1
2
SkinPad
ELECTRODE APPLICATION
ELECTRODE PLACEMENT
SKIN PREPARATION
Anterior-Anterior (Apex/Lateral-Sternum)
OPERATING TEMPERATURE:
SHORT TERM STORAGE TEMPERATURE:
LONG TERM STORAGE TEMPERATURE:
0°C to 50°C (32°F to 122°F)
-30°C to 65°C (-22°F to 149°F)
0°C to 35°C (32°F to 95°F)
Atrial Cardioversion
Anterior / Posterior Electrode Placement
(Recommended)
Recommended for defibrillation, ventricular cardioversion and ECG monitoring only. Not
recommended for noninvasive pacing. Noninvasive pacing with Anterior-Anterior electrode
placement can lead to decreased patient tolerance and increased capture thresholds.
Sternum:
Align the CPR sensor with the sternal notch. Grasp the
Sternum electrode and peel away the plastic liner.
Apply on the patient's upper right torso.
Placement of Apex electrode varies slightly in
anterior-anterior configuration. The more lateral
placement increases the likelihood that more of the
heart musculature will be within the current pathway.
Apex/Lateral:
Grasp the Apex/Lateral electrode at the bottom and peel
away the plastic liner. Apply so that the top of the gel
treatment area aligns with the bottom of the pectoral
muscle on a male patient. Position electrode under the
breast on a female patient.
Avoid any contact between nipple and gel
treatment area. Skin of the nipple area is more
susceptible to burning.
Back (Posterior):
Grasp the electrode at the wire exit and peel away the
plastic liner. Place to the left of the spine just below
the scapula at the heart level.
Front:
R2137-02 Rev. E
TM
OneStep
INDICATIONS FOR USE
Defibrillation Noninvasive Pacing Cardioversion ECG monitoring
PRECONNECTING THE ELECTRODES
Instructions for Use
CPR Feedback
Grasp the electrode and peel away the plastic liner.
Apply to upper right torso on the third intercostal space,
midclavicular line. If CPR sensor is on the electrode,
align the CPR sensor with the sternal notch.
CPR A/A Electrodes with Green Connector
For use on adult patients with ZOLL®R Series
®
and 8009-0020 connector, E Series
®
and 8000-0370 connector, and AED Pro
®
with
software version 2.86 or higher by trained personnel including Physicians, Nurses, Paramedics, Emergency Medical Technicians and
Cardiovascular Laboratory Technicians. The OneStep Adult electrodes are not indicated for use on a patient less than 8 years of age
or weighing less than 55lbs (25kg).
BACK
Anterior-Posterior (Apex/Front-Back)
Back (Posterior): Front:
APEX / FRONT
Separate CPR device from the sternum pad.
Grasp the Back/Sternum electrode at the red
tab, peel from the package liner. Place to the
left of the spine just below the scapula at the
heart level.
Always apply back electrode first.
If front electrode is already in place when
patient is being maneuvered for placement of
the back, the front may become partially lifted.
This could lead to arcing and skin burns.
Place CPR sensor in center of chest and
allign with sternal notch.
Grasp the Apex/Front electrode at the red
tab, peel from the package liner. and
apply over cardiac apex with the nipple
under adhesive area on a male patient.
Position under breast on a female patient.
Avoid any contact between nipple and gel
treatment area. Skin of the nipple area is
more susceptible to burning.
BACK FRONT
APEX/LATERAL
STERNUM
Recommended for defibrillation, noninvasive pacing, ventricular cardioversion, and ECG monitoring.
Optimal for noninvasive pacing because it increases patient tolerance and decreases capture thresholds.
Side 1 of 2
1. Do not open until ready to use.
2. Periodically inspect electrode packaging for integrity & expiration date.
3. Attach electrode connector to AED Pro or ZOLL Multi-Function cable and connector.
4. Open electrode package by pulling apart at yellow arrow.
5. Self test connection disengages after the electrode has been removed from the liner.
1. Remove excess chest hair. Clip if necessary to maximize gel to skin contact. Clipping is recommended
since shaving can leave tiny microabrasions that can lead to patient discomfort during pacing.
2. Ensure skin is clean and dry under electrode. Remove any debris,ointments, skin preps, etc.
with water (and mild soap if needed).Wipe off excess moisture/diaphoresis with dry cloth.
Instructions
Excessive hair can inhibit good
coupling (contact), which can
lead to the possibility of arcing
and skin burns.
1. Apply one edge of the electrode securely to the patient.
2. “Roll”the electrode smoothly from that edge to the other. Be careful not to trap
any pockets of air between the gel and skin.
Instructions
Poor adherence and/or air
under the electrodes can lead
to the possibility of arcing and
skin burns.
Rx ONLY
NOT MADE WITH
NATURAL RUBBER
LATEX.
32°F
0°C
50°C
122°F
Side 2 of 2
Noninvasive Temporary Pacing
1. After patient movement due to muscle contraction or patient
repositioning, press electrodes to skin to ensure good coupling
between electrodes and skin.
2. Do not conduct manual chest compressions through the
electrodes. Doing so may cause damage to the electrodes that
could lead to the possibility of arcing and skin burns. For
electrodes with the CPR sensor, place hands directly on the CPR
sensor when conducting chest compressions.
3. Electrodes should be replaced after 24 hours of use or 8 hours of
pacing to maximize patient benefit.
4. To avoid electrical shock, do not touch the electrodes, patient, or
bed when defibrillating.
5. Always apply electrodes to flat areas of skin. If possible, avoid
folds of skin such as those underneath the breast or those visible
on obese individuals.
6. Avoid electrode placement near the generator of an internal
pacemaker, other electrodes or metal parts in contact with the
patient.
7. Use only with ZOLL Pacemaker/Defibrillator products.
8. Do not use if gel is dry. Dried out gel can lead to skin burning.
Do not open pouch until ready to use. Do not use electrodes
past the expiration date printed on the pouch label.
9. Do not discharge standard paddles on or through electrodes or place
separate ECG leads under pads. Doing so could lead to arcing and/or skin
burning.
10. Some current generated by electrosurgical units (ESU) may concentrate in
the conductive gel of pacing / defibrillation electrodes, especially if an ESU
grounding pad other than that recommended by the ESU manufacturer is
used. Consult the ESU operator’s manual for further details.
11. Do not fold the electrodes or packaging. Any fold in or other damage to
the conductive element could lead to the possibility of arcing and/or skin
burns.
12. Electrode types equipped with ECG Electrodes embedded in the front pad
produce modified ECG vectors. Attach conventional ECG electrodes to the
patient for diagnostic purposes.
13. Device disposal should follow hospital protocol.
ZOLL International Holding B.V.
Newtonweg 18 6662 PV ELST
The Netherlands Tel. +31 (0) 481 366410
®
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA 01824-4105 USA
800-348-9011 Made in USA
OneStep Pacing is functional with
R Series only and is intended for
emergency pacing only. Electrodes
should be replaced after 24 hours
of use or 8 hours of pacing to
maximize patient benefit.
Transcutaneous pacing longer than
30 minutes may cause burns to the
skin. Periodically check the
electrode site to ensure that the
electrode is well adhered to the
skin. Press over the entire surface
of the electrode to ensure adhesion.
For pacing times in excess of
2 hours, ZOLL recommends
Pro-padz
®
with liquid gel
(part number 8900-2100-01
or 8900-2303-01).
WARNINGS

Other ZOLL Medical Equipment manuals

ZOLL IVTM User manual

ZOLL

ZOLL IVTM User manual

ZOLL 731 Series Quick start guide

ZOLL

ZOLL 731 Series Quick start guide

ZOLL Infusion Dynamics Power Infuser M100B-3A User manual

ZOLL

ZOLL Infusion Dynamics Power Infuser M100B-3A User manual

ZOLL Aed Plus Service manual

ZOLL

ZOLL Aed Plus Service manual

ZOLL AED PLUS Trainer2 Manual

ZOLL

ZOLL AED PLUS Trainer2 Manual

ZOLL Cool Line CL-2295A/8700-0781-01 User manual

ZOLL

ZOLL Cool Line CL-2295A/8700-0781-01 User manual

ZOLL M2 Manual

ZOLL

ZOLL M2 Manual

ZOLL Propaq M User manual

ZOLL

ZOLL Propaq M User manual

ZOLL 330 Manual

ZOLL

ZOLL 330 Manual

ZOLL M Series User manual

ZOLL

ZOLL M Series User manual

ZOLL aed pro User manual

ZOLL

ZOLL aed pro User manual

ZOLL M Series User manual

ZOLL

ZOLL M Series User manual

ZOLL aed pro Manual

ZOLL

ZOLL aed pro Manual

ZOLL PD 1400 Manual

ZOLL

ZOLL PD 1400 Manual

ZOLL M Series User manual

ZOLL

ZOLL M Series User manual

ZOLL Aed Plus Manual

ZOLL

ZOLL Aed Plus Manual

ZOLL aed pro User manual

ZOLL

ZOLL aed pro User manual

ZOLL ResQPUMP User manual

ZOLL

ZOLL ResQPUMP User manual

ZOLL E Series User manual

ZOLL

ZOLL E Series User manual

ZOLL AED PLUS Trainer 2 User manual

ZOLL

ZOLL AED PLUS Trainer 2 User manual

ZOLL CMCT Monitoring System User manual

ZOLL

ZOLL CMCT Monitoring System User manual

ZOLL M Series User manual

ZOLL

ZOLL M Series User manual

ZOLL Aed Plus Instruction Manual

ZOLL

ZOLL Aed Plus Instruction Manual

ZOLL AutoPulse 100 User manual

ZOLL

ZOLL AutoPulse 100 User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.