When to Replace:
Nozzle filters are made of a cellulose
material which prevents liquid from
entering the interior of the pipette. If
the filter becomes wet, the pipette will
not aspirate or dispense until the filter
is replaced.
Catalog No. 9060-5009
For electronic macro volume model 100–5000µL
Package includes:
•10 nozzle filters
• 1 plug
•Instructions
Electronic Macro Nozzle Filter
Replacement Instructions
N
NO
OT
TE
E:
:Wear gloves for this procedure.
1.
Place a macro pipette tip onto the nozzle
and push on ejector sleeve until click is
heard. Grip tip firmly near the nozzle,
and turn it in a counterclockwise
direction to unscrew and remove nozzle
from pipette.
2. Slide off ejector sleeve and set aside.
3. Remove the rubber plug at the tip end
of the nozzle by inserting a straightened
paperclip into the hole in the plug. Use
the inserted end of the paperclip to pry
the plug out of the end of the nozzle.
4. Insert paperclip into the threaded end
of the nozzle and push out the old filter.
Procedure:
Nozzle
Ejector Sleeve
MacroFilterReplacement 9/8/06 12:34 PM Page 1
L065-0004-001 Rev. A
2 Geneva Road
Brewster, NY 10509 USA
(888) 652-6520 (914) 244-6226 • www.vistalab.com
5.
Pick up a new nozzle filter and place it
into the tip end (non-threaded) of the
nozzle. The rounded end of the filter
should be inserted first. Use finger to
push the filter into position. Firmly
press until the filter is pushed as far as
it will go.
6.
Re-insert the rubber nozzle plug into
the front end of the nozzle. Press in
until plug is nearly even with surface
edge of the nozzle. If filter is not
completely seated, the plug will extend
beyond the end of the nozzle. Press
nozzle plug on countertop if necessary.
7. Place a gloved finger lightly against
the end of the air tube, and press and
release plunger to expel any liquid
that may be present in the end of the
tube. Gently wipe the end of the air
tube with a slightly dampened, lint-free
tissue, then dry it off. Replace the
ejector sleeve.
Note: The ejector sleeve is slightly
narrower on the end that is closest to
the pipette body.
8.
Screw the nozzle back onto the
extension tube. Tighten by hand, or use
a macro pipette tip to ensure the nozzle
is firmly attached to the extension tube.
MacroFilterReplacement 9/8/06 12:35 PM Page 2

Other VistaLab Laboratory Equipment manuals

VistaLab ali-Q 2 Installation instructions

VistaLab

VistaLab ali-Q 2 Installation instructions

VistaLab MLA Series User manual

VistaLab

VistaLab MLA Series User manual

VistaLab ali-Q User manual

VistaLab

VistaLab ali-Q User manual

VistaLab Ovation M Operating and safety instructions

VistaLab

VistaLab Ovation M Operating and safety instructions

VistaLab Ovation 9060-4005 Guide

VistaLab

VistaLab Ovation 9060-4005 Guide

VistaLab 1136 Manual

VistaLab

VistaLab 1136 Manual

VistaLab MLA Micro D-Tipper Series User manual

VistaLab

VistaLab MLA Micro D-Tipper Series User manual

VistaLab Ovation ESC User manual

VistaLab

VistaLab Ovation ESC User manual

VistaLab Ovation 9060-1011 Guide

VistaLab

VistaLab Ovation 9060-1011 Guide

VistaLab 1070-0002 User manual

VistaLab

VistaLab 1070-0002 User manual

VistaLab Ovation Manual

VistaLab

VistaLab Ovation Manual

VistaLab Ovation QS Manual

VistaLab

VistaLab Ovation QS Manual

VistaLab Ovation 9057-2009 Use and care manual

VistaLab

VistaLab Ovation 9057-2009 Use and care manual

VistaLab Ovation M Use and care manual

VistaLab

VistaLab Ovation M Use and care manual

VistaLab Ovation M Use and care manual

VistaLab

VistaLab Ovation M Use and care manual

VistaLab Ovation M User manual

VistaLab

VistaLab Ovation M User manual

VistaLab MLA pipette MLA 1061 Manual

VistaLab

VistaLab MLA pipette MLA 1061 Manual

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.