Dräger Polytron 8200 User manual

WARNING
Strictly follow the Instructions for Use.
The user must fully understand and strictly observe the
instructions. Use the product only for the purposes specified
in the Intended use section of this document.
!
Dräger Polytron 8200/8310
Instructions for use
i


Contents
Dräger Polytron 8200/8310 3
Contents
1 For your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.1 General safety statements. . . . . . . . . . . . . . . . . . . . .4
1.2 Definitions of alert icons. . . . . . . . . . . . . . . . . . . . . . .4
2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.2 Functional description . . . . . . . . . . . . . . . . . . . . . . . .6
2.3 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.4 Intended operating area and operating conditions. . .7
2.5 Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.6 Device marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
3.1 General information for the installation . . . . . . . . . . .7
3.2 Restrictions on the installation . . . . . . . . . . . . . . . . . .8
3.3 Impedance range of the signal loop. . . . . . . . . . . . . .8
3.4 Mechanical installation. . . . . . . . . . . . . . . . . . . . . . . .8
3.5 Electrical installation without e-Box . . . . . . . . . . . . . .9
3.6 Electrical installation with e-Box. . . . . . . . . . . . . . . .10
3.7 Connecting the device to a Dräger control unit . . . .10
3.8 Connecting the device to a PC . . . . . . . . . . . . . . . .10
3.9 PolySoft 8000 PC software (optional) . . . . . . . . . . .10
3.10 Installing software dongles. . . . . . . . . . . . . . . . . . . .10
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
4.1 Operating fundamentals. . . . . . . . . . . . . . . . . . . . . .11
4.2 Display and LEDs . . . . . . . . . . . . . . . . . . . . . . . . . .11
4.3 Activating the Info mode . . . . . . . . . . . . . . . . . . . . .12
4.4 Switching to the Quick Menu mode . . . . . . . . . . . . .13
4.5 Switching to Menu mode (password protected). . . .13
4.6 Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . .13
5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
5.1 Calibrating the device . . . . . . . . . . . . . . . . . . . . . . .14
5.2 Performing Auto Calibration. . . . . . . . . . . . . . . . . . .15
6 Menu overview . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
7 Information menu. . . . . . . . . . . . . . . . . . . . . . . . . .17
7.1 Device settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
7.2 Sensor information. . . . . . . . . . . . . . . . . . . . . . . . . .17
7.3 Datalogger information
(only with data dongle). . . . . . . . . . . . . . . . . . . . . . .17
8 Settings menu . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
8.1 Switching SIL lock on or off . . . . . . . . . . . . . . . . . . .18
8.2 Device settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
8.3 Communication settings. . . . . . . . . . . . . . . . . . . . . .20
8.4 Sensor settings . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
8.5 Datalogger settings (only with data dongle). . . . . . .24
9 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . .25
9.1 Fault reference. . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
9.2 Warning reference . . . . . . . . . . . . . . . . . . . . . . . . . .26
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
10.1 Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . .27
10.2 Changing the sensor . . . . . . . . . . . . . . . . . . . . . . . .27
10.3 Changing the main electronics. . . . . . . . . . . . . . . . .27
11 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
12 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
12.1 Torques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
12.2 Tightening torque and cable size for field wiring termi-
nals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
12.3 Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
13 Order list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
13.1 Dräger Polytron 8200/83X0 . . . . . . . . . . . . . . . . . . 31
13.2 Sensors for remote versions . . . . . . . . . . . . . . . . . 31
13.3 Replacement sensors
(except for remote versions) . . . . . . . . . . . . . . . . . 31
13.4 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
13.5 Replacement parts. . . . . . . . . . . . . . . . . . . . . . . . . 32
14 Declaration of Conformity. . . . . . . . . . . . . . . . . . 33

4Dräger Polytron 8200/8310
For your safety
1 For your safety
1.1 General safety statements
zBefore using this product, carefully read the Instructions for
Use.
zStrictly follow the Instructions for Use. The user must fully
understand and strictly observe the instructions. Use the
product only for the purposes specified in the Intended use
section of this document.
zDo not dispose of the Instructions for Use. Ensure that they
are retained and appropriately used by the product user.
zOnly fully trained and competent users are permitted to use
this product.
zComply with all local and national rules and regulations
associated with this product.
zOnly trained and competent personnel are permitted to
inspect, repair and service the product as detailed in these
Instructions for Use. Further maintenance work that is not
detailed in these Instructions for Use must only be carried
out by Dräger or personnel qualified by Dräger.
zProperly trained service personnel must inspect and
service this product as detailed in the Maintenance section
of this document.
zUse only genuine Dräger spare parts and accessories, or
the proper functioning of the product may be impaired.
zDo not use a faulty or incomplete product, and do not
modify the product.
zThe flameproof threads for the explosion proof enclosure
do not conform to the minimum/maximum values in EN/IEC
60079-1. The threads must not be modified by the user.
zThe measuring function of the gas measurement
transmitter for explosion protection according to Annex II,
Sections 1.5.5, 1.5.6 and 1.5.7 of Directive 94/9/EG is not
covered at present.
zExchanging components may compromise the intrinsic
safety. This only applies if the device is intrinsically safe.
zOnly operate the product within the framework of a risk-
based alarm signalling concept.
Safe coupling with electrical devices
zElectrical connections to devices which are not listed in
these Instructions for Use should only be made following
consultation with the respective manufacturers or an
expert.
Use in areas subject to explosion hazards
zDevices or components for use in explosion-hazard areas
which have been tested and approved according to
national, European or international Explosion Protection
Regulations may only be used under the conditions
specified in the approval and with consideration of the
relevant legal regulations. The devices or components may
not be modified in any manner. The use of faulty or
incomplete parts is forbidden. The appropriate regulations
must be observed at all times when carrying out repairs on
these devices or components.
1.2 Definitions of alert icons
The following alert icons are used in this document to provide
and highlight areas of the associated text that require a greater
awareness by the user. A definition of the meaning of each
icon is as follows:
WARNING
Indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not
avoided, could result in physical injury, or damage to
the product or environment. It may also be used to
alert against unsafe practices.
NOTICE
Indicates additional information on how to use the
product.
!
!
i
i

Description
Dräger Polytron 8200/8310 5
2 Description
2.1 Product overview
2.1.1 Explosion proof device
1 Housing cover
2 Clip
3 Enclosure with main electronics (and optional relay)
4 Lower part of housing
5 Sensor
6 Assembled device
2.1.2 Explosion proof device with e-Box for increased
safety
1 Housing cover
2 Clip
3 Enclosure with main electronics (and optional relay)
4 Lower part of housing
5 Sensor
6 Feed through cable
7e-Box
8 Assembled device
4
6
5
1
3
2
001233299.eps
4
8
1
6
7
3
5
2
002233299.eps

6Dräger Polytron 8200/8310
Description
2.2 Functional description
2.2.1 Principle of operation of the DrägerSensor DD, LC
The DrägerSensor is a transducer for measuring the partial
pressure of flammable gases and vapors in ambient air.
It operates on the heat of combustion principle.
The ambient air to be monitored diffuses through the flame
arrestor in the sensor. The mixture of flammable gases and
vapors is burned catalytically on the heated measuring element
(pellistor). The oxygen required for combustion is taken from
the monitored ambient air (must be greater than 12 vol.%).
The measuring element receives additional warming from the
heat arising from the combustion. This warming causes
a change in the resistance of the measuring element which is
proportional to the concentration of the mixture of flamable
gases and vapors in the air being monitored.
In addition to the catalytically activated measuring element, the
sensor also has a compensating element. Both elements are
part of a Wheatstone bridge circuit. Consequently, environmental
influences such as changes in temperature or humidity are
almost completely compensated for.
The DrägerSensor LC includes additional processing and
amplification of the measuring signal. As a result, the sensor is
suitable for the measuring range 0 - 10 %LEL.
2.2.2 Principle of operation of the DrägerSensor IR
The DrägerSensor IR infrared gas sensor is a gas transmitter
for determining the concentration of gases and vapors in
ambient air. The measuring principle is based on the
concentration-dependent absorption of infrared radiation in the
measured gases.
The ambient air to be monitored diffuses through a sintered
body into the pressure-tight housing of a measurement
cuvette. The broadband light produced by the emitter passes
through the gas in the cuvette, is reflected from the cuvette
walls and is focused on the entrance window of a dual-element
detector. One channel of the detector measures the gas-
dependent light transmission of the cuvette (measuring
channel) and the other is used as the reference. The gas
concentration in the cuvette can be determined from the ratio
of the measured signal to the reference signal. The cuvette is
heated to prevent condensation of atmospheric humidity.
The calculation of the concentration is performed by internal
electronics and software. For the output signal, the gas sensor
emulates the half bridge of a catalytic bead sensor.
2.2.3 Contaminating gases for the DrägerSensor DD, LC
It may be necessary to check the calibration if the sensor has
been exposed for a long time to a high concentration of
flammable gases or vapors or the above-mentioned
contaminating substances.
2.3 Intended use
The Dräger Polytron 8200/8310 is an explosion proof device for
continuous monitoring of the concentration of gases and vapors
containing hydrocarbons. The housing for the device consists
of robust, non-corroding stainless steel or aluminum and is
suitable for indoor or outdoor applications. The unit can be
connected to a Dräger monitoring system or a programmable
logic controller (PLC) by means of a sealed conduit or an
appropriately approved cable gland. With the optionally
integrated relay module, the device can be operated without a
central controller (with additional local alarm signalling). The
device is designed for fixed installation and is approved for use
in potentially explosive atmospheres, see chapter 2.5 on
page 7.
2.3.1 Dräger Polytron 8200 with DrägerSensor® 1 Ex DD
Catalytic bead sensor for monitoring flammable gases and
vapors containing hydrocarbons
{Detection range: 0 to 100 % LEL 2
2.3.2 Dräger Polytron 8200 with DrägerSensor Ex LC
Catalytic bead sensor for monitoring flammable gases and
vapors containing hydrocarbons
{Detection range: 0 to 10 % LEL 2)
NOTICE
Certain substances in the atmosphere to be measured
can impair the sensitivity of the sensors.
a) Polymerizing substances such as acrylonitrile,
butadiene and styrene
b) Corrosive compunds such as halogenated
hydrocarbons (which release halogens such as
bromine, chlorine or fluorine on oxidation) and halogen
hydracids, as well as gaseous acidic compounds such
as sulfur dioxide and oxides of nitrogen.
c) Catalyst poisons such as sulfur and phosphorus
compounds, silicon compounds (particularly silicones)
and organometallic vapors. The sensors used work with
measuring elements (pellistors) that are poison-resistant
(PR) and thus have a longer life than conventional
sensors when catalyst poisons are present.
WARNING
Not tested in oxygen enriched atmospheres
(>21% O2). High off-scale readings may indicate an
explosive concentration.
1 DrägerSensor®is a registered trademark of Dräger.
2 Lower Explosion Limit, depending on substance and the locally
applicable regulations.
i
i
!

Installation
Dräger Polytron 8200/8310 7
2.3.3 Dräger Polytron 8310 with DrägerSensor IR
Infrared sensor for monitoring flammable gases and vapors
containing hydrocarbons
{Detection range: 0 to 100 % LEL 2)
2.4 Intended operating area and operating
conditions
Hazardous areas classified by zones:
The device is only designed for use in Zone 1 or Zone 2
potentially explosive atmospheres for which a temperature
range as specified on the device must be observed and in which
gases of Explosion Groups IIA, IIB or IIC and Temperature
Classes T4 or T6 (depending on maximum ambient
temperature) or dust of Groups IIIA, IIIB or IIIC may occur.
Hazardous areas classified by divisions:
The device is only designed for use in potentially explosive
atmospheres of Classes I&II, Div. 1 or 2, for which
a temperature range as specified on the device must be
observed and in which gases or dust of Groups A, B, C, D or
E, F, G and Temperature Classes T4 or T6 (depending on
maximum ambient temperature) may occur.
2.5 Approvals
See printed version for approval markings.
2.6 Device marking
The device marking is supplied with the device as a copy on
a separate document.
Key to the Serial Number: The third letter of the Serial Number
specifies the year of manufacture: A = 2009, B = 2010, C = 2011,
D = 2012, E = 2013, F = 2014, H = 2015, J = 2016, K = 2017 etc.
Example: Serial Number ARBH-0054: The third letter is B,
i.e. the device was manufactured in 2010.
3 Installation
3.1 General information for the installation
The selection of a suitable mounting location is crucial for
the effectiveness and performance of the entire system. Every
detail of the installation must be thoroughly thought out.
The following must be noted in particular:
zthe local and national rules and regulations for the
installation of gas monitoring systems,
zthe applicable regulations for running and connecting
power and signal cables to gas monitoring systems,
zthe full extent of environmental influences to which the
device will be subjected,
zthe physical properties of the gases and vapors to be
measured,
zthe details of the particular application (e.g., potential
leaks, air movements/flows, etc.),
zaccessibility for required maintenance activities,
zthe geometric of the accessories that are used with the
system,
zall other limiting factors and stipulations that may affect the
installation of the system.
zFor installation without a conduit, an approved cable gland
(e.g. Hawke A501/421/A/¾" NPT or equivalent) must be
used (see chapter 12 on page 28). To increase the RF
interference immunity, it may be necessary to connect the
cable screen to the cable gland and to the control unit.
zThe explosion proof enclosure has three ¾" NPT
connections, which can be used for field wiring, the direct
attachment of a sensor or wiring an external sensor.
Unused openings must be closed off with a plug. For the
correct tightening torques for cable bushings, plugs,
sensors and connectors, see chapter 12 on page 28.
zThe secondary are supposed to be supplied from an
isolating source (does not apply to relay contacts).
zThe optional e-Box has up to four 20 mm connections,
which can be used for field wiring or wiring an external
sensor. The permissible cable diameter is 7 to 12 mm.
zIf the device is installed in locations where ambient
temperatures of over 55 °C prevail, appropriate cables
which are specified for use at temperatures of 25 °C above
the maximum ambient temperature must be used.
zStrip back the insulation on conductors by 5 to 7 mm.
zConnect the cable as shown in Chapter 3.5 on page 9
(shown here also with protective ground) oder Chapter 3.6
on page 10.
zThe connecting wires for the optional relay module must be
selected and fused according to the rated voltages,
currents and environmental conditions.
zWhen stranded conductors are used, an end ferrule must
also be used.
WARNING
Explosive. Not to be used in oxygen enriched
atmospheres. None of the Polytron 8000 transmitters
is certified and approved to be operated in oxygen
enriched atmospheres.
!
DPolytron 8200
ETR 0400
Dräger Safety 23560 Lübeck, Germany
II 2G
II 2D
WARNING: Do not open when energized
I
S
F
S
I
E
A
D
L
C
9N54
0158
Class I, Div 1, Groups A,B,C,D | Class II, Div 1, Groups E,F,G
Class I, Zone 1, Group IIC | T-Code T6/T4 | Type 4X
Supply: 16...30 VDC, 4...20 mA
CAUTION: Do not open cover. Opening
cover with circuits alive provokes risk of
Ignition of Hazardous Atmospheres.
WARNING: To reduce the risk of Ignition
of Hazardous Atmospheres, the conduit
must be sealed within 18’’ of the enclosure.
WARNING: Read Manual before operating.
Gas Detector for Use in Hazardous Locations as to
Fire, Electrical Shock and Explosion Hazards only
Part No: 4544440
Serial No. XXXX-9999
Ex d IIC T6/T4 Gb
Ex tb IIIC T80/130°C Db
-40°C ≤ Ta ≤ +40/+80°C
PTB 11 ATEX 1005X
IECEx PTB 11.0005X
IP6x | P ≤ 5 W
Sample

8Dräger Polytron 8200/8310
Installation
3.2 Restrictions on the installation
zThe device requires a DC voltage between 10 and 30 V.
The minimum supply voltage of 10 V and the cross-section
of the conductors used determine the distance of the
device from the supply or the central controller
(see chapter 12 on page 28). The device is designed for
cables of sizes 12 to 24 AWG (0.2 to 2.5 mm2). Use
a three-wire, shielded cable as a minimum.
zThe device must not be subjected to any radiant heat
(e.g. direct sunlight) as this will result in a temperature rise
above the specified thresholds (see chapter 12 on
page 28). The use of a reflective screen is recommended.
zThe housing is weather resistant and suitable for outdoor
applications. The use of the optional splash guard is
recommended to protect the sensor from water and dust.
zThe device must be installed and operated in an
environment which conforms to the stated specifications
(see chapter 12 on page 28).
zTo insure proper operation of the device, the impedance of
the 4 to 20 mA signal loop must not exceed 500 ohms.
Depending on the operating voltage and according to the
application (e.g. HART operation), certain minimum
impedances must be adhered to (see chapter 3.3 on
page 8). The conductors for the power supply must have
an adequately low resistance to insure the correct supply
voltage at the device.
3.3 Impedance range of the signal loop
Devices with a HART interface can be operated with HART
communication or in HART Multidrop mode.
3.4 Mechanical installation
zUse the supplied drilling template for mounting on a wall.
zThe mounting surface should be even and free of sharp
edges.
zDräger recommends using M6 Allen bolts.
zThe openings must be readily accessible to the surrounding
atmosphere.
NOTICE
A dust cap can be fitted to the cable entry of the device.
This cap is only used for transport purposes and must
be removed before the device is connected to a sealed
conduit.
i
i
Operating mode Impedance range of the
signal loop Supply
voltage range
Operation
without HART
communication
0 to 230 Ωat 10 V DC
Rising linearly with the
supply voltage from:
0 to 230 Ωat 10 V to
0 to 500 Ωat 16 V
10 to 18 V DC
0 to 500 Ω18 to 30 V DC
Operation with
HART
communication
(HART mode)
230 to 270 Ωat 13 V DC
Rising linearly with the
supply voltage from:
230 to 270 Ωat 11 V to
230 to 500 Ωat 16 V
11 to 16 V DC
230 to 500 Ω18 to 30 V DC
HART Multidrop
operation 230 to 500 Ω10 to 30 V DC

Installation
Dräger Polytron 8200/8310 9
3.5 Electrical installation without e-Box
Connection diagram for operation as a current source
Connection diagram for operation as a current sink
3.5.1 Power and signal cables
1. Release the set screw (1)
and unscrew the housing
cover from the device.
2. Lift the handle and remove
the enclosure with the main
electronics.
3. Connect the device
electrically to the protective
ground.
4. Rotate the main electronics
and disconnect the 5-way
plug.
5. Connect the three cables for the power supply and the
signal to the corresponding terminals (see following table
of connections for the 5-way plug).
Connections for 5-way plug (power supply and signal):
6. For operation without a central controller: connect Terminal 3
to Terminal 4.
7. Fit the plug back in the socket and tighten the screws.
8. The screen of the cable must only be connected to the
control unit.
9. Insert the main electronics back in the housing.
10. Screw the cover on again and tighten the set screw.
3.5.2 Version with relays
When the relay module is fitted, the wires for the alarm device
are connected to the 9-way plug.
1. Disconnect the 9-way plug on the rear of the main
electronics.
2. Connect the wires for Alarm Relay 1, Alarm Relay 2 and the
Fault relay to the corresponding terminals, (see following
table of connections for the 9-way plug).
3. The relays are energized when the default settings are used
and in the measurement mode. This insures "fail-safe"
operation. The terminal assignments in the connection table
below apply when the default settings are used and when in
measuring mode (see chapter 4.6 on page 13).
4. Fit the plug back in the socket and tighten the screws.
5. Fit the enclosed protective cover over the plug and secure
it with cable ties if necessary.
CAUTION
First connect the cables for the relays and make
the connections to the sensor before connecting the
device to the power supply.
!
01133300.eps
VDC
+
-
mA
01033300.eps
VDC
+
-
mA
00933300.eps
1
5-way plug (power supply and signal)
Terminal 1 2 3 4 5
Ident. PWR+ PWR- PWR- 4 to 20 mA signal PE
Function V+ V- V- 4 to 20 mA signal
(Operation as
source or sink)
PE
WARNING
At voltages >30 V AC or >42.4 V DC, the relay cables
must be enclosed in protective tubing, or double-
insulated cables must be used.
NOTICE
For clear and far-ranging detection of a device fault,
an alarm device should be switched by the fault relay.
!
i
i

10 Dräger Polytron 8200/8310
Installation
Connections for 9-way plug (relays):
(NO = Normally open, NC = Normally closed, C = Common)
3.5.3 Remote Sensor
The remote junction box (Order Number 45 44 098 for
stainless steel / Order Number 45 44 099 for aluminum)
permits the installation of the sensors remote from the device
electronics. This installation enables operation of the device in
difficult to access or unfavourable positions (see the
Instructions for Use for the Dräger Polytron 5000 Junction
Box). The maximum permissible distance between the
Polytron 8200/8310 and the sensors is 30 m.
3.6 Electrical installation with e-Box
zInstall the e-Box in accordance with the Installation
Instructions for the Polytron 5000/8000 e-Box.
3.7 Connecting the device to a Dräger
control unit
Electrical connections to the control unit
zConnect the screen for the wires to the grounding point in
the control unit (e.g., chassis, grounding rail).
3.8 Connecting the device to a PC
The separately available Polytron 5000/8000 IR connection kit
is intended for use with the Polytron 5000/8000 and enables
communication between the Polytron 5000/8000 and a PC.
3.9 PolySoft 8000 PC software (optional)
The PolySoft 8000 PC software is used to display device
information, edit configuration settings and read out from the
measurements memory.
3.10 Installing software dongles
The following SW dongles are available for the Polytron 8200:
To install a SW dongle:
1. Switch off the power to the unit or declassify the area
according to the local regulations.
2. Release the set screw and unscrew the housing cover from
the device.
3. Lift the handle and remove the enclosure with the main
electronics.
4. Insert the dongle in the slot with the Dräger lettering facing
upwards.
5. Insert the main electronics back in the housing.
6. Screw the cover on again and tighten the set screw.
The dongle is installed.
7. Switch on the power.
9-way plug (relays)
Fault relay A2 relay A1 relay
Terminal123456789
Ident. FLT
NO FLT
CFLT
NC A2
NO A2
CA2
NC A1
NO A1
CA1
NC
NOTICE
The e-Box can be pre-installed, wired and closed off
with the supplied cover. As soon as the installation is
ready for commissioning, the device is connected to
the e-Box and put into operation. This prevents the
device from being damaged during the construction
phase.
NOTICE
Information on connection can be found in the
Instructions for Use accompanying the particular
Dräger control unit (e.g. Regard, QuadGard).
i
i
i
i
Datalogger dongle
83 17 618
Enables the datalogger and the
graphic display of concentrations.
Sensor test dongle
83 17 619
Enables the sensor self test
(for certain versions only)
Diagnostic dongle
83 17 860
Enables the sensor self test,
the display of the remaining
sensor activity and the sensor
diagnostic function
(for certain functions only)
Combi-dongle (datalogger and sensor test dongle)
83 23 218
Combi-dongle (datalogger and diagnostic dongle)
83 23 219
00833300.eps

Operation
Dräger Polytron 8200/8310 11
4 Operation
4.1 Operating fundamentals
4.1.1 Menu navigation
It is possible to scroll throught he menu items by tapping the
and symbols with the magnetic wand (Part Number
4544101, blue). confirms a function.
Graphic symbols facilitate navigation in the various menus.
4.1.2 Changing parameter values or the parameter status
1. Select the desired menu item with and .
2. Confirm the menu item with .
The current value or status will be displayed.
3. Using and , the value of a numerical parameter can
be changed or preset values can be selected.
4. As soon as the desired value or the selection is displayed,
tap to validate/confirm the new parameter.
5. Tap to access the next higher level.
zIf exiting using Back to menu or Previous, all changes will
be discarded.
4.1.3 Exiting menus
zTap Back to measurement and to exit the menu.
4.2 Display and LEDs
In measurement mode, the current gas concentration, the name
of the gas and the measurement units are shown on the display.
The green LED lights.
In addition, the following symbols may be displayed:
z"↑↑↑↑", if the measurement range is exceeded
z" - - - - " and " X" in the event of a malfunction. The yellow
LED lights, (see chapter 4.2.2 on page 12).
z"SIL", when SIL status is activated.
When the optional relay PCB is used:
zIf the first alarm has been triggered, the red LED gives
single flashes.
zIf the second alarm has been triggered, the red LED gives
double flashes.
NOTICE
The housing cover should be fitted when using the
magnetic wand. If the housing cover is not fitted,
the magnetic wand might activate two or more buttons
at the same time (crosstalk).
Together with the text » Back «, » Menu «, etc., exit the
menu or go back 1 step.
Closed folder
Further functions or sub-menus are located under this item.
Opened folder
The functions and sub-menus available here are listed
under this item.
Function
When activated, functions can be executed in one or
more steps.
Selection activated
For functions that can be selected and activated,
the activation is initiated by taping » «.
List closed above / complete
There are no further functions, menus or sub-menus
listed above.
List can be scrolled upwards
There are further functions, menus or sub-menus
listed above.
00733300.eps
M
OK
i
i
OK
List closed below / complete
There are no further functions, menus or sub-menus
listed below.
List can be scrolled downwards
There are further functions, menus or sub-menus
listed below.
Next
Perform context-related action.
Number / total number of steps (screens) within
the function.
Entry
Enter data with » « and » «.
3/3
2/2
M
M
OK
M
OK
OK
OK

12 Dräger Polytron 8200/8310
Operation
4.2.1 Special symbols
The following special symbols, which indicate the device state,
may be displayed on the right hand side in measurement
mode.
4.2.2 LED symbols
When the optional relay PCB is used:
zIf the first alarm has been triggered, the red LED gives
single flashes. A1 relay is activated.
zIf the second alarm has been triggered, the red LED gives
double flashes. A2 relay is activated.
zIf an alarm can be acknowledged and is acknowledged,
the LED stops flashing and lights continuously instead until
the alarm condition is no longer present. The corresponding
relay is deactivated.
4.3 Activating the Info mode
The Info mode is used to present information about the central
device settings and states.
zTap and hold for 3 seconds.
The device information appears on several displays.
No changes can be made.
zTap or briefly to switch between the displays.
zThe Info mode can be ended at any time by selecting .
zIf no button is selected within 30 seconds, the device
reverts automatically to the measurement mode.
Example of the Info mode
(Displays vary depending on sensor):
1. Display - Device information:
zDate and Time
zSoftware version
zPart number of the device
zSerial number of the device
zDevice code
2. Display - Sensor information:
zSensor description
zPart number of sensor (only for certain sensors)
zSerial number of sensor (only for certain sensors)
zSoftware version (only for certain sensors)
zSoftware revision (only for certain sensors)
3. Display - Sensor configuration:
zGas description
zMaximum measurement limit and measurement units for
the sensor
zMeasurement range and measurement units for the analog
interface. Only displayed when the 4-20 mA interface is
enabled.
Warning message present –
to display warnings, see chapter 7.1.1 on page 17.
Information can be called up in Info mode, see chapter 4.3
on page 12.
Fault message present –
To display the fault, see chapter 7.1.2 on page 17
Maintenance signal will be output on the interface,
see chapter 8.3.2 on page 21.
Measurement range of analog interface exceeded
Measurement less than range of analog interface
Analog interface is set to a fixed value (e.g. Multidrop HART
communication) and is not transmitting any measurements.
"Preventive" maintenance: The sensor is ready for operation
"Preventive" maintenance: The sensor is ready for
operation but is close to the end of its life cycle.
"Preventive" maintenance: The sensor is still ready for
operation but should be changed as soon as possible.
The datalogger is active in rolling mode.
To activate/deactivate, see chapter 8.5.2 on page 24.
The datalogger is active in stacking mode.
To activate/deactivate, see chapter 8.5.2 on page 24.
SIL
SIL is activated (observe Safety Manual - Order Number
9033307).
Symbol LED Description
red Alarm triggered
yellow Malfunction / Warning
green Switched on
Measurement mode
NOTICE
The alarm triggering function is only available when
the optionally integrated relay module is used.
i
i
M
OK
A001.eps
07.11.2003 12:34
SW Version
: 1
Part No.
: 8317778
Serial No.
: ARUA0001
DeviceCode: 00006317
Instrument Info
1/3
A002.eps
2/3
Sensorname: O
2
Part No.
: 6809630
Serial No.
: XXXXXXX
EEPROM Typ
: 1
EEPROM Vers.: 1
Sensor Info
A003.eps
3/3
Gasname : O
2
Range : 25.00 Vo
4–20 SP : 25.00 Vo
Alarm A1 : 19.00 Vo
Alarm A2 : 23.00 Vo
Sensor Config.

Operation
Dräger Polytron 8200/8310 13
zA1 Alarm limit and measurement units1
zA2 Alarm limit and measurement units1
4.4 Switching to the Quick Menu mode
zTap and hold for 1 to 2 seconds.
The Quick Menu will open. This menu is displayed without
any password request. No changes can be made.
The selected information about status and device settings
can be retrieved (e.g. warning messages, error messages,
installed modules and calibration data).
4.5 Switching to Menu mode
(password protected)
1. Tap and hold for longer than 3 seconds.
The password request appears.
2. Enter the password and confirm.
The screen for password entry appears.
3. Set the password and confirm the entry.
The menu hierarchy corresponding to the entered password
will open.
zAn error message appears if the entered password is invalid.
Password presets:
Password for the Calibration menu: _ _ _ 1
Password for the Settings menu: _ _ _ 2
4.6 Measurement mode
zSwitch on the power.
{The device runs through a start-up sequence (LCD test
by inverting the video, LED test, software version and
initialization) and starts the warm-up phase. The display
indicates that the sensor will be ready to operate in
hh:mm:ss (the countdown begins) and the Polytron
8200/8310 sends the Maintenance signal.
{After the warm-up phase, the device passes into the
measurement mode.
{The current gas concentration, the selected gas and
the units of measure appear on the display.
{The LED lights green.
{The sensor is ready to take measurements.
4.6.1 Analog signals
In measurement mode the output current of the device is
between 4 and 20 mA and is proportional to the measured gas
concentration.
The device uses various current values to indicate different
operating modes. This conforms to the NAMUR NE43
recommendation. The 4-20 mA interface on the device can be
customized and configured to individual requirements.
The device is provided with suitable standard settings at the
factory (see chapter 12.3 on page 29).
1 Only displayed when relay is configured.
WARNING
The configuration and calibration settings must be
checked before the device is operated in measurement
mode.
NOTICE
The correct date and time settings are important for
many functions (see chapter 8.2.3 on page 19).
M
M
!
i
i

14 Dräger Polytron 8200/8310
Calibration
5 Calibration
Repeat the calibration of the device at regular intervals as
specified in the particular sensor data sheet.
5.1 Calibrating the device
1. Connect the pressure reducer to the calibration gas cylinder.
2. Attach the calibration adapter to the sensor.
3. The gas flow should be between 0.5 and 2.0 L/min.
4. Connect the hose to the calibration adapter.
1 Pressure reducer
2 Calibration adapter
3 Test gas cylinder
4Hose
For DSIR only (see also Instructions for Use for
DrägerSensor IR)
1. Position the magnetic wand on the sensor button marked
with and hold it there for at least 5 seconds.
A negative value will be displayed for as long as the
magnetic wand is held on the button (e.g. -10 %LEL).
This will start the automatic zeroing of the sensor.
2. Remove the magnetic wand.
The sensor ends the automatic zeroing after approx.
30 seconds. A negative value (e.g. -10 %LEL) is displayed
again as confirmation.
After the zero-point calibration of the DrägerSensor IR,
the instrument should show 0 %LEL in the display. If this is not
the case, a zero-point calibration must be performed on the
instrument (see chapter 5.1.1 on page 14).
5.1.1 Zero point calibration
1. Select Calibration > Zero Calibration and confirm.
The Maintenance signal is transmitted.
The message Apply zero gas appears.
2. Apply zero gas (synthetic air or nitrogen) to the sensor.
3. Select Next and confirm.
The current value is displayed.
After the displayed value is stable:
4. Set the value to 0 and confirm.
The message Please wait… appears.
The actual value is displayed.
zActual value within the permissible range:
{Select Next and confirm.
{Shut off the test gas and remove the calibration adapter
from the sensor or disconnect the hose.
CAUTION
Danger to health! Do not inhale the test gas.
Observe the hazard information in the corresponding
safety data sheets. Provide venting into a fume
cupboard or outside the building.
Always calibrate the zero point first before the
sensitivity. Otherwise the calibration will be faulty!
NOTICE
Calibration is not possible if the date and time are not
set (see chapter 8.2.3 on page 19).
Dräger recommends calibrating devices with the gas
that is to be measured operationally. This method of
measured gas calibration is more accurate than a
surrogate gas calibration. A surrogate gas calibration
may only be performed as an alternative if a measured
gas calibration is not possible.
!
i
i
00333299.eps
1
4
3
2
DSIR
DD
LC
NOTICE
Ambient air can also be used instead of nitrogen or
synthetic air to zero the sensor, provided it is insured
that the ambient air contains no measured gas or any
other gas for which the sensor may exhibit cross-
sensitivity (see the details in the sensor data sheet).
In this case, neither a gas cylinder nor a calibration
adapter is required for the zero point calibration.
i
i

Calibration
Dräger Polytron 8200/8310 15
5.1.2 Span calibration
1. Check which gas category is set (only for DrägerSensor IR,
see also the Instructions for Use for the DrägerSensor IR).
1. Select Calibration > Span calibration and confirm.
2. Apply calibration gas (e.g., 50 % LEL methane).
The current value is displayed.
After the displayed value is stable:
3. Set the value to 50 and confirm.
4. Shut off the test gas and remove the calibration adapter
from the sensor or disconnect the hose.
5.2 Performing Auto Calibration
The device offers the option of an automatically sequenced
calibration. This reduces the number of otherwise manual
interactions with the device. The auto calibration is only
suitable for experienced users, as calibration errors can occur
here as a result of tightened procedures.
1. Select Settings > Sensor > Autocal. set. to enable or
disable the function.
2. Supply fresh air or feed nitrogen or compressed air at
a flow rate of 0.5 - 2.0 L/min to the sensor.
3. Select Calibration > Auto cal. and confirm.
{Please wait... is displayed and a calibration of the zero
point is performed automatically.
{After a successful zero point calibration, the span
calibration is started.
{The calibration gas concentration, the units and, where
appropriate, the calibration gas are displayed and can
be changed.
4. Start the calibration with Next; Back to menu aborts the
calibration.
5. Feed the relevant calibration gas to the sensor at a flow
rate of 0.5 - 2.0 L/min.
{When the measured value is stable, the device performs
the calibration automatically.
{After a successful calibration, the measurement and
the message Value OK? are displayed
6. When the measurement is stable, shut off the calibration
gas and wait until the measurement is below any possible
alarm thresholds again.
zThe calibration is performed again with Redo.
zComplete the calibration with Accept value.
The device returns to the measurement mode.
NOTICE
Not all of the available sensors and gases support auto
calibration. If the function is not available, a manual
calibration must be performed.
WARNING
It must be insured that during the zero point calibration,
no measured gas is present in the supplied calibration
gas. Particularly with a fresh air calibration, Dräger
recommends performing a second, independent
measurement. Some sensors may only be calibrated
with synthetic air or nitrogen (observe the relevant
sensor data sheets).
NOTICE
After 10 minutes, without confirmation the device
reverts to measurement mode without performing
a calibration.
i
i
!
i
i

16 Dräger Polytron 8200/8310
Menu overview
6 Menu overview
◄Back to
measure
Information page 17 Instrument page 17 Warnings page 17
Sensor page 17 Faults page 17
Calibration page 14 Data logger page 17 Device codes page 17
Modules page 17
Settings page 18
Last cal. date page 17
Zero calibration page 14 Next cal. date page 17
Span calibration page 15
Auto CAL page 15 Logger status page 17
Graph page 17
SIL activation page 18
Instrument page 18 Alarm page 18 Alarm on/off page 18
Communication page 20 Passwords page 19 Relay A1 page 18
Sensor page 23 Date and time page 19 Relay A2 page 18
Data logger Language page 19 Alarm A1 page 18
Function key page 19 Alarm A2 page 18
Device init page 20 Set alarm A1 page 19
Datalogr. on/off SW dongle page 20 Set alarm A2 page 19
Set Datalogger Display page 20 Set fault page 19
Clear datalogr
Calibration PWD page 19
Sample time Settings PWD page 19
Peak/Average
Trigger on/off Display contrast page 20
Trigger value Operating mode page 20
Stack/Roll Display test page 20
HART interface page 20
Analog interface page 21 Polling address page 20
Profibus address page 23 Unique identifier page 21
Tag page 21
Auto CAL page 23 Dräger REGARD page 21
Sensor type page 23
Gas settings page 23 Fault current page 21
Reset sensor
required!
page 23 Warning page 21
Cal. interval page 23 Warning interval page 21
Warning current page 21
Maintenance signal page 21
Maintenance current page 22
Analog offset page 22
Set current page 22
Set concentr. page 22
Set fault page 22
Set warning page 22
Set mainten. page 22

Information menu
Dräger Polytron 8200/8310 17
7 Information menu
Specific data about the device, the sensor used and the
measured gas are summarized in this menu. The menu can be
accessed without a password. Changes to the data are not
possible.
7.1 Device settings
7.1.1 Displaying warning messages
To display warnings in plain text with warning numbers,
see chapter 9 on page 25.
The symbol is displayed when there is a warning present.
zSelect Information > Instrument > Warnings and confirm.
Warnings are displayed in plain text.
If several warnings are present, this is indicated in the
upper right-hand corner of the display, e.g., 1/3 = Screen
1 of 3.
7.1.2 displaying fault messages
To display faults present in plain text with fault numbers,
see chapter 9 on page 25.
The symbol is displayed when there is a fault present.
zSelect Information > Instrument > Faults and confirm.
Faults are displayed in plain text.
If several faults are present, this is indicated in the upper
right-hand corner of the display, e.g., 1/2 = Screen 1 of 2.
7.1.3 Displaying device codes
Display of codes for faults/warnings in a numerical table
(hexadecimal).
The symbol is displayed when there is a fault present.
zSelect Information > Instrument > Device codes and
confirm.
Faults are displayed as a hexadecimal numerical code in
tabular form. If all the numerical groups are displayed
as 00, there are no faults or warnings present.
7.1.4 Displaying installed modules
Display of the installed hardware modules.
zSelect Information > Instrument > Modules and confirm.
A list of all possible modules is displayed. The installed
modules are identified by a , those not installed by a .
zMark a module and request detailed information with .
7.2 Sensor information
This menu contains the functions for retrieving the
calibration data.
7.2.1 Display last calibration date
Display the date of the last calibration.
zSelect Information > Sensor > Last cal. date and confirm.
The date of the last calibration is displayed
7.2.2 Display next calibration date
Display the date of the next calibration.
zSelect Information > Sensor > Next cal. date and confirm.
The date when the next calibration of the sensor is due is
displayed.
7.3 Datalogger information
(only with data dongle)
This menu contains the functions for querying the datalogger.
7.3.1 Displaying the datalogger status
zSelect Information > Datalogger > Logger status and
confirm.
The current status of the datalogger is displayed: Datalogr. :
on or off (datalogger on or off)
zTo switch the datalogger on/off, see chapter 8.5.1 on page 24.
7.3.2 Displaying a graph of measured values
Show the measurements from the sensor on a 15-minute
time axis.
zSelect Information > Datalogger > Graph and confirm.
NOTICE
The fault code is very helpful for technical support and
should always be stated in full in the event of a fault.
i
i
OK

18 Dräger Polytron 8200/8310
Settings menu
8 Settings menu
This menu contains all the functions that are required for the
individual configuration of the device.
8.1 Switching SIL lock on or off
With this function, the device can be protected against
unauthorized changes to the configuration with a password.
Changing the configuration (e.g., changing the measurement
range from 100 %LEL to 50 %LEL) will cause the device to
display all the safety-relevant parameters to the user once more
before it returns to the measurement mode. The correctness of
the parameters and settings must be confirmed.
1. Select Settings > Instrument > Set SIL.
2. Select the desired state On or Off and confirm.
8.2 Device settings
8.2.1 Alarm settings (only with installed relay module)
Alarms issued via relays and LEDs.
Switching the alarm on or off
1. Select Settings > Instrument > Alarm > Alarm on/off.
2. Select the desired state On or Off and confirm.
Configuring Relay A1 or A2
This function is used to define whether the alarm relay is
energized in standard operation or in an alarm state. If the
configuration of the relay is set to "Normally energized",
the corresponding relay contact is closed in the normal case
and drops out in the event of an alarm. This configuration
results in an alarm being triggered in the event of a loss of
power to the transmitter (fail-safe). The fault relay is factory-
configured to "Normally energized" and cannot be changed.
1. Select Settings > Instrument > Alarm > Relay A1 or
Relay A2 and confirm.
2. Select Normal energy supply oder Energy supply
during alarm and confirm.
Configuring Alarm A1 or A2
1. Select Settings > Instrument > Alarm > Alarm A1 or
Alarm A2 and confirm.
2. Select the line for entering the alarm limit and confirm.
3. Set the threshold and confirm.
The setting for the alarm threshold is displayed.
4. Select Next and confirm.
The setting for the alarm direction is displayed.
5. Select Rising or Falling and confirm.
6. Select Next and confirm.
The setting for the alarm direction is displayed.
7. Select Latching or Non Latching and confirm.
8. Select Next and confirm.
The setting for the acknowledgement is displayed.
9. Select Acknowledgeable or NonAcknowledgeabl or
PreAcknowledgeabl and confirm.
10. Select Next and confirm.
The function for setting the A1 hysteresis is opened.
11. Select the line for entering the hysteresis and confirm.
12. Set the hysteresis for A1 and confirm.
The hysteresis setting for A1 is displayed.
13. Select Next and confirm.
The settings for A1 are displayed.
14. Select and confirm with Confirm.
The settings for the A1 alarm are now complete.
NOTICE
SIL lock is only available for certain sensors (see also
the sensor data sheets or the Dräger Polytron 8X00
Safety Manual).
NOTICE
When SIL is activated, the device displays all the
important parameters and setting before returning to
measurement mode. Check the parameters and
settings carefully and confirm.
CAUTION
In the normal state, the relays in the relay module are
energized. That is, the relays change their state if the
power supply is interrupted!
CAUTION
If the alarm is inactive, the alarm state will not be
issued by the LEDs or the relay interface! A warning
message will be shown on the display!
i
i
i
i
!
!
NOTICE
This function permits the setting of a zone within which
a triggered relay maintains its status until the gas
concentration is outside the zone. This function prevents
relays from chattering at an alarm threshold.
Example: A2 at 40 %LEL methane
Maximum possible hysteresis: 3 %LEL
Alarm activation at measured values ≥40 %LEL
Alarm deactivation at measured values ≤37 %LEL
(40 %LEL - 3 %LEL)
i
i

Settings menu
Dräger Polytron 8200/8310 19
The functional relationship between the various settings is
explained in the following table:
Testing Alarm A1
This function simulates the A1 alarm status.
1. Select Settings > Instrument > Alarm > Set Alarm A1
and confirm.
2. Select On or Off and confirm.
When the function is activated, the relay and the interface are
set to the A1 alarm status and the red LED gives single flashes.
Testing Alarm A2
This function simulates the A2 alarm status.
1. Select Settings > Instrument > Alarm > Set Alarm A2
and confirm.
2. Select On or Off and confirm.
When the function is activated, the relay and the interface are
set to the A2 alarm status and the red LED gives double flashes.
Testing Fault status
This function simulates the Fault alarm.
1. Select Settings > Instrument > Alarm > Set fault
and confirm.
2. Select On or Off and confirm.
When the function is activated, the Fault relay and the 4-20 mA
interface are set to the fault current and the yellow LED lights.
8.2.2 Changing passwords
The passwords for calibration (zero and span calibrations) and
for the settings (full configuration) can be defined in this menu.
1. Select Settings > Instrument > Passwords > Calibration/
Settings PWD and confirm.
2. Select the line for the password settings and confirm.
3. Set the password and confirm.
The setting for the password is displayed.
4. Select Confirm and confirm.
8.2.3 Setting the date and time
1. Select Settings > Instrument > Date and time and confirm.
2. Select the desired line and confirm in order to switch to the
Change mode.
3. Set each digit and confirm each setting.
4. Select Confirm and confirm.
8.2.4 Changing the display language
1. Select Settings > Instrument > Language and confirm.
2. Select the desired language from the list and confirm.
8.2.5 Configuring the function key
This function is used to assign a predefined function to the
key. The selected function is activated by briefly taping the
key.
1. Select Settings> Instrument > Function key and confirm.
2. Select the desired function and confirm.
zRepeating the confirmation deactivates the selected function.
zIf Bumptest is selected:
The maintenance symbol is shown on the right hand
side of the display and the maintenance signal is output.
Repeating the confirmation ends the bump test and the
device reverts to the normal measurement mode.
zIn measurement mode, the selected function can be
activated by briefly taping .
Latching and
acknowledgeable
The alarm must be acknowledged
manually. The alarm can be
acknowledged when the alarm
condition is still present.
Latching and not
acknowledgeable
The alarm must be acknowledged
manually. The alarm can only be
acknowledged when the alarm
condition is no longer present.
Latching and pre-
acknowledgeable
The alarm must be acknowledged
manually. The alarm can be
acknowledged when the alarm
condition is still present but will only be
reset when the alarm condition is no
longer present.
Non-latching and
acknowledgeable The alarm is acknowledged
automatically when there is no longer
any alarm condition present, or it can
be acknowledged manually.
Non-latching and not
acknowledgeable The alarm is acknowledged
automatically when there is no longer
any alarm condition present. The alarm
cannot be acknowledged manually.
Non-latching and pre-
acknowledgeable Corresponds to the alarm behavior
for "Non-latching" and "Not
acknowledgeable".
NOTICE
When the function Set Alarm A1 is exited in the menu,
the Polytron 8200/8310 reverts automatically to
measurement mode.
NOTICE
When the function Set Alarm A2 is exited in the menu,
the Polytron 8200/8310 reverts automatically to
measurement mode.
i
i
i
i
NOTICE
When the function Set fault is exited in the menu,
the Polytron 8200/8310 reverts automatically to
measurement mode.
i
i

20 Dräger Polytron 8200/8310
Settings menu
Selection options:
8.2.6 Resetting the device to factory settings
The following device parameters are reset to the factory
settings with this function (see chapter 12.3 on page 29).
zAlarm parameters
zPasswords
zLanguage
zFunction key
zHART interface
zData logger
zAnalog interface
zRelay configuration
1. Select Settings > Instrument > Device init. and confirm.
2. Select Confirm and confirm in order to reset the device to
the factory settings.
8.2.7 Deactivating the software dongle
This function allows the SW dongles to be deactivated before
they are removed or in the event of a fault.
A dongle can only be reactivated by switching the power to the
device off and on.
1. Select Settings > Instrument > SW dongle > Dongle
datalogger/Dongle sensortest/Dongle diagnostic and
confirm.
2. Select Deactivate func. and deactivate Dongle datalogger/
Dongle sensortest/Dongle diagnostic with .
8.2.8 Display settings
Changing the contrast
1. Select Settings > Instrument > Display > Display contrast
and confirm.
2. Change the contrast and confirm.
Changing the display mode
The display of measured values can be turned on or off with
this function.
1. Select Settings > Instrument > Display > Display mode
and confirm.
2. Select Standard or Non display and confirm.
If "Non display" has been selected, the display shows the start
screen and the symbols.
If the relay option is set and an alarm is triggered, the display
shows the current gas concentration and the red LED flashes,
regardless of the selected display mode.
Switching the display test on or off
The function of the display and the LEDs is tested with this
function.
1. Select Settings > Instrument > Display > Displaytest
and confirm.
2. Select On or Off and confirm.
When this function is activated, the display is shown inverted
and all the LEDs are lit.
8.3 Communication settings
8.3.1 Settings for the HART interface (only for devices
with the optional HART interface)
This group contains the setting functions for the HART interface.
Setting the polling address
The polling address configures the device for analog operation
(4 to 20 mA) or multidrop operation. Analog operation (4 to
20 mA) is activated by setting the polling address to 0.
For multidrop operation the polling address must be
configured to a value between 1 and 15, which will result in the
analog interface being deactivated and set to a constant
current of approx. 1 mA. To allow the control unit to request the
Unique Identifier (unique HART address) with HART command
#0, all the devices on a line must be configured with different
polling addresses. It is advisable to choose a sequence
starting with 1 and incrementing continuously.
This setting corresponds to HART command #6 (Write Polling
Address).
1. Select Settings > Communication > HART interface >
Polling address and confirm.
2. Select the address line for the entry and confirm in order to
switch to the Change mode.
3. Set the polling address and confirm.
4. Select Confirm and confirm.
Graph
(only with
datalogger)
The measurements from the sensor
are displayed graphically on a time axis
(see chapter 7.3.2 on page 17).
Faults Faults are displayed in plain text
(see chapter 7.1.2 on page 17).
Warnings Warnings are displayed in plain text
(see chapter 7.1.1 on page 17).
Bump test The bump test allows gas to be applied
without an alarm being issued. The
maintenance signal is output on the
4-20 mA output.
Sensor vitality Shows the remaining sensor vitality
(only available for certain sensors).
OK
This manual suits for next models
1
Table of contents
Other Dräger Accessories manuals

Dräger
Dräger Incubator 8000 IC Parts list manual

Dräger
Dräger X-pid 800 Series User manual

Dräger
Dräger Polytron SE Ex PR NPT1 DD User manual

Dräger
Dräger PID 5000 User manual

Dräger
Dräger Isolette 8000 plus User manual

Dräger
Dräger Polytron 8100 User manual

Dräger
Dräger SE Ex PR M User manual

Dräger
Dräger PEX 3000 User manual

Dräger
Dräger Incubator 8000 IC User manual

Dräger
Dräger caleo Configuration guide