VistaLab 1070-0002 User manual

To Install a Tip
1. Insert nozzle into tip.
2. Lightly press down until a
“click” is heard. This
indicates the tip is securely
sealed onto the nozzle.
M
Quick Reference Guide for
Mechanical Volume-Adjust Pipettes
Refer to Operator’s Guide at www.vistalab.com for complete instructions.
To Aspirate & Dispense
1. Depress plunger to the first stop.
2.
Place tip in sample and release
plunger,
wait one second, then
remove tip from sample.
3. Depress plunger to first stop,
wait one second and then fully
depress plunger.
To Eject a Tip
1.
Point tip into a waste container.
2. Press the tip eject button.
The ejector sleeve is activated,
effortlessly discarding the tip.
To Pick Up Ovation
1. Allow body to fill palm
2.
Rotate adjustable hook
to rest comfortably on
your forefinger
Ejec or Sleeve
Nozzle
Volume Display Body
Tip Ejec Bu on
Adjus able
Hook
Plunger
Volume Label
Buil -in Cap Opener
To Set olume
1.
Rotate the plunger to display the desired volume.
(clockwise to decrease, counter clockwise to increase).
Do not rotate beyond the pipette’s volume limits!
This will damage the pipette and void its warranty.
The Ovation M pipette is a variable volume, air displacement,
2 stroke
pipette intended to aspirate and dispense
precise fluid volumes.
Example Catalog number
4060-1002
VistaLab Tec nologies, Inc.
2 Geneva Road
Brewster, Y 10509 USA
(888) 652-6520 or (914) 244-6226 L070-0002-001 Rev. C
Calibra ion Ring
Tips and Accessories
VistaRak™
VistaStak™
Stacked Rack
0.2-2µL
Ovation M mechanical volume-adjust models: 1.0-10µL 10-100µL
2.0-20µL 20-200µL 100-1000µL
TIP SIZE: MICRO SMALL LARGE
Notes:
1. Not for use with Ovation M .02-2µL. 2. Ejector sleeve not available for Ovation M .02-2µL.
NON-STERILE
STERILEFILTERED
VistaRak
192 tips/rack, 5 racks 4060-1002 4060-2004 4060-3004
VistaRak, Low Retention
192 tips/rack, 5 racks
4070-1002LR 4070-2004LR 4070-3004LR
VistaStak, 192 tips/layer, 5 layers (small size)
or 3 layers (micro and large size) 4060-9024 4060-9025 4060-9026
Stacked Rack, 200 tips/layer,
5 layers (small size) or 3 layers (large size) 9025 9026
Stacked Rack, race Metal Certified, 200 tips/layer
5 layers (small size) or 3 layers (large size) 9022 9023
VistaBulk™, 1000 tips/bag, or
250 tips/bag (5mL) or 100 tips/bag (10mL) 4058-1000 4058-2000 4058-3000
Protectainer™Bulk Pack,
1000 tips (small) or 750 tips (large) 4025 4026
Econo-Pak™Bulk Pack,
1000 tips 4225 4226
4060-1032 4060-2132 4060-3132
4070-1032LR 4070-2132LR 4070-3132LR
2025 2026
2027 2028
4060-13321
4060-2332 4060-3332
4060-1333†
4070-1332LR1
4
070-2332LR 4070-3332LR
4070-1333LR†
Nozzl , 5/box (0.2–2µL) or 10/box 9057-1013† 9057-2010 9057-3017
† for Ovation 0.2-20µL only
9057-1010
Nozzl R plac m nt Filt rs, 25/bag n/a 9057-2009 9057-3009
Typ A Nozzl , 5/box n/a n/a 9057-1017
Nozzl Ext nd r ( xc pt 0.2-2µL) 9057-1012 9057-3012
Ej ctor Sl v 9057-10112 9057-3011
Calibration Lab ls, 5/box 9057-4002
Ovation Pip tt Stand 9057-4003
Calibration K y 9070-4007
ACCESSORIES
QRG_O ationM_Re C
VistaRak, Sterile,
non-pyrogenic, certified
RNase/DNase free,
192 tips/rack, 5 racks
VistaRak,
Sterile, non-pyrogenic, ow Retention,
certified
RNase/DNase free,
192 tips/rack,
5 racks
VistaTip™Individually Wrapped
Sterile,
200 tips
VistaTip Individually Wrapped, Sterile,
non-pyrogenic, 200 tips
VistaRak, Filtered, Sterile,
non-pyrogenic,
certified RNase/DNase free,
192 tips/rack,
5 racks
VistaRak, Filtered, Sterile,
non-pyrogenic,
Low Retention, certified RNase/DNase free
192 tips/rack, 5 racks
† 2µL tip for Ovation 0.2-2µL
† 2µL tip for Ovation 0.2-2µL
Pipetting Hints for Optimal Performance
•
When using viscous or volatile reagents, pre-wetting the
pipette tip may be appropriate. To pre-wet, aspirate and
dispense the liquid back into the original vessel. Then fill
the tip and dispense contents into receiving vessel.
• If an air bubble forms in the tip during aspiration,
return the sample, discard the tip, and use a fresh tip.
• Smoothly depress and release the plunger when
pipetting, maintaining the same speed of motion for all
samples. Do not let the plunger “snap back” to the
uppermost position.
•
When finished aspirating, touch the pipette tip against
the side of the vessel as it is being withdrawn to remove
any liquid that may have adhered to the outside of the
tip. Wipe the pipette tip O LY if there is liquid adhering
to the outside. Be careful not to “wick out” any of the
contents from the pipette tip.
•
Dispense against the side of the receiving vessel or
above the liquid surface. To remove any of the measured
liquid that may have adhered to the pipette tip, touch off
the tip against the side of the receiving vessel.
•
If aspirating from 15mL or 50mL centrifuge or other tall
tubes, an optional ozzle Extender accessory is
available which lengthens the pipette’s overall reach.
•
If aspirating from PCR or microtubes, the Ovation M has
a built-in cap opener. Simply insert the cap into the
pipette’s base.
Replacing a Nozzle
If the Ovation M is not aspirating and dispensing properly,
it could be caused by (a) poor fitting alternate source tip,
(b) nozzle is loose and needs to be tightened by turning it
clockwise, or (c) a nozzle or nozzle filter needs to be
replaced as liquid has been aspirated into it.
OTE: Wear gloves when doing this procedure.
1. Remove ejector sleeve by pulling it away from the
pipette body to expose the nozzle assembly.
2.
Place silicone tubing provided on nozzle, and unscrew
nozzle in a counter-clockwise direction. Gently wipe
the exposed end of the extension tube with a slightly
dampened, lint-free tissue, then dry it off.
ote: Do not block or dislodge anything at the end of
the extension tube.
3. Place silicone tubing on a new nozzle, and screw the
nozzle onto the extension tube in a clockwise
direction. Firmly tighten and remove silicone tubing.
4. Slide the ejector sleeve over the nozzle assembly.
5. Push the sleeve firmly into place while using a slight
rocking or twisting motion.
Taking Care of t e Ovation M
The Ovation pipette requires minimal routine maintenance.
Always store it in its “standing” position when not in use.
Clean outer surfaces as needed with a soft cloth dampened
with warm water. To decontaminate outer surfaces, wipe
with a 70% aqueous solution of ethanol or isopropanol, or
use a 10% bleach solution followed by water. Remove
ejector sleeve and wipe outside of nozzle and inside of
ejector sleeve with 70% alcohol to remove any
accumulated deposits. O LY WIPE THE VOLUME
DISPLAY WI DOW WITH WATER.
Pipette body chemical compatibility: Water, diluted
ethanol or isopropanol, diluted bleach. See the document
“Ovation Chemical Compatibility” in the support area at
www.vistalab.com about other specific chemical
formulations concerning the pipette or VistaLab tips.
Calibrating t e Ovation M
The Ovation M pipette can be easily in-lab calibrated. To
make a calibration adjustment, insert the calibration key
into the calibration ring at the base of the plunger rod,
and turn
clockwise to decrease volume delivery or
counterclockwise
to increase. Each 1/4 turn equals 1.7%
of the pipette’s volume range. (example: on 100µL
model, each 1/4 turn adjusts the olume 1.7µL)
Operating Temperature & Environment:
Temperature Range: 15˚– 35˚C
Relative Humidity Range, non-condensing: 10-85%
Atmospheric Pressure: 70-106 kPa
If the pipette is to be used with hazardous
fluids, safe laboratory practice should be
followed. Refer to manufacturer’s Material
Safety Data Sheets for proper handling instructions.
Always use VistaLab filtered tips (see tips table for catalog
numbers) when pipetting potentially caustic, corrosive or
volatile solutions. Failure to do so may result in premature
wear and damage to the internal seals and piston, and
void the pipette’s warranty. When organic or volatile
solutions are routinely used, VistaLab Technologies
recommends the use of VITO seals. See the chemical
compatibility chart in the Ovation documentation library at
www.vistalab.com for more information.
Warranty
VistaLab Technologies, Inc. warrants the Ovation M Pipette
against defects in materials and workmanship for two years
from the date of purchase. Register your pipette and
activate the warranty at www.vistalab.com.
This warranty is void if defects or damage result from
improper handling, unauthorized modification, or use of
ancillary products not supported by VistaLab
Technologies. This warranty is exclusive; no other warranty
is expressed or implied.
Should the pipette need to be returned for service or
calibration verification, contact Customer Service for a
return authorization. Repack the pipette in its original
packaging. Customer is responsible for shipping and
insurance charges. If original packaging is unavailable,
contact VistaLab Technologies for alternative packaging
instructions. Damage to the pipette as a result of improper
packaging is the responsibility of the customer.
To send in your pipette for factory maintenance and/or
calibration verification, see the Calibration & Repair area of our
website –
www.vistalab.com. For technical assistance,
contact us at:
istaLab Technologies, Inc. 1-914-244-6226
2 Geneva Road techservice@vistalab.com
Brewster, NY 10509 USA
Volume Dispensing
Range Increments
Model (µL) (µL) Accuracy Precision
Ovation M 2µL 0.2 – 2 .002 ±1.5% at 2µL 0.9% at 2µL
(1070-0002) ±2.7% at 1µL 1.8% at 1µL
±12% at 0.2µL 7.6% at 0.2µL
Ovation M 10µL 1 – 10 .01 ±1% at 10µL 0.4% at 10µL
(1070-0010) ±1.5% at 5µL 0.6% at 5µL
±2.5% at 1µL 1.2% at 1µL
Ovation M 20µL 2 – 20 .02 ±1% at 20µL 0.3% at 20µL
(1070-0020) ±1.5% at 10µL 0.5% at 10µL
±7% at 2µL 2% at 2µL
Ovation M 100µL 10 – 100 .1 ±0.8% at 100µL 0.15% at 100µL
(1070-0100) ±0.8% at 50µL 0.24% at 50µL
±3.5% at 10µL 1% at 10µL
Ovation M 200µL 20 – 200 .2 ±0.8% at 200µL 0.15% at 200µL
(1070-0200) ±0.8% at 100µL 0.25% at 100µL
±2.5% at 20µL 1% at 20µL
Ovation M 1000µL 100 – 1000 1.0 ±0.8% at 1000µL 0.15% at 1000µL
(1070-1000) ±0.8% at 500µL 0.2% at 500µL
±3% at 100µL 0.6% at 100µL
Ovation M Pipette Specifications
ote: Performance data is based on the use of Ovation tips from VistaLab Technologies. If using tips from
another manufacturer, the above performance claims may not be achieved. Specifications subject to change.
See www.vistalab.com for current information.
⤹
⤹
+
_
Autoclaving t e Ovation M
Set the Ovation M to its highest volume setting and place
it in the autoclave at 121 C for 15 minutes, followed by a
15 minute drying cycle. Allow the pipette to cool to room
temperature for at least 2 hours before use. Calibration
should be verified after each autoclave treatment.
ote: The Ovation M nozzle
should be replaced after 3
autoclave cycles as it will lose its liquid blocking capability.
Ejector Sleeve Nozzle
Extension Tube
Silicone Tubing
Service Options
VistaLab Technologies offers comprehensive calibration
and repair programs to meet all regulatory requirements
and ensure optimum pipette performance. Services can
be initiated through the Vistalab web site.
See www.vistalab.com/service.asp
QRG_O ationM_Re B_Layout 1 9/23/14 11:07 AM Page 2

This manual suits for next models

10

Other VistaLab Laboratory Equipment manuals

VistaLab Ovation M Use and care manual

VistaLab

VistaLab Ovation M Use and care manual

VistaLab 1136 Manual

VistaLab

VistaLab 1136 Manual

VistaLab Ovation Manual

VistaLab

VistaLab Ovation Manual

VistaLab MLA Series User manual

VistaLab

VistaLab MLA Series User manual

VistaLab Ovation ESC User manual

VistaLab

VistaLab Ovation ESC User manual

VistaLab ali-Q User manual

VistaLab

VistaLab ali-Q User manual

VistaLab Ovation 9060-5009 Guide

VistaLab

VistaLab Ovation 9060-5009 Guide

VistaLab Ovation M Operating and safety instructions

VistaLab

VistaLab Ovation M Operating and safety instructions

VistaLab Ovation M User manual

VistaLab

VistaLab Ovation M User manual

VistaLab Ovation 9060-1011 Guide

VistaLab

VistaLab Ovation 9060-1011 Guide

VistaLab Ovation 9060-4005 Guide

VistaLab

VistaLab Ovation 9060-4005 Guide

VistaLab Ovation 9057-2009 Use and care manual

VistaLab

VistaLab Ovation 9057-2009 Use and care manual

VistaLab Ovation M Use and care manual

VistaLab

VistaLab Ovation M Use and care manual

VistaLab ali-Q 2 Installation instructions

VistaLab

VistaLab ali-Q 2 Installation instructions

VistaLab Ovation QS Manual

VistaLab

VistaLab Ovation QS Manual

VistaLab MLA Micro D-Tipper Series User manual

VistaLab

VistaLab MLA Micro D-Tipper Series User manual

VistaLab MLA pipette MLA 1061 Manual

VistaLab

VistaLab MLA pipette MLA 1061 Manual

Popular Laboratory Equipment manuals by other brands

Sorvall Cellwasher 2 Service manual

Sorvall

Sorvall Cellwasher 2 Service manual

Talboys Advanced Series instruction manual

Talboys

Talboys Advanced Series instruction manual

Ametek USB Lab B.O.S.S. Operation manual

Ametek

Ametek USB Lab B.O.S.S. Operation manual

IKA H 44 Mounting instructions

IKA

IKA H 44 Mounting instructions

Drucker Diagnostics 642E Operator's manual

Drucker Diagnostics

Drucker Diagnostics 642E Operator's manual

Barnstead EASYpure RO 742 Series manual

Barnstead

Barnstead EASYpure RO 742 Series manual

SciCan Statim 2000 Operator's manual

SciCan

SciCan Statim 2000 Operator's manual

Biometra Whatman WT16 manual

Biometra

Biometra Whatman WT16 manual

Fluke FlexCam quick start

Fluke

Fluke FlexCam quick start

CytoTherm CT-DR owner's manual

CytoTherm

CytoTherm CT-DR owner's manual

Opticon NLV-5201 Specification manual

Opticon

Opticon NLV-5201 Specification manual

NACALAI TESQUE COSMOSIL quick start guide

NACALAI TESQUE

NACALAI TESQUE COSMOSIL quick start guide

Sibata 080130-042 Operation manual

Sibata

Sibata 080130-042 Operation manual

Opentrons Flex quick start guide

Opentrons

Opentrons Flex quick start guide

Wealtec HB-1 Installation and operation manual

Wealtec

Wealtec HB-1 Installation and operation manual

Thermo Scientific Neurobasal-A Medium user guide

Thermo Scientific

Thermo Scientific Neurobasal-A Medium user guide

Quantabio Q-qPCR quick start guide

Quantabio

Quantabio Q-qPCR quick start guide

Agilent Technologies varian 320-MS LC/MS Pre-installation instructions

Agilent Technologies

Agilent Technologies varian 320-MS LC/MS Pre-installation instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.