SPE Dry 96 and SPE Dry 96 Dual User Manual | © Biotage 2017
Operation
Once you have attached the gas supply and power cord as
described in “Installation” on page 3, the system is ready
for operation. Please take precautions when operating this
system, including the use of safety glasses, as consistent with
safe laboratory practice.
Plate Clearance and Adjusting Plate
Place the collection plate or plates (when using SPE Dry 96 Dual)
on the lower head of the system, making sure that it is/they
are properly seated in the recessed area. The needle assembly
is reversible for deep or shallow well plates. Choose the side
that best fits your plate's well height. Loosen the knurled knob,
located on the right side of the system, and raise the collection
plate or plates into the needles with the sliding arm to the
appropriate depth. Set the depth of the needles 3 to 4 mm above
the top of the sample. When the appropriate plate height is
reached, lock the plate in position by tightening the knurled knob.
For large volume assays where many plates of the same size
and volume will be processed, utilize the plate height stops
(located on the left front side of the system) to avoid plate-to-
plate sample contamination and damage to the needles. Simply
move the plate height stop pin to the location that approximates
the ideal plate height as determined above. Subsequent plates
can be raised until the stop is reached, and then locked into
position to maintain the appropriate plate height.
Figure 5.The plate height stop pin.
Adjusting Gas Flow Rate
Before applying inlet gas pressure, turn the flow meter knobs
clockwise until they stop to ensure that the flow meters are
closed. Set the pressure regulator on the inlet gas supply to
50 psi (SPE Dry 96) or 60 psi (SPE Dry 96 Dual). Slowly open the
knobs on the flow meters by turning them counterclockwise.
The flow meter bubbles will rise and then level off. The gas
flow rate is determined by placing the middle of the flow meter
bubble adjacent to the desired flow rate value. Remember to
turn on the gas flow before turning on the system power switch.
Adjust the gas flow to the desired rate for your method. A good
starting point for the flow rate of the upper needle assembly
is 50 L/min (SPE Dry 96) or 80 L/min (SPE Dry 96 Dual) and
25 L/min for the lower needle assembly. Optimum drying results
can usually be obtained by maintaining the 25 L/min flow rate
for the lower needle assembly. This allows the lower assembly
to be heated while redirecting the majority of the airflow across
the sample via the upper needle assembly.
Temperature Control
Turn the power switch on at the lower left side near the rear of
the SPE Dry 96 system or the lower rear near the left side of
the SPE Dry 96 Dual system. The temperature controllers will
illuminate, the system will perform a short self-diagnostic,
and within a few seconds the lower value will indicate the user
set point and the upper value will indicate the temperature
of the gas exiting the needles. Choose the desired operating
temperature by pressing the up or down arrows. Allow
approximately 15 minutes for the system to equilibrate before
beginning sample processing.
The small
Control
indicator on the face of the temperature
controller is factory set and calibrated. The
Control
indicator
shows when power is being supplied to the heating elements.
The small
Alarm
indicator has been deactivated, as its function
is not required for the operation of the system. Therefore,
please disregard the
Alarm
indicator.
The circular arrow keys on the temperature controllers are used
to calibrate the system and change internal electronic settings.
These keys have been deactivated on the keypad to avoid
accidental changes of these settings.
All systems are factory set and calibrated for temperature and
flow rate. If there is a suspected problem with these functions,
please contact Biotage®1-Point Support™for service.
Note: The small red indicator light on the upper right face of
the system should not be illuminated. This light indicates that
internal diagnostic systems have shutdown the system due to an
overtemp with the heaters. If this situation is encountered, turn
off the power switch and let the system cool down. Then turn the
power switch back on to reset the temperature controllers.
Operation