Biotage®Extrahera™LV-200 User Manual | © Biotage 2021
Set Up a Method
Method Parameters
Note: For DFE method parameters, see page 17.
Available method parameters are listed in alphabetical order:
»
Advanced pressure settings: Press Edit... to set up a
pressure gradient. For more information, see page 16.
»
Air push after last elution?: Whether to apply positive
pressure with a gas flow of approximately 10 mL/min
after the last elution (Yes) or not (No). SLE only.
»
Air push time (s): The amount of time, in seconds, that
positive pressure with a gas flow of approximately 10 mL/min
will be applied to the extract
ion
plate
or columns to trigger
gravity loading or elution. This field is only enabled for
elution
when Air push after last elution? is set to Yes. SLE
only. Note that this field is disabled when a pressure gradient
is enabled; see “Set Up a Pressure Gradient” on page 16.
»
Clog detection?: GLP feature. For more information,
see the Extrahera GLP User Manual (P/N 417250).
»
Clog settings: GLP feature. For more information,
see the Extrahera GLP User Manual (P/N 417250).
»
Collect in position: In which position the
collection platerack or the flow-through plate
(waste) is to be loaded onto the carousel.
»
Collection plate height (%): How high, in percentage of the
maximum height, the collection
plate
is moved up by the
lifter. See Table 3 on page 16 for guidelines. Filtration only.
»
Conditioning solvent: The solvent to be used for
pipette tip conditioning. SPE only. This field is only
visible when Tip conditioning? is set to Yes.
Note: The solvent must include parameters for the pipette tip
it will be used with; see “Manage Solvents” on page 6.
»
Dispose solvent tips?: Whether the pipette tips
will be disposed of after the solvent load (Yes)
or reused (No). PPT/PLD (solvent first) only.
»
Dispose solvent tips after each step?: Whether the pipette
tips will be disposed of after each step (Yes) or reused (No).
Note that it is not possible to reuse a pipette tip after sample
dispensation. Not available for Filtration and PPT/PLD.
»
Extraction media: The type of extraction
plate
or columns.
»
Method comment: Optional information about the method.
»
Method name: The name of the method.
»
Mix number of times: The number of mixing iterations when
mixing the sample with a solvent following pretreatment
(except for PPT/PLD). Not available for Filtration.
»
Mix volume (µL): The aspiration volume when mixing
the sample with a solvent following pretreatment
(except for PPT/PLD). Not available for Filtration.
»
Move pretreatment step: Use the arrow buttons to move
the pretreatment operation to the desired position,
before conditioning (default) or before load. SPE only.
»
Number of steps: The number of steps the operation
contains. Not available for Filtration and PPT/PLD.
»
Number of times: The number of mixing iterations
in the sample plate. This field is only enabled when
Premix? is set to Yes. Not available for Filtration.
»
Pause after this step/last step/each load? Whether
to pause after a certain step or operation (Yes) or not
(No). When the system is paused, it is possible to open
the door to, for example, check for clogged wells or
columns. Not available for Filtration and PPT/PLD.
»
Plate dry after last elution?: Whether to remove the last
residual solvent from the sorbent bed using positive
pressure at 5 bar with a gas flow of approximately
600 mL/min after the last elution (Yes) or not (No). SPE only.
»
Plate dry after last wash?: Whether to remove the last
residual solvent from the sorbent bed using positive
pressure at 5 bar with a gas flow of approximately
600 mL/min (Yes) or not (No). SPE only.
»
Plate dry time (s): The amount of time, in seconds, that
positive pressure at 5 bar with a gas flow of approximately
600 mL/min will be applied to remove the last residual
solvent from the sorbent bed. This field is only enabled when
Plate dry after last elution/wash? is set to Yes. SPE only.
»
Positive pressure time (s): The amount of time, in seconds,
that positive pressure will be applied to the extraction
plate
or columns. Not available for SLE. Note that this
field is disabled when a pressure gradient is enabled;
see “Set Up a Pressure Gradient” on page 16.
»
Premix?: Whether the sample will be premixed (Yes)
or not (No) before it is loaded into the extraction
plate
or columns. Not available for Filtration.
»
Pressure (bar): The pressure, in bar, that will be applied to
the extraction
plate
or columns. Not available for SLE. Note
that this setting will not be used when a pressure gradient is
enabled; see “Set Up a Pressure Gradient” on page 16.
»
Repeat (number of times): The number of times
this step will be repeated. SPE and SLE only.
»
Reuse tips for pretreatment mix and sample load?: Whether
to reuse a sample pipette tip when mixing following
pretreatment, when loading a sample in aliquots, or when
rinsing the wells/tubes in the sample plate (Yes) or not (No).
Not available for Filtration, PPT/PLD, and Hydro DME+.
»
Rinse repeat number of times: The number of rinsing
iterations in the sample plate. This field is only
enabled when Premix? is set to Yes. SPE only. This
field is only visible when Rinsing? is set to Yes.
»
Rinse solvent: The solvent to be used to rinse the
wells/tubes in the sample plate. SPE only. This field
is only visible when Rinsing? is set to Yes.
Note: The solvent must include parameters for the pipette tip
it will be used with; see “Manage Solvents” on page 6.