22. Lift up the foot bolster and place the power cable
in the left cavity.
23. For all bolsters, make sure:
• The bolsters are the same height as the
mattress.
• The print on the top cover is visible.
• All handles on the bolsters are on the outside.
• All buckles are securely attached to the
mattress.
• All locking clips are attached, on both the head
and foot end of the mattress.
• All zip aps are folded down, covering the zips.
24.
Place the patient at the centre of the mattress.
25. Make sure all side rails are up.
Transport mode
The side bolsters and/or the foot bolster might need
to be removed and reassembled during transport, for
example to t in an elevator.
Make sure there is a caregiver present at the open
side of the bed at all times and make sure that the
patient’s arms and feet are clear of the bolsters.
Never leave extended sides open. Retract the foot
end or the side of the bed frame once each bolster
has been removed.
Patient leaving bed
When the patient needs to leave the bed, make sure
there is a caregiver present at the open side at all
times.
WARNING
To avoid risk of falling or entrapment, make
sure that the patient is placed at the centre
of the mattress.
WARNING
To avoid risk of falling or entrapment, make
sure there is a caregiver present on the open
side of the bed frame, when assembling or
removing the bolsters.
WARNING
To avoid risk of falling when a patient is
leaving the bed, make sure there is a caregiver
present on the open side of the bed frame.
08.AAV.03EN_01 • 2023-11 • Page 4/5