Start, Monitor, and Control a Purification
Manually Fractionate
At any time during the gradient run, you can switch to a new
collection vessel by pressing in the chromatogram.
Line Flush, Decompression,
and Detector Flush
At the end of a run, i.e. after the gradient purification stage
is completed, the system performs a line flush (if enabled),
system decompression (purge), and a detector flush (if
enabled). To enable or disable flushes and specify whether
enabled flushes are collected or not, see page 27.
Auto Extend
If the collection criteria are still met (any signal is above
threshold level) when the system reaches the end of a
purification, the system extends the gradient purification
stage of the run with 25% of the total gradient length using
the final conditions of the run.
Pause, End, or Abort a Purification
Note: Pausing a purification may cause gradient inconsistency
due to heat, solvent and sample diffusion, etc.
Pause and Resume a Purification Run
1. To pause the purification in progress, press .
The collection arm returns to its home position (in the
inner right corner) and the system is paused.
2. To resume the run from the point at which it was paused,
press .
End the Equilibration Step or
End or Abort the Purification Run
1. Press . The collection arm returns to its home position
(in the inner right corner) and the system is paused.
2. Press .
3. When the System Paused dialog opens, select one of the
available options:
»
Skip Equilibration: End the equilibration
step and move on to the gradient.
»
Abort: End the purification run without
performing any flushes.
»
Cleanup: End the purification run and perform the
flushes that are enabled in the system settings
and a system decompression (purge).
Edit a Purification
Edit on the Fly
The following run parameters can be changed without pausing
the purification run:
»
The run comment in the Information panel.
»
The flow rate in the Column panel.
»
All collection parameters in the Collection panel
except for the UV baseline correction; see “Specify
the Collection Parameters” on page 11.
»
The gradient except for adding new solvent combinations.
»
The modifier percentage.
All changes except for changes to the gradient are instant.
When editing the gradient, the modified gradient is purple and
the ongoing gradient is black (see Figure 32). The modified/
purple gradient is applied when you press .
Figure 32.When editing the gradient while the system is processing,
the modified gradient is purple and the ongoing gradient is black.
The modified/purple gradient is applied when you press .
Pause to Edit a Purification in Progress
With the system paused, it is not only possible to change all the
run parameters that can be changed without pausing (see “Edit
on the Fly” above), but it is also possible to add new solvent
combinations to the gradient and change the Max Fraction
Volume setting in the Rack panel.
1. Press . The collection arm returns to its home position
(in the inner right corner) and the system is paused.
2. Edit the run settings; see “Set Up a Purification” on
page 8. To undo all changes, press .
Note: To add a solvent combination to the gradient,
expand the chromatogram to full size by pressing .
3. To apply the changes and resume the run, press .