GE SENOGRAPHE DMR+ User manual

Technical
Publications
0459
5128689–2–100
Revision 1
SENOGRAPHE DMR+
om
Operator Manual
do not duplicate
CopyrightE2005 GE Medical Systems
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

WARNING
X–RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR
UNLESS MEASURES OF PROTECTION ARE STRICTLY OBSERVED
Though this equipment is built to the highest standards of electrical and mechanical safety, the useful x–ray
beam becomes a source of danger in the hands of the unauthorized or unqualified operator. Excessive
exposure to x–radiation causes damage to human tissue.
Therefore,adequateprecautionsmustbetakentopreventunauthorizedorunqualifiedpersonsfromoperating
this equipment or exposing themselves or others to its radiation.
Before operation, persons qualified and authorized to operate this equipment should be familiar with the
RecommendationsoftheInternationalCommissiononRadiologicalProtection,containedinAnnalsNumber
60 of the ICRP, and with applicable national standards.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Services Offices of GE Medical Systems
For service issues, contact the Service Office for your country.
AMERICAS
UNITED STATES OF AMERICA
GE Medical Systems Headquarters Tel: +1–414–544–3011
3000 North Grandview Blvd.
WAUKESHA, WI 53188
Mail:P.O. Box 414
MILWAUKEE, WI 53201
LATIN AMERICA
GE Medical Systems Tel: +1–305–497–1200
Latin America Headquarters
5101 NW 21st Avenue, Suite 350
FORT LAUDERDALE, FL 33309
USA
ASIA
NIPPON (Japan)
GEMSA Headquarters Tel: 81–426–48–2940
GE Medical Systems Asia Fax: 81–426–48–2905
67–4, Takakura, Hachioji
TOKYO, 192–0033, JAPAN
REST OF ASIA
GE Pacific Pte. Ltd (GEMS Asia) Tel: +65–291–8528
South East Asia Pacific Operations (SEAPO) Fax: +65–291–7006
298 Tiong Bahru Road
#15–01/06 Tiong Bahru Plaza
SINGAPORE 0316
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

EUROPE
BELGIE/BELGIQUE (Belgium)
GE Medical Systems Benelux NV/SA (NL) Tel:+32 2 644 09 38
Manhattan Center (F) Tel:+32 2 644 08 42
Bolwerklaan 21 b 9 Avenue du Boulevard (LUX) Tel:0800 2973
B–1210 BRUSSEL/BRUXELLES Fax: +32 2 207 73 33
DANMARK (Denmark)
GE Medical Systems Danmark Tel: +45 45 51 00 55
Skovlytoften 4 Fax: +45 42 42 59 89
DK–2840 HOLTE
DEUTSCHLAND (Germany)
GE Medical Systems Deutschland GmbH & Tel: 49–6102–36–0
Co. KG Fax: +49 610–36–2588
Martin–Behaim–Strasse 10
D–63263 NEU_ISENBURGS
ESPAÑA (Spain)
GE Medical Systems España Tel: +34 1 663 25 00
Avda. de Europa 22 Fax: +34 1 663 25 01
Parque Empresarial la Moraleja
E–28100 ALCOBENDAS
FRANCE (France)
GE Medical Systems Tel: +33 (0)1 30 70 40 40
283 rue de la Minière
BP34
F–78533 BUC CEDEX
HELLAS (Greece)
GE Medical Systems Hellas Tel: +30 1 93 24 582
41 Nikolaou Plastira Street Fax: +30 1 93 58 414
GR–171 21 NEA SMYRNI
ITALIA (Italy)
GE Medical Systems Italia Tel: +39 02 64 22 01
Viale Fulvio Testi 28–B Fax: +39 02 64 22 0401
20126 MILANO
NEDERLAND (Netherlands)
GE Medical Systems Nederland B.V. Tel: +31 73 6 457 457
Hambakenwetering 1 Fax: +31 73 6 441 233
NL–5231 DD S HERTOGENBOSCH
ÖSTERREICH (Austria)
GE GesmbH Medical Systems Austria Tel: 0660 8651 (gebührenfrei)
Prinz Eugen Strasse 8/8 Fax: +43 1 505 38 74
A–1140 WIEN Tlx: 136314
POLSKA (Poland)
GE Medical Systems Polska
Krzywickiego 34 Tel: +48 2 625 59 62
P–02–078 WARSZAWA Fax: +48 2 615 59 66
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PORTUGAL (Portugal)
GE Medical Systems Portuguesa S.A.
Rua Sà da Bandeira, 585 Tel: +351 2 2007696/97
Apartado 4094 Fax: +351 2 2084494
P–4002 PORTO CODEX Tlx: 22804
ROSSIYA (Russia)
GE Medical Systems Tel:+7 095 935 72 41
Kosmodamianskaya nab. 52, Bldg 1, 6th Floor Fax:+7 095 935 73 46 and 48
113054 MOSCOW Tel (satellite): +7 502 220 30 39
Fax (satellite): +7 502 220 32 59
Tlx:613020 GEMED SU
SVERIGE (Sweden)
GE Medical Systems Tel: +46 8 457 95 20
Box 6768 Fax: +46 8 457 95 47
St. Eriksgatan 117 Tlx: 12228 CGRSWES
S–113 85 STOCKHOLM
SCHWEIZ/SUISSE (Switzerland)
GE Medical Systems (Schweiz) AG Tel: 155 6958 (gebührenfrei)
Sternmattweg 1 Fax: +41 41 421859
CH–6010 KRIENS
TÜRKIYE (Turkey)
GE Medical Systems Turkiye A.S. Tel: +90 212 75 5552
Mevluk Pehliran Sodak Fax: +90 212 211 2571
Yilmaz Han, No 24 Kat 1
Gayretteppe
ISTANBUL
UNITED KINGDOM
IGE Medical Systems
Coolidge House Tel: +44 753 874000
352 Buckingham Avenue Fax: +44 753 696067
SLOUGH
Berkshire SL1 4ER
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Blank page.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

REGULATORY REQUIREMENTS
This product complies with the regulatory requirements of the following:
DCouncil Directive 93/42/EEC concerning medical devices: the 0459 label affixed
to the product testifies compliance to the Directive.
For a system, the location of the CE marking label is described in the system manual.
European registered place of business:
GE Medical Systems Europe
Quality Assurance Manager
BP 34
F 78533 BUC CEDEX France
Tel: +33 (0)1 30 70 40 40
DMedical Device Good Manufacturing Practice Manual issued by the FDA (Food and
Drug Administration, Department of Health, USA).
DUnderwriters’ Laboratories, Inc. (UL), an independent testing laboratory.
DCanadian Standards Association (CSA).
DInternational Electrotechnical Commission (IEC), international standards organiza-
tion, when applicable.
DUSA/HHS:
United States Federal law restricts this device to use by or on the order of a
physician.
DGeneral Electric Medical Systems is ISO 9001 and EN 46001 certified.
DThe original document was written in English.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

ELECTROMAGNETIC COMPATIBILITY (EMC)
This equipment complies with IEC60601–1–2 Edition 2 EMC standard for medical
devices.
This equipment generates, uses, and can radiate radio frequency energy. The equipment
may cause radio frequency interference to other medical and non-medical devices and
radio communications. To provide reasonable protection against such interference, this
product complies with radiated emissions asperCISPR11 Group 1, Class A standard limits.
Detailed requirements and recommendations about power supply distribution and
installation are listed in the Pre–Installation Manual (pim) shipped with your system.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment is found to cause interference (which may be determined by turning the
equipment on and off), the user (or qualified service personnel) should attempt to correct
the problem by one or more of the following measure(s):
– reorient or relocate the affected device(s)
– increase the separation between the equipment and the affected device
– power the equipment from a source different from that of the affected device
– consult the point of purchase or service representative for further suggestions
The manufacturer is not responsible for any interference caused by using other than
recommended interconnect cables or by unauthorized changes or modifications to this
equipment. Unauthorized changes or modifications could void the users’ authority to
operate the equipment.
All interconnect cables to peripheral devices must be shielded and properly grounded. Use
of cables not properly shielded and grounded may result in the equipment causing radio
frequency interference.
Do not use devices which intentionally transmit RF Signals (Cellular Phones, Transceivers,
or Radio Controlled Products) in the vicinity of this equipment as it may cause performance
outside the published specifications. Recommended separation distances are detailed in
the Pre–Installation Manual (pim) shipped with your system. Keep the power to this type of
devices turned off when near this equipment.
The medical staff in charge of this equipment is required to instruct technicians, patients,
and other people who may be around this equipment to comply fully with the above
requirement.
Further data and recommendations for meeting Electromagnetic Compatibility
requirements for a typical installation are given in the Pre–Installation Manual (pim)
shipped with your system. Note that the magnetic field of an MRI device located nearby
may cause a risk of interference. Magnetic field amplitude limits are specified in the
Pre–Installation Manual (pim) shipped with your system.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

RECYCLING:
Machines or accessories at end–of–life:
The elimination of machines and accessories must be in accordance with national
regulations for waste processing.
All materials and components that could pose a risk to the environment must be removed
from the end–of–life machines and accessories (examples: dry and wet cell batteries,
transformer oil, etc...).
Please consult your local GEMS representative before discarding these products.
Packing materials:
The materials used to pack our equipment are recyclable. They must be collected and
processed in accordance with the regulations in force for the country where the machinesor
accessories are unpacked.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Blank page
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

ix
TABLE OF CONTENTS
CHAPTER TITLE PAGE
REVISION HISTORY xv. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
LIST OF EFFECTIVE PAGES xvi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 INTRODUCTION 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 1 – PRESENTATION 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 2 – DESCRIPTION 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1 X-ray System 1–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–1 Tube-Support Arm 1–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–2 Receptor-Support Arm 1–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–3 Film-Holder Assembly 1–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2 Power Supply Cabinet 1–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3 Accessories 1–18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–4 Optional Accessories 1–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–5 Control Console 1–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–5–1 Control Keyboard 1–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 3 – CONTROL KEYBOARD 1–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–1 Lefthand Section of Control Keyboard 1–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–2 Central Section of Control Console Keyboard 1–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3 Righthand Section of Control Keyboard 1–33. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 PREPARING AN EXAMINATION 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 1 – PREPARING THE EQUIPMENT 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 2 – PATIENT POSITIONING 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 3 – EXPOSURE MODE 2–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–1 AOP Mode 2–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–2 AEC Mode 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–2–1 Standard Exposures 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–2–2 Magnification 2–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–2–3 Examination of Chest Wall 2–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3 Manual Mode 2–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3–1 Standard Exposures 2–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3–2 Magnification 2–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3–3 Examination of Chest Wall 2–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

x
TABLE OF CONTENTS
CHAPTER TITLE PAGE
3 MAINTENANCE 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 1 – CLEANING & DISINFECTION 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GENERAL INFORMATION 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
EQUIPMENT CLEANING INSTRUCTIONS 3–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
LOW LEVEL OR INTERMEDIATE LEVEL DISINFECTION 3–2. . . . . . . . . . . . . . . . . .
HIGH LEVEL DISINFECTION 3–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RECOMMENDED GERMICIDES 3–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 2 – PREVENTIVE MAINTENANCE 3–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 3 – MESSAGES 3–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 4 – MEDICAL PROGRAMMING 3–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–1 GENE (language) 3–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–1–1 Access to Programming Mode 3–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–1–2 Selecting a Function 3–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–1–3 Confirm, & Return to APPLICATION Menu 3–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2 Arm: Speed of Compression 3–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2–1 Access to Programming Mode 3–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2–2 Selecting a Function 3–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2–3 Confirm, & Return to APPLICATION Menu 3–18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3 Arm: Compressive Force 3–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3–1 Access to Programming Mode 3–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3–2 Selecting a Function 3–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3–3 Confirm, & Return to APPLICATION Menu 3–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–4 Arm: Automatic Decompression 3–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–4–1 Access to Programming Mode 3–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–4–2 Selecting a Function 3–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–4–3 Confirm, & Return to APPLICATION Menu 3–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–5 Arm: Beep 3–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–5–1 Access to Programming Mode 3–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–5–2 Selecting a Function 3–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–5–3 Confirm, & Return to APPLICATION Menu 3–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–6 Decompression Height 3–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–6–1 Access to Programming Mode 3–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–6–2 Selecting a Function 3–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–6–3 Confirm, & Return to APPLICATION Menu 3–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

xi
TABLE OF CONTENTS (CONT.)
CHAPTER TITLE PAGE
4–7 Exposure Interlock Disable 3–27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–7–1 Access to Programming Mode 3–27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–7–2 Selecting a Function 3–27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–7–3 Confirm, & Return to APPLICATION Menu 3–28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 SPECIFICATIONS 4–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 1 – PHYSICAL SPECIFICATIONS 4–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 2 – TECHNICAL SPECIFICATIONS 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1 Electrical Specification 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–1 Line voltage specifications 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–2 kVA load characteristics 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–3 Input impedance 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–4 Generator Output (excluding tube) 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–5 Duty Cycle
The generator is always limited by the tube: 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–6 Line Conditioner Resonance Frequency 4–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–7 Generator Power (excluding tube) 4–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–8 Maximum Tolerance of Displayed Constants (with and without AEC or AOP) 4–7. . . . . .
2–1–9 Measurement Conditions 4–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2 Filters 4–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3 Beam-limiting Devices 4–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–4 Special Material Specifications 4–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 3 – ENVIRONMENT 4–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SECTION 4 – MEANING OF SYMBOLS 4–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 REGULATIONS 5–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

xii
Blank page
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

xiii
FOREWORD
Thismanual is provided for SENOGRAPHE DMR operators. Itis designedto supplyall the
information required for the correct use of this equipment.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

xiv
Blank page
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

!
! ! "#
!#$
%&&& '($ '!( ) *(" * ) +# "
*, # , " "
-(
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

"
" " "
. /# ! !
! 0 ! 0
! 0) ) ! )
Note for
the Technical Writer
This document is using the Conditional selection. It is a slave
document.
To maintain this document it is mandatory to use the “Master File”
2309392–2–800 as the Interleaf source.
This “Master File” contains a “Read me” which explains how to
proceed.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

# $
#
The SENOGRAPHE DMR is an x-ray system used primarily for mammography
examinations in the standing or sitting position and under various views and magnification
factors.
Breast puncture by tridimensional localization is an additional function by using one of the
following options:
. Stereotix 2,
. Senovision.
If required, the system can provide radiographs of the hand and foot of high image quality.
The major features of the SENOGRAPHE DMR are the dual track x-ray tube, entirely
automatic exposure modes, and its new ergonomic design.
Each of the two anode tracks is composed of a different metal. This allows the best
compromise between contrast and dose given to the patient.
The rhodium (Rh) track provides improved penetration with a reduced radiation dose to the
very dense breast. The molybdenum (Mo) track provides high quality, mammography
images in all other cases.
Standard or magnification examinations can be performed with each track. Focal spot sizes
are: 0.3 for contact exams and 0.1 for magnification exams.
Two modes of automatic exposure are available:
Automatic Exposure Control (AEC) Mode controls density (mAs) and provides
exposures of constant optical density.
Automatic Optimization Parameters (AOP) Mode controls radiation. For a given
priority (dose reduction, contrast quality or compromise of dose and contrast), the AOP
Mode selects the track, filter, and kV. It includes the AEC mode, and provides an automatic
selection of the radiologic parameters (track, filter, kV, and mAs).
The new design ensures speed and convenient positioning, with examination comfort for
the patient.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

1223*.45.16
" !
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Other manuals for SENOGRAPHE DMR+
1
Table of contents
Other GE Medical Equipment manuals

GE
GE MiCOM P40 Agile User manual

GE
GE Corometrics 250cx Series User manual

GE
GE MAC 2000 User manual

GE
GE Definium 5000 User manual

GE
GE Proteus XR/a User manual

GE
GE LOGIQ E9 User manual

GE
GE Corometrics 250cx Series User manual

GE
GE OPTIMA MR360 Manual

GE
GE SENOGRAPHE DMR+ User manual

GE
GE MAC 3500 User manual
Popular Medical Equipment manuals by other brands

Shimadzu
Shimadzu MUX-100D installation manual

MedNet
MedNet HeartTrends user manual

DERMAVIT
DERMAVIT Diamond Shield Crystal instruction manual

Moretti
Moretti SKEMA MO351N-MO361N instruction manual

Jedmed
Jedmed PHOENIX III Product Information / Technical Service Manual

Siemens Healthcare
Siemens Healthcare MULTIX Impact C Operator's manual