
4
HANDPIECE CONT OLS
The Aseptico ADU-12DCEB includes a manual
handpiece control system. The system includes
one highspeed control with water coolant and one
lowspeed control. The handpiece selection switch
allows the user to select highspeed or lowspeed
operation, and the handpiece on/off toggle switch
turns the selected handpiece on or off.
TANK P ESSU E TOGGLE SWITCH
Turns the pressure to the water supply tank on or
off. Turn switch off before removing water supply
tank.
WATE COOLANT ON/OFF TOGGLE SWITCH
Turns water coolant to highspeed handpiece on or
off.
WATE COOLANT FLOW CONT OL
Adjusts the flow of water coolant to the highspeed
handpiece control from off to full spray when the
water coolant on/off toggle switch is in the “ON”
position.
CAVIT ON WATE OUTLET
Provides water via standard 1/4” quick disconnect
to independent Cavitron or scaler units requiring a
water supply.
WATER COOLANT
FLOW CONTROL
WATER COOLANT
ON/OFF
TOGGLE SWITC
P ESSU IZED WATE SUPPLY SYSTEM
The Aseptico ADU-12DCEB incorporates a self
contained pressurized water system. This consists
of a 900 mL white plastic tank dispensing water
through the 3-way syringe and highspeed
handpiece control. Before removing the tank for
refilling, it must be depressurized. Turn off the tank
pressure toggle switch to release pressure in the
water system. Then unscrew the cap from tank
and fill tank with water. Screw cap back onto tank.
Sterilization of tank is recommended between
patient use.
SALIVA EJECTO VACUUM SYSTEM
The Saliva Ejector is powered by the
compressor/vacuum pump and is a self contained
oral evacuation system. It is supplied with a plastic
waste container and a saliva ejector hose.
To operate, attach the saliva ejector hose to the
top of the waste container and insert a standard
saliva ejector tip. To activate the saliva ejector,
depress the foot switch.
Waste from the vacuum system will collect in the
clear plastic waste container. When the waste
container is over 1/2 full, it must be emptied to
prevent a back-up in the system. After each use,
the vacuum system should be cleaned by running
clean water or vacuum system cleanser through
the saliva ejector hose and by thoroughly cleaning
the waste container. CAUTION: Use only NON-
foaming cleansers in the vaccum lines. Use a new
saliva ejector tip for each patient.
ANDPIECE
SELECTION
SWITC
ANDPIECE
ON/OFF
TOGGLE SWITC
CAVITRON
WATER OUTLET
TANK PRESSURE
TOGGLE SWITC
T E WATER SUPPLY TANK IS
PRESSURIZED DURING
OPERATION. DEPRESSURIZE
WATER TANK BY TURNING OFF
TANK PRESSURE TOGGLE
SWITC BEFORE REMOVAL.
WA NING:
The waste container has a ball float shutoff to
prevent accidental overflow. If the waste
container becomes full and the ball float
activates, the vacuum motor may not restart,
and could potentially blow the fuse.
Immediately empty waste container to
continue operation.
NOTE: