SUTO S 120 Manual

English
Instruction and operation manual
S120
Oil Vapor Monitor

Dear Customer,
Thank you for choosing our product.
Before you start up the device, please read this manual in full and
carefully observe instructions stated. The manufacturer cannot be held
liable for any damage that occurs as a result of non-observance or non-
compliance with this manual.
Should the device be tampered with in any manner other than a
procedure that is described and specified in the manual, the warranty is
void and the manufacturer is e empt from liability.
The device is designed e clusively for the described application.
SUTO offers no guarantee for the suitability for any other purpose.
SUTO is also not liable for consequential damage resulting from the
delivery, capability or use of this device.
2 S120

Table of Contents
1 Safety instructions......................................................................5
2 Registered trademarks.................................................................7
3 Application.................................................................................8
4 Features....................................................................................8
5 Technical data.............................................................................9
5.1 Measuring parameter.............................................................9
5.2 General data.........................................................................9
5.3 Electrical data.....................................................................10
5.4 Output signals.....................................................................10
6 Dimensional drawing..................................................................11
7 Installation ..............................................................................12
7.1 Installation requirements......................................................12
7.2 Wall mounting instructions ...................................................12
7.3 Installation procedure..........................................................14
7.3.1 Installation requirements................................................14
7.3.2 Installation steps...........................................................16
7.4 Electrical connection............................................................17
7.4.1 USB port.......................................................................17
7.4.2 RJ45 connector..............................................................17
7.4.3 M12 connectors.............................................................17
8 Configuration ...........................................................................20
8.1 Integrated display...............................................................20
8.2 E ternal display device ........................................................20
8.3 Service kit..........................................................................20
9 Operations using the integrated display........................................21
9.1 User interface ....................................................................21
9.1.1 Main screen..................................................................21
9.1.2 Quick buttons................................................................22
9.1.3 Status bar....................................................................22
9.2 Main menus........................................................................23
9.3 Sensor settings ..................................................................24
9.3.1 Basic setting.................................................................24
9.3.2 Analog output ..............................................................25
9.3.3 Modbus/RTU settings......................................................25
9.3.4 Modbus/TCP settings......................................................26
9.3.5 Alarm settings...............................................................26
9.3.6 Status..........................................................................26
9.4 Logger settings...................................................................27
9.5 Files...................................................................................28
9.6 Service info .......................................................................28
9.7 System settings .................................................................29
S120 3

10 Troubleshooting.......................................................................30
10.1 LED indicators...................................................................30
10.2 Error indications................................................................31
11 Signal outputs.........................................................................32
11.1 Analog output ..................................................................32
11.2 Digital output ...................................................................32
11.3 Alarm output ....................................................................33
12 Optional accessories................................................................35
12.1 Sensor display...................................................................35
12.2 Service kit .......................................................................35
13 Calibration..............................................................................35
14 Maintenance...........................................................................35
15 Disposal or waste....................................................................35
16 Warranty................................................................................36
4 S120

1 Safety instructions
1 Safety instructions
Please check if this instruction manual matches the
product type
Please observe all notes and instructions indicated in this
manual. It contains essential information which must be
observed before and during installation, operation and
maintenance. Therefore this instruction manual must be read carefully
by the technician as well as by the responsible users and qualified
personnel.
This instruction manual must be available at the operation site of the oil
vapor sensor at any time. In case of any obscurities or questions,
regarding this manual or the product, please contact the manufacturer.
WARNING!
Compressed air!
Any contact with quickly escaping air or bursting
parts of the compressed air system can lead to
serious injuries or even death!
•Do not e ceed the ma imum permitted pressure range (see
sensors label).
•Only use pressure tight installation material.
•Avoid that persons get hit by escaping air or bursting parts of the
instrument.
•The system must be pressure-less during maintenance work.
WARNING!
Voltage used for supply!
Any contact with energized parts of the product, may
lead to an electrical shock which can lead to serious
injuries or even death!
•Consider all regulations for electrical installations.
•The system must be disconnected from any power supply during
maintenance work.
•Any electrical work on the system is only allowed by authorized
qualified personal.
S120 5

1 Safety instructions
ATTENTION!
Permitted operating parameters!
Observe the permitted operating parameters, any
operation exceeding this parameters can lead to
malfunctions and may lead to damage on the
instrument or the system
•Do not e ceed the permitted operating parameters.
•Make sure the product is operated in its permitted limitations.
•Do not e ceed or undercut the permitted storage and operation
temperature and pressure.
•The product should be maintained and calibrated frequently, at
least annually.
General safety instructions
•It is not allowed to use the product in e plosive areas.
•Please observe the national regulations before/during installation
and operation.
Remarks
•It is not allowed to disassemble the product.
ATTENTION!
Measurement values can be affected by malfunction!
The product must be installed properly and
frequently maintained, otherwise it may lead to
wrong measurement values, which can lead to
wrong results
Storage and transportation
•Make sure that the transportation temperature of the device is
between -30 ... +70°C.
•For transportation it is recommended to use the packaging which
comes with the device.
•Please make sure that the storage temperature of the device is
between -20 ... +50°C.
•Avoid direct UV and solar radiation during storage.
•For the storage the humidity must be <90%, no condensation.
6 S120

3 Application
3 Application
The S120 Oil Vapor Monitor is designed to monitor oil contents and dew
point (option) in compressed air and gases within the permissible
operating parameters. These parameters can be found in the technical
data section.
The S120 is mainly used in compressed air systems in industrial
environment. The S120 is not developed to be used in e plosive areas.
To evaluate its applicability in e plosive areas, please contact the
manufacturer.
4 Features
•Measures oil vapor contents and pressure dew point (option) in
compressed air and other gases.
•Easy connection through sampling hose and quick connect.
•Applicable in the permanent or portable applications.
•Measures down to 0.001 mg/m3.
•PID sensor for the highest accuracy.
•Optional dew point sensor with dual-sensor technology for high
accuracy over the whole range from –100 ... +20 °C Td (option).
•Service and alarm indication through LEDs.
•Connectable to display and data logger of SUTO as well as third-
party display and control units.
•IP65 casing provides robust protection in rough industrial
environment.
•Optional local display for showing actual readings without e tra
cable connection.
•Supports multiple interfaces for signal outputs and electrical
connections.
8 S120

5 Technical data
5 Technical data
5 1 Measuring parameter
Parameter Unit Range Resolution Accuracy
Oil vapor mg/m30.001 ... 5.000
mg/m3 *
0.001 mg/m³ 5% of reading ± 0.003
mg/m3
Pressure bar(g) 0 … 16 bar(g) 0.01 bar(g) 0.5% FS
Temperature °C 0 … 50°C 0.1°C 0.5°C
Dew point
(Option)
°C Td -100 ... +20°C Td 0.1°C Td ±1°C Td (0 ... 20°C Td)
±2°C Td (-70 ... 0°C Td)
±3°C Td (-100 ... -70°C
Td)
* Based on 1000 hPa(a), 20°C, 0% relative humidity
5 2 General data
Principle of measurement •Photo ionization
•Oscillating crystal (option)
Sensor •PID (photo ionization detector)
•Piezzo resistive pressure sensor
•QCM + Polymer (option)
Measuring medium Compressed air and gases free of corrosive,
aggressive, caustic and flammable
constituents
Sample flow range < 2 l/min, measuring gas is released to
ambient
Operating temperature -20 ... +50°C
Gas humidity < 40% rel. humidity, no condensation
Operating pressure 3 ... 15 bar(g) (higher pressure on request)
Low pressure (optional) 0.5 ... 2 bar(g)
Housing material PC, Al alloy
Protection class IP65
Dimensions See dimensional drawing on page 11.
S120 9

5 Technical data
Display (optional) 5” color touch screen with a data logger of
100 million measurement values
Weight •2.4 kg
•2.58 kg (with option A1250)
UV lamp lifetime 6,000 working hours or 1 year, whichever
comes first
5 3 Electrical data
Power supply 24 VDC ± 5%, 10 W
5 4 Output signals
Analog output 4 ... 20 mA
Alarm output Relay, NO, 40 VDC, 0.2 A
Digital interface RS-485, Modbus/RTU,
Ethernet, Modbus/TCP, USB (available for
display version)
10 S120

Installation
7 Installation
Please make sure that all components listed below are included in your
package.
Qty Description Item No
1 S120 Oil Vapor Monitor, or S604 1201
S120 Oil Vapor Monitor portable, or S604 1202
S120 Oil Vapor Monitor with display S604 1203
Option: Integrated dew point sensor,
-100 ... +20 °C Td (only for S604 1203
and P604 1205)
A1250
3 M12 connectors or M12 cables
(depending on orders)
Connector: C219 0059
Cable: A553 0104 /
A553 0105
1 Mains unit (100 … 240 VAC) A554 0107
1 1.5 m teflon hose with a quick
connector and a compressed air
coupling at the ends
A554 3316
1 Mounting brackets No
1 Instruction manual No
1 Calibration certificate No
7 1 Installation requirements
The S120 comes with two versions:
•S120 for stationary use. The stationary version comes with four
mounting brackets which can be mounted from the backside of the
instrument at each corner. This allows an easy installation at a
wall.
•S120-P for portable use. The portable version comes in a transport
case.
7 2 Wall mounting instructions
The device can be mounted on the wall using the supplied brackets.
Please use one of the following dimensions to prepare your holes.
12 S120

Installation
7 3 Installation procedure
7 3 1 Installation requirements
Please consider the following recommendations for a successful
measurement result:
•All components from the sampling point to the S120 must be oil
and grease free.
•Ambient and gas temperature must be within the specified ranges
stated in section General data.
•The inlet gas must be pressurized with the valid ranges.
•The sampling gas mus be dry (< 40% RH) and clean.
•Ensure that valves at the sampling point are not lubricated.
ATTENTION!
Avoid contamination with oil or grease!
It will lead to very slow measurement or impossible
measurement results!
•Make sure that there is no rough contaminants at the point of
measurement. Steps are as follows:
1. Connect the purge filter test kit onto your measuring point
first. Open the purge valve on the test kit and purge air for a
short period.
2. Check the filter in the test kit to see whether it shows high
contamination of water, oil or dust.
3. If the filter is contaminated severely, stop using the S120 for
measurement because this may lead to serious damage to the
device. In case you are not sure, please contact the
manufacturer.
14 S120

Installation
7 3 2 Installation steps
The following steps e plain the procedure of an appropriate installation.
Most importantly, before you connect S120 to the compressed air,
purge air out from the measuring point to remove any residual
contamination using the purge filter test kit.
1. Connect the teflon hose with
the inlet of the S120 as
shown in the picture.
2. Connect the other end of the
teflon hose with a quick
connector. The teflon hose
with quick connector is used
to connect the S120 to the
process.
16 S120

Installation
7 4 Electrical connection
The S120 provides 3 kinds of electrcal connections by the following
interfaces:
•One USB port
•One RJ-45 connector
•Three M12 connectors
7 4 1 USB port
The USB port is used to import or e port files.
Through the USB port, the S120 can be connected with:
•An OTG memory stick: To import firmware for upgrade and to
e port data.
•A PC where data analysis software such as S4A is installed: To
e port data to the PC for analysis.
7 4 2 RJ-45 connector
The RJ-45 connector is used as an Ethernet port for IP networking.
Through this connector, the S120 connects to the TCP/IP network over
the Modbus/TCP protocol.
7 4 3 M12 connectors
The 3 M12 connectors are used to connect to display units from SUTO
or the third-party displays and control units.
Also, the S120 is powered through these connectors.
Connection to the following e ternal display units from SUTO.
S120 17

Installation
S120 without display - M12 pin assignment
Connector Pin 1 Pin 2 Pin 3 Pin 4 Pin 5
A SDI -Vb+VbD+ D-
B PE -Vb+Vb+I -I
C Relay Relay GND D+ D-
Color Brown White Blue Black Gray
S120 with display - M12 pin assignment
Connector Pin 1 Pin 2 Pin 3 Pin 4 Pin 5
A SDI -Vb+VbNA NA
B PE -Vb+Vb+I -I
C Relay Relay GND D+ D-
Color Brown White Blue Black Gray
Legend to pin assignment
SDI Digital signal (internal use)
-VBNegative supply voltage
+VBPositive supply voltage
+l Positive 4 … 20 mA signal
-I Negative 4 … 20 mA signal
D+ RS-485, Modbus/RTU
D- RS-485, Modbus/RTU
Relay Alarm output
PE Earth connection
GND Communication ground
NA Not available
S120 19

8 Configuration
8 Configuration
The S120 is delivered with standard e -work configuration or with
specific customer settings according to the order.
Standard ex-work configuration
Scaling : 4 mA = 0.000 mg/m3
20 mA = 5.000 mg/m3
Alarm : 1.000 mg/m3, up
Oil type : Isobutene
Modbus : Device address = Last two digits of the serial number
Baud rate = 19200
Framing/parity/Stop bit = 8, N, 1
Transmission mode = RTU
You can use one of the following ways to configure S120.
8 1 Integrated display
See Chapter Operations using the integrated display.
8 2 External display device
See the S330/S331 Instruction Manual.
8 3 Service kit
Please ensure that S120 or the service kit is connected with the power
supply because the USB port cannot supply enough power for both of
them.
For more information, refer to the instruction manual of Service Kit.
20 S120