement.........................................7
4 Application.................................................................................8
5 Features....................................................................................8
6 Technical data.............................................................................9
6.1 General data.........................................................................9
6.2 Electrical data.....................................................................10
6.3 Output signals.....................................................................10
6.4 Accuracy ...........................................................................10
7 Dimensional drawing..................................................................11
8 Installation...............................................................................12
8.1 Installation considerations....................................................12
8.2 LED indicators.....................................................................12
8.3 Electrical connection ..........................................................13
9 Sensor signal outputs................................................................14
9.1 Analog output.....................................................................14
9.2 Pulse output.......................................................................14
9.2.1 Pulse Connection Diagram...............................................16
9.3 Modbus output....................................................................17
10 Configuration..........................................................................18
11 Calibration..............................................................................19
12 Disposal or waste....................................................................19
13 Appendi A - Specifications.......................................................20
13.1 Flow ranges......................................................................20
13.2 Error code.........................................................................21
14 Appendi B - Modbus communication e ample............................22
S415 3

1 Safety instructions
1 Safety instructions
Please check if this instruction manual matches with the
product type.
Please observe all notes and instructions indicated in this
manual. It contains essential information which must be
observed before and during installation, operation and
maintenance. Therefore this instruction manual must be read carefully
by the technician as well as by the responsible user / qualified
personnel.
This instruction manual must be available at the operation site of the
flow sensor at any time. In case of any obscurities or questions,
regarding this manual or the product, please contact the manufacturer.
WA NING!
Compressed air!
Any contact with quickly escaping air or bursting
parts of the compressed air system can lead to
serious injuries or even death!
•Do not e ceed the ma imum permitted pressure range (see
sensors label).
•Only use pressure tight installation material.
•Avoid that persons get hit by escaping air or bursting parts of the
instrument.
•The system must be pressureless during maintenance work.
WA NING!
Voltage used for supply!
Any contact with energized parts of the product, may
lead to a electrical shock which can lead to serious
injuries or even death!
•Consider all regulations for electrical installations.
•The system must be disconnected from any power supply during
maintenance.
•Any electrical work on system is only allowed by authorized
qualified personal.
4 S415

1 Safety instructions
ATTENTION!
Permitted operating parameters!
Observe the permitted operating parameters, any
operation exceeding this parameters can lead to
malfunctions and may lead to damage on the
instrument or the system.
•Do not e ceed the permitted operating parameters.
•Make sure the product is operated in its permitted limitations.
•Do not e ceed or undercut the permitted storage and operation
temperature and pressure.
•The product should be maintained and calibrated frequently, at
least annually.
General safety instructions
•It is not allowed to use the product in e plosive areas.
•Please observe the national regulations before/during installation
and operation.
emarks
•It is not allowed to disassemble the product.
ATTENTION!
Measurement values can be affected by malfunction!
The product must be installed properly and
frequently maintained, otherwise it may lead to
wrong measurement values, which can lead to
wrong results.
•Always observe the direction of the flow when installing the
sensor. The direction is indicated on the housing.
•Do not e ceed the ma imum operation temperature at the sensors
tip.
•Avoid condensation on the sensor element as this will affect
accuracy enormously.
Storage and transportation
•Make sure that the transportation temperature is between
-30 ... +70°C.
S415 5

1 Safety instructions
•For storage and transportation it is recommended to use the
packaging which comes with the sensor.
•Make sure that storage temperature of the sensor is between
-10 ... +50°C.
•Avoid direct UV and solar radiation during storage.
•For the storage the humidity must be <90% with no condensation.
2 egistered trademarks
SUTO®
Registered trademark of SUTO iTEC
MODBUS®
Registered trademark of the Modbus Organization, Hopkinton, USA
HART®
Registered trademark of the HART Communication Foundation, Austin,
USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe,
Germany
Bluetooth® word mark and logos
Registered trademarks of Bluetooth SIG, Inc.
Android™, Google Play
Trademarks of Google LLC
6 S415

3 RF exposure information and statement
3 F exposure information and statement
This equipment complies with FCC RF radiation e posure limits set forth
for an uncontrolled environment. This equipment should be installed
and operated with minimum distance of 20 cm between the radiator and
your body.
This device complies with part 15 of the FCC rules. Operation is subject
to the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
NOTE: The manufacturer is not responsible for any radio or TV
interference caused by unauthorized modifications to this equipment.
Such modifications could void the user’s authority to operate the
equipment.
NOTE: This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or
more of the following measures:
•Reorient or relocate the receiving antenna.
•Increase the separation between the equipment and receiver.
•Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
•Consult the dealer or an e perienced radio/TV technician for help
•This device and its antenna(s) must not be co-located or operating
in conjunction with any other antenna or transmitter.
S415 7

4 Application
4 Application
The S415 is the thermal mass flow sensor that is designed to measure
the volumetric flow and consumption of compressed air and nitrogen
within the permitted operating parameters(Chapter 6 Technical data).
The default unit settings are: volumetric flow in l/min and total
Consumption in m3. Other units can be configured using the service app
S4C-FS, which can be downloaded from the Google Play store or our
website. For more information see chapter 10.
5 Features
•Inline thermal mass flow sensor virtually independent of pressure
and temperature changes
•Process connection G-type thread, DN8, DN15, DN20 and DN25
•Very short response time
•Particularly suitable for measuring at point-of-use flow and
consumption of compressed air or N2
•Integrated display showing volumetric flow
•Choices of output signals:
◦Analogue 4 … 20 mA and pulse output
◦Modbus interface
◦M-Bus interface
•Bluetooth interface for sensor settings
•Configurable through Android devices
8 S415

6 Technical data
6 Technical data
6.1 General data
FCC ID: 2ASK2-SUTO-002
Parameters Standard unit flow: l/min
Consumption units: m³ (default)
Reference conditions ISO1217 20°C, 1000 hPa (Standard-Unit)
DIN1343 0°C , 1013.25 hPa (Norm-Unit)
Principle of measurement Thermal mass flow
Sensor Glass coated resistive sensor
Ambient temperature
/Transport temperature
0 ... +50°C / -30 ... +70°C
Medium conditions 0 ... +50°C, rH<90% no condensation
Operating pressure 0 ... 1.0 MPa
Pressure drop Ma imum pressure drop at the ma imum
flow of the Standard (S) flow range:
•DN8 type : 30 hPa
•DN15 type: 100 hPa
•DN20 type: 100 hPa
•DN25 type: 200 hPa
Casing Process connection: aluminum alloy
Wetted parts: aluminum alloy
Top casing: PC + ABS
Protection class IP54
Dimensions See dimensional drawing on page 11.
Display 4-digit LED display
Tube diameter DN8, DN15, DN20, DN25
Process connection G inner thread ISO 228-1
Weight 0.45 kg (DN8), 0.44 kg (DN15)
0.97 kg (DN20), 0.94 kg (DN25)
S415 9

6 Technical data
6.2 Electrical data
Power supply 15 ... 30 VDC, 120 mA @ 24 VDC
6.3 Output signals
Analogue output Signal: 4 ... 20 mA, isolated
Scaling: 0 to ma flow
Ma load: 250R
Pulse output 1 pulse per m3 , isolated switch, ma . 30
VDC, 200 mA
(pulse length: 10 ... 120 ms, depends on flow
rate)
Modbus output See section 9.3.
6.4 Accuracy
Accuracy*
(at 6 bar, 20°C, rH<40%)
± (3.0% of reading + 0.3% Full Scale)
Temperature coefficient < 0.1%/K Full Scale
Pressure coefficient <5% / 1 MPa
Turndown ratio 50 : 1
Stated accuracy at Ambient/process temperature 23°C ± 3°C
Ambient/process humidity <90%
Process pressure at 0.6 MPa
Repeatability ± 1% of reading
Sampling rate 3 samples per second
* The specified accuracy is valid only within the minimum and ma imum
flow rates that are stated in section 13.1.
10 S415

8 Installation
8 Installation
S415 is delivered with following components:
Qty. Description Item no.
1 S415 Thermal mass flow meter
S695 4150
S695 4151
S695 4152
S695 4153
1 5 m cable with M8 connector and open ends A553 0136
1 Instruction manual No P/N
1 Calibration certificate No P/N
8.1 Installation considerations
To maintain the accuracy stated in the technical data, the sensor must
be installed inline and fitted to tubes with the same diameter. Please
note the following:
•The sensor is for indoor use only! At an outdoor installation, the
sensor must be protected from solar radiation and rain.
•It is strongly recommend not to install S415 permanently in wet
environment which e ists usually right after a compressor outlet.
8.2 LED indicators
The S415 provides four LED indicators on the display panel, which
indicate the status of the data logger, display, and Bluetooth connection
separately.
12 S415

8 Installation
LED Function Status
Rec Indicator for the logger state On: NAND flash is formatting.
Flashing: Data logger is
working.
lpm Indicator for the display state
(volumetric flow)
On: The display is showing the
volumetric flow rate.
Flashing: N/A
stat. Indicator for the Bluetooth
state
On: The Bluetooth is
connected.
Flashing: N/A
bar Indicator for the display state
(pressure)
On: The display is showing the
pressure value.
Flashing: N/A
8.3 Electrical connection
The flow sensor comes with two connector plugs “A” and “B”. By default
the sensor is delivered with one 5 m cable with a M8 connector on one
side and open wires on the other side. To make the S415 work, one
cable connection is sufficient. However, if the pulse output is to be used
or the supply and the signal need to be on separate cables, a second
connection cable must be ordered.
Legend to pin assignment
GND: Ground for Modbus
-VB: Negative supply voltage
+VB: Positive supply voltage
I+: Positive 4 … 20 mA
signal
I-: Negative 4 ... 20 mA
signal
D+: Modbus data +
D-: Modbus data -
P: Pulse signal
M-Bus: M-Bus data
NA: Not Applicable
S415 13

8 Installation
Pin assignment connector plug M8
Output Version Connector Pin 1 Pin 2 Pin 3 Pin 4
Modbus A D- -VB +VB D+
B D- GND NA D+
Pulse and analog A I- -VB +VB I+
B I- P P I+
M-Bus A M-Bus -VB +VB M-Bus
B M-Bus NA NA M-Bus
Wire color brown white blue black
ATTENTION!
Do not screw the M8 plug using force. Otherwise, it may
damage the connecting pins.
9 Sensor signal outputs
9.1 Analog output
The sensor has an analog output of 4 ... 20 mA. This output can be
scaled to match a desired measuring range. Standard scaling is from 0
to the ma flow. The corresponding flow in different pipe sizes can be
found in section 13.1.
9.2 Pulse output
The sensor outputs one pulse per a consumption unit. This pulse output
can be connected to an e ternal pulse counter to count the total
consumption. The number of m3 per second are summed up and
indicated after one second. Pulse length depends on flow rate.
14 S415

9 Sensor si nal outputs
9.3 Modbus output
Mode : RTU
Baud rate : 19200
Device address : Last two digits of the serial number
Framing / parity / stop bit : 8, N, 1
Response timeout : 1 second
Response delay : 0 ms
Inter-frame spacing : 7 char
emark: Modbus communication settings can be changed by the
service app S4C-FS.
Channel description Resolution Format Length
Modbus
register
address
Flow 0.1 FLOAT 4-Byte 6
Consumption 1 UNIT32 4-Byte 8
emark: The physical units of measurement channels can be
configured using the service APP S4C-FS.
In the response message that the device returns to the master:
•Function code: 03
•Byte order (32-bit data): MID-LITTLE-ENDIAN.
To properly decode the 4-byte float and unsigned integer data in
the response message, the master must change the byte order
from MID-LITTLE-ENDIAN to the order that it is using (LITTLE-
ENDIAN or BIG-ENDIAN).
Byte sequencing of byte orders
Byte order Byte sequencing (HEX) Example
MID-LITTLE-ENDIAN
(Read from the device)
A B C D 0 0A 11 42 C5
LITTLE-ENDIAN B A D C 0 11 0A C5 42
BIG-ENDIAN C D A B 0 42 C5 0A 11
S415 17

10 Confi uration
10 Configuration
To change any settings on the S415, please download and install the
service App S4C-FS from the Google Play store or our Website. This
App works on any Android system with Bluetooth supported.
To be allowed to change settings, the App needs to scan the QR code
attached on the side of the sensor head or on the calibration certificate.
This ensures that only authorized users can access the sensor settings.
For more information about instructions, see the S4C-FS Instruction and
operation manual.
ATTENTION!
Changes on the settings may lead to wrong
measurement results! Contact manufacturer in case
you are not familiar with the settings.
18 S415

11 Calibration
11 Calibration
The sensor is calibrated e work. The e act calibration date is printed
on the certificate which is supplied together with the sensor. The
accuracy of the sensor is regulated by the on site conditions,
parameters like oil, high humidity or other impurities can affect the
calibration and furthermore the accuracy. However we recommend to
calibrate the instrument at least once per year. The calibration is
e cluded from the instruments warranty. For this please contact the
manufacturer.
12 Disposal or waste
Electronic devices are recyclable material and do not
belong in the household waste.
The sensor, the accessories and its packing must be
disposed according to your local statutory requirements.
The dispose can also be carried by the manufacturer of the
product, for this please contact the manufacturer.
S415 19

13 Appendix A - Specifications
13 Appendix A - Specifications
13.1 Flow ranges
The measuring ranges of air are stated under the following conditions:
•Standard flow in air
•Reference pressure: 1000 hPa
•Reference Temperature: +20°C
DN8 DN15 DN20 DN25
Min Ma Min Ma Min Ma Min Ma
Standard
range (S) in
l/min
50 250 20 1000 40 2000 70 3500
Low range (L)
in l/min
1 50 4 200 8 400 14 700
The measuring ranges of N2 are stated under the following conditions:
0°C and 1013.25 hPa:
DN8 DN15 DN20 DN25
Min Ma Min Ma Min Ma Min Ma
Standard
range (S)
4.44 222 17.8 890 35.6 1780 62.2 3110
Low range
(L)
0.89 44.5 3.56 178 7.12 356 12.44 622
Order Table (Air and N2 only)
Order no. Size ange Output Description
S695 415 S415, thermal mass flow meter,
3% o.RDG., 24 VDC
0 DN8 G inner thread
1 DN15 G inner thread
2 DN20 G inner thread
3 DN25 G inner thread
20 S415
Other manuals for S415
2
This manual suits for next models
4
Table of contents
Other SUTO Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Rhopoint Instruments
Rhopoint Instruments IQ Flex 20 operating instructions

JENCO
JENCO VisionPlus 6177 Operation manual

PASCO
PASCO SF-9069 quick start guide

König Electronic
König Electronic SEC-DVR416 manual

Ruskin
Ruskin TDP05K Technical bulletin

Dwyer Instruments
Dwyer Instruments SLR-01 Installation and operating instructions