3
Intended use
The intended use of the activ L instruments are to
allow successful implantation of the Aesculap activ L
device. These instruments aid in selecting, placing, and
revising the activ L device. The activ L device comprises
of superior and inferior endplates that are offered in
four footprint sizes (S, M, L, and XL), two footprint
shapes (Normal and S1), and two fixation options
(spiked and keeled) as well as a polyethylene core that
results in 4 possible implant heights, (8,5 mm, 10 mm,
12 mm, and 14 mm). This instructions for use docu-
ment refers to these variables, and the instruments
themselves are clearly labelled, but additional infor-
mation regarding use can be found in the surgical
technique guide.
activ L parallel distractor FW970
The activ L parallel distractor is used for distraction
and determining the correct size and height of the
intervertebral disk prosthesis.
activ L handle (for chisel guide) FW980R,
double chisels FW985R–FW988R, single
chisels FW989R–FW992 & chisel guides
FW981R–FW984R/FW989R-FW996R
activ L chisels and chisel guides are used to prepare the
keel bed in one or two vertebral bodies for anchoring
the activ L disk prosthesis with keel.
A double or single chisel is required depending on
whether the keel or combined version is being used.
activ L insertion instrument FW961R–
FW964R
The activ L insertion instrument is used for implanting
the intervertebral disc prosthesis through anterior (0°)
approach. The appropriate color-coded and labeled
insertion instrument is used according to the implant
height.
activ L revision instruments and distrac-
tion forceps
The activ L revision instruments and distraction forceps
are used for the distraction/mobilization of two adja-
cent lumbar vertebral bodies or prostheses plates, as
well as for the repositioning and revision of activ L
intervertebral disk prostheses.
activ L angled distraction forceps FW960R
The activ L angled distraction forceps is used for the
distraction/mobilization of two adjacent vertebral
bodies.
activ L angled revision distraction forceps
FW965R
The activ L angled revision distraction forceps is used
for distracting the two activ L metal plates of an
implanted activ L prosthesis in order to remove the PE
inlay in corrective surgery.
activ L revision instrument for PE
inlay FW968R
The activ L revision instrument for PE inlay is used for
the revision of the activ-L PE inlay.
activ L revision instrument for
endplates FW966R/FW967R
The activ L revision instrument for endplates and the
corresponding handle are used for the repositioning
and revision of activ L intervertebral disk prostheses.